DEN220041 · Tristel Solutions Limited · QWS · Jun 2, 2023 · General Hospital
Device Facts
Record ID
DEN220041
Device Name
Tristel Duo ULT
Applicant
Tristel Solutions Limited
Product Code
QWS · General Hospital
Decision Date
Jun 2, 2023
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 880.6886
Device Class
Class 2
Indications for Use
Tristel Duo ULT is a high level disinfect foam for reprocessing ultrasound probes. It can be used to high level disinfect endocavity transvaginal and transrectal probes, and skin surface transducers that may contact non-intact skin during use. Tristel Duo ULT is a high level disinfectant when used in accordance with the directions for use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC). Tristel Duo ULT must be applied on the surface of an ultrasound probe using Duo Wipes. Duo Wipes are intended to apply Tristel Duo ULT foam on the surface of ultrasound probes and to remove residue of the foam after high level disinfection. Each dose of Tristel Duo ULT foam and each Duo Wipe are single use. The semi-critical ultrasound probes reprocessed by Tristel Duo ULT must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions. Tristel Duo ULT foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC. Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel Duo ULT may be verified using Duo Test Strips.
Device Story
Tristel Duo ULT is a point-of-care high-level disinfectant foam for ultrasound probes. The system consists of two precursor solutions (Base and Activator) housed in a dual-chamber dispenser; depressing the pump mixes the solutions to generate chlorine dioxide in situ. A mesher in the pump head converts the liquid into a ready-to-use foam. Healthcare personnel apply the foam to probes using single-use, non-woven Duo Wipes. The process requires a 2-minute contact time at room temperature. The device includes semi-quantitative Duo Test Strips to verify the Minimum Recommended Concentration (MRC) of ~280ppm. Used in procedure rooms, exam rooms, or central processing units, the device eliminates the need for transport to automated reprocessing equipment, potentially increasing throughput. Clinical benefit includes rapid, effective disinfection of semi-critical devices, reducing infection risk when used according to validated protocols.
Clinical Evidence
Non-clinical performance testing required, including storage and transport stability, potency testing (sporicidal, mycobactericidal, fungicidal, bactericidal, virucidal), simulated use testing (demonstrating 10^6 kill of resistant mycobacteria), in-use testing on clinically-relevant microorganisms, and material compatibility. Biocompatibility evaluation and human factors testing required.
Technological Characteristics
Foam-based chemical high-level disinfectant. Includes foam delivery bottle and applicator wipes. Features chemical indicator (Duo Test Strips) for MRC verification. Requires specific contact time (2 minutes) and temperature (room temperature).
Indications for Use
Indicated for high-level disinfection of endocavity (transvaginal/transrectal) ultrasound probes and skin surface transducers that may contact non-intact skin.
Regulatory Classification
Identification
Tristel Duo ULT is a high level disinfectant foam intended to disinfect cleaned, reusable, non-lumened ultrasound probes, including endocavity transvaginal and transrectal probes, and skin surface transducers that may contact non-intact skin. It generates chlorine dioxide foam from two precursor solutions (Base and Activator) dispensed via a pump mechanism. It is intended for use by qualified healthcare personnel at the point of care, requiring a 2-minute contact time at room temperature at or above the Minimum Recommended Concentration (MRC).
Special Controls
In combination with the general controls of the FD&C Act, the foam or gel chemical sterilant/high level disinfectant is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR TRISTEL DUO ULT HIGH LEVEL DISINFECTANT
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Foam or gel chemical sterilant/high level disinfectant. A foam or gel chemical sterilant/high level disinfectant is a germicide in the form of a foam or gel that is intended for use as the terminal step in high level disinfection of medical devices prior to patient use.
NEW REGULATION NUMBER: 21 CFR 880.6886
CLASSIFICATION: Class II
PRODUCT CODE: OWS
#### BACKGROUND
DEVICE NAME: TRISTEL DUO ULT
SUBMISSION NUMBER: DEN220041
DATE DE NOVO RECEIVED: July 13, 2022
#### SPONSOR INFORMATION:
Hyman, Phelps, & McNamara, P.C. 700 13th St NW. Suite 1200. Washington, DC 20005 USA On behalf of: Tristel Solutions Limited, Lynx Business Park, Fordham Road, Snailwell, CB8 7NY, United Kingdom
#### INDICATIONS FOR USE
The Tristel Duo ULT high level disinfectant is indicated as follows:
Tristel Duo ULT is a high level disinfect foam for reprocessing ultrasound probes. It can be used to high level disinfect endocavity transvaginal and transrectal probes, and skin surface transducers that may contact non-intact skin during use.
Tristel Duo ULT is a high level disinfectant when used in accordance with the directions for use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).
Tristel Duo ULT must be applied on the surface of an ultrasound probe using Duo Wipes.
{1}------------------------------------------------
Duo Wipes are intended to apply Tristel Duo ULT foam on the surface of ultrasound probes and to remove residue of the foam after high level disinfection.
Each dose of Tristel Duo ULT foam and each Duo Wipe are single use.
The semi-critical ultrasound probes reprocessed by Tristel Duo ULT must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions.
Tristel Duo ULT foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.
MRC: ~90% v/v (~280ppm) at point of use. MRC of Tristel Duo ULT may be verified using Duo Test Strips.
## LIMITATIONS
Tristel Duo ULT is intended to be marketed for prescription use.
## PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
Tristel Duo ULT is a high level disinfectant foam intended to disinfect cleaned, reusable, nonlumened ultrasound probes. Tristel Duo ULT is intended to be used by qualified healthcare personnel. Tristel Duo ULT does not sterilize.
Tristel Duo ULT generates chlorine dioxide foam from two "precursor" solutions (Base Solution and Activator Solution), which are packaged separately within two individual containers and secured together within a single dispenser apparatus known as the "pump" shown in Figure 2. When depressed, the pump dispenses these solutions at equal volumes.
Image /page/1/Figure/11 description: The image shows two bottles connected at the top. The bottle on the left is labeled "Duo Activator", and the bottle on the right is labeled "Duo Base". Both bottles appear to be empty and have a pump mechanism at the top.
Figure 2 - Schematic Representation of Duo Bottle
{2}------------------------------------------------
The components of the Base and Activator solutions chemically react. This reaction generates chlorine dioxide in liquid solution, which is the active germicidal ingredient. A mesher present in the applicator pump causes the formulation to be dispensed as a ready-to-use foam.
The active ingredient in Tristel Duo ULT high level disinfectant is chlorine dioxide, which is generated in situ from the Activator and Base solutions. Tristel Duo ULT also contains other (D) CI (b) 49 ingredients, such as sodium chlorite, citric acid monohydrate. (0) 49)
### (b)(4) (b)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4)(4
The Duo bottle configuration consists of the two bottles (containers), a single pump head with two thread neck closures and a cap. Each Duo bottle contains 125 milliliters (mL)/4.40 fl oz of Base Solution and 125 mL/4.40 fl oz of Activator Solution. The left and right containers are made to the same specification, with the difference in the orientation of the neck thread. The bottle is handheld. Shown in Figures 3 and 4.
Image /page/2/Picture/4 description: The image shows a bottle of "TriGel Duo ULT" hand sanitizer. The bottle is white with a pink and purple label. The label has the words "TriGel Duo ULT" in large, bold letters. Below the name is the phrase "Proper Patient Decontamination System for today's clinical environment."
Figure 3 - Tristel Duo Handheld Dispenser
Figure 4 - Tristel Duo in Use
An 80% V/V dilution of Duo was determined as the Minimum Effective Concentration (MEC) for Duo.
Duo Wipes are dry, non-woven, low-linting wipes. They are specifically designed to apply the Tristel Duo ULT foam onto the surface of a medical device. A user lays the Duo Wipe in one hand and applies 4 doses of the Tristel Duo ULT foam before closing their hand around the wipe and waiting for 10 seconds. The wipe can then be used to spread the foam in a massaging motion from one end of the device to the other 4 times. After the device is wiped, it is left undisturbed for 2 minutes.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a container of Duo Wipes. The container is white and has the words "Duo Wipes" printed on it in a bold, sans-serif font. Below the words, there is a colorful graphic of swirling lines in shades of blue, pink, and yellow. The container appears to be made of plastic and has a label on the side with additional information.
Image /page/3/Picture/1 description: The image shows a package of Duo Wipes. The package has a white background with colorful streaks of blue, pink, yellow, and green. The words "Duo Wipes" are printed in a white oval in the center of the package.
Figure 7 - Duo Wipes in Tub and Flow Wrap Commercial Packaging
The Duo Test Strips are semi-quantitative chemical indicators used to determine whether the concentration of the Tristel Duo ULT high level disinfectant foam is above or below the minimum recommended concentration (MRC) (~90% v/v /~280 ppm).
Tristel Duo ULT foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.
The Duo Test Strips, when used in accordance with directions for use detect the MRC and above giving a color indication of a "PASS" and if the concentration is below the MRC a color indication of a "FAIL".
Image /page/3/Picture/6 description: The image shows a container of Duo Test Strips next to three individual test strips. The container is white and has the words "Duo Test Strips" printed on it. The test strips are thin and white with a small square at the end. The background is white.
Figure 8 - The Duo Test Strips
Tristel Duo ULT, according to the directions of use, has passed the AOAC Sporicidal Activity Test in 12 hours and additional potency tests against mycobacteria, fungi, bacteria and viruses in 2 minutes at 20°C at a worst case concentration below the MRC at 80% v/v of the nominal concentration of Duo. The test results qualify Tristel Duo ULT as a high level disinfectant.
Tristel Duo ULT has been tested on a large number of materials and found to be compatible after a total of 120 hours immersion. Tristel Duo ULT is labeled for use on ultrasound probes only and is not intended for disinfection of other semi-critical medical devices.
Tristel Duo ULT is intended for use by qualified healthcare professionals (HCP) and reprocessing technicians. The HCP's intended use environments include a procedure room and
{4}------------------------------------------------
exam room. The reprocessing technicians' intended use environment includes a central processing unit. Each environment should include a designation between contaminated and clean work surfaces, a sink, a clock or another device to track time, and personal protective equipment such as gloves, masks with eye protection, and gowns.
### SUMMARY OF BENCH STUDIES
#### REPROCESSING, STERILITY AND SHELF-LIFE
The Tristel Duo ULT high level disinfectant foam is not intended for either cleaning or sterilization, it is intended to disinfect cleaned, reusable, non-lumened ultrasound probes. Performance related to the high level disinfection was reviewed as a part of the performance testing. The device includes instructions for reprocessing ultrasound probes. Specifically, the sponsor has included reprocessing instructions that they have validated according to the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".
The storage stability study is provided to support: (i) a two-year shelf life of the product from the date of manufacture of the unopened product; and (ii) a use period of six months of the opened container from the date of opening under the prescribed storage conditions of room temperature (15-25°C).
The transport stability study is provided to support the resilience to transport conditions (such as temperature, pressure, and humidity), environmental factors (such as freeze and thaw), and mechanical impact (such as drop effect on package integrity) and the capability of maintaining its composition (physical and chemical integrity).
## BIOCOMPATIBILITY
The Tristel Duo ULT is categorized as a surface disinfectant, with contact to mucosal membranes, A-limited (<24hour), in accordance with Guidance for Industry and Food and Drug Administration Staff -Use of International Standard ISO10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
Tristel Duo ULT (Duo) is based on Tristel's proprietary chlorine dioxide chemistry. The sponsor provided the following studies to assess the potential toxicity of Duo to a user (100% solution), Duo with Duo Wipes to the user, and Duo to a patient (i.e., residue levels). The results of these tests are deemed acceptable.
- Duo Working Solution (Genotoxic tests, Skin Irritation tests, Acute Eye Irritation . tests. Duo Skin Sensitization tests. Acute Oral tests)
- . Duo with Duo Wipes (Cytotoxicity, Dermal Irritation, Skin Sensitization)
- . Duo Residues (Vaginal Mucosa Irritation Test, Rectal Mucosa Irritation Test. Dermal Irritation Test, Residues Cytotoxicity)
{5}------------------------------------------------
The sponsor provided the following studies on the exhaustive extractable and leachable to evaluate the potential adverse health risk from patient exposure to residues remaining on ultrasound probes:
- . Trace metals.
- Volatile Organic Compounds (VOC), 0
- Semi-Volatile Organic Compounds (SVOC), and .
- . Non-Volatile Organic Compounds (NVOC).
The sponsor also provided the exposure and risk assessment for chlorine dioxide's inhalation toxicity, which met the acceptance criteria in accordance with the U.S. Occupational Safety and Health Administration (OSHA) legal airborne permissible exposure limit (PEL) for chlorine dioxide of 0.1 ppm averaged over eight hours and California Division of Occupational Safety and Health PEL of averaged 0.3 ppm over a 15-minute period.
#### SOFTWARE
The Tristel Duo ULT does not require software.
#### ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
The Tristel Duo ULT does not require equipment or installation.
#### Human Factors
A simulated-use human factors (HF) study was provided in accordance with the guidance "Applying Human Factors and Usability Engineering to Medical Devices." It included participants representative of the intended healthcare professionals and reprocessing technicians. During the session, participants were presented with a simulated-use scenario where they were tasked with using the product to simulate reprocessing a reusable medical device (either an ophthalmic tonometer, a flexible nasoendoscope, or a transvaginal ultrasound probe). Following the session, participants answered questions and provided feedback regarding use-related issues and the product itself. The study included 27 assessments which evaluated the critical tasks for using the product and found that users understood.
The provided human factors validation testing results, analyses and conclusions established that Tristel Duo ULT is safe and effective for the intended users, its intended uses, and use environments.
#### PERFORMANCE TESTING - BENCH
The sponsor conducted the following performance tests to support that the device can achieve its intended use:
- Potency tests were provided, including sporicidal efficacy, mycobactericidal . efficacy, fungicidal efficacy, bactericidal efficacy, virucidal efficacy, to qualify
{6}------------------------------------------------
the Tristel Duo ULT as a high level disinfectant, which has been confirmed by an independent laboratory.
- Simulated-use tests were provided. with a total of six skin surface ultrasound . probes and a total of seven endocavity ultrasound probes were tested in triplicate. It demonstrated the penetrating capability of this high-level disinfectant along with other factors that may prevent or limit contact and effectiveness of the germicide.
- In-use tests were conducted under clinical condition with wild type bacteria and . fungi from patient contaminated BK Medical endocavity ultrasound probes. It confirmed the results of simulated-use testing and high level disinfectant performance in clinical use conditions.
- . The material compatibility testing with ultrasound device manufacturers was conducted and supportive of compatibility on various ultrasound probes and equipment. Manufacturers of ultrasound svstems have evaluated compatibility of Tristel Duo ULT with their ultrasound devices assessing the impact Duo high level disinfectant has on the materials and functionality.
- A chemical indicator validation test was provided, with adequate performance . testing and shelf life / stability testing, to support that the Duo test strip is appropriately designed to respond with a characteristic chemical reaction to the concentration of the germicide active ingredients of Tristel Duo ULT.
The device has been appropriately evaluated for performance on the bench.
## LABELING
The labeling consists of Instructions for Use in user guide and packaging labels for all three components of Tristel Duo ULT. Duo Wipe, and Duo Test Strip. The instructions for use include the indications for use; a description of the device, contraindications, warnings, precautions; information on use-life: instructions for the safe use of the device (Duo ULT. Duo wipe, and Duo test strip), storage condition and stability information, user training, emergency and/or additional information, disposal; germicide classification scheme for labeling purposes; general information. on selection and use of germicides for medical device reprocessing; material and device compatibility: mode of action of germicidal activity; precleaning agent compatibility; toxicology and adverse reactions.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a high level disinfectant foam:
| Identified Risks to Health and Mitigation Measures | |
|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Identified Risks to Health | Mitigation Measures |
| Patient cross-contamination due to high<br>level disinfectant (HLD) foam lacking<br>adequate potency | Potency efficacy testing<br>Simulated-use testing<br>In-use testing<br>Shelf life / Stability testing |
#### Identified Risks to Health and Mitigation Measures
{7}------------------------------------------------
| | Reliability testing (device design verification/validation)<br>Labeling |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Disinfected device malfunction due to<br>its incompatibility with HLD foam | Device and Material compatibility testing<br>Labeling |
| Adverse tissue reaction to patient due to<br>HLD foam's toxicity | Biocompatibility evaluation |
| Adverse event to end user due to<br>use/user error caused by inappropriate<br>labeling | Human factors testing, simulated-use testing, and in-use<br>testing<br>Labeling |
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the foam or gel chemical sterilant/high level disinfectant is subject to the following special controls:
- (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated under challenging conditions:
- (i) Storage stability testing must demonstrate the real time stability and dynamics of the device formulation within the expiration date (shelf life) of the unopened product and within a use period of the opened container from the date of opening under the proposed storage conditions.
- (ii) Transport stability testing must demonstrate device resilience to transport conditions (such as temperature, pressure, and humidity), environmental factors (such as freeze and thaw), and mechanical impacts (such as the effect of drops on package integrity).
- (iii) Potency testing must demonstrate sporicidal, mycobactericidal, fungicidal, bactericidal, and virucidal activities of the device.
- Simulated use testing must use the mycobacterium species most resistant to the (iv) germicide as the test organism on inoculated instruments to demonstrate a kill of at least 106 inoculated mycobacteria under the labeled contact time.
- (v) In-use testing must test clinical-relevant microorganism on clinically used instruments, in accordance with the labeled contact conditions for device high level disinfection, must confirm the results of simulated use testing.
- (vi) Compatibility testing with labeled devices and materials.
- (vii) Chemical indicator validation must demonstrate a characteristic chemical reaction to the concentration of the germicide active ingredients.
- (2) The device must be demonstrated to be biocompatible.
- Usability / human factors validation testing must demonstrate that the device can be (3) used correctly, based solely on the directions for use.
- (4) Labeling must include:
- Instructions for personal protective equipment to be used with the device. (i)
- Directions for use including: (ii)
- (A) Instructions for preparation and use of the germicide; cleaning steps in preparation for high level disinfection; high level disinfections of cleaned
{8}------------------------------------------------
devices; rinsing, neutralizing, and removing residues, when needed; and reuse of the solution, if applicable.
- (B) Chemical indicator for monitoring the Minimum Effective Concentration (MEC) or Minimum Recommended Concentration (MRC) of the product's active ingredient(s).
- (iii) Storage conditions and expiration date information for stock solution, opened container, activated solution, and use-dilution.
- (iv) A statement that the end user should be trained in the reprocessing (decontamination and sterilization or disinfection) of medical devices and in the handling of toxic substances, such as liquid chemical germicides.
- (v) Instructions for disposal of the germicide and any neutralizers, including an instruction to check local and state regulations.
- Germicide classification scheme. (vi)
- (vii) General information on selection and use of germicides for medical device reprocessing.
- (viii) Material and device compatibility and incompatibility information.
- (ix) Microbial mode of action of germicidal activity.
- Precleaning agent / method compatibility and incompatibility with this device. (x)
- (xi) Toxicology profile of the final product formulation and information on adverse reactions following exposure to the product.
#### BENEFIT-RISK DETERMINATION
#### Benefit
The current paradigm for high level disinfectants is that they support, through testing, that the high level disinfectants will allow for the semi-critical devices to be high level disinfected to at least 6-log reduction of a mixed suspension of vegetative organisms and 6-log reduction of an appropriate mycobacteria. From an end user perspective. this may decrease reprocessing time, as this is a point of care product and does not require transport to a reprocessing room. The subject high level disinfectant, a chlorine dioxide based foam with two-minute contact time, is intended to be used at the point of care without the need for equipment or installation. With appropriate testing and controls in place, the probable benefit of this feature is that it will provide a new option to healthcare providers for high level disinfection of ultrasound devices. Proper use of this device is also likely to improve the time taken to reprocess ultrasound devices between patients and enable healthcare providers to serve more patients. These probable benefits were supported through non-clinical performance testing including potency testing, simulated-use, and in-use testing.
## Risk
The identified risks have been mitigated with appropriate performance testing (for risks of inadequate potency), biocompatibility testing (for risks of solution/residue toxicity to healthcare provider/patient), as well as addressing human factors and the risk related to active ingredient concentration measurement.
{9}------------------------------------------------
## Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
## Benefit/Risk Conclusion
Based on the performance testing provided (in particular potency test, simulated-use test, in-use test, stability / shelf life test, biocompatibility test, human factor study, compatibility test, reliability test, and labeling, chemical indicator validation), the sponsor has adequately supported that the subject device is a high level disinfectant foam for manually reprocessing ultrasound probes. This feature will provide a new option for high level disinfection of ultrasound probes and is likely to improve the time needed to reprocess ultrasound probes between patients. This testing has also mitigated the probable risk of inadequate potency and the subsequent risk of infection resulting from exposure to inadequately disinfected ultrasound probes. Biocompatibility testing has mitigated the risk of adverse tissue reaction. Chemical indicator validation testing and labeling have mitigated the risk of using high level disinfectant at a concentration lower than MEC/MRC.
In conclusion, given the testing provided above, for the following indication statement:
Tristel Duo ULT is a high level disinfect foam for reprocessing ultrasound probes. It can be used to high level disinfect endocavity transvaginal and transrectal probes, and skin surface transducers that may contact non-intact skin during use.
Tristel Duo ULT is a high level disinfectant when used in accordance with the directions for use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).
Tristel Duo ULT must be applied on the surface of an ultrasound probe using Duo Wipes.
Duo Wipes are intended to apply Tristel Duo ULT foam on the surface of ultrasound probes and to remove residue of the foam after high level disinfection.
Each dose of Tristel Duo ULT foam and each Duo Wipe are single use.
The semi-critical ultrasound probes reprocessed by Tristel Duo ULT must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions.
Tristel Duo ULT foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.
Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel Duo ULT may be verified using Duo Test Strips.
Tristel Duo ULT is intended to be marketed for prescription use.
{10}------------------------------------------------
The probable benefits outweigh the probable risks for the Tristel Duo ULT high level disinfectant. The device benefits patients and the risks can be mitigated by the use of general controls and the identified special controls.
## CONCLUSION
The De Novo request for the Tristel Duo ULT high level disinfectant is granted and the device is classified as follows:
Product Code: QWS Device Type: Foam or gel chemical sterilant/high level disinfectant Regulation Number: 21 CFR 880.6886 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.