DEN220014 · Reflexion Medical, Inc. · QVA · Feb 1, 2023 · Radiology
Device Facts
Record ID
DEN220014
Device Name
RefleXion Medical Radiotherapy System (RMRS)
Applicant
Reflexion Medical, Inc.
Product Code
QVA · Radiology
Decision Date
Feb 1, 2023
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 892.5060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RefleXion Medical Radiotherapy System is indicated for treatment planning and precise delivery of image-guided radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery for tumors or other targeted tissues anywhere in the body when radiation treatment is indicated, while minimizing the delivery of radiation to vital healthy tissue. The megavoltage X-ray radiation is delivered in a rotational, modulated, image-guided format in accordance with the physician approved plan. The RefleXion Medical Radiotherapy System is also indicated for FDG-guided treatment which includes modeling, planning and precise delivery of FDG-guided radiation therapy, a type of Biology-guided Radiotherapy (BgRT), in five or fewer fractions for adults. It is indicated for tumor volumes in lung and bone subject to potential motion and positional uncertainty that have each been assessed with on-board PET/CT prior to delivery for adequate localization, sufficient FDG metabolic activity, local contrast and consistent biodistribution to meet the RMRS requirements, while minimizing the delivery of radiation to vital healthy tissue. BgRT involves the detection of signals from F18 during active beam delivery as a guide to deliver megavoltage X-ray radiotherapy in a rotational, modulated format in accordance with a physician approved treatment plan. For complete fludeoxyglucose F18 prescribing information, refer both to the current medical imaging agent labeling and to this device labeling under "FDG Medical Imaging Agent Information".
Device Story
System combines emission computed tomography (PET) and linear accelerator; enables Biology-guided Radiotherapy (BgRT). Inputs: F18-FDG signals detected during active beam delivery via on-board PET/CT. Operation: PET system acquires images of positron-emitting F18 to guide megavoltage X-ray delivery; system performs rotational, modulated radiation delivery based on physician-approved plan. Used in clinical settings by trained professionals. Output: Precise radiation dose to targeted tumor volumes. Benefit: Enables real-time tracking of tumors with motion/positional uncertainty using metabolic signals, potentially improving targeting accuracy while sparing healthy tissue. Workflow includes pre-treatment eligibility assessment, image interpretation, and dose delivery tracking.
Clinical Evidence
Prospective, single-arm IDE study (n=17 total across two cohorts) evaluated BgRT workflow feasibility in adults with lung or bone tumors. Cohort I (n=8) established 15 mCi as the recommended FDG dose. Cohort II (n=9) used in silico emulation to confirm BgRT dose distribution consistency with approved plans. Primary endpoint: 95% of DVH points for BTZ/OAR fell within bounded DVH. Results: 100% of fractions met gamma index (3mm/3%) in phantom testing; 94% of PET evaluations passed. No device-related safety concerns reported.
Technological Characteristics
System includes a compact linear accelerator, primary collimation, PET scanner, kVCT, and MV x-ray detectors. PET subsystem uses two arcs on a ring gantry. Materials/reprocessing unchanged from K190978. Complies with IEC 60601-1, 60601-2-1, 60601-2-44, 60601-2-68, and NEMA NU 2-2018. Software is major level of concern. Connectivity: DICOM/RT structure set import.
Indications for Use
Indicated for adults with lung or bone tumors subject to motion/positional uncertainty requiring image-guided radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery. Requires assessment via on-board PET/CT for adequate localization, FDG metabolic activity, local contrast, and biodistribution.
Regulatory Classification
Identification
A fludeoxyglucose F18-guided radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage x-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
Special Controls
In combination with the general controls of the FD&C Act, the fludeoxyglucose F18-guided radiation therapy system is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR REFLEXION MEDICAL RADIOTHERAPY SYSTEM (RMRS)
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Fludeoxyglucose F18-guided radiation therapy system. A fludeoxyglucose F18guided radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage x-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
NEW REGULATION NUMBER: 21 CFR 892.5060
CLASSIFICATION: Class II
PRODUCT CODE: QVA
#### BACKGROUND
DEVICE NAME: RefleXion Medical Radiotherapy System (RMRS)
SUBMISSION NUMBER: DEN220014
DATE DE NOVO RECEIVED: February 23, 2022
#### SPONSOR INFORMATION:
RefleXion Medical Inc 25841 Industrial Boulevard, Suite 275 Hayward. California 94545
#### INDICATIONS FOR USE
The RefleXion Medical Radiotherapy System (RMRS) is indicated as follows:
The RefleXion Medical Radiotherapy System is indicated for treatment planning and precise delivery of image-guided radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery for tumors or other targeted tissues anywhere in the body when radiation treatment is indicated. while minimizing the delivery of radiation to vital healthy tissue. The megavoltage X-ray radiation is delivered in a rotational, modulated, image-guided format in accordance with the physician approved plan.
{1}------------------------------------------------
The RefleXion Medical Radiotherapy System is also indicated for FDG-guided treatment which includes modeling, planning and precise delivery of FDG-guided radiation therapy, a type of Biology-guided Radiotherapy (BgRT), in five or fewer fractions for adults. It is indicated for tumor volumes in lung and bone subject to potential motion and positional uncertainty that have each been assessed with on-board PET/CT prior to delivery for adequate localization, sufficient FDG metabolic activity, local contrast and consistent biodistribution to meet the RMRS requirements, while minimizing the delivery of radiation to vital healthy tissue. BgRT involves the detection of signals from F18 during active beam delivery as a guide to deliver megavoltage X-ray radiotherapy in a rotational, modulated format in accordance with a physician approved treatment plan.
For complete fludeoxyglucose F18 prescribing information, refer both to the current medical imaging agent labeling and to this device labeling under "FDG Medical Imaging Agent Information".
### LIMITATIONS
- . The sale, distribution, and use of the RMRS are restricted to prescription use in accordance with 21 CFR 801.109.
- . The clinical study did not include treatment of patients with the RMRS device, and instead assessed effectiveness through the use of data simulated in silico.
- . The quality of the positron emission tomography (PET) images obtained using the RMRS device are not of diagnostic quality. The PET system within the RMRS device is not intended to be used as a stand-alone diagnostic device.
- Clinical performance testing did not include an evaluation of the treatment of multiple . targets in a single treatment session.
- . The RMRS is not intended for use with radiopharmaceuticals other than fludeoxyglucose F18.
- . The biology-guided radiotherapy function requires a use of fludeoxyglucose F18 that is not in accordance with the currently approved prescribing information.
- . The exact dose to be delivered to the target prior to treatment is unknown. Bounded dose volume histogram (DVH), which is calculated using the planning PET scan, includes a range of uncertainty to account for possible variations in dosimetric outcomes during treatment.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
{2}------------------------------------------------
# DEVICE DESCRIPTION
The RMRS is a linear accelerator capable of delivering intensity-modulated radiation therapy, stereotactic body radiation therapy, stereotactic radiotherapy, stereotactic radiosurgery, and biology-guided radiotherapy (BgRT).
The RMRS is comprised of six major subsystems as shown in Figure 1.
Image /page/2/Figure/3 description: The image shows a large, complex piece of machinery, possibly a medical imaging device like a CT scanner or MRI machine. The machine has a circular opening in the center, surrounded by various electronic components, wiring, and mechanical parts. The numbers 1 through 6 are overlaid on the image, pointing to different sections of the machine, likely for identification or labeling purposes. The overall impression is one of advanced technology and intricate engineering.
Figure 1. The major subsystems of the RefleXion Medical Radiotherapy System are: 1) Compact linear accelerator, 2) primary collimation, 4) PET scanner, 5) kVCT, and 6) MV x-ray detectors.
The linear accelerator portion of the device was previously cleared under a 510(k) premarket notification (K190978). The subject device in K190978 included the PET hardware, but the PET hardware and associated software was inactive and outside the scope of K190978. The subject of the current submission is the addition of the BgRT functionality and the activation of the PET hardware.
### Biology-guided Radiation Therapy
BgRT is a type of volumetric modulated arc therapy that allows radiation dose delivery in a slice-by-slice fashion using a multi-pass couch motion based on the collection and processing of
{3}------------------------------------------------
PET data from the radiotracer fludeoxyglucose F18 (FDG). BgRT also allows the dose delivery to be adjusted in real-time to account for target motion in the lung and bone.
The PET system is comprised of two PET detector arcs mounted on the ring gantry that rotate with the system to generate complete tomographic samples of radiotracer distribution required for the BgRT algorithm.
Rapidly acquired "limited time sampled" (LTS) PET imaging data are used to form accumulated lines of response (projection data). These data are then used to guide the beam using the mapping calculated during BgRT planning, maintaining the ability to track the target in real-time as it moves within a pre-defined volume called the biology-tracking zone (BTZ). The BTZ is the volume within which primary dose delivery is allowed. It encompasses the motion extent (internal target volume or ITV), biology-guidance margin (BgM) and the set-up margin (SM) (Figure 2). Unlike standard radiotherapy, the planning target volume (PTV) expansion for BgRT only includes the biology-guidance margin (BgM) and not the motion extent with internal margin (IM) or the set-up margin since the delivery tracks the PTV. The clinical target volume (CTV) and the gross tumor volume (GTV) are the same between BgRT and standard RT. The treatment delivery system uses the BTZ as a limiting factor in the delivery of the prescription dose.
Image /page/3/Figure/3 description: The image compares standard radiation therapy (RT) to biology-guided RT. The standard RT diagram shows nested regions labeled GTV, CTV, ITV, IM&SM, IM, and PTV. The biology-guided RT diagram shows nested regions labeled GTV, CTV, PTV, BgM, BgM&SM, BTZ, and motion extent. The diagrams illustrate how the target volumes and margins differ between the two approaches.
Figure 2. Comparison of volumes drawn for standard radiation therapy and BgRT.
In order to ensure sufficient FDG signal during planning and on the day of treatment, the treatment planning software calculates two quantitative parameters from the tumor PET image to determine whether BgRT is a viable option:
- . Activity Concentration (AC) - a minimum FDG activity concentration calculated by estimating the net signal concentration from the target.
- Normalized Target Signal (NTS) Calculated by taking the Activity Concentration . (focused on the target signal) normalized by the noise in the background around the BTZ.
{4}------------------------------------------------
An AC of > 5 kBq/ml at the time of treatment planning and treatment delivery ensures sufficient counts in the LTS PET image to translate to accurate fluence (hence accurate dose). An NTS of ≥ 2.7 at the time of treatment planning and ≥ 2.0 at the time of treatment delivery ensures adequate PET tumor contrast (tumor signal to background) for the system to calculate the fluence for both the planning and treatment delivery.
Image /page/4/Figure/1 description: This image is a flowchart that outlines the steps involved in a medical simulation and treatment process. The flowchart is divided into five main stages: Simulation, Plan set-up, Imaging-only, Optimization & Evaluation, and Delivery. Each stage consists of several steps, such as importing patient data, setting optimization constraints, and performing PET injections. The flowchart also indicates which steps are performed by an external system and which are unique to BgRT.
# Clinical Workflow
Figure 3. Clinical workflow using BgRT
Step 1: Using patient data obtained from an external system PET/CT, users draw contours on a simulation CT on third-party software. Contours include the BTZ in addition to other target contours. The RefleXion treatment planning station allows the user to import the resulting CT DICOM image set and the DICOM RT structure set, which includes contours for the appropriate volumes.
Step 2: The detailed treatment plan is developed (typically by a medical physicist or dosimetrist) using the RefleXion treatment planning software, providing estimated dose volume histograms and other dosimetric measures. The radiation oncologist must approve the treatment plan, which is then saved in the RefleXion database and available at the operator console. The medical physicist conducts patient-specific quality assurance of the treatment plan and must approve the treatment plan prior to proceeding.
- . Step 2a (BgRT only): If the patient is a candidate for BgRT, an initial plan will be created and approved for an "Imaging-only" session on the RMRS. This plan is used to collect the pre-treatment planning PET image (a RefleXion planning PET image) that will be used in the remaining treatment planning process.
- . Step 2b (BgRT only): FDG is administered at an appropriate time prior to the Imagingonly session. FDG has a half-life of 109.7 minutes, and an injection of 15 mCi is recommended. Typically, the injection is administered in an "uptake" room, where the
{5}------------------------------------------------
patient sits quietly for approximately one hour after injection. The patient then proceeds to the radiotherapy delivery suite.
- . Step 2c (BgRT only): The patient is positioned on the treatment couch using lasers for initial alignment and then the kVCT subsystem is used to check patient positioning. The planning PET image is acquired.
- . Step 2d (BgRT only): With the planning PET image available in the treatment planning system, the radiation oncologist will decide, based on the metabolic characteristics (which include AC and NTS) of the target volume. whether radiotherapy delivery with PET guidance is warranted. If so, the planning PET image is used to create a BgRT plan.
Step 3 (BgRT only): The plan is then optimized and evaluated using DVHs that are specific to the BgRT workflow called bounded DVHs (bDVHs).
- Step 3a (BgRT only): If the plan is considered satisfactory, and the BgRT metrics . (activity concentration and normalized target signal) are met, the plan can then go through patient-specific plan QA, be approved, and used for treatment delivery.
Step 4 (BgRT only): Patients receiving BgRT will have the FDG iniection (15 mCi) prior to each treatment fraction. Typically, the injection is administered in an "uptake" room, where the patient sits quietly for approximately one hour after iniection. The patient then proceeds to the radiotherapy delivery suite.
Step 5: At the time of treatment, the patient is positioned on the treatment couch. Lasers assist in the initial patient alignment. The kVCT system will be used for every fraction to obtain kVCT images for use in checking patient positioning against the treatment plan. After the clinical staff confirms satisfactory alignment of the kVCT images with the planning CT image, any necessary couch shifts can be applied.
Step 5 (BgRT Only): For cases where PET information is being used to guide the radiotherapy, after the patient is positioned using the kVCT system, a PET pre-scan of the patient is conducted immediately prior to delivering radiation to confirm that the PET information acquired on the day of treatment is consistent with the PET information used to develop the treatment plan. The PET evaluation must pass for the patient to proceed to treatment delivery. The physician also must confirm the alignment of the PET signal with the appropriate contours before moving to treatment delivery.
Step 6: After the kVCT and PET pre-scan imaging is complete, the couch moves longitudinally through the bore of the gantry to the starting position at the therapy plane and then the radiotherapy begins according to the treatment plan.
- Step 6a (BgRT Only): During BgRT delivery, LTS PET images are used to provide . tracked dose delivery to the PTV, which can account for positional offsets or tumor motion during treatment (for example, patient respiration, peristalsis, or voluntary movement).
{6}------------------------------------------------
Step 7: After the therapy is completed, the delivered dose is recorded and verified. The delivered dose distribution (calculated from the delivered beamlet sequence) can be generated and reviewed on the Treatment Planning system. The post-treatment PET image (i.e., the PET image generated from all emission data collected during the treatment delivery) can also be reviewed on the Treatment Planning system at this time.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
### BIOCOMPATIBILITY/MATERIALS
The patient contacting components are unchanged from the cleared linear accelerator compatible with this system (K190978).
## SHELF LIFE/STERILITY
No components of the device are provided sterile. Reprocessing instructions are unchanged from the cleared linear accelerator compatible with this system (K190978).
## ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
Electrical safety and electromagnetic compatibility testing were performed per the following standards and was found to be acceptable:
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2 Edition 4.0 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-3 Edition 2.1 2013 Medical electrical equipment Part 1-3: General . requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic x-ray equipment
- . IEC 60601-1-6 Edition 3.1 2013 Medical electrical equipment - Part 1-6; General requirements for basic safety and essential performance - Collateral standard: Usability
- . IEC 60601-2-1 Edition 3.1 2014 Medical electrical equipment - Part 2-1: Particular requirements for the basic safety and essential performance of electron accelerators in the range 1 MeV to 50 MeV
- EC 60601-2-44 Edition 3.2: 2016 Medical electrical equipment Part 2-44: Particular . requirements for the basic safety and essential performance of x-ray equipment for computed tomography
- IEC 60601-2-68 Edition 1.0 2014 Medical electrical equipment Part 2-68: Particular . requirements for the basic safety and essential performance of x-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam therapy equipment
{7}------------------------------------------------
# SOFTWARE
The RMRS software documentation and testing provided demonstrate that the device meets all requirements outlined in the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for software of major level of concern.
| Test | Purpose | Method | Performance Metrics<br>and Acceptance<br>Criteria | Results |
|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Moving target<br>with 3-D<br>respiratory and<br>3-D non-<br>respiratory<br>motion; organs<br>at risk (OARs)<br>independent<br>motion | To demonstrate the<br>BgRT dose delivery<br>accuracy to target(s)<br>and measure dose to<br>OARs with respiratory<br>or non-respiratory<br>continuous motion<br>pattern:<br>• BgRT dose delivery<br>accuracy to target<br>and nearby normal<br>tissues (OARs)<br>• BgRT dose delivery<br>accuracy for varying<br>degrees of clinical<br>motion patterns<br>• Comparison with<br>stereotactic body<br>radiation therapy<br>(SBRT) treatment<br>delivery | Tests were performed<br>with a large<br>anthropomorphic<br>phantom. The<br>background consisted<br>of water with FDG to<br>simulate an FDG-active<br>background. Two arms<br>were included that<br>could be programmed<br>with distinct motion<br>patterns. Inserts of<br>varying shape/size were<br>injected with FDG to<br>simulate targets or<br>OARs. The shapes<br>included sphere, C-<br>shape, and ovoid. The<br>C-shape and sphere<br>OARs were used for<br>respiratory motion tests,<br>and the ovoid was used<br>for the non-respiratory<br>motion test. | Coverage margin loss -<br>The maximum of<br>(CMP-CMD* in a<br>given plane) for all<br>planes in 3D space. The<br>acceptance criterion for<br>coverage margin loss is<br>≤ 3mm.<br>Dosimetric coverage -<br>Dose at all the points<br>measured on film<br>within the CTV ≥ 97%<br>of prescription dose,<br>but ≤ 130% of<br>maximum planned<br>dose.<br>For the OAR, two<br>measurements were<br>made. A point dose<br>measurement was made<br>by a calibrated ion<br>chamber and was<br>required to be less than<br>or equal to the expected<br>plan dose as a<br>quantitative dose check.<br>In addition, the<br>maximum dose<br>measured on the<br>radiochromic film<br>inserted into the OAR<br>structure was required<br>to be less than the<br>maximum plan dose as<br>a qualitative dose<br>check. | Both dosimetric<br>coverage criteria as<br>well as OAR dose<br>criteria were met for all<br>BgRT test cases.<br>Under the same test<br>circumstances, SBRT<br>plans with an ITV<br>approach for the target<br>were not able to<br>successfully meet the<br>same criteria as the<br>BgRT plans. |
| Target or OAR<br>with a single<br>step shift - | To demonstrate the<br>BgRT dose delivery<br>accuracy to target(s) | Same methodology as<br>described above. A<br>specific OAR shape | Same acceptance<br>criteria as described<br>above. | Both dosimetric<br>coverage criteria as<br>well as OAR dose |
| (positional<br>uncertainty) | OARs in the case of a<br>single step shift of the<br>target:<br>• Effectiveness of<br>BgRT treatment<br>providing conformal<br>dose to the target<br>while sparing nearby<br>normal tissues<br>(OARs)<br>• BgRT dose delivery<br>accuracy<br>• Comparison with<br>SBRT treatment<br>delivery | was not utilized for this<br>test. | criteria were met for all<br>BgRT test cases.<br><br>Under the same test<br>circumstances, SBRT<br>plans with an ITV<br>approach for the target<br>were not able to<br>successfully meet the<br>same criteria as the<br>BgRT plans. | |
| Test clinical<br>variations of<br>targets<br>(stationary) | To demonstrate the<br>BgRT dose delivery<br>accuracy for various<br>clinical situations:<br>• Various target-to-<br>background (T:B)<br>contrast<br>• Different FDG<br>biodistribution<br>patterns between<br>planning and<br>delivery while<br>keeping the same<br>anatomic shape of<br>the target<br>• Test homogenous or<br>heterogenous<br>backgrounds around<br>the target | The phantom used in<br>the stationary tests was<br>the ArcCHECK Model<br>(b)(4) | PET evaluation results<br>(AC/NTS/bDVH).<br><br>Bounded DVH – 95%<br>of points in the<br>calculated dose DVHs<br>of all static structures<br>must be within the<br>bounds of the bDVH<br>from the treatment plan.<br><br>Gamma Index - This<br>method compares the<br>plan dose extrapolated<br>to all points in space<br>represented by the<br>(calibrated) detector<br>diodes. The acceptance<br>criteria chosen for these<br>tests required that 90%<br>of the comparison<br>points (that were above<br>10% of the prescription<br>dose) were within either<br>3% in dose or closer<br>than 3mm in 3D space<br>(DTA or distance to<br>agreement) to a point<br>with the same dose.<br><br>Acceptance criteria<br>requires both the PET<br>Evaluation to pass and<br>either post-delivery<br>DVH to have ≥ 95% of<br>the points fall within<br>the bounds of the<br>treatment plan and/or<br>the Gamma index to be<br>above 90% | For all tests where the<br>acceptance criteria<br>required both the PET<br>evaluation to pass and<br>either the post-<br>treatment dose<br>evaluation against the<br>plan and/or the Gamma<br>index to pass met the<br>dose accuracy criteria.<br>For cases where there<br>was a significant<br>change in<br>biodistribution in the<br>target, low contrast<br>between the target and<br>background, or a<br>change in the<br>background signal, the<br>PET evaluation did not<br>pass, and the resulting<br>dose accuracy showed<br>mixed results. |
| Tests evaluating<br>special edge<br>cases<br>(stationary) | To demonstrate<br>robustness of the BgRT<br>system under<br>challenging situations:<br>• Variations between<br>planning and<br>delivery such as<br>changes in PET<br>biodistribution<br>patterns<br>• PET-avid OAR in the<br>BTZ prior to delivery<br>• Dose delivery<br>accuracy to target in<br>the presence of<br>nearby PET-avid<br>OAR.<br>• PET Evaluation step<br>challenge with large<br>changes between<br>planning and<br>delivery | Same methodology as<br>described above with<br>stationary tests. | Same acceptance<br>criteria as described<br>above with stationary<br>tests. | For cases where there<br>was a significant<br>change in<br>biodistribution between<br>planning and delivery,<br>or a situation when an<br>FDG-avid target moves<br>into the BTZ, the PET<br>evaluation did not pass. |
| BgRT ON/OFF<br>with Motion<br>ON/OFF | To demonstrate the<br>benefit of BgRT in the<br>same test setup by<br>running with BgRT ON<br>compared to BgRT<br>OFF.<br>• BgRT dose delivery<br>accuracy to target<br>and measure dose to<br>OAR with<br>respiratory<br>continuous motion<br>pattern for the target<br>• BgRT dose delivery<br>accuracy to target<br>and measure dose to<br>OAR in the case of a<br>single shift of the<br>target | Same phantom as used<br>with the motion tests<br>described above. The<br>C-shape was used for<br>respiratory motion tests,<br>and the sphere was used<br>for the non-respiratory<br>motion test. | Coverage margin loss -<br>The maximum of<br>(CMP-CMD* in a<br>given plane) for all<br>planes in 3D space. The<br>acceptance criterion for<br>coverage margin loss is<br>≤ 3mm.<br>Dosimetric coverage -<br>Dose at all the points<br>measured on film<br>within the CTV ≥ 97%<br>of prescription dose,<br>but ≤ 130% of<br>maximum planned<br>dose.<br>For nearby OARs, the<br>maximum dose will be<br>measured using<br>radiochromic film over<br>all valid dose points<br>and the acceptance<br>criteria is that the<br>maximum dose<br>measured on the film<br>must be below a value<br>that is 3% higher than<br>the bDVH maximum<br>dose as calculated by<br>the plan. | BgRT ON Motion ON -<br>The tracked dose<br>delivery was<br>successfully provided<br>where the CTV was<br>fully covered (100%) in<br>all cases. In cases for<br>which the PET<br>evaluation did not pass,<br>some degradation to<br>margin loss was<br>observed.<br>BgRT OFF Motion ON<br>- In all cases dose<br>accuracy failed the<br>criteria. The CTV dose<br>coverage was less than<br>100% and margin loss<br>higher than 3.0mm.<br>BgRT ON Motion OFF<br>- All test cases met the<br>dose accuracy criteria.<br>BgRT OFF Motion<br>OFF - All test cases<br>met the dose accuracy<br>criteria. |
| Multi-target<br>testing | To determine accuracy<br>of delivery with cases<br>of lower target to<br>background contrast | Same methodology as<br>motion testing. | Coverage margin loss -<br>The maximum of<br>(CMP-CMD* in a<br>given plane) for all<br>planes in 3D space. The<br>acceptance criterion for<br>coverage margin loss is<br>≤3mm.<br><br>Dosimetric coverage -<br>Dose at all the points<br>measured on film<br>within the CTV ≥ 97%<br>of prescription dose,<br>but ≤ 130% of<br>maximum planned<br>dose | For all 3-target, moving<br>tests, including with<br>reduced target:<br>background for the<br>second and third<br>targets, all dosimetric<br>criteria (CTV dose<br>coverage and Margin<br>Loss) were met. |
# PERFORMANCE TESTING - BENCH
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
* Coverage margin planning (CMP) - The distance (in any plane) from the CTV contour to the 97% isodose contour in a treatment plan. The target was considered "covered" in the treatment plan if 100% of the PTV receives 97% of the dose. Coverage margin delivery (CMD) - The distance (measured in any plane, but usually LR, SI, AP) from the CTV contour to 97% delivered dose contour measured (usually from film) in the reference frame of the target (that is, on a moving phantom)
Verification testing was performed for the PET subsystem to assess the following:
- PET NEMU NU-2 performance tests for spatial resolution, scatter fraction and count rate . measurements, sensitivity, and image quality
- PET System Performance Checks .
- . PET image reconstruction and attenuation correction
All tests passed the pre-defined acceptance criteria.
Verification testing was performed to assess the performance and integration of the following subsystems:
- . kVCT
- . MV x-ray detectors (MVD)
- Beamgen .
- Collimation .
- Couch .
- Treatment planning .
- Treatment delivery .
Testing was performed in accordance with the following standards and found to be acceptable:
- IEC 62368-1 / UL 62368-1:2014 Ed.2 Audio/video, information and . communication technology equipment - Part 1: Safety requirements
- IEC 61217 Edition 2.0 2011 Radiotherapy equipment Coordinates, movements. . and scales
- IEC 62083 Edition 2.0 2009 Medical electrical equipment Requirements for the . safety of radiotherapy treatment planning systems
- . IEC 62274 First Edition 2005 Medical electrical equipment - Safety of radiotherapy record and verify systems
{11}------------------------------------------------
- IEC 60976 Edition 2.0 2007 Medical electrical equipment Medical electron . accelerators - Functional performance characteristics
- NEMA NU 2-2018 Performance Measurements of Positron Emission . Tomographs
### SUMMARY OF CLINICAL INFORMATION
A single-arm, prospective, IDE clinical trial was conducted at two sites in the U.S. The study was conducted using two cohorts, and the population included adults 21 years or older with at least 1 active tumor in the bone or lung.
### Cohort I
- Primary Objective: To identify the Recommended RefleXion FDG Dose (RRFD) that . enables the use of BgRT on the RMRS.
- · Primary Endpoint: The FDG dose that results in Activity Concentration necessary for BgRT functioning (i.e., 5 kBq/ml or higher).
- Secondary Objective: To assess the performance of the BgRT PET imaging-only session, . treatment planning and quality assurance at the studied dose level.
- o Secondary Endpoints
- 1. Percent of cases where there was an agreement between a site investigator (SI) and agreement standard (AS) for the BgRT PET imaging-only session localization decision (overall percent agreement). Positive percent agreement and negative percent agreement are also reported.
- 2. Percent of cases where there was concordance of the positive "plan proceed" decision between the BgRT imaging-only session PET and a cleared, thirdparty diagnostic PET/CT (positive percent agreement). Overall percent agreement and negative percent agreement are also reported.
- 3. Percent of cases where RefleXion PET data could be used to generate an acceptable BgRT plan such that relevant dosimetric parameters for the target and the nearby normal anatomy were met based on Investigator assessment.
- 4. Percent of cases where the intended dose distribution of the BgRT plan was achieved in a physical phantom, defined as meeting a standard gamma index for external beam radiotherapy quality assurance, i.e., whether 90% of pixels meet the 3 mm/3% deviation standard.
- Study design .
- This cohort sought to identify the Recommended RefleXion FDG dose (RRFD). o which is the dose of administered FDG that allows for functioning of the RMRS. This cohort also sought to assess RMRS PET imaging performance in comparison to a third-party diagnostic PET/CT.
- The dose levels of 15 mCi and 20 mCi were to be assessed sequentially in an o escalation protocol (note that 15 mCi was found sufficient and thus the study never progressed to the 20 mCi dose level). Patients with at least one known FDG avid tumor (i.e., SUVmax ≥ 6 on diagnostic PET/CT) in the bone or lung were enrolled into this cohort.
{12}------------------------------------------------
- These patients underwent a CT simulation in an acceptable radiotherapy treatment o position and with immobilization devices as needed. After acquisition of (4D)CT images, contours for targets, OARs, and BTZ were be generated by the investigator.
- Next, the patient underwent back-to-back PET scans on the RMRS device and a thirdo party diagnostic PET/CT device after a single injection of FDG at the studied FDG dose level
- Quantitative metrics were collected for each lesion to assess the performance of the o RMRS PET subsystem.
- The demographics of Cohort I are listed in Table 1. .
| Demographic | Cohort I (n=8) |
|-------------------------------------|----------------------------------------|
| Age (years) | $64.6 \pm 14.01$ |
| Gender | Female: 12.5% (1) |
| Ethnicity | Not Hispanic or Latino: 100% (8) |
| Race | Asian: 25% (2) |
| | Black or African American: 12.5% (1) |
| | White: 50% (4) |
| | Unknown: 12.5% (1) |
| Baseline ECOG<br>performance status | ECOG 1: 62.5% (5)<br>ECOG 2: 37.5% (3) |
| Baseline height (cm) | $176.6 \pm 8.33$ |
| Baseline weight (kg) | $67.43 \pm 15.257$ |
| Baseline BMI (kg/m²) | $21.47 \pm 3.803$ |
| Cancer type | Bone: 62.5% (5)<br>Lung: 37.5% (3) |
| Tumor size (cm) | $2.71 \pm 0.831$ |
| Cancer stage | IV: 75% (6)<br>Unknown: 25% (2) |
Table 1. Cohort I Demographics
- The results from the primary endpoint are displayed in Table 2. .
| Tumor type | Planned Dose<br>(mCi) | Actual Dose*<br>(mCi) | Activity<br>Concentration<br>(kBq/ml) | Met the Threshold<br>for BgRT? |
|------------|-----------------------|-----------------------|---------------------------------------|--------------------------------|
| Lung | 15 | 15.46 | 8.87 | Yes |
| Lung | 15 | 15.12 | 7.46 | Yes |
| Lung | 15 | 16.37 | 21.05 | Yes |
| Bone | 15 | 15.52 | 14.61 | Yes |
| Bone | 15 | 14.64 | 11.27 | Yes |
| Bone | 15 | 17.70 | 18.47 | Yes |
#### Table 2, Cohort I Primary Endpoint Results
* A variance of 10% was pre-defined as acceptable based on typical practice.
{13}------------------------------------------------
- The results from the secondary endpoints are displayed in Table 3. Of the 8 patients enrolled, . 6 had evaluable images.
| Endpoint # | Description | Result |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| 1 | Percent of cases where there was an agreement<br>between a SI and AS for the BgRT PET imaging-<br>only session localization decision (overall<br>percent agreement). | Overall % agreement = 83.3% (5/6)<br>Positive % agreement = 83.3%<br>Negative % agreement = N/A |
| 2 | Percent of cases where there was concordance of<br>the positive “plan proceed” decision between the<br>BgRT imaging-only session PET and a cleared,<br>third-party diagnostic PET/CT (positive percent<br>agreement). | Positive % agreement = 100% (6/6)<br>Overall % agreement = 83.3%*<br>Negative % agreement = 0 |
| 3 | Percent of cases where RefleXion PET data<br>could be used to generate an acceptable BgRT<br>plan such that relevant dosimetric parameters for<br>the target and the nearby normal anatomy were<br>met based on Investigator assessment. | 67% (4/6) |
| 4 | Percent of cases where the intended dose<br>distribution of the BgRT plan was achieved in a<br>physical phantom, defined as meeting a standard<br>gamma index for external beam radiotherapy<br>quality assurance, i.e., whether 90% of pixels<br>meet the 3 mm/3% deviation standard. | 67% (4/6)<br>100% (4/4) for all cases where an<br>acceptable plan was created. |
Table 3. Cohort I Secondary Endpoint Results
* The overall % agreement for the aggregate radiation oncologist review was 83.3% (5/6) due to one case in which localization was not established on the RMRS device even though it was established on the diagnostic PET/CT.
- The safety results from Cohort I are shown in Table 4. The Safety Population were the . participants who received any FDG dose for RefleXion Imaging-only session.
Table 4. Cohort I Safety Results
| System Organ Class<br>Preferred Term | Dose Level 1<br>(15 mCi)<br>(n=8)<br>n (%) |
|-------------------------------------------------|--------------------------------------------|
| Any TEAEs | 5 (62.5) |
| Musculoskeletal and connective tissue disorders | 4 (50.0) |
| Back pain | 2 (25.0) |
| Arthralgia | 1 (12.5) |
| Neck pain | 1 (12.5) |
| Pain in extremity | 1 (12.5) |
| Gastrointestinal disorders | 3 (37.5) |
| Nausea | 2 (25.0) |
| Oral dysaesthesia | 1 (12.5) |
| Metabolism and nutrition disorders | 1 (12.5) |
| Decrease appetite | 1 (12.5) |
{14}------------------------------------------------
| Nervous system disorders | 1 (12.5) |
|--------------------------|----------|
| Somnolence | 1 (12.5) |
% = 100 x n/N, where N = number of participants in the given study group in the population and n = number of participants in a specified category.
A TEAE was defined as any AE that started on or after the first dose of FDG or occurred prior to the first dose and worsened in severity on or after the first dose of FDG, during the treatment period and follow-up period. AE = adverse event: FDG = fludeoxyglucose. PT = preferred term: SOC = system organ class; TEAE = treatment emergent adverse event
### Cohort II
- Primary Objective: To determine whether BgRT dose distributions generated from LTS PET . images obtained at the time of treatment delivery are consistent with the approved BgRT plan.
- Primary Endpoint: The percent of radiotherapy fractions where the emulated BgRT o dose distribution in silico was shown to be consistent with the approved BgRT treatment plan (i.e., 95% of DVH Delivered points for the BTZ and organ-at-risk [OAR] fell within bounded DVH of the approved BgRT plan).
- . Secondary Objective: To emulate and confirm deliverability of the fluence associated with the BgRT dose distribution generated from LTS PET images obtained at the time of treatment delivery as well as to assess imaging, process and safety characteristics of the endto-end workflow.
- O Secondary Endpoints:
- 1. Percent of fractions where there was concordance between the physical and digital phantoms of emulated BgRT delivery derived from human participant PET emissions. Concordance was defined as a standard gamma index with a goal that 90% of pixels met the 3 mm/3% deviation standard.
- 2. Percent of cases where there was agreement between a SI and the AS for the BgRT PET PreScan localization decision (overall percent agreement). Positive percent agreement and negative percent agreement are also reported.
- 3. Percent of cases where there was concordance of a positive localization decision between the short duration PET PreScan and a third-party diagnostic PET/CT scan (positive percent agreement). Overall percent agreement and negative percent agreement are also reported.
- 4. Safety of multiple FDG administrations and toxicity rates of bladder and bone marrow assessed by complete blood counts (CBCs), urinalysis, and adverse events (AEs) specific to bladder and bone marrow assessed by Common Terminology Criteria for Adverse Events (CTCAE) version 5.0 at 72 ± 24 hours after final FDG injection.
- 5. Workflow characterization:
- a. Percent of PET imaging sessions at RRFD that met the AC threshold for BgRT (5 kBq/mL);
- b. Percent of PET imaging sessions which led to acceptable BgRT plans, with acceptability based upon meeting user-defined coverage goals for tumor targets and avoidance goals for OARs:
{15}------------------------------------------------
- c. Percent of approved BgRT plans that went on to pass physics quality assurance, as defined by 90% of pixels meeting the 3 mm/3% deviation standard: and
- d. Percent of PET evaluations on the day of fraction delivery that elicited a "Pass" signal.
- Study design .
- · This cohort sought to confirm that the machine-deliverable fluence generated by applying the BgRT firing filter to PET LTS images obtained at the time of a radiotherapy delivery does in fact result in an anatomic dose distribution that is consistent with the approved BgRT plan. It also sought to confirm that the linear accelerator subsystem hardware is able to deliver the received machine instructions. The purpose was to emulate and assess (without actually delivering the radiation therapy to the patient) the entire end-to-end BgRT workflow from simulation to treatment planning to, finally, dose delivery.
- Patients dispositioned to undergo conventional SBRT for a single bone tumor or a o single lung tumor were enrolled. For each patient, RMRS PET collections were added to the SBRT workflow at 3 timepoints representing the steps when the RMRS PET subsystem would be utilized during the BgRT workflow. Specifically, these timepoints included a RMRS PET imaging-only session prior to the start of SBRT delivery that was used to create a BgRT plan as well as RMRS PET collections before the first and final fractions of their planned course of SBRT. A single comparison third-party diagnostic PET/CT image was obtained (utilizing the same FDG injection) on the day of the final fraction.
- In order to emulate BgRT delivery, each of these fractional PET collections consisted of two phases:
- 1. Short-duration PET collection that corresponds to the duration of a PET PreScan Evaluation
- 2. Long-duration PET collection that corresponds to the duration of an active BgRT delivery fraction
- · The long-duration PET collection data mimics RefleXion LTS PET image acquisition during live BgRT and was therefore used to determine the hardware instructions that would be transmitted to the RMRS delivery hardware (linear accelerator, gantry, MLC, etc.) based upon the algorithmic interaction between the incoming LTS images and the approved BgRT treatment plan.
- The demographics of Cohort II are listed in Table 5. .
| Demographic | Bone Tumors (n=4) | Lung Tumors (n=5) | Total (n=9) |
|-------------|-------------------------------------|------------------------------------|--------------------------------------|
| Age (years) | 69.5 ± 9.81 | 69.6 ± 12.44 | 69.6 ± 10.65 |
| Gender | Female: 0% | Female: 40% (2) | Female: 22.2% (2) |
| Ethnicity | Not Hispanic or Latino:<br>100% (4) | Not Hispanic or Latino:<br>80% (4) | Not Hispanic or Latino:<br>88.9% (8) |
Table 5. Cohort II Demographics
{16}------------------------------------------------
| | | Hispanic or Latino: 20%<br>(1) | Hispanic or Latino: 11.2%<br>(1) |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Race | Black or African<br>American: 25% (1)<br>White: 75% (3) | Black or African<br>American: 40% (2)<br>White: 40% (2)<br>American Indian or<br>Alaskan Native: 20% (1) | Black or African<br>American: 33.3% (3)<br>White: 55.6% (5)<br>American Indian or<br>Alaskan Native: 11.1% (1) |
| Baseline ECOG<br>performance<br>status | ECOG 0: 75% (3)<br>ECOG 1: 25% (1) | ECOG 0: 40% (2)<br>ECOG 1: 40% (2)<br>ECOG 2: 20% (1) | ECOG 0: 55.6% (5)<br>ECOG 1: 33.3 (3)<br>ECOG 2: 11.1 (1) |
| Baseline height<br>(cm) | 177.3 ± 5.91 | 170.8 ± 7.60 | 173.7 ± 7.31 |
| Baseline weight<br>(kg) | 92.05 ± 14.563 | 85.78 ± 17.208 | 88.57 ± 15.444 |
| Baseline BMI<br>(kg/m $^{2}$ ) | 29.21 ± 3.389 | 29.58 ± 6.350 | 29.42 ± 4.950 |
| Cancer type | Primary bone: 25% (1)<br>Metastatic bone: 25% (1)<br>Primary lung: 25% (1)<br>Metastatic lung: 0%<br>Other: 25% (1) | Primary bone: 0%<br>Metastatic bone: 0%<br>Primary lung: 80% (4)<br>Metastatic lung: 20% (1)<br>Other: 0% | Primary bone: 11.1% (1)<br>Metastatic bone: 11.1%<br>(1)<br>Primary lung: 55.6 (5)<br>Metastatic lung: 11.1% (1)<br>Other: 11.1% (1) |
| Tumor size (cm) | 2.70 ± 0.735 | 2.46 ± 0.541 | 2.57 ± 0.604 |
| Cancer stage | IV: 100% (4) | II: 20% (1)<br>IV: 20% (1)<br>Unknown: 60% (3) | II: 11.1% (1)<br>IV: 55.6% (5)<br>Unknown: 33.3% (3) |
- The results from the primary endpoint are displayed in Table 6. .
| Tumor type | Imaging-Only Session Metrics | | Fraction | Emulated<br>Delivery<br>bDVH% (>95) | Result |
|------------|------------------------------|-------|----------|-------------------------------------|--------|
| | AC kBq/ml | NTS | | | |
| Lung | 24.52 | 22.75 | First | 99.9 | Pass |
| Lung | | | Last | 100 | Pass |
| Lung | 5.6 | 2.76 | First | 94.52 | Fail |
| Lung | | | Last | * | * |
| Lung | 41.58 | 18.59 | First | 96.37 | Pass |
| Lung | | | Last | 100 | Pass |
| Lung | 5.43 | 2.97 | First | 100 | Pass |
| Lung | | | Last | 100 | Pass |
| Lung | 6.67 | 4.37 | First | 99.7 | Pass |
| Lung | | | Last | 97.6 | Pass |
| Bone | 13.15 | 7.03 | First | 100 | Pass |
| Bone | | | Last | 99.61 | Pass |
| Bone | 9.82 | 3.93 | First | 100 | Pass |
| Bone | | | Last | 100 | Pass |
| Bone | 8.67 | 4.42 | First | 99.6 | Pass |
{17}------------------------------------------------
| | | | Last | 100 | Pass |
|------|-------|------|-------|-----|------|
| Bone | 10.94 | 6.24 | First | 100 | Pass |
| | | | Last | 100 | Pass |
* Last fraction was not evaluable as the PET Evaluation did not pass.
- The results from the secondary endpoints are displayed in Table 7. .
| Table 7. Cohort II Secondary Endpoint Results |
|-----------------------------------------------|
|-----------------------------------------------|
| # | Description | Result |
|----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Percent of fractions where there was concordance<br>between the physical and digital phantoms of emulated<br>BgRT delivery derived from human participant PET<br>emissions. Concordance was defined as a standard<br>gamma index with a goal that 90% of pixels met the 3<br>mm/3% deviation standard. | 100% (18/18) |
| 2 | Percent of cases where there was agreement between an<br>SI and the AS for the BgRT PET PreScan localization<br>decision (overall percent agreement). | Overall % agreement = 72.2%<br>(13/18)<br>Positive % agreement = 80%<br>Negative % agreement = 33% |
| 3 | Percent of cases where there was concordance of a<br>positive localization decision between the short-duration<br>PET PreScan and a third-party diagnostic PET/CT scan<br>(positive percent agreement). | Positive % agreement = 100%<br>(7/7)<br>Overall % agreement = 100%<br>Negative % agreement = N/A |
| 4 | Safety of multiple FDG administrations and toxicity<br>rates of bladder and bone marrow. | The reported events were<br>documented by the investigator<br>as not being related to the study<br>device or FDG injection. None<br>resulted in discontinuation of the<br>subject in the study. |
| Sa | Workflow: Percent of PET imaging sessions at RRFD<br>that met the AC threshold for BgRT (5 kBq/mL) | 100% (18/18 fractions) |
| 5b | Workflow: Percent of PET imaging sessions which led<br>to acceptable BgRT plans. | 100% (9/9 cases) |
| 5c | Workflow: Percent of approved BgRT plans that went<br>on to pass physics quality assurance. | 100% (9/9 cases) |
| રવ | Workflow: Percent of PET valuations on the day of<br>fraction delivery that elicited a "Pass" signal | 94% (17/18 fractions) |
In addition to the analyses above, the sponsor also compared the BgRT and SBRT PTV . volumes (Table 8) to determine if there was a reduction in PTV by using the BgRT treatment paradigm.
{18}------------------------------------------------
| Target tumor<br>location | BgRT (PTV)<br>(cc) | SBRT (PTV or<br>IPTV) (cc) | Absolute<br>Volume<br>Reduction<br>with BgRT<br>(cc) | % Volume<br>Reduction<br>with BgRT |
|--------------------------|--------------------|----------------------------|------------------------------------------------------|------------------------------------|
| Bone | 23.18 | 42.42 | 19.24 | 45.36% |
| Bone | 17.00 | 17.40 | 0.40 | 2.30% |
| Bone* | 53.35 | 96.60 | 43.25 | 44.77% |
| Bone | 14.90 | 15.00 | 0.10 | 0.67% |
| Lung | 8.76 | 14.19 | 5.43 | 38.27% |
| Lung | 12.40 | 19.97 | 7.57 | 37.91% |
| Lung | 41.30 | 47.30 | 6.00 | 12.68% |
| Lung | 49.20 | 49.00 | -0.20 | -0.41% |
| Lung | 12.50 | 16.90 | 4.40 | 26.04% |
| | | | Average | 23.1% |
#### Table 8. Comparison of BgRT and SBRT volumes
* In accordance with the protocol, a single lesion with a circumferential margin was used for the BgRT plan 101-015. However, for the SBRT plan, the site clinicians opted to treat to a larger target volume to cover more of the pelvic bone housing the lesion based on their clinical judgment (as opposed to the strict circumferential margin applied to BgRT). As a result, the difference in PTV volumes is especially marked in this case (96.6 cc for SBRT v. 43.25 cc for BgRT). Also, two nearby lesions in the pelvis were also treated during the SBRT course for this patient, resulting in an aggregate PTV volume of 143.2 cc when the contributions from all 3 lesions were summed.
- The safety results from Cohort II are shown in Table 9. The Safety Population (i.e., all . participants who received any amount of FDG dose for the RefleXion Imaging-only session) included 9 participants for Cohort II.
| | Bone Tumors | Lung Tumors | Total |
|------------------------------------------------------|-------------|-------------|----------|
| System Organ Class | (N=4) | (N=5) | (N=9) |
| Preferred Term | n (%) | n (%) | n (%) |
| Any TEAEs | 3 (75.0) | 2 (40.0) | 5 (55.6) |
| Investigations | 2 (50.0) | 1 (20.0) | 3 (33.3) |
| Blood creatinine increased | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Lymphocyte count decreased | 0 (0.0) | 1 (20.0) | 1 (11.1) |
| Platelet count decreased | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| White blood cell count decreased | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| General disorders and administration site conditions | 2 (50.0) | 0 (0.0) | 2 (22.2) |
| Fatigue | 2 (50.0) | 0 (0.0) | 2 (22.2) |
| Musculoskeletal and connective tissue disorders | 1 (25.0) | 1 (20.0) | 2 (22.2) |
| Arthralgia | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Back pain | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Pain in extremity | 0 (0.0) | 1 (20.0) | 1 (11.1) |
| Blood and lymphatic system disorders | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Anaemia | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Renal and urinary disorders | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Pollakiuria | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Reproductive system and breast disorders | 1 (25.0) | 0 (0.0) | 1 (11.1) |
Table 9. Cohort II Safety Results
{19}------------------------------------------------
| Prostatic obstruction | 1 (25.0) | 0 (0.0) | 1 (11.1) |
|----------------------------------------|----------|---------|----------|
| Skin and subcutaneous tissue disorders | 1 (25.0) | 0 (0.0) | 1 (11.1) |
| Pain of skin | 1 (25.0) | 0 (0.0) | 1 (11.1) |
- For both Cohort I and Cohort II, the shifts applied during alignment of the PET scan with the . CT were calculated (Table 9). Overall. 5 out of 16 had PET alignment shifts (31.3%). The largest shift applied in any of the standard directions (lateral, anterior-posterior, or superiorinferior) was 4.7 mm. The largest total shift was 6.0 mm.
| Applied PET Shifts | | | | | |
|--------------------|-----------------------------|-----------------|------------------|-----------------|-----------------------------------|
| Cohort | Target<br>tumor<br>location | Lateral<br>(mm) | Ant-Post<br>(mm) | Sup-Inf<br>(mm) | Calculated<br>Total shift<br>(mm) |
| Cohort I | Bone | 0 | -3.0 | 2.0 | 3.6 |
| Cohort I | Lung | -3.4 | 2.3 | 1.7 | 4.4 |
| Cohort I | Lung | -0.9 | 3.6 | 4.7 | 6.0 |
| Cohort II | Lung | 3.0 | 0 | 1.1 | 3.2 |
| Cohort II | Lung | 1.0 | 0 | 3.0 | 3.2 |
#### Table 10. Cohort I and II PET alignment shifts
## Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# TRAINING
Training is required for the BgRT treatment paradigm. The sponsor provided a training program, which includes these key elements:
- . Determining patient eligibility
- Treatment planning .
- . Treatment delivery
# LABELING
The labeling consists of Instructions for Use and packaging labels. The Instructions for Use include the indications for use; a description of the device including an Imaging Agent section to describe the new use of fludeoxyglucose F18; contraindications, warnings, precautions; a detailed summary of the non-clinical and clinical data collected in support of the device: and instructions for the safe use of the device. The labeling satisfies the requirements of 21 CFR 801.109.
{20}------------------------------------------------
# RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the fludeoxyglucose F18-guided radiation therapy system and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device-specific modifications of<br>fludeoxyglucose F18 use compared to the<br>current approved drug label that affect safety<br>and effectiveness of fludeoxyglucose F18 | Clinical performance testing<br>Labeling<br>Analysis of drug and device label differences |
| Postmarket modifications to fludeoxyglucose<br>F18 labeling that affect safety and<br>effectiveness when used with the device | Design verification and validation activities |
| Inaccurate therapeutic radiation dose delivery<br>due to intra- or inter-fractional changes of<br>fludeoxyglucose F18 biodistribution | Non-clinical performance testing<br>Clinical performance testing<br>Labeling |
| Incompatibility of the linear accelerator and<br>the PET scanner leading to machine failures<…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.