The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. The device includes the Atrial Fibrillation Detection Feature. The feature analyzes pulse pressure wave data captured during the user's blood pressure measurement to identify episodes of atrial fibrillation (AFib) and provides a notification on the display screen to the user. This feature is intended to be used for home screening for the possibility of AFib. The Atrial Fibrillation Detection Feature is not intended to provide a notification on every episode of AFib and the absence of detection is not intended to indicate the absence of atrial fibrillation. The feature is not intended to replace traditional methods of diagnosis or treatment or to be used in the management of patients with AFib. The device is intended for over-the-counter (OTC) use in the home environment. The device is intended for use in adults (22 years or older); it has not been tested for and is not intended for use in pregnant women or people with implanted pacemakers or defibrillators.
Device Story
Upper arm oscillometric blood pressure monitor; battery or AC powered. Captures pulse pressure wave data during standard BP measurement. Integrated software algorithm analyzes pulse wave data to identify potential AFib episodes. If AFib detected, device displays 'Possible AFib' icon and 'AF' text; BP/pulse readings are suppressed. Used in home environment by adult patients. Optional Omron Connect mobile app syncs data via Bluetooth Low Energy (BLE) for storage/display. Provides early AFib screening; does not replace clinical diagnosis or manage known AFib. Benefits include high sensitivity/specificity for AFib detection in a familiar, easy-to-use format. Risks include false positives/negatives and potential for user over-reliance or inappropriate treatment changes.
Clinical Evidence
Prospective, non-randomized, multi-center study (n=573). Compared device AFib detection against physician-adjudicated 12-lead ECG ground truth. Sensitivity: 95.09% (BP7460) and 94.72% (BP5460); Specificity: 98.64% (BP7460) and 98.30% (BP5460). All p-values < 0.0001. No adverse events reported.
Technological Characteristics
Oscillometric upper arm blood pressure monitor; includes monitor and arm cuff. Battery-powered with optional AC adapter. Connectivity via Bluetooth Low Energy (BLE) to mobile app. Software-based AFib detection algorithm. Basic software documentation level. Complies with FDA guidance for software and human factors engineering.
Indications for Use
Indicated for adults (22+ years) for home blood pressure and pulse rate measurement and screening for possible atrial fibrillation. Contraindicated for pregnant women and individuals with implanted pacemakers or defibrillators. Not for use in patients with peripheral arterial disease (PAD).
Regulatory Classification
Identification
The Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP7460 (HEM-7371T1-AZ) and BP5460 (HEM-7372T1-AZAZ)) is an oscillometric-based, over-the-counter, upper arm blood pressure monitor intended for use in adults (22 years or older) to measure blood pressure and pulse rate. It includes an AFib detection feature that analyzes pulse pressure wave data captured during blood pressure measurement to identify episodes of atrial fibrillation and provide a notification on the display screen. It is intended for home screening for the possibility of AFib and is not intended to replace traditional diagnosis or treatment or to be used in the management of patients with known AFib.
Special Controls
In combination with the general controls of the FD&C Act, the oscillometric-based, over-thecounter, atrial fibrillation notification feature is subject to the following special controls:
- 1) Clinical performance testing must compare the device algorithm performance against readings from patients with and without AFib that have been clinically adjudicated and evaluate the sensitivity and specificity of the output measurement. Justification for the clinical adjudication procedures to determine ground truth must be provided.
- 2) Non-clinical performance testing of the algorithm must include:
- a. Evaluation and qualification of interoperability with the oscillometric hardware components; and
- b. Validation of the device to detect adequate motion/noise artifacts during measurement.
- 3) Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the diagnostic algorithm and its inputs and outputs.
- 4) Human factors and usability testing must demonstrate the following:
- a. The user can correctly use the device based solely on reading the device labeling;
- The user can correctly interpret the device output and understand when to seek b. medical care; and
- c. The user understands situations during which the device should not be used.
- 5) Labeling must include:
- a. A description of what the device measures and outputs to the user;
- b. Situations in which the device may not operate at an expected performance level and when its use may pose risks to the user;
- c. A summary of the clinical performance testing conducted with the device, including user demographics;
- d. Guidance on interpretation of any measurements, detection results, and error messages: and
- e. Hardware platform requirements.
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR
# OMRON BLOOD PRESSURE MONITOR WITH ATRIAL FIBRILLATION (AFIB) DETECTION FEATURE (BP7460 (HEM-7371T1-AZ)); OMRON BLOOD PRESSURE MONITOR WITH ATRIAL FIBRILLATION (AFIB) DETECTION FEATURE (BP5460 (HEM-7372T1-AZAZ))
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Oscillometric-based, over-the-counter, atrial fibrillation notification feature. An oscillometric-based, over-the-counter, atrial fibrillation notification feature is a software algorithm for analysis of signals from an oscillometric, non-invasive blood pressure measurement system to screen for possible atrial fibrillation in the device user.
NEW REGULATION NUMBER: 21 CFR 870.1135
CLASSIFICATION: Class II
PRODUCT CODE: OXY
#### BACKGROUND
DEVICE NAME: Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP7460 (HEM-7371T1-AZ)); Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP5460 (HEM-7372T1-AZAZ))
SUBMISSION NUMBER: DEN230076
DATE DE NOVO RECEIVED: November 13, 2023
#### SPONSOR INFORMATION:
Omron Healthcare, Inc. % Veranex, Inc. 224 Airport Parkway San Jose, California 95110
#### INDICATIONS FOR USE
The Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP7460 (HEM-7371T1-AZ)); Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP5460 (HEM-7372T1-AZAZ)) is indicated as follows:
{1}------------------------------------------------
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
The device includes the Atrial Fibrillation Detection Feature analyzes pulse pressure wave data captured during the user's blood pressure measurement to identify episodes of atrial fibrillation (AFib) and provides a notification on the display screen to the user. This feature is intended to be used for home screening for the possibility of AFib.
The Atrial Fibrillation Detection Feature is not intended to provide a notification on every episode of AFib and the absence of detection is not intended to indicate the absence of atrial fibrillation. The feature is not intended to replace traditional methods of diagnosis or treatment or to be used in the management of patients with known AFib.
The device is intended for over-the-counter (OTC) use in the home environment. The device is intended for use in adults (22 years or older); it has not been tested for and is not intended for use in pregnant women or people with implanted pacemakers or defibrillators.
# LIMITATIONS
The possible AFib detection feature evaluates for a possibility of AFib Only. It will not detect other potentially life-threatening arrhythmias or diseases, such as a possibility of other cardiac arrhythmias or heart attack.
Do NOT adjust medication (including altering your use of any drug or treatment) based on readings from this blood pressure monitor.
Do NOT postpone/stop regular checkups or physician visits based on the results that you are getting from this monitor.
DO NOT use this monitor in areas containing high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, computerized tomography (CT) scanners. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
Not intended for use by individuals under age 22.
Not intended for use if you have an implanted defibrillator or pacemaker.
Not intended for use if you are pregnant.
Safety and effectiveness in patients with peripheral arterial disease (PAD) have not been established.
Not intended for the diagnosis, treatment, or management of AFib.
{2}------------------------------------------------
It does NOT continuously monitor your heart and therefore cannot alert you if AFib happens at any other time. This monitor cannot detect all forms of AFib. If the irregularity of the heart rhythm is too small, it may not be detected.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The Omron Blood Pressure Monitor with AFib Detection Feature is an OTC, at home use, upper arm blood pressure monitor. The device consists of a monitor and an arm cuff designed to be used on the upper arm. It is battery powered and can optionally be connected to an AC adapter. The proposed device measures blood pressure (systolic [SYS] and diastolic [DIA]) and pulse rate (PR), detects the presence of irregular heartbeats, and can also detect the possible presence of AFib by analyzing pulse pressure wave data captured during the user's blood pressure measurement to identify episodes of AFib and provides a notification on the display screen to the user.
Image /page/2/Picture/4 description: The image shows an Omron blood pressure monitor and cuff. The monitor displays a systolic reading of 118, a diastolic reading of 78, and a pulse of 70. The monitor also displays the time as 7:30 and indicates that the reading is the latest. The cuff is black and has the Omron logo on it.
Figure 1
Additionally, the Monitor may be used with the Omron Connect mobile application ("App"), an optional software-only accessory residing on a consumer smart device which connects to the device via Bluetooth Low Energy (BLE) and transfers information on completed device recordings for storage and redisplay.
When the measurement is complete, if atrial fibrillation has not been detected in the reading, and the time/date appear alternately.
If atrial fibrillation is detected in the reading, the device displays the "Possible AFib" icon and furthermore displays the characters "AF" in place of blood pressure readings. No blood pressure or pulse rate readings are displayed to the user as shown below.
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image shows a digital display with the time 8:00 AM at the top left. There is a small icon of a person with the number 1 next to it at the top right. In the center of the display, the letters 'AF' are shown in a large digital font. At the bottom left, there is a message that says 'Possible AFib' with a heart icon.
Figure 2
### SUMMARY OF NONCLINICAL/BENCH STUDIES
### SOFTWARE
The monitor has a Basic Documentation Level. Appropriate documentation was provided to support the validation of the software for Basic Documentation in accordance with the FDA Guidance Document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," (issued June 14, 2023). Documentation describing the software/firmware, software specifications, architecture design, software development environment, revision level history, unresolved anomalies, and cybersecurity provide the foundation that the software will operate in a manner as described in the specifications. Risk Management/Hazard Analysis was performed to characterize software risks including device malfunction and measurement related errors. The submission included verification and validation (V&V) testing to ensure the mitigation measures were successful.
### PERFORMANCE TESTING - BENCH
| Test | Purpose | Method | Acceptance Criteria | Results |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------------|---------|
| Bench Testing for<br>confirming the<br>influence of body<br>movement on<br>possible Afib<br>detection | The purpose of this test is<br>to demonstrate that low<br>levels of motion below the<br>error threshold of the<br>device do not have a<br>clinically significant effect<br>on the atrial fibrillation<br>detection accuracy of the<br>proposed device. | (b) (4) | | PASS |
# USABILITY TESTING
Usability Testing was provided in accordance with the FDA Guidance Document, "Applying Human Factors and Usability Engineering to Medical Devices- Guidance for Industry and Food and Drug Administration Staff" (issued February 02, 2016).
{4}------------------------------------------------
Sixty (60) participants participated in the human factors validation study to demonstrate the usability of the blood pressure monitor. The testing was performed in a simulated-use environment associated with self-selecting, setting up, and using the blood pressure monitor. All subjects responded to critical tasks by responding to the output of the monitor.
Testing identified critical tasks as those tasks where the user does not understand the output from the device, fails to seek medical care despite symptoms or modifies their treatment plan based on the output. Each task was assessed for completion and success criteria were clearly defined. Overall, usability testing demonstrated that the AFib Detection Feature is safe and has appropriate performance for the intended users, uses, and use environments.
## SUMMARY OF CLINICAL INFORMATION
The study objective is to assess the AFib classification performance of OMRON BPMs with the AFib Detection Feature. The performance of the BPM-A (OMRON BPM with cuff HEM-FL31: BP7460) and BPM-B (OMRON BPM with cuff HEM-RML31: BP5460) was evaluated by sensitivity and specificity as calculated by comparing output data of the subject devices to ground truth. Ground truth for AFib identification was established from ECG adjudication performed by multiple independent expert adjudicators.
### Methods
The overall study was a prospective, non-randomized, multi-center study. The study enrolled 267 subjects with and 292 without a known diagnosis of atrial fibrillation into two separate cohorts. Key exclusion criteria included the presence of a pacemaker or implantable cardioverterdefibrillator, heart failure class III or IV, and arterio-venous shunt. Subjects in the without cohort must not have any known diagnosis of AF. To be enrolled in the AF cohort, the subject must be in atrial fibrillation at the time of data collection.
Upon enrollment, the participant had the 12-lead ECG electrodes applied by the Investigator, and participants used the requisite cuff of the Omron BP Monitor with AFib screening on their upper arm. Then the participant would sit in a chair with their feet on the floor, rest their arm on a table so that the cuff is same level as their heart. Lastly, the investigator started the 12-lead ECG recording device and Omron monitor simultaneously. The participant was coached to remain still and do not talk until the measurement process was complete.
Three independent board-certified cardiologists reviewed all ECG recordings and assigned a classification of "Interpretable" or "Uninterpretable" classification means that it is difficult to perform interpretation of AFib due to low signal quality in the ECG reading. If the ECG is "Interpretable", the adjudicators classified the ECG readings into "AFib is observed (AFib)" or "AFib is not observed (non-AFib)". If "AFib is observed (AFib)", the adjudicator will indicate which ECG Lead(s) the AFib is observed. If the readers disagreed on the diagnosis, the final interpretation was determined by the simple majority rule.
{5}------------------------------------------------
# Study Endpoints
## Primary Endpoint
Sensitivity and specificity of the AFib Detection Feature in detecting AF compared with physician-adjudicated 12-lead ECG. The acceptance criteria that the sensitivity and specificity should be statistically significant to the target goal (85%).
## Secondary Endpoint
- . Subgroup analyses of sensitivity and specificity (Age, Gender, Race, Weight/BMI, Arm size and Site)
- Comparison of ECG reading results of the clinical study and ECG reading results by . independent adjudicators.
# Results
Subject characteristics
The study enrolled a total of 573 subjects at 5 investigational sites. Subject disposition is provided in Figure 4.
Image /page/5/Figure/9 description: The image is a flowchart titled "Figure 3" that shows the flow of subjects through a study. The study started with 654 subjects screened, and 80 screen failures and 1 withdrawal resulted in 573 subjects enrolled. 14 subjects were excluded from mITT due to protocol deviations, device deficiency, and ECG interpretation issues, resulting in 267 subjects in the mITT subpopulation AFib and 292 subjects in the mITT subpopulation Non-AFib.
The median age was 69 years, ranging from 22-93. Comparing to the Non-Affb cohort, subjects in the AFib cohort were older (mean age 73.0 vs 57.4) and less likely to be female (32.2% vs 67.1%).
{6}------------------------------------------------
There were 14 cases excluded from the study and their subject information included:
- . Three (3) cases where the quality of the ECG waveform was poor, and the reader was not confident in interpreting them.
- One (1) case where measurement was not possible due to device deficiency in the ECG . device.
- Four (4) cases of protocol deviation where BP measurement was not performed. ●
- . Four (4) cases of protocol deviation where ECG data was missing.
- One (1) case of protocol deviation where the subject was a patient with an implanted ● pacemaker.
- . One (1) case of protocol deviation of inappropriate cuff size.
The results indicate that the study met the protocol specified primary endpoint as shown below.
| Device | Sensitivity | | | Specificity | | |
|-----------------------------------------|-------------|-----------------|----------|-------------|-----------------|----------|
| | Result | 95% CI | p-value | Result | 95% CI | p-value |
| OMRON BP Monitor with<br>HEM-FL31 cuff | 0.9509 | 0.9249 - 0.9769 | < 0.0001 | 0.9864 | 0.9655 - 0.9963 | < 0.0001 |
| OMRON BP Monitor with<br>HEM-RML31 cuff | 0.9472 | 0.9202 - 0.9741 | <0.0001 | 0.9830 | 0.9608 - 0.9945 | < 0.0001 |
| | BP7460 (HEM-FL31-B cuff) | | | BP5460 (HEM-RML31-B cuff) | | |
|-----------------------------------------------------------|--------------------------|---------|-------|---------------------------|---------|-------|
| | Afib Reference | | | | | |
| Afib Final Alert | Afib | No Afib | Total | Afib | No Afib | Total |
| Monitor does not output “Afib alert” with error output | 1 | 0 | 1 | 1 | 0 | 1 |
| Monitor does not output “Afib alert” without error output | 12 | 290 | 302 | 13 | 289 | 302 |
| Monitor output “Afib alert” | 252 | 4 | 256 | 251 | 5 | 256 |
| Total | 265 | 294 | 559 | 265 | 294 | 559 |
# Safety
There were no adverse events reported for this study.
### Pediatric Extrapolation
The device is indicated for use only in adults - that is, persons aged 22 years and older. The Federal Food. Drug. and Cosmetic Act defines pediatric patients as persons up to but not including the 22nd birthday. In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
The labeling for the device is sufficient and satisfies the requirements of 21 CFR 801.109. The labeling consists of Instructions for Use. The Instructions for Use include the indications for use; a description of the device, precautions; a detailed summary of the clinical data collected in support of the device; a list of potential adverse events; and instructions for the safe use of the device.
Please see the Limitations section above for important contraindications, warnings and precautions presented in the device labeling. RISKS TO HEALTH
{7}------------------------------------------------
The table below identifies the risks to health that may be associated with use of the oscillometricbased, over-the-counter, atrial fibrillation notification feature and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| False positive resulting in unnecessary medical<br>intervention | Clinical performance testing |
| | Non-clinical performance testing |
| | Software verification, validation, and hazard<br>analysis |
| | Human factors testing |
| | Labeling |
| | False negative resulting in failure to detect<br>atrial fibrillation and delayed treatment. |
| Non-clinical performance testing | |
| Human factors testing | |
| Software verification, validation, and hazard<br>analysis | |
| Labeling | |
| Misinterpretation and/or over-reliance on<br>device output, leading to failure to seek<br>treatment despite symptoms, or inappropriate<br>treatment changes. | |
| | Software verification, validation, and hazard<br>analysis |
| | Labeling |
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the oscillometric-based, over-thecounter, atrial fibrillation notification feature is subject to the following special controls:
- 1) Clinical performance testing must compare the device algorithm performance against readings from patients with and without AFib that have been clinically adjudicated and evaluate the sensitivity and specificity of the output measurement. Justification for the clinical adjudication procedures to determine ground truth must be provided.
- 2) Non-clinical performance testing of the algorithm must include:
- a. Evaluation and qualification of interoperability with the oscillometric hardware components; and
- b. Validation of the device to detect adequate motion/noise artifacts during measurement.
- 3) Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the diagnostic algorithm and its inputs and outputs.
- 4) Human factors and usability testing must demonstrate the following:
- a. The user can correctly use the device based solely on reading the device labeling;
- The user can correctly interpret the device output and understand when to seek b. medical care; and
- c. The user understands situations during which the device should not be used.
{8}------------------------------------------------
- 5) Labeling must include:
- a. A description of what the device measures and outputs to the user;
- b. Situations in which the device may not operate at an expected performance level and when its use may pose risks to the user;
- c. A summary of the clinical performance testing conducted with the device, including user demographics;
- d. Guidance on interpretation of any measurements, detection results, and error messages: and
- e. Hardware platform requirements.
## BENEFIT-RISK DETERMINATION
The device is a digital monitor intended for use in analyzing pulse pressure wave data captured during the user's blood pressure measurement to identify episodes of AFib and provides a notification on the display screen to the user. As a blood pressure cuff, there is a possibility of bruising, due to blood flow interference. Overall, there are minimal safety concerns.
The risks of the device include unnecessary medical attention, over-reliance on the device outputs, and misinterpretation of the device output. A false positive result can lead to additional unnecessary medical procedures, which uses unnecessary medical resources leading to the clinician's workload being disrupted. A false negative result may falsely reassure the user and cause delay or inappropriate changes in medical evaluation and treatment. There is a potential risk of a patient self-adjusting AF treatments.
For asymptomatic undiagnosed AF patients, the device's probable benefit is to provide early detection of the disease. The device can identify AF with high sensitivity and specificity with an easy to use, familiar device to most patients, while obtaining blood pressure data.
Overall, the probable benefits outweigh the probable risks given the available information. There is reasonable assurance of the safety and effectiveness for this device for the intended use.
### PATIENT PERSPECTIVES
This submission did not include specific information on patient perspectives for this device.
### BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
The device includes the Atrial Fibrillation Detection Feature analyzes pulse pressure wave data captured during the user's blood pressure measurement to identify
{9}------------------------------------------------
episodes of atrial fibrillation (AFib) and provides a notification on the display screen to the user. This feature is intended to be used for home screening for the possibility of AFib.
The Atrial Fibrillation Detection Feature is not intended to provide a notification on every evisode of AFib and the absence of detection is not intended to indicate the absence of atrial fibrillation no disease process is present. The feature is not intended to replace traditional methods of diagnosis or treatment or to be used in the management of patients with known AFib.
The device is intended for over-the-counter (OTC) use in the home environment. The device is intended for use in adults (22 years or older); it has not been tested for and is not intended for use in pregnant women or people with implanted pacemakers or defibrillators.
The probable benefits outweigh the probable risks for the Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP7460 (HEM-7371T1-AZ)); Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP5460 (HEM-7372T1-AZAZ)). The device provides benefits, and the risks can be mitigated by the use of general and the identified special controls.
## CONCLUSION
The De Novo for the Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP7460 (HEM-7371T1-AZ)); Omron Blood Pressure Monitor with Atrial Fibrillation (AFib) Detection Feature (BP5460 (HEM-7372T1-AZAZ)) is granted, and the device is classified as follows:
Product Code: QXY
Device Type: Oscillometric-based, over-the-counter, Atrial Fibrillation Notification
Regulation Number: 21 CFR 870.1135
Class: Class II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.