DEN210037 · Soclean, Inc. · QXQ · Aug 12, 2024 · General Hospital
Device Facts
Record ID
DEN210037
Device Name
SoClean 3+ Bacterial Reduction Device
Applicant
Soclean, Inc.
Product Code
QXQ · General Hospital
Decision Date
Aug 12, 2024
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 880.6993
Device Class
Class 2
Indications for Use
The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (ATCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established. The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed ClimateLine Air (tubing), and SlimLine (tubing) for ResMed AirSense 10 CPAP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established.
Device Story
SoClean 3+ is an over-the-counter, home-use device designed to reduce bacterial populations on CPAP masks and hoses. It functions as an adjunct to standard cleaning procedures; it does not replace them. The device processes compatible respiratory accessories (e.g., ResMed Mirage FX, ClimateLine Air, SlimLine) through a complete cycle to achieve a 3-log (99.9%) reduction of specific bacteria. The device is operated by the patient in a home setting. It requires adherence to specific compatibility guidelines to ensure safety and efficacy. The device provides a secondary microbial reduction step, potentially benefiting the patient by lowering bacterial load on respiratory equipment, though clinical outcomes related to this reduction have not been established.
Clinical Evidence
No clinical data. Evidence is based on in vitro bench testing demonstrating a 3-log (99.9%) reduction of eight clinically relevant bacteria (S. aureus, K. aerogenes, S. haemolyticus, E. coli, S. hominis, K. pneumoniae, P. aeruginosa, S. pyogenes) after a complete processing cycle. Performance testing also confirmed the device produces the minimum effective dose of the microbicidal agent and that accessories remain functional after processing.
Technological Characteristics
Ozone generator, air pump, fan, humidifier, catalyst filter, microbial reduction chamber. UI uses physical buttons and LED indicators. Connectivity: BLE (disabled for end-users), UART (manufacturing only). Materials: tested for compatibility with ResMed Mirage FX, ClimateLine Air, SlimLine. Standards: IEC 60601-1, ISO 17510, ISO 80601-2-70, ISO 5356-1, CISPR11, IEC 61000 series.
Indications for Use
Indicated for single-patient home use as an adjunct to reduce bacterial populations on compatible CPAP masks and ventilation hoses after standard cleaning. Not for use as a replacement for manufacturer-recommended cleaning procedures.
Regulatory Classification
Identification
A respiratory accessory microbial reduction device is a home-use device intended to be used as an adjunct for microbial reduction of compatible respiratory accessories, such as hoses and masks, after cleaning. This device is not intended to replace the original accessory manufacturer's cleaning instructions.
Special Controls
In combination with the general controls of the FD&C Act, the respiratory accessory microbial reduction device is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR SOCLEAN 3+ BACTERIAL REDUCTION DEVICE
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Respiratory accessory microbial reduction device. A respiratory accessory microbial reduction device is a home-use device intended to be used as an adjunct for microbial reduction of compatible respiratory accessories, such as hoses and masks, after cleaning. This device is not intended to replace the original accessory manufacturer's cleaning instructions.
NEW REGULATION NUMBER: 21 CFR 880.6993
CLASSIFICATION: Class II
PRODUCT CODE: QXQ
### BACKGROUND
DEVICE NAME: SoClean 3+ Bacterial Reduction Device
SUBMISSION NUMBER: DEN210037
DATE DE NOVO RECEIVED: March 23, 2022
#### SPONSOR INFORMATION:
SoClean, Inc. 1 Vose Farm Road Peterborough, New Hampshire 03458
#### INDICATIONS FOR USE
The SoClean 3+ bacterial reduction device is indicated as follows:
The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (ATCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after
{1}------------------------------------------------
a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established.
The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed ClimateLine Air (tubing), and SlimLine (tubing) for ResMed AirSense 10 CPAP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established.
### DEVICE DESCRIPTION
SoClean 3+ is an ozone-generating device used as an adjunct for bacterial reduction of certain compatible continuous positive airway pressure (CPAP) mask and hoses after performing the cleaning steps as recommended per the procedures of the original mask and hose manufacturer.
To use the device, the CPAP Hose is disconnected from the CPAP machine and inserted into the Hose Port on the SoClean 3+. The other end of the CPAP Hose (near where the CPAP Masks attaches) is routed through the SoClean 3+ Hose Slot and the CPAP Mask is placed inside the SoClean 3+ Microbial Reduction Chamber. The lid of the SoClean 3+ is then closed, and a "Microbial Reduction Cycle" is initiated which generates and conveys humidified ozone through the CPAP Hose and CPAP Mask and into the Microbial Reduction Chamber. Prior to its exhaust, ozone generated from SoClean 3+ is converted to molecular oxygen via an Ozone Catalyst Filter located on the SoClean 3+ device. SoClean 3+'s general processing architecture is diagramed in Figure 1 below. and Figure 2 provides a graphical illustration of the device's use configuration. Figure 3 shows how CPAP Masks rest within the SoClean 3+ Microbial Reduction Chamber during the processing cycles.
{2}------------------------------------------------
Figure 1: SoClean 3+ System Function Diagram
Image /page/2/Picture/1 description: The image shows three different views of a medical device, possibly a CPAP machine, rendered in a technical drawing style. The top view shows the device with a connected hose, highlighting its interface and overall design. The bottom left view presents a compact, angled perspective of the device, focusing on its control panel and form factor. The bottom right view shows the device with the hose attached, emphasizing the connection point and the hose's flexibility.
(b)(4)
Figure 2: SoClean 3+ In Use Configuration
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image shows two SoClean 3+ units, one is a humidification unit. The image labels the ozone catalyst filter and PAP mask, which is inside a closed chamber. The image also labels the hose slot, pap hose, and hose port. The image is a diagram of the SoClean 3+ units and their components.
Figure 3: CPAP Mask Location During SoClean 3+ processing cycles
SoClean 3+ treatment is accomplished via a user-initiated and pre-programmed Microbial Reduction Cycle. During this cycle, SoClean 3+ produces ozone which is humidified in the SoClean 3+ Humidification Unit (Figure 4) and conveyed through the CPAP Hose to the CPAP Mask and into the SoClean 3+ Microbial Reduction Chamber. A strong oxidant, ozone reduces bacteria on CPAP Mask and Hose surfaces by reacting with and damaging cell membranes and other biomolecules. This damage can occur either via a direct reaction between ozone and/or through reactions with the radical by-products of ozone decomposition.
Image /page/3/Picture/3 description: The image shows a gray rectangle with a red border. At the top of the rectangle, the text "Humidification Unit" is visible. Below this text, the text "(b)(4)" is visible.
Figure 4: SoClean 3+ Humidification Unit
{4}------------------------------------------------
The SoClean 3+ Microbial Reduction Cvcle takes approximately 90 minutes to complete and encompasses the following stepwise phases:
- Step 1 Buildup Phase (builds up ozone levels in the system. 6-minute duration). .
- Step 2 Pulsing Phase (pulses ozone into the system for 10-second periods interspersed . with 65-second dwell periods; 72-minute duration),
- . Step 3 - Purge Phase (pumps fresh air without ozone through the system to remove residual ozone; 12-minute duration).
To initiate a Microbial Reduction Cycle, the following conditions must be met:
- A CPAP Hose must be inserted and detected by the SoClean 3+ (2 places: connection to . the Hose Port and the Hose Slot).
- A valid Ozone Catalyst Filter must be inserted and detected, .
- The Filter Door must be closed, and .
- The SoClean 3+ Microbial Reduction Chamber lid must be closed. .
Once these conditions are satisfied, a Microbial Reduction Cycle can be initiated via the User Interface (UI) by pressing the Microbial Reduction Cycle Button. Alternatively, the UI can allow for a Microbial Reduction Cycle to be started following a 2, 4, or 8-hour delay. When the Microbial Reduction Cycle is successfully completed, the Microbial Reduction level icon will light up green. The Microbial Reduction Level Icon remains illuminated until the lid is opened. If the Microbial Reduction Level Icon is orange, the system has detected either ozone concentrations outside of the specified limits or humidity levels in the bucket below a specified value. The ozone and relative humidity (RH) specifications for SoClean 3+ are 270 ± 15 ppm and > 70%, respectively.
SoClean 3+ can also perform a Fresh Air Cycle, where only the Air Pump and the Exhaust Fan operate for 5 minutes to move fresh air through the CPAP airway. This freshens the air within the CPAP airway before CPAP use. Ozone is not generated during the Fresh Air Cycle. No bactericidal efficacy claims, or other indications are made for the Fresh Air Cycle. Preprogrammed settings allow for the running of only 3 Fresh Air Cycles before a Microbial Reduction Cycle must be run, and LED indicators allow user to distinguish between whether a Microbial Reduction Cycle or a Fresh Air Cycle was last run.
SoClean 3+ includes an Ultraviolet (UV) ozone sensor that monitors ozone concentrations, providing feedback to the Ozone Generator itself for ozone production adjustment and to the SoClean 3+ user via LED lights on the UI when measured ozone concentrations are above or below the prespecified 270 ± 15 ppm range.
Based on in vitro testing. SoClean 3+ demonstrated a 3-log bacterial reduction on the following bacteria: S. aureus ATCC 6538. K. aerogenes ATCC 13048. S. haemolyticus ATCC 29970. E. coli ATCC 11229, S. hominis ATCC 27844, K. pneumoniae ATCC 4352, P. aeruginosa ATCC 15442, and S. pyogenes ATCC 14289. Correlation between in vitro results and any clinical outcome has not been established.
The above in vitro 3-log bacterial reduction was established. according to the laboratory condition, on test article (Climateline Air Hose, SlimLine Hose, and Mirage FX Nasal Mask)
{5}------------------------------------------------
that had either been conditioned (have undergone five (5) SoClean 3+ ozone pre-treatment cycles) or conditioned at the End of Useful Life (EOUL) (have undergone no less than 183 SoClean 3+ ozone pre-treatment cycles).
SoClean 3+ is intended for home use by CPAP user or lay person. The intended use environments include an electrical outlet in a well-ventilated room. Once the microbial reduction cycle has been initiated, the users should leave the room for the cycle, which takes 90 minutes to complete.
## SUMMARY OF BENCH STUDIES
### BACTERIAL REDUCTION AND SHELF LIFE
The SoClean 3+ bacterial reduction device is not intended for cleaning, disinfection, sanitization, or sterilization: it is intended to be used as an adjunct for bacterial reduction of certain compatible home use CPAP masks and ventilation hoses after the original CPAP accessory manufacturer's recommended cleaning procedures are performed. Performance related to bacterial reduction was reviewed as a part of the performance testing. The device includes instructions for bacterial reduction of CPAP accessories. Specifically, the sponsor has included bacterial reduction instructions that SoClean 3+ can be used with one specific ResMed CPAP Mask (Mirage FX) and two specific ResMed CPAP Hoses (ClimateLine Air, SlimLine).
The storage stability study is provided to support:
- a six-month shelf life of the product from the date of manufacture of the (i) unopened product; and
- (ii) an in-use period of two months of the opened container from the date of opening under the storage/use conditions at 10-32 °C and 15-65 % relative humidity (RH).
#### COMPATIBILITY
CPAP Mask and Hose compatibility testing was conducted in accordance with the following standard scientific practices with respect to design, documentation, etc.
- · IEC 60601-1:2005+A1:2012+A2:2020. Medical electrical equipment Part 1: General requirements for basic safety and essential performance, Clause 11.1.1
- ISO 17510:2015. Medical devices -- Sleep apnea breathing therapy -- Masks and o application accessories, Annex B, Annex C, Annex F, Annex G
- o ISO 80601-2-70:2020
- Medical Electrical Equipment Part 2-70: Particular requirements for basic safety o and essential performance of sleep apnea breathing therapy equipment. Clause 201.11.1.2.2
- · ISO 5356-1, Anesthetic and respiratory equipment -- Conical connectors -- Part 1: Cones and sockets
The results from compatibility testing, which also included visual and mechanical evaluation, show that using the SoClean 3+ as the bacterial reduction device for
{6}------------------------------------------------
processing ResMed CPAP Mask (Mirage FX) and ResMed CPAP Hoses (ClimateLine Air and SlimLine) for up to 6 months (183 days at 1 bacterial reduction cycle/day) does not adversely affect their safety or functionality. However, the compatible CPAP Mask and Hoses may still undergo limited performance change (due to memory foam, natural rubber and/or nylon degradation), and discoloration (yellowing on silicone portions of mask) by the end of their use life.
## BIOCOMPATIBILITY
The SoClean 3+ does not directly contact the patient. However, SoClean 3+ processed PAP accessories contact patients via the skin (PAP Masks) and through gas pathway (PAP Masks and Ventilation Hoses). Therefore, SoClean 3+ is categorized as both a surface device and an externally communicating device with tissue contact and permanent contact duration (>30 days), in accordance with Guidance for Industry and Food and Drug Administration Staff -Use of International Standard ISO10993-1. 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'.
The sponsor provided the following studies to assess the potential toxicity of SoClean 3+ processed PAP accessories to a user.
- Cytotoxicity, irritation/intracutaneous reactivity, and sensitization on CPAP . accessories after conditioning with three back-to-back cycles and conditioning to EOUL. The test article extracts met the requirements of the tests and showed no evidence of cytotoxicity, irritation, or sensitization under the testing condition.
- . Volatile Organic Compounds (VOCs) and Particulate Matter (PM) tests on CPAP accessories after conditioning with three (3) back-to-back cycles and conditioning to EOUL. The test article extracts met the requirements of the tests and the acceptance criteria of ISO 18562-2 Test for emission of particulate matter and ISO 18562-3 Test for emission of volatile organic compounds were met.
The sponsor provided exhaustive chemical extractable and leachable testing for Trace Metals, Volatile Organic Compounds (VOC). Semi-Volatile Organic Compounds (SVOC), and Non-Volatile Organic Compounds (NVOC). Toxicological risk assessment for systemic toxicity, genotoxicity, and carcinogenicity of the chemical extractables and leachables showed that the probable risk for patient exposure to SoClean 3+ processed CPAP accessories is low.
The sponsor also provided testing to demonstrate that the amount of ozone released into the environment during processing and remaining on the accessories after processing are below 0.05 ppm as specified in 21 CFR 801.415.
## SOFTWARE AND CYBERSECURITY
The SoClean 3+ device contains an air pump, fan, humidifier, ozone generator, catalyst filter, microbial reduction chamber, humidification unit, and a User Interface (UI) made up primarily of physical push buttons and Light-Emitting Diode (LED) indicators lights.
{7}------------------------------------------------
The device has physical buttons and LEDs to implement the user interface, as well as a progress bar, consisting of a sequence of LEDs.
Certain SoClean 3+ functionality is controlled by firmware.
- The internal ozone generator. .
- . Signals from optical sensors on the device.
- . Logging to a memory chip to store device logs for major system events.
- Capable of communicating with Bluetooth Low Energy (hereafter BLE). Note . that BLE functionality is specifically designed to be utilized by the manufacturer and is disabled for end-users.
- . Capable of communicating via a Command-Line Interface (CLI) over a serial (UART) interface during manufacturing and testing.
The sponsor provided the following assessments, demonstrating that the probable risk is low:
- . Level of concern as "Major"
- . Device hazard analysis
- Software requirements specification (SRS) .
- . Architecture design chart
- Software design document (SDS) .
- . Traceability analysis
- . Software Development Environment Description
- . Verification and Validation documentation (V&V)
- 0 Revision level history
- . Unresolved anomalies
- . Cybersecurity
## ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
The EMC and Electrical Safety was evaluated to mitigate the risk of electrical fault resulting in injury to patient or user. The following Electrical/ Mechanical/Thermal Safety, and electromagnetic compatibility (EMC) testing has been performed:
- Radiated Emissions: CISPR11 ed6.2 (2015 +A1:2016 +A2:2019), FCC 47CFR . Part 15 Subpart B (2023-08)
- AC Mains Conducted Emissions: CISPR11 ed6.2 (2015 +A1:2016 +A2:2019) .
- Harmonics: IEC 61000-3-2 ed3.2 (2005 +A1:2008 +A2:2009) .
- Flicker: IEC 61000-3-3 ed3.0 (2013-05) .
- . Electro-Static Discharge Immunity Test: IEC 61000-4-2 ed2.0 (2008-12), CENELEC EN 61000-4-2 (2009)
- Radiated. Radio-Frequency, Electromagnetic Immunity: IEC 61000-4-3 ed3.0 . (with A1:2007+A2:2010), CENELEC EN 61000-4-3 (2006), A1 (2008) and A2 (2010)
- Immunity to Proximity Fields from FR Wireless Communications Equipment .
- Electrical Fast Transient/Burst Immunity Test: IEC 61000-4-4 ed3.0 (2012-04), . CENELEC EN 61000-4-4 (2012)
{8}------------------------------------------------
- Immunity to Surges: IEC 61000-4-5 ed3.1 (2014 +A1:2017), CENELEC EN . 61000-4-5 (2014)
- . Radio-Frequency, Electromagnetic Immunity Test: IEC 61000-4-6 ed4.0 (2013). CENELEC EN 61000-4-6 (2014)
- Power Frequency Magnetic Field Immunity Test: IEC 61000-4-8 ed2.0 (2009-09) .
- Voltage Dips/Interruptions Immunity Test: IEC 61000-4-11:2014 +A1:2017. . CENELEC EN 61000-4-11 (2004)
- . Immunity to Proximity Magnetic Fields in the Frequency Range of 9 kHz to 13.56 MHz IEC 61000-4-39 ed1.0 (2017-03)
All testing and results were adequate and met the above standards requirements.
## HUMAN FACTORS
A simulated-use human factors (HF) study was provided in accordance with FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices." It included participants representative of the positive airway pressure (PAP) patient users, and caregivers who may facilitate set-up and daily use of the device on behalf of the PAP patient user. During the session, participants (16 patient users and 15 caregivers) were provided with SoClean 3+ that were targeted in this validation study. Specifically, participants first engaged in simulated first-time use of the SoClean 3+. divided between a setup scenario (first 13 study tasks) and a use scenario (next 7 study tasks), and then by knowledge tasks (last 14 tasks). Following the session, participants answered questions and provided feedback regarding use-related issues and the product itself. The study included 34 assessments which evaluated the critical tasks for using the product and found that users understood.
The provided human factors validation testing results, analyses, and conclusions established that SoClean 3+ is safe and effective for the intended users, its intended uses, and use environments.
# PERFORMANCE TESTING - BENCH
The sponsor conducted the following performance conditions and tests to support that the device can achieve its intended use:
- A six-month shelf life (unopened) and two-month in-use life have been . demonstrated.
- . The test article from compatible ResMed Mirage FX mask and ResMed ClimateLine and SlimLine (ventilation hoses) had either been conditioned (have undergone five (5) SoClean 3+ ozone pre-treatment cycles) or conditioned until End of Useful Life (EOUL, have undergone >= 183 SoClean 3 + ozone pretreatment cycles). Each test configuration contained three replicates.
- . The tested microorganisms included the following eight bacteria: Staphylococcus aureus ATCC 6538, Staphylococcus haemolyticus ATCC 29970, Pseudomonas aeruginosa ATCC 15442, Staphylococcus hominis ATCC 27844, Streptococcus pyogenes ATCC 14289, Klebsiella pneumoniae ATCC 4352, Klebsiella
{9}------------------------------------------------
aerogenes ATCC 13048, and Escherichia coli ATCC 11229. The target inoculum amount was equal to or above 1 x 106 CFU per device. Appropriate extraction control, recovery factor, positive control, negative control, and neutralization control have been included for each test configuration.
- As a worse-case condition, all SoClean 3+ units utilized for testing were . characterized to produce ozone at or below the minimum specification of 255 ppm over a shorter Microbial Reduction Cycle time (70 minutes). A 3-log (99.9%) bacterial reduction for all tested bacteria has been demonstrated on the above conditions.
- . The firmware has been modified, the Design Verification Evaluation and Testing Support reports are able to demonstrate that the ozone sensor and feedback loop. (a) consistently generates ozone within the required concentration specification (270 ± 15 ppm), and (b) accurately informs users when ozone concentrations are outside of this range.
# LABELING
The labeling consists of a user guide and package labels. The user guide contains a description of the device (including the mechanism of action of bacterial reduction; in vitro bacterial logreduction test results), contraindications, warnings, precautions for processing three compatible CPAP accessories |SoClean 3+ has been tested for use with ResMed Mirage FX (mask), ResMed ClimateLine Air (tubing), and SlimLine (tubing) for ResMed AirSense 10], and information on shelf-life, use-life, maintenance, storage, transport conditions, and disposal. The user guide includes a statement that SoClean 3+ should not be used as a replacement for routine cleaning as recommended by the CPAP accessory manufacturer.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a respiratory accessory microbial reduction device:
| Risks to Health | Mitigation Measures |
|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Respiratory tract infection due to<br>device failure leading to inadequate<br>microbial reduction | Non-clinical performance testing<br>Software verification, validation, and hazard analysis<br>Shelf life validation<br>In-use life validation<br>Labeling |
| Damage to respiratory accessories<br>resulting in reduced performance of<br>respiratory accessories | Non-clinical performance testing<br>Labeling |
| Adverse events, e.g., respiratory<br>mucous membrane, pulmonary, or skin<br>irritation, due to chemical exposure | Non-clinical performance testing<br>Biocompatibility evaluation<br>Human factors/usability evaluation<br>Labeling |
#### Identified Risks to Health and Mitigation Measures
{10}------------------------------------------------
| Respiratory tract infection due to user<br>error leading to inadequate cleaning of<br>accessory device | Human factors/usability evaluation<br>Labeling |
|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Injury from electrical, mechanical, or<br>thermal hazards | Electrical safety testing<br>Electromagnetic compatibility testing<br>Mechanical and thermal safety testing<br>Software verification, validation, and hazard analysis<br>Labeling |
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the respiratory accessory microbial reduction device is subject to the following special controls:
- (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- Microbial reduction testing must demonstrate at least 2 log or greater reduction (i) of the indicated, clinically relevant microorganisms inoculated on the intended respiratory device accessories after the labeled processing (exposure) time;
- (ii) Performance testing must demonstrate that the device can produce the minimum effective dose or concentration of the microbicidal agent that is necessary to achieve the intended 2 log or greater microbial reduction;
- Device intercompatibility testing must demonstrate that the labeled respiratory (iii) device accessories (e.g., masks and hoses) and materials function as intended after worst-case processing with the microbial reduction device: and
- (iv) Performance testing must demonstrate that any release of the microbicidal agent into the use environment is within safe limits for human exposure.
- (2) The respiratory device accessories intended to be processed with the microbial reduction device must be demonstrated to be biocompatible.
- (3) Performance data must support the shelf life of the device by demonstrating continued device functionality over the labeled shelf life.
- (4) Performance data must support the in-use life of the device by demonstrating continued device functionality over the labeled in-use life.
- Usability/human factors evaluation must demonstrate that intended users can correctly (5) use the device, based solely on reading the instructions for use.
- (6) Software verification, validation, and hazard analysis must be performed.
- Performance data must demonstrate the electromagnetic compatibility (EMC) and (7) electrical safety of the device.
- (8) Performance data must demonstrate thermal and mechanical safety of the device.
- (9) Labeling must include:
- Directions for use including: (i)
- (A) A statement that the device is not intended to replace the respiratory accessory device manufacturer's recommended cleaning procedures;
- (B) Instructions for how to determine the minimum effective dose or concentration is achieved to ensure the device is functioning as intended:
{11}------------------------------------------------
- (C) Instructions for how to detect and respond to electrical, mechanical, thermal, or chemical hazardous operating conditions;
- (D) Instructions for periodic maintenance and cleaning/reprocessing of the microbial reduction device;
- (ii) An expiration date or shelf life:
- Use life information; (iii)
- Information on respiratory device accessories or materials that can and cannot be (iv) used with the microbial reduction device. This information must also be disclosed in the package label; and
- (v) Safety information regarding exposure to hazardous chemicals or compounds, process residues, or degraded materials from the use of the microbial reduction. device and related adverse reactions.
# BENEFIT-RISK DETERMINATION
## Benefit
The bench studies performed support bacterial reduction using the SoClean 3+ device for processing respiratory accessory devices by demonstrating that bacteria on the washed respiratory accessory devices were reduced by at least 3-log on eight (8) different pure-cultured vegetative microorganisms. The subject bacterial reduction device, an ozone-generating device with 72-minute processing time, is intended to be used by patient user and/or caregiver in a home environment. The probable benefit of SoClean 3+ is to serve as an adjunct for bacterial reduction of certain compatible CPAP respiratory accessories, on the condition that the manufacturer's recommended cleaning procedures were performed before the use of SoClean 3+ device. Limitations of the testing include that the bacterial reduction of SoClean 3+ device must be in addition to CPAP accessory manufacturer's cleaning procedures. The clinical benefit of added bacterial reduction from the device use after CPAP accessory manufacturer's cleaning procedures is unknown. No clinical testing was performed and as such, the in vitro results should not be correlated with any clinical outcome.
# Risk
The identified risks include but are not limited to inadequate microbial reduction causing patient respiratory tract infection. CPAP accessory incompatible with microbial reduction device reducing CPAP therapy benefits, adverse events caused by exposure to microbicidal agent, residues, condensate, and/or degraded materials, and inappropriate labeling causing use/user errors. These risks can be mitigated with appropriate performance testing, material compatibility testing, biocompatibility testing and toxicological risk assessment, as well as human factors studies and by the special controls and labeling.
## Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
{12}------------------------------------------------
### Benefit/Risk Conclusion
In conclusion, given the available information above, the following indications for use statement is supported:
The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (ATCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established.
The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed ClimateLine Air (tubing), and SlimLine (tubing) for ResMed AirSense 10 CPAP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established.
The probable benefits outweigh the probable risks for the SoClean 3+ bacterial reduction device. The device provides benefits, and the risks can be mitigated using general controls and the identified special controls.
## CONCLUSION
The De Novo request for the SoClean 3+ Bacterial Reduction Device is granted and the device is classified as follows:
Product Code: OXQ Device Type: Respiratory accessory microbial reduction device Regulation Number: 21 CFR 880.6993 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.