The allay Nerve Cap is indicated for use in adults aged 22 years or older as a physical barrier to separate the peripheral nerve end from the surrounding environment to reduce the risk of the development of a symptomatic neuroma.
Device Story
The allay Nerve Cap is an in situ polymerizing peripheral nerve cap. It consists of precursor materials delivered to the end of a peripheral nerve; these materials polymerize in situ to form a physical barrier. This barrier isolates the nerve end from the surrounding in vivo environment, thereby reducing the risk of symptomatic neuroma formation. The device is intended for use by clinicians in a surgical setting. By preventing nerve end interaction with surrounding tissues, the device aims to mitigate pain and nerve sensitivity associated with neuroma development.
Clinical Evidence
No clinical data provided. The safety and effectiveness are supported by non-clinical performance testing, including in vivo performance testing, biocompatibility evaluation, polymerization process characterization, and human factors/usability testing as required by the special controls.
Technological Characteristics
In situ polymerizing peripheral nerve cap. Composed of precursor materials that polymerize upon delivery. Technical parameters include chemical formulation, viscosity, gel time, and reaction temperature. Requires characterization of polymerization mechanism, polymer structure, degradation pathways, and intermediates. Must meet biocompatibility, sterility, and shelf-life standards. Mechanical integrity evaluated via elastic modulus, compression, swelling, and rebound testing.
Indications for Use
Indicated for adults aged 22+ to reduce risk of symptomatic neuroma development by providing a physical barrier at the peripheral nerve end.
Regulatory Classification
Identification
An in situ polymerizing peripheral nerve cap is a prescription use only device composed of precursor materials that polymerize when delivered to the end of a peripheral nerve to function as a physical barrier to the surrounding in vivo environment to reduce the risk of formation of a symptomatic neuroma.
Special Controls
In combination with the general controls of the FD&C Act, the in situ polymerizing peripheral nerve cap is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR ALLAY NERVE CAP
## REGULATORY INFORMATION
FDA identifies this generic type of device as:
In situ polymerizing peripheral nerve cap. An in situ polymerizing peripheral nerve cap is a prescription use only device composed of precursor materials that polymerize when delivered to the end of a peripheral nerve to function as a physical barrier to the surrounding in vivo environment to reduce the risk of formation of a symptomatic neuroma.
NEW REGULATION NUMBER: 21 CFR 882.5260
CLASSIFICATION: Class II
PRODUCT CODE: SBG
## BACKGROUND
DEVICE NAME: allay Nerve Cap
SUBMISSION NUMBER: DEN230061
DATE DE NOVO RECEIVED: September 19, 2023
Tulavi Therapeutics CONTACT: 160 Knowles Avenue Los Gatos, California 95032
### INDICATIONS FOR USE
The allay Nerve Cap is indicated for use in adults aged 22 years or older as a physical barrier to separate the peripheral nerve end from the surrounding environment to reduce the risk of the development of a symptomatic neuroma.
### LIMITATIONS
The allay Nerve Cap is not designed, sold, or intended for use except as described in the indications for use and is contraindicated for use in:
- . Areas of active surgical site infection.
- Areas of active blood flow. .
- Areas of excessive movement or over a joint. .
{1}------------------------------------------------
- . Patients with a known allergy to poly(ethylene glycol) (PEG) or the color additives FD&C Yellow No. 5 Dye (tartrazine) or FD&C Blue No. 1 Dye (brilliant blue FCF).
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS
{2}------------------------------------------------
## DEVICE DESCRIPTION
The allay Nerve Cap is a sterile, absorbable, in situ formed, hydrogel composed of water and polyethylene glycol (PEG). The hydrogel forms in seconds after delivery of the precursor solutions around a nerve seated in a temporary silicone Cap Form (Figure 1). The hydrogel provides a transparent, compliant nerve cover that conforms to and provides non-constricting encasement of the nerve. The Cap Form is removed and discarded after the implantation procedure. The hydrogel nerve cap is absorbed within 8 months.
The allay Nerve Cap system is provided in a plastic tray sealed in a sterile, peelable outer pouch. The product is available in two sizes of a Small Nerve Set, for nerves less than 4 mm in diameter, and a Large Nerves greater than 4 mm in diameter and less than 7 mm in diameter. The allay Nerve Cap system includes a Powder Vial/Vial Adapter. Diluent Solution. Acceleration Solution, Dual Applicator and Adapter, Delivery Tip with Blunt Needle, and the Cap Forms [Small Nerve Set (1, 2, 3, and 4 mm) and Large Nerve Set (5, 6, and 7 mm)].
Image /page/2/Picture/3 description: The image shows a syringe with a needle and several small, white, origami-like objects arranged in a row. The syringe is filled with a blue liquid and has a clear barrel. The needle is attached to the syringe and is pointing towards the left side of the image. The origami-like objects are arranged in a row in front of the syringe. They are all white and have a similar shape. The background is a plain white surface.
Figure 1: Prepared hydrogel precursors ready for delivery system (applicator) with available cap form sizes.
The delivery mechanism of the starting materials to the site of application. the polymerization. mechanism and polymer structure, the intermediate and side produced, and degradation pathway and degradants were described in the device description of the allay Nerve Cap. The characterization of the polymerization reaction to form PEG-based hydrogels was supported by the published scientific literature. 1-23,45,6 Additionally, the material safety data sheets (MSDS), certificates of analysis, and incoming inspection results were provided for the device raw and precursor materials.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
Reid, et al. 2015. "PEG Hydrogel Degradation and the Role of the Surrounding Tissue Environment." J Tissue Eng Regen Med, 9(3), 315-318.
<sup>2</sup> Fruitier-Polloth (2005). "Safety assessment of polyethylene glycols (PEGs) and their derivatives as used in cosmetic products." Toxicology 214: 1-38.
<sup>3</sup> Yamaoka, et al. (1994). "Distribution and tissue uptake of poly (ethylene glycol) with different molecular weights after intravenous administration in mice." J. Pharm. Sc. 83: 601.
<sup>4</sup> Sharda, N. et al. (2021) "Pharmacokinetics of 40 kDa polyethylene glycol (PEG) in mice, rats, cynomolgus monkeys and predicted pharmacokinetics in humans." European Journal of Pharmaceutical Sciences. 165, 105928.
{3}------------------------------------------------
### BIOCOMPATIBILITY/MATERIALS
The allay Nerve Cap is a long-term contact implant (> 30 davs) in contact with tissue/bone. As such, its biocompatibility evaluation was conducted in accordance with ISO 10993-1:2018. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," and the FDA guidance, "Use of International Standard ISO-10993. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"." For a long-term contact implant in contact with tissue/bone, the allay Nerve Cap was evaluated for the following: physical and chemical information, cytotoxicity, sensitization, irritation or intracutaneous reactivity, material-mediated pyrogenicity, acute systemic toxicity, subacute/subchronic toxicity, chronic toxicity, implantation, hemocompatibility, genotoxicity, carcinogenicity, and neurotoxicity. As part of this biocompatibility evaluation, the materials of construction (direct, indirect) including the hydrogel and degradation products, the Applicator and the Cap Forms, the packaging materials including the tray, and any additives that might arise from the manufacturing process were assessed (see Table 1).
| Test | Test Method | Results |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| allay Nerve Cap | | |
| Cytotoxicity | ISO 10993-5:2009, ISO Elution Method | Non-cytotoxic |
| Sensitization | ISO 10993-10:2010, Maximization Test<br>in Guinea Pigs | Non-sensitizing |
| Intracutaneous<br>Irritation | ISO 10993-10:2010, Intracutaneous<br>Reactivity in Rabbits | Non-irritant |
| Acute Systemic<br>Toxicity | ISO 10993-11:2017, Acute Systemic<br>Toxicity in Mice | Non-toxic |
| Implantation | ISO 10993-6:2016, Local Subcutaneous<br>Implantation, 2 Weeks, 13 Weeks, 26<br>Weeks | Non-irritant |
| In vivo Animal<br>Study<br>Efficacy - Local | Hydrogel Formed in situ Around<br>Transected Nerve in Cap Form, 4 Weeks,<br>12 Weeks | Non-irritant<br>Non-compressive<br>No migration or<br>adverse tissue<br>responses or effects<br>observed |
| Implantation at<br>Clinically Relevant<br>Site | Hydrogel Formed in situ Around Sciatic<br>Nerve, 6 Months, 8 Months | |
| Neurotoxicity | ISO 10993-1 and ASTM F2901-19, 4<br>Weeks, 12 Weeks, 6 Months, 8 Months | Non-irritant<br>No adverse<br>responses were<br>observed |
| Repeated Exposure<br>Systemic Toxicity | Chemical Characterization and<br>Toxicological Risk Assessment (TRA) | Non-toxic |
| Test | Test Method | Results |
| (Subacute,<br>Subchronic,<br>Chronic) | | |
| Hemolysis<br>(Indirect Contact) | ISO 10993-4:2017 and ASTM F756 | Non-hemolytic |
| Pyrogenicity | ISO 10993-11:2017 and USP 34 <151>,<br>USP Rabbit Pyrogen Study, Material<br>Mediated | Non-pyrogenic |
| Genotoxicity | ISO 10993-3:2014, Bacterial Reverse<br>Mutation Test (Ames Assay), in situ<br>Formed<br>ISO 10993-3:2014, Bacterial Reverse<br>Mutation Test (Ames Assay), in Cap<br>Form<br>ISO 10993-3:2014, Bacterial Reverse<br>Mutation Test (Ames Assay), Fully<br>Degraded<br>ISO 10993-3:2014, Mouse Lymphoma<br>Assay, Dose Finding, in situ Formed<br>ISO 10993-3:2014, Mouse Lymphoma<br>Assay, Dose Finding, in Cap Form<br>ISO 10993-3:2014, Mouse Lymphoma<br>Assay, Dose Finding, Fully Degraded | Non-genotoxic |
| Reproductive<br>Toxicity | Chemical Characterization and TRA | Non-toxic |
| Carcinogenicity | Chemical Characterization and TRA | Non-carcinogenic |
| Sterile Cap Forms | | |
| Cytotoxicity | ISO 10993-5:2009, ISO Elution Method | Non-cytotoxic |
| Intracutaneous<br>Irritation | ISO 10993-10:2010, Intracutaneous<br>Reactivity in Rabbits | Non-irritant |
| Acute Systemic<br>Toxicity | ISO 10993-11:2017, Acute Systemic<br>Toxicity in Mice | Non-toxic |
| Pyrogenicity | ISO 10993-11:2017 and USP 34 <151>,<br>USP Rabbit Pyrogen Study, Material<br>Mediated | Non-pyrogenic |
| Sterile Applicator | | |
| Cytotoxicity | ISO-10993-5, ISO Elution Method | Non-cytotoxic |
| Sterile Tray Packaging | | |
| Cytotoxicity | ISO-10993-5, ISO Elution Method | Non-cytotoxic |
| Infrared<br>Spectroscopy | Infrared Analysis per ASTM F2475-20<br>and ISO 10993-18:2020 | No adverse effects<br>observed |
| Nonvolatile Residue | USP Physiochemical Tests for Plastic<br>(Aqueous), Non Volatile Residue (USP | Residue ≤ 15 mg |
| Test | Test Method | Results |
| | 43, NF 38, General Chapter <661>) and<br>ASTM F2475-20, Appendix XI | |
| Particulates | Particulate Analysis, Light Obscuration<br>Method per USP 43, NF 38, General<br>Chapter <788> | No adverse<br>particulates observed |
Table 1: Biocompatibility Evaluations for the allay Nerve Cap System
<sup>5</sup> Webster, et al. (2007). "PEGylated Proteins: Evaluation of their safety in the absence of Definitive Metabolism Studies." Drug Metab Disposition, 35(1), 9-16.
<sup>6</sup> Longley, et al. (2013). "Biodistribution and excretion of radiolabeled 40 kDa polyethylene glycol following intravenous administration in mice." J Pharm Sc 10,2(7), p. 2362-2370.
{4}------------------------------------------------
{5}------------------------------------------------
Chemical Characterization and Toxicological Risk Assessment (TRA):
In order to assess the risks associated with the allay Nerve Cap resulting in an adverse tissue reaction or systemic toxicity effects, chemical characterization and subsequent. TRA of the hvdrogel under worst-case clinical conditions was performed, including identification and quantification of all detected entities extracted from the in situ formed hydrogel, the hydrogel after formation in the Cap Form, as well as the partially and fully degraded hydrogel. The characterization of the entire final finished product ensures that any potential residual chemicals from the manufacturing, packaging, or sterilization processes that could leach from the device during clinical use are not present at levels capable of causing adverse biological responses in patients implanted with the device. The TRA concluded that the extractables from the test article, the allay Nerve Cap, are unlikely to pose a toxicological safety concern. Based on the safety of the allay Nerve Cap as assessed by chemical characterization and TRA, the systemic toxicity (subacute, subchronic, and chronic), reproductive toxicity and carcinogenicity endpoints were evaluated using this testing approach in lieu of the biological testing recommended in ISO 10993-1, annex A.
# SHELF LIFE/STERILITY
The allay Nerve Cap and its packaging has been validated for a shelf life of 2 years supported by performance testing using both accelerated aged and real-time aged test samples that were subjected to preconditioning and packaging distribution conditions. The results showed the final product packaging protects the product and maintains sterility under worst-case shipping, handling, and storage conditions.
The allay Nerve Cap is provided sterile using electron beam (E-beam) sterilization and is intended for single use only. The final finished product is sterilized to a sterility assurance level (SAL) of 10-6 using E-beam radiation. E-beam sterilization achieves a minimum internal dose of 25 kGy using the VDmax25 method for the allay Nerve Cap.
Bacterial endotoxin testing using the limulus amebocyte lysate (LAL) kinetic chromogenic test method was evaluated for the allay Nerve Cap to meet the endotoxin limit of < 20 endotoxin units (EU)/device. Bacterial endotoxin testing will be performed on each lot of the allay Nerve Cap as part of the manufacturing process.
# PERFORMANCE TESTING - BENCH
Non-clinical bench testing was performed to demonstrate that the device can function as
{6}------------------------------------------------
intended under clinical conditions of use and mitigate the risks to health (e.g., adverse tissue reaction, tissue injury, use error). The non-clinical bench testing performed are presented in Table 2. All tests met pre-defined acceptance criteria.
| Test | Test Purpose and Description |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Color Additive<br>Exposure<br>Contact<br>Category | The color additive is provided in the allay Nerve Cap to permit intraoperative visualization of the hydrogel relative to the tissue and in vivo background. The dissolution of color additive from the allay Nerve Cap within 72 hours was confirmed through in vitro dissolution testing on hydrogel cylinders and in situ formed hydrogel, which leaves behind a colorless, transparent hydrogel. The extracts show no change in color or turbidity and there were no visible particulates. |
| Determination<br>of pH and<br>Osmolarity<br>Hydrogel<br>Deliverability | Performance benchtop evaluation of hydrogel injectability was conducted to demonstrate that in situ crosslinking of the allay Nerve Cap is reproducible and results in homogenous hydrogel delivered across the full range of Cap Form sizes. Study design included an assessment of hydrogel deliverability under worst-case conditions including delivery at different angles and locations of injection and after the maximal introduction of bubbles into the precursor solutions prior to delivery. Hydrogel delivery was successful across all worst-case conditions in all Cap Form sizes. In addition to the worst-case deliverability assessments, performance benchtop evaluations of the range of suitable nerve sizes for each allay Nerve Cap was performed. The study was designed to ensure that the available range of Cap Form sizes are suitable for delivery of the hydrogel around each respective nerve size. The hydrogel was successfully delivered around the range of nerve sizes per the Instructions for Use and the Cap Form was successfully removed without breaking or damaging the in situ formed allay Nerve Cap. |
| Dimensional<br>Evaluation | Dimensional evaluation was conducted to assess the hydrogel physical dimensions and volume including confirming nerve encapsulation within the hydrogel for each nerve size per the Instructions for Use, measurement of gel dimensions (length, width, volume) for each Cap Form size, and assessment of the appropriate nerve sizes for each Cap Form. |
| Dimensional<br>Stability | To achieve the intended use of the allay Nerve Cap, the hydrogel must maintain integrity as a protective layer around the end of the nerve without significant loss of mechanical strength for at least three months. In vitro dimensional stability testing was conducted to demonstrate the integrity and persistence of the hydrogel under real-time physiologic conditions (37 °C, phosphate buffered saline (PBS)) |
| Test | Test Purpose and Description |
| | pH 7.4) for at least four months (16 weeks), including the ability of<br>the hydrogel to maintain a protective layer around the nerve and<br>maintain lumen patency past the three-month time frame during<br>which neuroma formation could occur after nerve transection. As part<br>of this evaluation, maintenance of hydrogel integrity without cracks<br>or collapse, mechanical strength (compressive modulus), and lumen<br>patency with no substantial dimensional loss were demonstrated. In<br>addition, minimal swelling of the hydrogel was demonstrated.<br>The dimensional stability of the allay Nerve Cap, as assessed by<br>visual observation and dimensional measurements, was also<br>demonstrated at clinically relevant but worst-case physiologic<br>extremes of osmolarity. At an extreme range of 240 mOsm/kg to 320<br>mOsm/kg, there were no significant changes to the hydrogel integrity<br>(no evidence of cracks or collapse), dimensions, or swelling (within<br>product specifications). |
| Dissolution<br>Time | The time between injection of the Diluent Solution into the Powder<br>Vial and the complete dissolution of the powder is defined as the<br>'dissolution time.' This test demonstrated that the dissolution time is<br>maintained under 5 minutes. |
| Gel Time | The time between the start of advancing the contents of the<br>Applicator through the mixer and the onset of gelation is defined as<br>the 'gel time.' The test was performed to demonstrate that the<br>hydrogel can consistently fill the Cap Form with a gel time of less<br>than 10 seconds to ensure there is no significant delay in the<br>procedure. |
| Percent<br>Swelling | The allay Nerve Cap is formed into cylinders (6 mm diameter, 6 mm<br>long). The hydrogels are weighed and equilibrated in a PBS solution<br>(pH 7.4) at 37 °C for 24 hours. After 24 hours, the cylinder is<br>removed from the solution and weighed. This test was performed to<br>characterize the maximal percent swelling of the hydrogel. The<br>acceptance criteria for this test was the swelling must be between 0%<br>and 45% swelling of the crosslinked hydrogel as measured by weight<br>24 hours after immersion in PBS @ 37 °C. |
| Compression<br>and Rebound<br>Testing | Transverse compression testing is performed on the hydrogel to<br>demonstrate the hydrogel can withstand compressive forces greater<br>than 0.25 N/cm and then rebound following removal of the force. A<br>high threshold was established to ensure that the hydrogel rebounds ><br>95% ensuring good crosslinking of the polymer. |
| Compressive<br>Modulus | The compressive properties of the allay Nerve Cap are measured to<br>ensure the hydrogel has sufficient integrity to prevent nerve<br>outgrowth in the gel and sufficient cross-linking density to persist at<br>the site around the nerve for 8 months while remaining soft and<br>compliant on delivery. The acceptance criteria for this test was the<br>device must have a minimum of 30 KPa compressive modulus of the<br>crosslinked hydrogel after 24 hours |
| Test | Test Purpose and Description |
| | in PBS. |
| Ease of<br>Preparation | Ease of preparation is defined by the time it takes to assemble the<br>allay Nerve Cap from the initial step of the hydrogel preparation to<br>the time at which the Applicator is ready for delivery of the hydrogel.<br>A preparation time longer than 10 minutes indicates PEG melting or<br>that the PEG powder was stored in inappropriate conditions. The<br>device must be able to be prepared in less than 10 minutes. |
| Pressure<br>Testing | Benchtop pressure testing was conducted to demonstrate that there<br>was no increase in pressure on an artificial nerve using a pressure<br>transducer as a function of time after formation and equilibration of<br>the allay Nerve Cap under physiologic conditions. The study<br>demonstrated that no pressure was exerted on the artificial nerve at<br>any time during hydrogel equilibration. Therefore, the allay Nerve<br>Cap is non-compressive. |
| Device<br>Migration | Ex vivo benchtop testing was conducted on a range of freshly<br>harvested ovine nerves to confirm that the allay Nerve Cap adheres to<br>a clinically relevant range of nerve sizes (diameters) for up to one<br>week under physiologic conditions to mitigate against device<br>migration both off of the nerve and subsequently from the implant<br>site. As the in vivo animal testing demonstrated no device migration<br>on smaller nerves (1 mm rat sciatic nerve, 2 mm rabbit sciatic nerve<br>for up to 6 months), a range of larger nerves were evaluated, at<br>minimum establishing the ability to deliver the allay Nerve Cap<br>around small to medium (1 to 4 mm nerves, 1 to 4 mm Cap Forms,<br>Small Nerve Set) and up to the maximum 7 mm nerve (7 mm Cap<br>Form, Large Nerve Set). Testing established that it was not possible<br>to remove the allay Nerve Cap after firmly pulling on the hydrogel<br>with forceps in an attempt to remove the hydrogel, consistent with<br>the in vivo animal testing results demonstrating that the hydrogel<br>maintains adherence to the transected nerve stump. Therefore, the<br>hydrogel does not migrate off of the nerve end as established both<br>through in vitro and in vivo animal testing. |
| Wear<br>Resistance<br>Testing | Accelerated wear testing was performed on the allay Nerve Cap to<br>evaluate the ability of the device to resist wear. The allay Nerve Cap,<br>delivered around an artificial nerve and placed in a clinically relevant<br>orientation between two muscle layers (ex vivo chicken muscle),<br>underwent <span style="background-color:#F0F0F0">(b)(4)</span> cm passes to evaluate any wear debris generated<br>from the hydrogel implants under compressive, shear, and dynamic<br>mechanical loads. The device passed the acceptance criteria that there<br>was no mass loss and no visible particulates released from the<br>hydrogel. |
| Mass Loss | An evaluation of the hydrogel mass loss over the course of<br>accelerated degradation of the implant was performed in vitro. The<br>mass of the allay Nerve Cap is characterized by three phases of mass<br>loss. The study established minimal mass loss occurred over the 3- |
| Test | Test Purpose and Description |
| | month period post implantation (2.8%, 6 days). In the first phase,<br>spanning 8 days (comparable to 4 months in vivo), only 6.6% mass<br>loss occurs. In the second phase of mass loss, between 8 and 15 days<br>(between 4 months and ~7.5 months), substantial degradation is<br>occurring, over 39.8% of the mass is lost. In the third and final stage<br>of rapid mass loss, between day 15 and 16, the remaining 44.2% of<br>mass loss occurs. |
| Exaggerated<br>Clinical Use<br>Conditions | In addition to the in vivo assessment of delivery of the in situ<br>forming hydrogel around the rat and rabbit sciatic nerves after<br>transection, an ex vivo evaluation of the hydrogel performance after<br>delivery under exaggerated clinical use conditions was performed.<br>This represents a worst-case clinical scenario as the proposed<br>Instructions for Use direct the surgeon not to use the device in a<br>region of active blood flow. Hydrogel integrity was assessed after<br>delivery directly into a Wrap Form (identical to a Cap Form but with<br>an entrance and an exit) containing either a) fresh harvested rat blood<br>or b) saline (mimicking physiologic fluids and/or irrigation solution).<br>The hydrogel formed successfully in the presence of blood or saline,<br>integrating with the blood and saline to form an intact allay Nerve<br>Cap. No visible particulates were formed. |
| Use Errors | Benchtop testing was performed to evaluate whether use errors<br>relating to hydrogel preparation and delivery that might result in poor<br>quality hydrogel formation could impact device performance. The<br>study was designed to evaluate the impact of each use error on device<br>performance both immediately after hydrogel preparation and after<br>the equivalent of 3 months degradation in vivo (6 to 8 days<br>accelerated degradation). Hydrogels were prepared according to each<br>worst-case use error and compared with hydrogels formed as<br>intended per the proposed Instructions for Use. The use errors,<br>including bubble generation, incomplete dissolution of the powder,<br>exceeding the delivery window, or fluids did not result in<br>significantly earlier degradation of the hydrogel, the hydrogel quality,<br>or affect the durability of the hydrogel as measured through<br>engineering performance testing. |
| Human Factors<br>and Usability | A human factors and usability study was conducted to evaluate the<br>design and usability of the allay Nerve Cap for the intended users,<br>uses and use environment. The study evaluated the device in the<br>sterile final finished form, including the final commercial packaging<br>and tray configuration following the proposed Instructions for Use.<br>The study design was informed by the FDA guidance, "Applying<br>Human Factors and Usability Engineering to Medical Devices," ISO<br>14971:2019, "Application of risk management to medical devices,"<br>ISO 13485:2015, "Medical Devices - Quality management systems –<br>Requirements for regulatory purposes," clause 7.3.3a, IEC 62366-1:<br>2015, "Medical devices - Part 1: Application of usability engineering |
| Test | Test Purpose and Description |
| | to medical devices," and IEC/TR 62366-2:2016, "Medical devices - Part 2: Guidance on the application of usability engineering to medical devices." |
| | A total of 20 participants were included in the human factors and usability study, including fifteen (15) surgeons with relevant surgical experience on nerve repair, including hand surgeons, plastic surgeons, orthopedic surgeons, and podiatrists and five (5) surgical scrub nurses. The surgeons were independently recruited through a market research agency and had no prior knowledge or relationship with the sponsor and were evaluating the product for the first time. The surgeons assembled the allay Nerve Cap and then performed the simulated surgical procedure from removal of the device from the packaging, preparation of the dual-component applicator system, delivery of the hydrogel in situ into the Cap Form around an artificial nerve, and removal of the Cap Form with two different nerve sizes following the Instructions for Use. The scrub nurses were evaluated for their ability to assemble the device in preparation for delivery in the surgical site. The human factors and usability assessment demonstrated that the allay Nerve Cap and proposed Instructions for Use could be followed without any use errors and the use time of < 5 minutes was achieved with all users. All surgeons could perform all critical tasks (100%, n=15/15) related to preparing the device and delivering the hydrogel around different nerve sizes without any failures or use-related errors. |
Table 2: Summary of Non-Clinical Bench Testing for the allay Nerve Cap
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
# PERFORMANCE TESTING - ANIMAL
Two animal studies were performed to evaluate the in vivo safety and effectiveness and biocompatibility of the allay Nerve Cap.
# Good Laboratory Practice (GLP) Rat Sciatic Nerve Transection Study
A GLP rat study was designed to evaluate the safety and performance of the sterile, final finished allay Nerve Cap after delivery in situ in a clinically relevant nerve transection injury model at 4 weeks and 3 months following implantation. Following nerve transection, the allay Nerve Cap was delivered in situ around the cut end of the rat sciatic nerve per the proposed Instructions for Use. The safety and performance of the allay Nerve Cap was compared with the cleared Polyganics Innovation BV NeuroCap Nerve Capping Device (K152684) in the control group. In addition, a negative control. procedure, a transected nerve alone was included in the GLP rat study as a third group. The study captured outcomes both in-life and terminally to evaluate the safety and performance of the device in comparison to the two control groups.
The study demonstrated a benefit of the allay Nerve Cap on histologic, gross, and
{11}------------------------------------------------
behavioral assessments to support the intended use of the product to reduce the risk of neuroma formation (0%) and to help restore normal sensorimotor function. Overall, the hvdrogel preparation and delivery were adequate with a total procedure time under two minutes. The hydrogel provided a transparent cover around the end of the nerve for the duration of the study with no evidence of axonal escape from the nerve stump into the surrounding muscle and scar tissue. There was no evidence of device migration.
The allay Nerve Cap has comparable safety to the Polyganics Innovation BV NeuroCap Nerve Capping Device (K152684) control for tissue response (non-irritant, nonneurotoxic), animal health, body weights, clinical and neurological observations, autotomy scores, and clinical pathology. The allay Nerve Cap also has comparable safety to the negative control (transection only) for these outcomes as well, with the exception that it is a slight irritant relative to the negative control group, consistent with the presence of a minimally degrading biomaterial implant at the site of implantation. No device related adverse effects were observed throughout the study.
## Non-GLP Implantation Study on Intact Rabbit Sciatic Nerve
A chronic study was designed to evaluate the safety and performance of the sterile, final finished allay Nerve Cap after delivery in situ in a clinically relevant intact nerve model in a rabbit at 6 months and 8 months following implantation. The study objective was to evaluate the biocompatibility of the allay Nerve Cap over a chronic time frame during device absorption and clearance. The acceptance criteria are that the hydrogel, including partially or substantially degraded hydrogel, has (a) acceptable biocompatibility and does not cause an adverse inflammatory response to the nerve or the surrounding tissue and (b) the device should have similar biocompatibility to the comparator control group, recognizing that the control article is non-degradable and therefore a degradationmediated response to the material is absent. The acceptance criteria were evaluated through clinical assessments, gross pathology of the implant, histopathologic assessment, and systemic assessment (clinical chemistry, hematology, and urinalysis).
A total of fourteen (14) New Zealand white rabbits received bilateral implants with the allay Nerve Cap (n=14) around the left sciatic nerve and a silicone cuff with slit (n=14) around the right sciatic nerve. In brief, after sciatic nerve exposure, the nerve was placed inside the temporary silicone Wrap Form (identical to a Cap Form but with an entrance and an exit) and the allay Nerve Cap hydrogel was delivered around the nerve to form a compliant hydrogel wrap. The silicone Wrap Form was then removed and discarded. The rabbit sciatic nerve model was selected as an acceptable small animal model for ISO 10993-6 implantation testing that provides a larger diameter (2 to 3 mm) clinically relevant nerve in which to assess the long-term in vivo persistence and degradation behavior of the hydrogel, the local tissue response at the site of implantation of the degrading hydrogel, the absence of neurotoxicity, the absence of compression of the nerve, and an assessment of the systemic toxicity as measured through clinical pathology, hematology, and urinalysis.
All animals were in good health over the course of the study and survived to the
{12}------------------------------------------------
scheduled study terminal time points. The allay Nerve Cap was successfully delivered in one or more layers around the rabbit sciatic nerve in the Wrap Form in all animals (n = 14). The hydrogel formed a transparent circumferential layer around the nerve. No adverse neurologic or behavioral effects such as lameness, dropped hocks, gait abnormalities were observed at any time. All animals gained an appropriate amount of weight over the study and were in good body condition at the time of termination. There were no apparent clinical signs related to the allay Nerve Cap or silicone cuff observed during the study. The animals were able to ambulate normally and there were no signs of pain, autotomy or overt signs of infection. Similarly, no neurologic abnormalities were noted post-implant or any other observations that could be attributed to the device or procedure. The clinical pathology and urinalysis supported the clinical findings of normality at study term. Hematology, serum chemistry and coagulation parameters were all normal and there were no trends indicative of a pathological pattern.
## LABELING
The labeling includes instructions for use for the physician and satisfies the requirements of 21 CFR § 801.109 for prescription devices. The labeling also includes:
- · Detailed description of the device technical parameters and all components.
- Detailed instructions on proper device preparation and implantation. 0
- O A shelf life.
#### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of an in situ polymerizing peripheral nerve cap and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pain<br>• Decrease or increase in nerve<br>sensitivity<br>• Ineffective treatment leading to<br>symptomatic neuroma formation | In vivo performance testing |
| Adverse tissue reaction | Biocompatibility evaluation<br>Polymerization process characterization |
| Tissue injury (thermal/mechanical)<br>resulting from:<br>• Thermal/mechanical effects<br>• Use error<br>• Device migration | In vivo performance testing<br>Polymerization process characterization<br>Human factors/usability testing<br>Non-clinical performance testing<br>Labeling |
| Infection | Sterilization validation<br>Shelf life testing<br>Labeling |
{13}------------------------------------------------
# SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the in situ polymerizing peripheral nerve cap is subject to the following special controls:
- (1) A characterization of the following chemical characteristics of the polymerization process must describe how the in situ application of the precursor materials will result in a consistent final device. All chemically relevant changes to parts (iii)-(vi) below are determined to significantly affect th…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.