DEN220048 · Invictus Medical, Inc. · QWX · Jun 16, 2023 · General Hospital
Device Facts
Record ID
DEN220048
Device Name
Neoasis
Applicant
Invictus Medical, Inc.
Product Code
QWX · General Hospital
Decision Date
Jun 16, 2023
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 880.5405
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Neoasis® device is intended to reduce noise levels inside an infant incubator in the neonatal intensive care unit (NICU). The attenuation performance of the device is for noises in the NICU with frequencies between 250 Hz to 1,000 Hz.
Device Story
Neoasis is an active noise attenuation system for infant incubators. System components include an outside noise sensor, control unit, two speakers, and a residual noise sensor. The outside sensor captures environmental noise; the control unit processes this input via a software algorithm to calculate a deconstructive interference waveform. This waveform is played through speakers inside the incubator to attenuate noise. The residual noise sensor monitors the resulting sound levels inside the incubator. The device also features a voice pass-through for parents or clinicians to communicate with the infant. Used in the NICU, the device is operated by healthcare staff. It aims to improve infant sleep, development, and health by reducing noise exposure, which is known to impact physiological and behavioral states. The device is intended for use only with the GE Healthcare Giraffe OmniBed Incubator.
Clinical Evidence
No clinical testing performed. Evidence relies on literature regarding the developmental and health benefits of noise reduction for NICU infants and bench testing comparing Neoasis to the MiniMuffs (K935160) using an infant mannequin. Bench testing evaluated attenuation across various NICU noise sources (ventilators, patient monitors, syringe pumps) and infant positions. Results showed approximately 7 dB noise reduction, comparable to the predicate. Human Factors testing with 15 nurses confirmed the ability to set up the system, place the infant, and manage device features.
Technological Characteristics
System includes external noise sensor, control unit, two speakers, and residual noise sensor. Materials evaluated for biocompatibility per ISO 10993-1 for transient contact with intact skin. Operates via active noise cancellation (deconstructive interference). Electrical safety and EMC conform to IEC 60601-1 and IEC 60601-1-2. Software is moderate level of concern. Not provided sterile; 6-month shelf life. Designed for use with GE Healthcare Giraffe OmniBed Incubator.
Indications for Use
Indicated for infants in the neonatal intensive care unit (NICU) to reduce incubator noise levels between 250 Hz and 1,000 Hz.
Regulatory Classification
Identification
Active noise attenuation system for infant incubators. A device system which captures the environmental noise and outputs noise cancelling acoustic sound waves to attenuate noise in infant incubators in the healthcare environment. The Neoasis is comprised of an outside noise sensor, a control unit, two speakers, and a residual noise sensor, intended to reduce noise levels inside an infant incubator in the NICU for frequencies between 250 Hz to 1,000 Hz.
Special Controls
In combination with the general controls of the FD&C Act, an active noise attenuation system for infant incubators is subject to the following special controls:
- (1) Non-clinical performance testing under anticipated conditions of use must demonstrate that the device performs as intended, including:
- (i) Verification and validation of critical acoustic parameters, including the maximum output of the device:
- (ii) Verification and validation of the attenuation performance of the device, including:
- Testing with compatible incubator model(s) and dimensions; (A)
- Attenuation performance testing simulating different infant locations and (B) orientations within the incubator; and
- (C) Testing with relevant noise sources and room configurations.
- Software validation, verification, and hazard analysis must be performed. (2)
- Electrical safety and electromagnetic compatibility (EMC) testing must be performed for (3) any electrical components of the device.
- The patient- or user-contacting components of the device must be demonstrated to be (4) biocompatible.
- Labeling for the device must include: (5)
- Instructions for infant placement and the expected attenuation performance of the (1) device:
- (ii) Warnings regarding the risks of exposure to the potential maximum output of the device:
- Methods and instructions for cleaning and disinfection; and (iii)
- (iv) Identification of the incubator(s) that the device is intended to be used with.
Submission Summary (Full Text)
{0}------------------------------------------------
# DE NOVO CLASSIFICATION REQUEST FOR NEOASIS
### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Active noise attenuation system for infant incubators. A device system which captures the environmental noise and outputs noise cancelling acoustic sound waves to attenuate noise in infant incubators in the healthcare environment.
NEW REGULATION NUMBER: 21 CFR 880.5405
CLASSIFICATION: Class II
PRODUCT CODE: OWX
### BACKGROUND
DEVICE NAME: Neoasis
SUBMISSION NUMBER: DEN220048
DATE DE NOVO RECEIVED: July 25, 2022
## SPONSOR INFORMATION:
Invictus Medical. Inc. 16601 Blanco Road, Suite 120 San Antonio. TX 78232
#### INDICATIONS FOR USE
The Neoasis® device is intended to reduce noise levels inside an infant incubator in the neonatal intensive care unit (NICU). The attenuation performance of the device is for noises in the NICU with frequencies between 250 Hz to 1,000 Hz.
## LIMITATIONS
- The sale, distribution, and use of the Neoasis is restricted to prescription use in . accordance with 21 CFR 801.109.
- The Neoasis® is only intended for use in the GE Healthcare Giraffe OmniBed Incubator. .
- All disposable accessories are intended to be used on a single patient only. None of the . items that are used inside the incubator should be used with more than one patient.
- All incubator cables must be secured and away from the infant. .
- . Use only the power supply provided by manufacturer
- The accessories including the speakers should never be connected to any device except as . provided by Invictus Medical.
{1}------------------------------------------------
### DEVICE DESCRIPTION
The Neoasis is comprised of an outside noise sensor, a control unit, two speakers, and a residual noise sensor. All components are connected to the control unit with different cables. The outside noise sensor captures the environmental noise and transmits the information to the control unit. The control unit determines if the noise should be attenuated and contains a software algorithm which calculates the sound wave necessary to cancel out the environmental noise. If cancelation is appropriate, the control unit plays the corresponding sound wave through the speakers inside of the incubator, attenuating the noise via deconstructive interference. The residual noise sensor then captures the resulting noise inside of the incubator and communicates it to the control unit. Additionally, the device contains a voice pass through feature which allows parents or healthcare practitioners to communicate with the infant from the outside using the control unit and speakers.
Image /page/1/Figure/2 description: The image shows a diagram of an incubator shell with various components labeled. The diagram includes an outside noise sensor, a residual noise sensor, speakers, a rod, a mattress, and a control unit. Cables connect the control unit to the speakers and other components, and the incubator shell is depicted as a transparent enclosure.
{2}------------------------------------------------
## SUMMARY OF NONCLINICAL/BENCH STUDIES
### SOFTWARE
The Neoasis device contains software to control and determine if an external noise is appropriate for the device to attenuate based on noise characteristics (duration, wavelength, sound level), calculates the appropriate waveform needed to match the external noise waveform to reduce noise and incorporates safeguards which ensure that an incorrect waveform or an unnecessarily high volume noise is not produced. The device outputs the calculated waveform through the speakers to cancels out the environmental noise. Software verification and analysis was performed, and the results demonstrate that the software functions as intended, and the required specifications are met. The software/firmware was reviewed according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005. The software has a moderate level of concern.
A cybersecurity evaluation was not needed as the device does not connect to external networks or devices.
## ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY
The Neoasis device conforms to the requirements for electrical safety and electromagnetic compatibility of the following standards:
- IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) Medical . electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014. Medical electrical equipment Part 1-2: General requirements . for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60950-1 Information technology equipment Safety Part 1: General . requirements
#### BIOCOMPATIBILITY
The Neoasis device is classified as intact skin contacting with a transient contact duration. Biocompatibility was evaluated in accordance with ISO 10993-1:2018, Biological evaluation of medical devices and FDA Guidance: Use of International Standard ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Invictus Medical provided material information and a description of the manufacturing processes in lieu of biocompatibility testing to support the biological safety of the device. The results of these evaluations support that the biological risks of the device have been adequately mitigated.
{3}------------------------------------------------
### PERFORMANCE TESTING-BENCH
The performance of the Neoasis was compared against a 510k cleared passive noise attenuation device, the MiniMuffs (K935160) in a nonclinical bench testing which simulated use with both devices on an infant mannequin. To assess the noise attenuation performance under simulated clinical use conditions. Invictus Medical used various sound sources that are present in a typical NICU setting and a mannequin with noise sensors (microphones) to evaluate the noise attenuation. The testing also evaluated situations where the infant could move to different positions within the infant incubator by providing attenuation mapping across representative locations inside of the incubator. The table below (Table 1) provides a summary of the test results and is ordered in terms of best attenuation to worst attenuation sequences. Noise attenuation is represented by the number of decibels where a positive number represents a reduction of noise and negative numbers represent an increase/amplification of noise at the indicated frequency bands. Note that the alarm priority denotes the alarm sound from each devices play different sounds based on the alarm level, which is determined by the severity of issues. Testing the various alarm priority levels is representative of the possible alarms that the device could play in a clinical setting.
| Alarm<br>Device | Alarm<br>Brand | Alarm<br>Priority | Octave<br>Band<br>(Hz) | Neoasis<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears and<br>Human Hair<br>Attenuation<br>(dB) |
|--------------------|----------------|-------------------|------------------------|--------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Patient<br>Monitor | Philips | High | 125 | 0.49 | -0.72 | -1.19 |
| | | | 250 | 0.3 | -0.05 | -0.12 |
| | | | 500 | 1.32 | 1.37 | 0.17 |
| | | | 1000 | 11.68 | 3.44 | -1.27 |
| | | | 2000 | 1.23 | 1.8 | 0.2 |
| | | | 4000 | 0.44 | 1.73 | 1.23 |
| | | | 8000 | -0.19 | 0.02 | 0.01 |
| Ventilator | Maquet | Medium | 125 | 0.35 | -1.59 | 0.1 |
| | | | 250 | 3.18 | 3.18 | 1.01 |
| | | | 500 | 9.38 | 7.45 | 2.54 |
| | | | 1000 | 0.14 | 2.35 | 0.6 |
| | | | 2000 | -0.02 | 0.33 | -1.39 |
| Table 1: Performance Testing - Attenuation Comparison | |
|-------------------------------------------------------|--|
|-------------------------------------------------------|--|
De Novo Summary (DEN220048)
Page 4 of 11
{4}------------------------------------------------
| Alarm<br>Device | Alarm<br>Brand | Alarm<br>Priority | Octave<br>Band<br>(Hz) | Neoasis<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears and<br>Human Hair<br>Attenuation<br>(dB) |
|---------------------------------------|--------------------------------|----------------------|------------------------|--------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| | | | 4000 | 0.01 | 0.15 | -0.04 |
| | | | 8000 | -0.02 | 0.02 | 0.01 |
| | | High | 125 | -0.12 | -0.69 | -0.16 |
| | | | 250 | 5.07 | 1.93 | 1.06 |
| Ventilator | Maquet | | 500 | 8.64 | 2.43 | 1.36 |
| | | | 1000 | 0.6 | 2.64 | 0.45 |
| | | | 2000 | 0.16 | 0.76 | -1.64 |
| | | | 4000 | 0.07 | 0.3 | -0.31 |
| | | | 8000 | -0.05 | 0.02 | -0.13 |
| | | High<br>High | 125 | -0.66 | -0.39 | -0.13 |
| | | | 250 | 3.5 | 1.39 | 0.44 |
| Syringe<br>Pump | Medfusion<br>Maquet | | 500 | 7 | 1.14 | 0.58 |
| Ventilator | | | 1000 | 0.35 | 4.36 | -0.05 |
| | | | 2000 | 0.3 | 3.03 | -0.09 |
| | | | 4000 | 0.11 | 0.88 | 0.27 |
| | | | 8000 | -0.21 | 0.09 | -0.15 |
| | Medfusion<br>Maquet<br>Philips | High<br>High<br>High | 125 | -0.76 | -1.74 | -0.96 |
| Syringe<br>Pump | | | 250 | 1.44 | 0.27 | 0.31 |
| Ventilator<br>Patient | | | 500 | 6.52 | 5.03 | 5.39 |
| | | | 1000 | 1.49 | -0.29 | 0.53 |
| | | | 2000 | 0.45 | 0.58 | -0.27 |
| Monitor | | | 4000 | -0.13 | 1.86 | -0.66 |
| | | | 8000 | -0.43 | 0.06 | 0.04 |
| | Medfusion<br>Maquet | Low<br>High | 125 | -0.09 | -0.26 | -0.83 |
| | | | 250 | 2.12 | 0.89 | 0.3 |
| Syringe<br>Pump<br>Ventilator | | | 500 | 5.07 | 0.2 | 0.5 |
| | | | 1000 | -0.18 | 2.42 | 1.06 |
| | | | 2000 | -0.27 | 1.94 | 0.2 |
| | | | 4000 | -0.05 | 0.42 | 0.39 |
| | | | 8000 | -0.24 | 0.04 | -0.11 |
| Male and | N/A | N/A | 125 | -0.05 | -0.06 | 0.35 |
| Alarm<br>Device | Alarm<br>Brand | Alarm<br>Priority | Octave<br>Band<br>(Hz) | Neoasis<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears and<br>Human Hair<br>Attenuation<br>(dB) |
| Female<br>Voices | | | 250 | -0.02 | 1.2 | -0.18 |
| | | | 500 | -0.01 | 4.7 | 0.98 |
| | | | 1000 | 0.13 | 1.66 | 0.03 |
| | | | 2000 | 0 | 0.31 | -1.16 |
| | | | 4000 | 0.02 | 0.09 | 0.02 |
| | | | 8000 | 0.01 | 0.03 | 0.01 |
| Syringe<br>Pump<br>Patient<br>Monitor | Medfusion<br>Philips | High<br>Medium | 125 | -0.31 | -0.55 | -0.56 |
| | | | 250 | 0.75 | 1.04 | 0.07 |
| | | | 500 | 3.31 | 2.69 | 1.64 |
| | | | 1000 | -0.92 | 0.22 | -0.78 |
| | | | 2000 | -0.83 | 0.08 | -1.35 |
| | | | 4000 | 0.37 | 0.09 | -0.86 |
| | | | 8000 | -0.07 | 0.06 | 0.02 |
| Patient<br>Monitor | Philips | Medium | 125 | -0.21 | 0.07 | 0.04 |
| | | | 250 | -0.15 | 0.35 | 0.31 |
| | | | 500 | -0.13 | 2.04 | 1.97 |
| | | | 1000 | 0.13 | -0.12 | 0.65 |
| | | | 2000 | 0.04 | -0.07 | 0.55 |
| | | | 4000 | -0.03 | 0.03 | 0.02 |
| | | | 8000 | -0.12 | 0.02 | 0 |
| Syringe<br>Pump | Medfusion | Low | 125 | 0.12 | -0.99 | -0.63 |
| | | | 250 | -0.1 | -0.05 | -0.04 |
| | | | 500 | 0 | 1.46 | 1.11 |
| | | | 1000 | -0.04 | 0.44 | -0.05 |
| | | | 2000 | -0.01 | -0.01 | -1.6 |
| | | | 4000 | 0 | 0.04 | -0.64 |
| | | | 8000 | 0.02 | 0.04 | 0.03 |
| Syringe<br>Pump | Medfusion | High | 125 | 0.07 | -0.59 | -0.31 |
| | | | 250 | 0.01 | 0.41 | 0.04 |
| | | | 500 | 0.03 | 1.95 | 1.15 |
| | | | 1000 | -0.04 | 0.84 | -1.51 |
| Alarm<br>Device | Alarm<br>Brand | Alarm<br>Priority | Octave<br>Band<br>(Hz) | Neoasis<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears<br>Attenuation<br>(dB) | Earmuffs<br>Adhered Over<br>Mannequin<br>Ears and<br>Human Hair<br>Attenuation<br>(dB) |
| | | | 2000 | -0.3 | -0.3 | -2.03 |
| | | | 4000 | -0.15 | 0.24 | -0.9 |
| | | | 8000 | -0.01 | 0.02 | 0.03 |
De Novo Summary (DEN220048)
Page 5 of 11
{5}------------------------------------------------
De Novo Summary (DEN220048)
Page 6 of 11
{6}------------------------------------------------
Additionally, Invictus Medical moved the noise sensing mannequin into different locations within the infant incubator to account for the potential movement of real infants. The attenuation performance study described above evaluates the changes in performance within different locations of the incubator. This resulted in an attenuation map which demonstrates the device performance at different locations in the incubator. See below for a summary of locations tested.
Figure 1: Attenuation Mapping Results by Octave Band
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains four plots showing attenuation in decibels at different frequencies. The frequencies are 500 Hz, Lower 1,000 Hz, Upper 1,000 Hz, and 2,000 Hz. Each plot shows several data points with corresponding attenuation values, such as 8.3 and 10.6 at 500 Hz, and 0.9 and 0.6 at 2,000 Hz. Each plot also contains labels for "Speaker" and "Residual Noise Sensor (RNS)".
As shown in the results table above, within the 250-1000 hz range, the Neoasis and MiniMuffs device demonstrate similar attenuation performance. Representative positions were also tested in order to accommodate infant movement and account for performance differences based on location using a mannequin. These different positions tested are represented by the different colored circles in the incubator above. The numbers above the open circles represent the sound attenuation measured in decibels for the loudest ear of the mannequin across different locations when using the Neoasis device. The results indicate that the device performs as intended, with similar attenuation levels compared to the MiniMuffs. Based on the above testing results, the attenuation performance of the Neoasis device is similar to that of the MiniMuffs device.
## REPROCESSING AND SHELF LIFE
The Neoasis is not provided sterile. Performance testing with accelerated aging confirmed the Neoasis maintained device performance over the 6 month labeled shelf life. Invictus Medical provided reprocessing and disinfection instructions to the end user which were evaluated per the 2015 FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" under clinical worst-case conditions. Cleaning and disinfection validation was provided and met the recommendations of the guidance above.
{8}------------------------------------------------
# SUMMARY OF CLINICAL INFORMATION
No clinical testing was performed to support the submission. Instead, literature was relied on to support the developmental. sleep. and health benefits associated with noise reduction for infants in the NICU from passive noise attenuation devices, like earmuffs. In addition, the American Academy for Pediatrics (AAP) recommends keeping sound levels below 45 dB and notes that higher noise environments can lead to adverse effects on the growth and neurodevelopment of neonates. Overall, it is well understood that high noise environments can lead to hearing damage, reduced sleep, and can negatively impact various physiological and behavioral states.
Invictus Medical conducted Human Factors testing by recruiting 15 nurses to assess the user's ability to set up the system and place the infant correctly. The Human Factors study was carried out following FDA 2016 guidance "Applying Human Factors and Usability Engineering to Medical Devices". The participants were successful in setting up and using the Neoasis system with infants. Major tasks included attaching/setting up the device, placing the infant correctly inside the incubator. being able to read the current and past noise levels. using the talk to infant feature and handling various errors. All participants were able to complete all of the identified tasks.
# Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
Physician labeling includes the device indications for use, a description of the device, warnings and precautions, the expected benefit of the device, and instructions for the safe and effective use of the device. The labeling instructs users to clean or, for components placed inside the incubator, clean and disinfect all parts of the device prior to use and at least weekly during use. In addition. reusable components of the device are also cleaned and disinfected between patients. The labeling satisfies the requirements of 21 CFR 801.109 Prescription devices. Per the special controls for this generic type of device, labeling includes a summary of the attenuation performance.
## RISKS TO HEALTH
Table 2 identifies the risks to health that may be associated with use of an active noise attenuation system for infant incubators, and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hearing loss from high device<br>output or ineffective device<br>attenuation | Non-clinical performance testing<br>Software validation, verification, and hazard analysis<br>Electrical safety and electromagnetic compatibility testing |
| Infection | Labeling |
| Adverse tissue reaction | Biocompatibility evaluation |
| Table 2 Identified Risks to Health and Mitigation Measures | | | | | | | | | |
|------------------------------------------------------------|--|--|--|--|--|--|--|--|--|
|------------------------------------------------------------|--|--|--|--|--|--|--|--|--|
De Novo Summary (DEN220048)
{9}------------------------------------------------
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, an active noise attenuation system for infant incubators is subject to the following special controls:
- (1) Non-clinical performance testing under anticipated conditions of use must demonstrate that the device performs as intended, including:
- (i) Verification and validation of critical acoustic parameters, including the maximum output of the device:
- (ii) Verification and validation of the attenuation performance of the device, including:
- Testing with compatible incubator model(s) and dimensions; (A)
- Attenuation performance testing simulating different infant locations and (B) orientations within the incubator; and
- (C) Testing with relevant noise sources and room configurations.
- Software validation, verification, and hazard analysis must be performed. (2)
- Electrical safety and electromagnetic compatibility (EMC) testing must be performed for (3) any electrical components of the device.
- The patient- or user-contacting components of the device must be demonstrated to be (4) biocompatible.
- Labeling for the device must include: (5)
- Instructions for infant placement and the expected attenuation performance of the (1) device:
- (ii) Warnings regarding the risks of exposure to the potential maximum output of the device:
- Methods and instructions for cleaning and disinfection; and (iii)
- (iv) Identification of the incubator(s) that the device is intended to be used with.
## BENEFIT-RISK DETERMINATION
Nonclinical laboratory studies as well as literature were used to evaluate the safety and effectiveness of the Neoasis.
## Summary of Benefits
The device reduces noise at the 250-1000 Hertz range and the benefit of noise reduction for infants is well understood. Invictus Medical cites literature which concludes that there are benefits to infants from passive noise attenuation devices like earmuffs. These benefits relate to development, healing, sleep and overall health of infants in high noise environments such as the NICU. Furthermore. NIOSH guidance recommends 85 dBA for <8 hours and 100 dBA for <15 minutes to avoid hearing damage for adults. The American Academy of Pediatrics recommendation of noise level is <45db. Invictus Medical performed testing with some expected De Novo Summary (DEN220048) Page 10 of 11
{10}------------------------------------------------
noise sources and compared it to earmuffs. Invictus Medical has demonstrated through performance testing that the reduction in noise is comparable to passive noise canceling devices such as the MiniMuffs. Although the attenuation performance varies depending on the acoustic waveform and the environmental volume, performance testing shows approximately a 7 decibels noise reduction, which is similar to MiniMuffs.
#### Summary of Risks
The device is overall a low risk device. There are two probable safety risks which were identified and subsequently mitigated by Invictus Medical. Firstly, in order to attenuate noise, the device has a maximum output of 98 decibels. During use, this output would be activated when the environmental noise is high which is needed to reduce the noise levels. In addition, there are software safeguards in place to prevent the device from malfunctioning and producing noise inappropriately. Secondly, based on the way the acoustic waveforms are matched, FDA was concerned the device use may result in noise amplification. If this occurred, it would lead to the infant experiencing a higher noise than the environmental noise. However, due to the geometry and movement of the wayeforms inside the incubator, it would be improbable that a noise waveforms match would lead to amplification.
#### Benefit/Risk Conclusion
The Neoasis performance testing demonstrates a meaningful noise attenuation, similar in impact to the cleared MiniMuffs device. Considering that device risks are mitigated through software and device performance controls, there is an overall favorable benefit-risk profile that supports granting this De Novo request.
#### CONCLUSION
The De Novo request for the Neoasis is granted and the device is classified as follows:
Product Code: OWX Device Type: Active noise attenuation system for infant incubators Regulation Number: 21 CFR 880.5405 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.