AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN210015 · Jan 7, 2022
Sunrise Sleep Disorder Diagnostic Aid
Sunrise
Retrospective clinical study datasets (mandibular movement signals recorded via head position sensor)
Retrospective clinical data were used to evaluate the performance of the Sunrise SDDA device algorithms for sleep structure parameters and OSA severity compared to in-lab polysomnography (PSG).
Retrospective, comparative and open study (Belgium); Retrospective, comparative and open study
Patients 18 years and older; Number of Sites: 1
In-lab polysomnography (Somnoscreen, Somnomedics)
Sleep structure parameters (TST, SOL, WASO, SE, ArI)
Independent clinical study (Belgium); Retrospective, comparative and open study
Patients 18 years and older; Number of Sites: 1
In-lab polysomnography (Somnoscreen, Somnomedics)
OSA severity, sleep structure parameters (TST, SOL, WASO, SE, ArI)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Obstructive Sleep Apnea (OSA) severity
Machine Learning algorithms
—
Sensitivity and specificity at three different ORDI cut-offs
Retrospective study in Belgium: mandibular movement signals recorded using a head position sensor.
>1 (experienced sleep technicians)
Prospective study in France: ambulatory at-home PSG system (NOX A1, ResMed). Independent retrospective study in Belgium: in-lab PSG (Somnoscreen, Somnomedics).
>1 (experienced sleep technicians)
Sleep structure parameters
Machine Learning algorithms
—
RMSE values for TST, SOL, WASO, SE, and ArI
Retrospective study in Belgium: mandibular movement signals recorded using a head position sensor.
>1 (experienced sleep technicians)
Prospective study in France: ambulatory at-home PSG system (NOX A1, ResMed). Independent retrospective study in Belgium: in-lab PSG (Somnoscreen, Somnomedics).
>1 (experienced sleep technicians)
Indications for Use
The Sunrise SDDA device is a non-invasive home care aid in the evaluation of obstructive sleep apnea (OSA) in patients 18 years and older with suspicions of sleep breathing disorders.
Device Story
Sunrise Sleep Disorder Diagnostic Aid (SDDA) is a non-invasive home-use device for evaluating obstructive sleep apnea (OSA). It functions by analyzing sensor readings of mandibular movement during sleep. The device captures movement data, which is processed to assist in the evaluation of sleep breathing disorders. It is not intended as a substitute for full polysomnography or as an apnea monitor. The output is provided to healthcare providers to aid in clinical decision-making regarding OSA. The device is intended for use in a home environment by patients suspected of having sleep breathing disorders.
Clinical Evidence
Three clinical studies (one retrospective, one prospective, one retrospective) compared Sunrise SDDA to gold-standard in-lab or ambulatory polysomnography (PSG). Studies included adults 18+ with varying BMIs and neck circumferences. Performance metrics included OSA severity classification (sensitivity/specificity) and sleep structure parameters (RMSE). Results demonstrated clinically acceptable agreement with PSG. No adverse events or device deficiencies were reported. The device is not a substitute for PSG.
Technological Characteristics
Sensor: Inertial measurement unit (accelerometer and gyroscope). Materials: 3M SC Spunlace Extended Wear Nonwoven Tape and 3M Tegaderm film 16004 (ISO 10993 compliant). Connectivity: Mobile app to cloud-based analysis. Power: Battery-operated. Form factor: Chin-mounted, single-use, non-sterile. Software: Machine learning-based analysis. Safety: Compliant with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients 18 years and older with suspected sleep breathing disorders to aid in the evaluation of obstructive sleep apnea (OSA).
Regulatory Classification
Identification
A device for sleep apnea testing based on mandibular movement is a prescription device intended to aid in evaluation of sleep apnea during sleep in patients suspected of having sleep breathing disorders by analyzing sensor readings of mandibular movement. The device is not intended as a substitute for full polysomnography nor intended to be used as an apnea monitor.
Submission Summary (Full Text)
{0}------------------------------------------------
# DE NOVO CLASSIFICATION REQUEST FOR SUNRISE SLEEP DISORDERS DIAGNOSTIC AID
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Device for sleep apnea testing based on mandibular movement. A device for sleep apnea testing based on mandibular movement is a prescription device intended to aid in evaluation of sleep apnea during sleep in patients suspected of having sleep breathing disorders by analyzing sensor readings of mandibular movement. The device is not intended as a substitute for full polysomnography nor intended to be used as an apnea monitor.
NEW REGULATION NUMBER: 21 CFR 868.2376
CLASSIFICATION: Class II
PRODUCT CODE: QRS
#### BACKGROUND
DEVICE NAME: Sunrise Sleep Disorder Diagnostic Aid (SDDA)
SUBMISSION NUMBER: DEN210015
DATE DE NOVO RECEIVED: April 2, 2021
#### SPONSOR INFORMATION:
Sunrise SA Chaussée de Marche 598/02 5101 Namur BELGIUM
#### INDICATIONS FOR USE
The Sunrise SDDA device is a non-invasive home care aid in the evaluation of obstructive sleep apnea (OSA) in patients 18 years and older with suspicions of sleep breathing disorders.
#### LIMITATIONS
The sale, distribution, and use of the Sunrise SDDA are restricted to prescription use in accordance with 21 CFR 801.109.
Sunrise sensor is only to be placed on the chin. If the sensor is not placed on the chin, it will affect the accuracy of test results.
{1}------------------------------------------------
Patients suffering from conditions affecting the rotation of the condyle in the temporomandibular joint cannot use the Sunrise SDDA since it is not compatible with detection of mandibular movements.
The device is not intended as a substitute for full polysomnography (PSG) nor intended to be used as an apnea monitor.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The Sunrise SDDA device consists of a Sunrise sensor and a cloud-based software device that analyzes data from the sensor when placed on the patient's mandible. The device also includes a mobile application to record patient's responses to questions about their sleep quality and transfer sensor data to the cloud. By analyzing patient's mandibular movements, the device also detects obstructive respiratory disturbances, identifies sleep states, notifies about the Obstructive Sleep Apnea (OSA) severity in a categorical format (non-OSA, mild-OSA, moderate-OSA, severe-OSA), generates sleep structure information (namely, total sleep time, sleep onset latency, wake after sleep onset, sleep efficiency, arousal index) and head position discrete states. Data collected by the device is integrated in a report for further interpretation and diagnosis by the healthcare provider.
Image /page/1/Picture/5 description: The image shows a woman with a white sensor placed on her mandible. The woman is wearing a red turtleneck sweater. The background is a light purple color. The text at the bottom of the image says "Figure 1 Sunrise Sensor Placed on Mandible".
Figure 1 Sunrise Sensor Placed on Mandible
## SUMMARY OF NONCLINICAL/BENCH STUDIES
#### BIOCOMPATIBILITY / MATERIALS
The Sunrise SDDA sensor includes two patient-contacting components which are in contact with the skin of the patient during the test overnight: skin adhesive (3M SC Spunlace Extended Wear Nonwoven Tape) and film dressing (3M Tegaderm film 16004) which are classified as skin-surface-devices with ≤ 30 days contact duration ('prolonged exposure'). Biocompatibility evaluation has been completed according to FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
{2}------------------------------------------------
(https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guidan ceDocuments/UCM348890.pdf). The test articles were found non-cytotoxic per ISO 10993-5. The test articles were found to be non-sensitizing in tests for skin sensitization per ISO 10993-10. The test articles were found to be non-irritating in tests for irritation per ISO 10993-10.
All other components of the device system are not direct patient-contacting.
# PACKAGING AND SHELF LIFE
Testing was conducted to evaluate the packaging and labeling of the Sunrise SDDA sensor to withstand anticipated shipping conditions while preserving the functionality of the device. This testing included freefall drop testing from | | resistance to rain and humidity, and label integrity evaluations. Each of these tests demonstrated that the packaging provides appropriate protection for the Sunrise SDDA sensor device, ensuring it will remain functional with legible labeling following distribution.
The Sunrise SDDA sensor is a non-sterile and single-use device intended to be placed on the chin. The shelf-life of the device is 2 years. This shelf-life was determined based on the shelf-life of the skin adhesive and specific storage conditions defined by the supplier.
## ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY
The following Electrical/Mechanical/Thermal Safety and Electromagnetic Compatibility (EMC) testing has been performed and results support electrical safety and electromagnetic compatibility:
- IEC 60601-1:2005+ CORR. 1:2006 + CORR. 2:2007 + A1:2012. General . requirements for basic safety and essential performance for medical electrical equipment
- . IEC 60601-1-2:2014. Medical electrical equipment Part 1-2 - General requirements for basic safety and essential performance - Electromagnetic compatibility.
## SOFTWARE
The Sunrise SDDA device has a Moderate Level of Concern (LOC) software component, Appropriate documentation was provided to support the validation of the software for a Moderate LOC in accordance with FDA's 2005 guidance titled. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The sponsor also provided details on how the Machine Learning algorithms detect sleep disordered breathing events to estimate output parameters. The sponsor also demonstrated an overview of algorithm workflow, software handling of hardware error or failures and a summary of algorithms development and training steps.
{3}------------------------------------------------
Appropriate documentation was provided in accordance with FDA's 2014 guidance document titled "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" to support adequate cvbersecurity measures have been taken and will be monitored and updated throughout the device life cycle.
#### PERFORMANCE TESTING - BENCH
The majority of performance testing for the Sunrise SDDA device are encapsulated in software section above and through validation of the overall system in clinical studies. Additionally, the sponsor performed bench testing to demonstrate equivalent technical performance between the Sunrise sensor and a head position sensor that was used in two retrospective clinical validation studies. Both head position and Sunrise sensors use the exact same inertial measurement unit and therefore, technical specifications including resolution, sensitivity, output data rate of accelerometer and gyroscope sensors are the same. To demonstrate dynamic performance equivalency between two sensors, a bench set up was prepared using a dummy head and a rotation disk. Sunrise and head position sensors were attached to the dummy head and movement data were recorded at different disk rotation speeds and directions from the chin area. The measured signals of the accelerometer and of the gyroscope between the two sensors were compared and dynamic performance found to be equivalent. Therefore, the retrospective clinical data collected with the head position sensor can used for performance evaluation of the Sunrise SDDA device.
## HUMAN FACTORS/USABILITY
A Formative usability engineering testing in accordance with IEC 62366-1: 2015, "Medical devices - Part 1: Application of usability engineering to medical devices" was conducted in Belgium using adult users. [of which were tech savvy, and who were not considered to be tech savvy. The report describes the risk-based approach to assessment of use related risks. Based on the assessment, three risks were identified related to use of the Sunrise SDDA device: false negative, choking, and no result. Each user was required to complete distinct tasks for use of the device using the provided Instructions for Use. The majority of participants in both user groups completed all tasks correctly, providing strong evidence for the appropriate design of the use interface and proposed labeling for the Sunrise SDDA device. None of these tasks were considered critical tasks, because Sunrise SDDA device is a prescription use device for aid in diagnosis of OSA and even a false negative result may delay the final diagnosis by few weeks and is not associated with a user task which if performed incorrectly or not performed could result in serious harm or serious compromise in medical care. The sponsor also stated that Sunrise SDDA device has previously been placed on the European market and no issues have been reported. The outcome of the usability study was assessed as satisfactory, because the testing provided adequate assurance that all tasks linked to a safety mitigation could be successfully performed by more than 1% of participants.
## SUMMARY OF CLINICAL INFORMATION
{4}------------------------------------------------
The sponsor provided three clinical study protocols and reports to support the safety and effectiveness of the Sunrise SDDA device. The primary objective of all clinical studies was to compare the agreement of sleep parameters between Sunrise SDDA device and the gold-standard polysomnography (PSG). No adverse event, adverse device effects or device deficiencies were reported in conducted clinical studies. Patient recruitment of these studies was set up to ensure collection on a wide range of ages (18 years and older), sleep disordered breathing problems, body mass indices, and neck circumferences comparable to the US intended patient population.
The first clinical study was a retrospective, comparative and open study performed in Belgium in patients. The mandibular movement signals were recorded using a head position sensor and its equivalency to the Sunrise sensor was demonstrated using bench testing. The Sunrise Machine Learning algorithms were used to analyze sensor data and evaluate the performance of the device only for sleep structure parameters compared to in-lab PSG (i.e., Somnoscreen, Somnomedics), The PSG data were visually scored by an experienced sleep technician that was blind to the study protocol and hypothesis according to the 2012 American Academy of Sleep Medicine (AASM) recommendations. The performance of the device for sleep structure parameters was calculated as the root mean square error (RMSELin a statistical non-inferiority test across all measurements when compared to the PSG. [10]4] confidence intervals (Cls) were calculated for these measurements using the bootstrap method. The RMSEs were found to be | | (0) 0) (DX4) (b(4) 10/4) (p) ch (D)(4) (CI (CI = 1 (014) and [b]4] events/h (CI [6] - [107(0) |events/h) for TST, SOL, WASO, SE and Arl, respectively. This clinical study, on its own, was not sufficient to demonstrate a reasonable assurance of safety and effectiveness use of the Sunrise SDDA device as the same datasets were used for both optimizing diagnostic thresholds (training) and performance evaluation (validation)
The second study was a prospective, comparative and open clinical study independent from the first one and performed in France in patients. The final Sunrise sensor was used to record mandibular movements from patients. The clinical report focused on evaluating the performance of the device in terms of all output parameters (i.e., OSA severity and sleep structure measurements) compared to data recorded by an ambulatory at-home PSG system (i.e., NOX A1. ResMed). The PSG data were visually scored by experienced sleep technicians from two different sleep centers (Université Grenoble Alpes, Grenoble, France and Imperial College London, London. United Kingdom) according to the 2012 AASM recommendations. The performance of the device for OSA severity output was calculated as standard diagnostic metrics compared to the PSG at three different ORDI cut offs as following: sensitivity off (DIGI) and specificity and ORDI> ORDI> ORDI> and ORDI> Tor respectively. For sleep structure of 015 CI 007 -(b)(4) (D(4) parameters, the RMSEs were found to be | 0014) (CI (b)Gr (D)(E) (b)(4D) (CI DK-) (b)(4) 1. 19% (CIBK4) - (b)(4) and [ 0)4 levents/h (CI events/h) for TST, SOL, WASO, SE and ArI, respectively. (5) (1)
The third study was an independent clinical study with similar design as the first one, and was a retrospective, comparative and open study performed in Belgium in PM patients. The head position sensor was used to collect mandibular movement signals and the Sunrise SDDA algorithms analyzed sensor data and evaluated the performance of the device compared to in-lab PSG (Somnoscreen, Somnomedics). The PSG data were visually scored by an experienced sleep
of OSA severity.
{5}------------------------------------------------
technician that was blinded to the study protocol and hypothesis according to the 2012 AASM recommendations. The performance of the device for OSA severity output was calculated similarly to the second study as follows: sensitivity of (6)(4) and specificity of (b)(4) and bit for ORDI>= ORDI>= and ORDI>= or respectively. For sleep structure parameters, the RMSEs were found to be (6)(4) (C) 1074 -(15)(4) (D)(4) (CI (D)(4) -10 (4) (CI Did) and (0)(4) (b)[4) events/h (C) events/h) for TST, SOL, WASO, SE and ArI, respectively.
# Pediatric Extrapolation
The Sunrise SDDA device is indicated for patients aged 18 and older. For medical devices, the FD&C Act defines patients before their 22nd birthday as pediatric patients. In this De Novo request, data from patients between 18-22 were used to support the use of the device in patients over the age of 18. It was appropriate to indicate the device for individuals 18 and older because of this data and patients aged 18 to 21 do not carry additional differences or risks relative to the patient population studied.
# LABELING
The labeling meets the requirements of 21 CFR 801.109 for prescription devices and includes a description of what the device measures and outputs to the user. A summary of the clinical validation data provides expected performance for intended use populations and environments. As certain conditions affecting the rotation of the condyle in the temporo-mandibular joint may confound accurate readings, patients suffering from such conditions should not use this device. Other patient conditions that may affect performance of the device are listed in the labeling. The labeling provides instructions on the system set-up, taking measurements, device maintenance, and troubleshooting. The Sunrise SDDA system is not intended as a substitute for full polysomnography nor intended to be used as an apnea monitor. Installation will be carried out by the patient, according to installation instructions and verification procedures provided by Sunrise.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a device for sleep apnea testing based on mandibular movement and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Delayed or incorrect treatment due<br>to erroneous output as a result of<br>software malfunction or algorithm<br>error | Software verification, validation, and hazard analysis<br>Clinical performance testing<br>Labeling |
| Delayed or incorrect treatment due<br>to user misinterpretation | Labeling |
| Delayed or incorrect treatment due<br>to sensor failing to provide inputs<br>for software to adequately analyze | Software verification, validation, and hazard analysis<br>Clinical performance testing<br>Labeling |
| Electrical shock, burn, or<br>interference with other devices | Electrical safety testing<br>Electromagnetic compatibility (EMC) testing |
{6}------------------------------------------------
| | Labeling |
|-------------------------|-----------------------------|
| Adverse tissue reaction | Biocompatibility evaluation |
## SPECIAL CONTROLS
In combination with the general controls of the FD&C Act. the devices for sleep apnea testing using physiological signals not standard in clinical sleep studies is subject to the following special controls:
- (1) Clinical data must be provided. This assessment must fulfill the following:
- (i) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
- (ii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment.
- The assessment must compare device performance with a clinical comparator device (iii) (e.g., polysomnography).
- The patient-contacting components of the device must be demonstrated to be biocompatible. (2)
- (3) The performance data must be provided to demonstrate the electromagnetic compatibility (EMC) and electrical, mechanical, and thermal safety of the device.
- (4) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must include:
- (i) A full characterization of the software technical parameters, including algorithms;
- A description of the expected impact of all applicable sensor acquisition hardware (ii) characteristics and associated hardware specifications; and
- (iii) A description of all mitigations for failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy.
- (5) Labeling must include:
- (i) A description of what the device measures and outputs to the user;
- Warnings identifying sensor acquisition factors or subject conditions or characteristics (ii) (e.g., conditions affecting the anatomy of the recording site, or subject conditions that may affect mandibular movement) that may impact measurement results;
- (iii) Guidance for interpretation of the measurements, including a statement that the device is not intended as a substitute for full polysomnography nor intended to be used as an apnea monitor; and
- The expected performance of the device for all intended use populations and (iv) environments.
## BENEFIT-RISK DETERMINATION
The risks of the device are based on data collected in clinical studies described above.
{7}------------------------------------------------
The consequences of a false negative result would be a delay in patient management and may lead to deterioration of the patient's health. Since OSA is a chronic condition that affects quality of life in the long term, the risk severity of false negative results is significant. However, per AASM guidelines, it is recommended that if a single home sleep apnea test is negative, inconclusive or technically inadequate, PSG be performed for the diagnosis of OSA. Sunrise SDDA device is a prescription use device and results are only intended to be used by the clinician as an aid in diagnosis of OSA. Therefore, the Sunrise SDDA system is to be used in combination with all available information to make patient management decisions and a false negative result would not likely lead to patient injury. Also, since OSA is usually diagnosed after several years, a short delay of a few weeks can be insignificant. The Sunrise SDDA system utilizes a new technology for aiding in automated evaluation of OSA using mandibular movement signals. While movement signals from the mandible are not among conventional PSG channels, and clinical interpretation of raw data may not be feasible by physicians, the device has been validated to have clinically acceptable accuracy and the probability of a patient harm related to false results is minimal.
The probable benefits of the device are also based on data collected in clinical studies as described above.
The device can benefit the patient by providing an easy-to-use and more accessible tool to aid in OSA diagnosis in order to initiate treatment or seek further diagnosis. In certain settings and circumstances, a home sleep apnea testing (HSAT) device like Sunrise SDDA device offers a more practical, and faster alternative to the conventional way of overnight in-lab PSG. Compared to the standard in-lab PSG, Sunrise SDDA device provides users with favorable sleeping conditions at home to overcome in-lab PSG issues such first-night effects and undesired sleep position because of electrodes and wires, etc. Sunrise SDDA device can also shorten the management time for referring patients to the specialists by providing automated notification of OSA severity and other sleep structure parameters within 48 hours of testing.
#### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
#### Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The Sunrise SDDA device is a non-invasive home care aid in the evaluation of obstructive sleep apnea (OSA) in patients 18 years and older with suspicions of sleep breathing disorders.
{8}------------------------------------------------
The probable benefits outweigh the probable risks for the Sunrise SDDA device. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.
## CONCLUSION
The De Novo request for the Sunrise SDDA device is granted, and the device is classified as follows:
Product Code: QRS Device Type: Device for sleep apnea testing based on mandibular movement Regulation Number: 868.2376 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.