DEN210014 · Appliedvr, Inc. · QRA · Nov 16, 2021 · Neurology
Device Facts
Record ID
DEN210014
Device Name
EaseVRx
Applicant
Appliedvr, Inc.
Product Code
QRA · Neurology
Decision Date
Nov 16, 2021
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 890.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EaseVRx is a prescription-use immersive virtual reality system intended to provide adjunctive treatment based on cognitive behavioral therapy skills and other evidencebased behavioral methods for patients (age 18 and older) with a diagnosis of chronic lower back-pain (defined as moderate to severe pain lasting longer than three months). The device is intended for in-home use for the reduction of pain and pain interference associated with chronic lower back pain.
Device Story
EaseVRx is an immersive virtual reality (VR) system for in-home adjunctive treatment of chronic lower back pain. Device delivers cognitive behavioral therapy (CBT) and other behavioral methods via VR display. Patient interacts with software-based therapeutic content to learn pain management skills. Intended for use by patients 18+; prescription required. System aims to reduce pain intensity and pain interference. Clinical benefit derived from behavioral therapy engagement within immersive environment. Safety profile managed through biocompatibility, electrical/mechanical/thermal testing, and labeling regarding potential nausea, motion sickness, or discomfort.
Clinical Evidence
Single-cohort, double-blinded, randomized, placebo-controlled trial (N=188). Subjects with chronic lower back pain received either EaseVRx or a sham VR device (2D non-immersive footage) for 8 weeks. Primary endpoints: pain intensity and pain interference (activity, mood, sleep, stress) measured via DVPRS. Results: 66% of EaseVRx vs 41% of sham participants achieved >30% reduction in pain intensity. Statistically significant improvements observed across all five primary endpoints for EaseVRx compared to sham. Adverse events included nausea, motion sickness, and headset discomfort, which were temporary and resolved by discontinuing or adjusting the device.
Technological Characteristics
System includes GoerTek A7510/Pico G2 4K HMD (3840x2160 resolution, 72 fps, 3-DOF tracking) and mechanical Breathing Amplifier. Materials: polyurethane (face pad, battery compartment), silicone/nylon/spandex/PET (straps), polycarbonate/ABS (controller). Powered by 3V Li-ion battery. Connectivity: standalone. Software: preloaded behavioral therapy content. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 62133.
Indications for Use
Indicated for patients age 18+ with chronic lower back pain (moderate to severe, duration >3 months) as an adjunctive treatment for pain reduction and pain interference.
Regulatory Classification
Identification
A virtual reality behavioral therapy device for pain relief is a device intended to provide behavioral therapy for patients with pain. Therapy is administered via a virtual reality display that utilizes a software program containing the behavioral therapy content.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under the labeled conditions for use must validate the model of behavioral therapy as implemented by the device and evaluate all adverse events.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(5) Labeling must include the following:
(i) A warning regarding the risk of nausea and motion sickness;
(ii) A warning regarding the risk of discomfort from the device; and
(iii) A summary of the clinical testing with the device.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR EASEVRX
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Virtual reality behavioral therapy device for pain relief. A virtual reality behavioral therapy device for pain relief is a device intended to provide behavioral therapy for patients with pain. Therapy is administered via a virtual reality display which utilizes a software program containing the behavioral therapy content.
NEW REGULATION NUMBER: 21 CFR 890.5800
CLASSIFICATION: Class II
PRODUCT CODE: ORA
#### BACKGROUND
DEVICE NAME: EaseVRx
SUBMISSION NUMBER: DEN210014
DATE DE NOVO RECEIVED: March 31, 2021
#### SPONSOR INFORMATION:
AppliedVR. Inc. 16760 Stagg St. Ste 216 Van Nuys, CA 91406
#### INDICATIONS FOR USE
The Ease VRx System is indicated as follows:
EaseVRx is a prescription-use immersive virtual reality system intended to provide adjunctive treatment based on cognitive behavioral therapy skills and other evidencebased behavioral methods for patients (age 18 and older) with a diagnosis of chronic lower back-pain (defined as moderate to severe pain lasting longer than three months). The device is intended for in-home use for the reduction of pain and pain interference associated with chronic lower back pain.
#### LIMITATIONS
{1}------------------------------------------------
The sale, distribution, and use of EaseVRx are restricted to prescription use in accordance with 21 CFR 801,109.
Safety and effectiveness have not been demonstrated in patients with moderate to severe depression.
If a user experiences motion sickness, dizziness, headache, or eve strain when using the device, stop use of the device and resume therapy per your doctor's advice.
If you suffer from the following, please consult your doctor before use:
- Hearing and visual impairment .
- . Hypersensitivity to flashing light or motion
- Injury to eyes, face, or neck that prevents comfortable use of VR .
- Have a history of epilepsy, suffer from physical, mental or heart disease .
- Have any serious medical condition .
## PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
EaseVRx is an immersive virtual reality (VR) system which delivers 3-D VR treatment that incorporates principles of cognitive behavioral therapy (CBT), other behavioral methods, and mindfulness strategies to treat patients diagnosed with chronic pain. It is a prescription-use device which uses preloaded software content on a proprietary hardware platform to deliver treatment. The EaseVRx device, shown in the figures below, consists of an off-the-shelf VR head-mounted display (Figure 1) with added Breathing Amplifier (Figure 2) and AppliedVR developed software. The Breathing Amplifier is a mechanical attachment added to the commercially available headset which enables diaphragmatic breathing exercises designed to enhance the user's engagement by amplifying the user's exhalation into the on-board microphone. The device is also marketed with an optional hand-held controller to aid in navigating the user interface.
Image /page/1/Picture/12 description: The image shows a front view of a black virtual reality headset. The headset has a sleek, rounded design with a dark, matte finish. A head strap is attached to the headset, which is also black. The text "Front View" is at the bottom of the image.
Front View
Image /page/1/Picture/14 description: The image shows a person wearing a VR headset. The headset is white and has a strap that goes around the head. The person's face is visible through the headset. There is a diagram showing how to adjust the headset to fit the face.
Figure 1: The EaseVRx Headset Figure 2: Breathing Apmplifier Component of EaseVRx
De Novo Summary (DEN210014)
{2}------------------------------------------------
EaseVRx is intended for single patient use in the patient's home while the patient is seated. Technical specifications of the EaseVRx head-mounted display are presented in the table below.
| Headset manufacturer and model number | GoerTek Technology Inc.<br>Model: A7510/Pico G2 4K |
|---------------------------------------------------------------|---------------------------------------------------------|
| Rating | 5V DC, 2A |
| Weight | 278 (w/o Band), 470 (total) |
| Frame-rate | 72 fps |
| Screen Resolution | 3840 x 2160 |
| Number of pixels horizontally and vertically per<br>eye | Horizontally: 1907<br>Vertically: 1964 |
| Luminance | Maximum: 0.06nits<br>Minimum: 63.4nits |
| Interpupillary distance (IPD) and IPD range of<br>the headset | Default 63mm, optical adaptive<br>range is from 55~71mm |
| Tracking degrees of headset x/y/z 360° | 3 degrees of freedom |
| Field of view per eye | Horizontally: 98<br>Vertically: 101 |
| Eye relief for prescription lenses | 17mm |
| Range in depths of the virtual content in the<br>software | 2m for optics; 3m for launcher<br>software |
Table 1: Technical Specifications
EaseVRx is designed to follow an 8-week treatment program which delivers a multifaceted combination of pain relief skills training through a sequence of daily sessions ranging from 2-16 minutes in length (average of 6 minutes). Sessions are designed to minimize triggers of emotional distress and motion sickness, as well as to induce relaxation and activation of the parasympathetic nervous system. Similar to multisession behavioral treatments, each week of the treatment program focuses on a specific theme and each daily treatment session is designed to align with this weekly theme in terms of the clinical messaging and content purpose. Initial themes in EaseVRx are focused on developing the skills and understanding of therapeutic principles in VR. including understanding the body, attention and distraction, relaxation strategies, and sleep and pain management, while later themes are focused on the transferring of learning outside of VR, including steps for how to move forward using acceptance and mindfulness while incorporating the behavioral skills learned to daily life. Treatment is presented chronologically such that informational content (cognitive) is delivered in conjunction with experiential content (behavioral), which allows the user to grow educated in the techniques and skills necessary to mitigate their pain.
Each week provides multiple types of VR experiences, capitalizing on principles of VR design, including immersion, gamification, and interactivity for increased session engagement and improved learning. Different types of experiences, shown in the table below, not only reinforce various pain relief skills but also provide a variety of content that appeals to different user preferences.
{3}------------------------------------------------
| Category | Title | Information or Skills Provided | |
|----------|------------------------------|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| I | Relaxation/<br>Interoceptive | Image: Relaxation/Interoceptive | Relaxing scenes that change from busy/active<br>to calm reflecting a user's progressively<br>enhanced state of relaxation. These sessions<br>train users to understand and perceive what's<br>going on inside the body as they engage in<br>relaxation. |
| II | Education | Image: Education | Brief visual and voice-guided lessons about<br>the central nervous system as it relates to pain<br>and breathing. These experiences use<br>anatomical animation to facilitate learning and<br>establish a medical and scientific rationale for<br>the program. |
| III | Mindful<br>Escapes | Image: Mindful Escapes | Immersive 360° videos with therapeutic<br>narration, guided breathing, music, and visual<br>effects to reinforce mindfulness-based pain<br>relief skills and increase engagement. |
| IV | Pain<br>Distraction | Image: Pain Distraction | Interactive games to train pain relief skills<br>related to shifting attentional focus away from<br>pain. |
| V | Dynamic<br>Breathing | Image: Dynamic Breathing | Breathing-based interactive environments to<br>train pain relief skills and induce relaxation.<br>These sessions become increasingly<br>challenging as users increase their skill with<br>diaphragmatic breathing and parasympathetic<br>control. |
# Table 2: Categories of Experiences in EaseVRx Treatment
## SUMMARY OF NONCLINICAL/BENCH STUDIES
#### BIOCOMPATIBILITY/MATERIALS
The face pad, straps, and battery compartment of the headset are considered patientcontacting components of the device, in addition to the optional hand-held controller. The face pad and battery compartment are composed of polyurethane; the straps are composed of silicone, polyurethane polyethylene terephthalate, spandex, and nylon; and the controller is composed of polycarbonate and acrylonitrile butadiene styrene (ABS). All patient-contacting components are categorized as surface contacting with limited contact duration (less than 24 hours cumulative use) of intact skin.
EaseVRx is subject to biocompatibility evaluation in accordance with the International Standard ISO 10993-1: 2009 "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing Within a Risk Management Process." According to the contact
{4}------------------------------------------------
classification and duration of the patient-contacting materials, assessment of the device should include the following tests:
- Cytotoxicity (ISO 10993-5:2009) .
- Sensitization Test (ISO 10993-10:2010) .
- Intracutaneous Reactivity (ISO 10993-10:2010) .
However. considering that these materials have a history of safe use in medical devices. the total cumulative contact duration is low, and that the manufacturing process complies with Quality System Regulations (21 CFR 820.50. 21 CFR 820.80, 21 CFR 820.100, 21 CFR 820.198, and 21 CFR 803), the risk associated with cytotoxicity, irritation, and sensitization are low. Thus, biocompatibility data was not evaluated for Ease VRx.
## ELECTROMAGNETIC CAPABILITY & ELECTRICAL SAFETY
Ease VRx was tested according to the following FDA-recognized consensus standards:
- · IEC 60601-1:2005 (Modified to be equivalent to (AAMI/ANSI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and, A2:2010/(R)2012) "Medical Electrical Equipment; Part 1: General requirements for basic safety and essential performance." Results demonstrated that the device is compliant to this standard.
- . IEC 60601-1-2:2014 "Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance-Collateral Standard: Electromagnetic disturbances - Requirements and Tests."
- IEC 60601-1-11:2015 "Medical electrical equipment: General requirements for . basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment."
#### Battery testing
Ease VRx is powered by a 3V rechargeable lithium ion battery, which complies with IEC 60601-1:2005 + A1:2012 and IEC 62133 Edition 2.0 2012-12 "Secondary cells and batteries containing alkaline or other non-acid electrolytes - safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications."
#### SOFTWARE
The sponsor provided documentation acceptable for software and firmware with a "Minor" Level of Concern (LoC), as outlined in the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." issued May 11, 2005. The primary risk to patients is delay of treatment due to software/firmware malfunction or failure. Adequate documentation describing the software, firmware, software requirements specification, traceability, revision level history, and cybersecurity provides the foundation that the software will operate in a manner as described in the specifications. A hazard analysis was performed to characterize software risks including device malfunction and measurement related errors.
{5}------------------------------------------------
The submission included verification and validation (V&V) testing to address the potential hazards with satisfactory result.
## SUMMARY OF CLINICAL INFORMATION
EaseVRx was evaluated in a single-cohort, double-blinded (participant and analysts), crosssectional, placebo-controlled, randomized clinical trial. The primary objective of this study was to assess the effectiveness of the Ease VRx therapeutic program in patients with chronic low back pain (cLBP).
The trial consisted of 188 subjects with cLBP randomly assigned to either an EaseVRx treatment group or a sham group. Subjects were first screened for inclusion/exclusion criteria. Once enrolled in the study, subjects participated in a 2-week baseline assessment period where they were required to complete a baseline assessment and one of three pain surveys in order to progress to the treatment phase. The treatment phase consisted of 8-weeks of home use and biweekly surveys. The treatment group received EaseVRx devices and 3-D VR cognitive behavioral based treatment. The sham group received VR devices displaying 2-D non-immersive nature footage with neutral music intended to be distracting and not relaxing or aversive. The sham content was designed not to have any treatment effect. Although each VR device contained software specific to the individual participant's assigned VR treatment group, all device packaging and directions for use were common to both treatment groups. While it was encouraged, participants were not required to select a fixed time to complete each VR session. Participants and study statisticians were blinded to treatment group assignment. In order to get long term data, post-treatment assessments were conducted at 1, 2, 3, and 6 months. Data up to and including the 3-month post assessment were included and reviewed in the De Novo request.
Five primary effectiveness endpoints were defined for this study: average pain intensity and pain interference on activity, mood, sleep, and stress at 8 weekly time points across the 8-week treatment phase. Each endpoint was measured using the Defense and Veterans Pain Rating Scale (DVPRS). A number of secondary endpoints were defined as well, including Patient's Global Impression of Change, various pain coping mechanisms, physical function, sleep disturbance, usability, and treatment satisfaction. Adverse event information was collected to characterize the safety profile of EaseVRx. Patients were encouraged to contact study staff to report any problems encountered during treatment, and a cybersickness was assessed at the end of treatment using a motion sickness and nausea survey.
Results for the pain intensity endpoint demonstrated that 66% of EaseVRx participants and 41% of sham participants achieved >30% reduction in pain intensity. For the Ease VRx group. 46% of participants achieved >50% pain reduction while 26% of the sham participants reached that threshold. In terms of pain interference with activity, 71% of EaseVRx participants and 57% of Sham VR participants achieved >30% reduction, and 56% of the EaseVRx participants achieved >50% reduction. 74% of EaseVRx participants and 60% of the Sham VR participants achieved >30% reduction in pain interference with mood, and 60% of the EaseVRx participants achieved >50% reduction. For pain interference with sleep. 70% of EaseVRx participants and 60% of Sham VR participants achieved >30% reduction, and 60% of the EaseVRx participants achieved >50% reduction. Lastly. 76% of EaseVRx participants and 56% of the Sham VR participants
{6}------------------------------------------------
achieved >30% reduction in pain interference with stress, and 63% of EaseVRx participants achieved >50% reduction.
Results are also shown in the tables below which describe the percent reduction from baseline and the number of patients exhibiting a >30% and >50% change for each primary endpoint. In conclusion, a clinically meaningful (>30%) improvement was observed in the EaseVRx group alone and compared to the sham group.
Table 3: Percent Reduction for Each Primary Endpoint from Pre-treatment to End of Treatment for Ease VRx and Sham Groups
| | Pain<br>Intensity | Activity<br>Interference | Sleep<br>Interference | Mood<br>Interference | Stress<br>Interference |
|----------|-------------------|--------------------------|-----------------------|----------------------|------------------------|
| Ease VRx | 41.6 | 48.9 | 51.7 | 55.7 | 57.0 |
| Sham VR | 23.5 | 32.1 | 38.1 | 36.9 | 37.4 |
# Table 4: Number (and Proportion) of EaseVRx Participants Showing 30% and 50% Pain Reductions, and the Average Pain Reduction for Each Group
| | | Pain | Pain Interference | | | |
|-----------------------------------------------|--------------------|-----------|-------------------|--------|--------|--------|
| | | Intensity | Activity | Mood | Sleep | Stress |
| Number (and<br>proportion) of<br>Participants | 30% +<br>Reduction | (b)(4) | (b)(4) | (b)(4) | (b)(4) | (b)(4) |
| | 50% +<br>Reduction | | | | | |
| Ave Reduction | 30% +<br>Reduction | | | | | |
| | 50% +<br>Reduction | | | | | |
Table 5: Number (and Proportion) of Sham VR Participants Showing 30% and 50% Pain Reductions, and the Average Pain Reduction for Each Group
| | | Pain | Pain Interference | | | | |
|-----------------------------------------------|--------------------|-----------|-------------------|--------|--------|--------|--|
| | | Intensity | Activity | Mood | Sleep | Stress | |
| Number (and<br>proportion) of<br>Participants | 30% +<br>Reduction | (b)(4) | (b)(4) | (b)(4) | (b)(4) | (b)(4) | |
| | 50% +<br>Reduction | | | | | | |
| Ave Reduction | 30% +<br>Reduction | | | | | | |
| | 50% +<br>Reduction | | | | | | |
{7}------------------------------------------------
The table below shows that a statistically significant difference was observed for all five primary endpoints.
| | ITT (N=188) and mITT (N=179) Analysis Sets | | | |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Analysis Set | ITT (N= 188) | | mITT (N= 179) | |
| Model | condition, time, condition x time | | condition, time, condition x time | |
| Covariates | Pre Specified<br>age, gender, race (dichotomized as<br>caucasian and non-caucasian),<br>ethnicity (dichotomized as hispanic<br>and non-hispanic), baseline PROMIS<br>sleep, baseline PROMIS physical<br>function | | Pre Specified<br>age, gender, race (dichotomized as<br>caucasian and non-caucasian).<br>ethnicity (dichotomized as hispanic<br>and non-hispanic), baseline PROMIS<br>sleep, baseline PROMIS physical<br>function | |
| Handling of Missing Data | Multiple Imputation | | Observed Data | |
| Statistical Test | Unadj.<br>p-value | Adjusted<br>p-value§ | Unadj.<br>p-value | Adjusted<br>p-value§ |
| Pain Average | (b)(4) | (b)(4) | (b)(4) | (b)(4) |
| Pain Interference with Activity | | | | |
| Pain Interference with Mood | | | | |
| Pain Interference with Sleep* | | | | |
# Table 5: Primary Endpoint Overall Efficacy Results
* Condition effect p-value for sleep is shown;
& Hochberg Step-up corrected p-value
Source FDA Submit - Import and Analysis.sas; Analyzed: 2021-08-31
In terms of safety, no participants contacted study staff during the trial to report adverse events of any type. Due to an error with the electronic survey administration, the electronic cybersickness survey was captured at one-month post-treatment instead of at the end of treatment. Additionally, only 147 participants completed the survey. Seven (9.7%) participants from the Ease VRx group and 5 (6.7%) participants from the sham group reported experiencing nausea and motion sickness during the treatment phase of the study. In addition, 15 (20.8%) participants from the EaseVRx group and 4 (5.3%) participants from the sham group reported discomfort with the headset. These adverse events are common and to be expected for VR devices. Furthermore, they were temporary and resolved by discontinuing use or adjusting the device.
## Pediatric Extrapolation
EaseVRx is indicated for patients age 18 and older. For medical devices, the FD&C Act defines patients before their 22nd birthday as pediatric patients. In this De Novo request, complete data from patients between 18-81 (mean age 51.4) were used to support the use of the device in adult patients. Because only one patient was evaluated within the age range of 18-22 it was unclear
{8}------------------------------------------------
whether there was enough data to support effectiveness in this population. However, it is appropriate to indicate the device for individuals 18 and older because patients aged 18 to 21 do not carry additional differences or risks relative to the adult patient population studied, and this device has a likely benefit for this group.
## LABELING
The labeling (User Manual) meets the requirements of 21 CFR Part 801.109 for prescription devices.
The labeling provides information to users describing the clinical data showing the safety and effectiveness of the EaseVRx device in the intended patient population. It also includes instructions for operating the device and navigating the user interface. Appropriate warnings and precautions are included to avoid hazardous situations and ensure safe use of the device as intended.
The labeling also outlines appropriate cleaning methods for home use.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the virtual reality behavioral therapy device for pain relief and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|---------------------------------------------------------------------|-----------------------------------------------------------|
| Adverse tissue reaction | Biocompatibility evaluation |
| Electric shock or burn or interference<br>with other devices | Electromagnetic compatibility (EMC) testing |
| | Electrical, mechanical, and thermal safety testing |
| Nausea and motion sickness | Clinical performance testing |
| | Labeling |
| Discomfort | Clinical performance testing |
| | Labeling |
| Ineffective treatment | Clinical performance testing |
| | Software verification, validation, and hazard<br>analysis |
| | Labeling |
| Use error or improper device use<br>leading to a delay in treatment | Labeling |
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the virtual reality behavioral therapy device for pain relief is subject to the following special controls:
- (1) Clinical performance testing under the labeled conditions for use must validate the model of behavioral therapy as implemented by the device and evaluate all adverse events.
{9}------------------------------------------------
- (2) The patient-contacting components of the device must be demonstrated to be biocompatible.
- (3) Software verification, validation, and hazard analysis must be performed.
- (4) Electromagnetic compatibility and electrical, and thermal safety testing must be performed.
- (5) Labeling must include the following:
- (i) A warning regarding the risk of nausea and motion sickness;
- (ii) A warning regarding the risk of discomfort from the device; and
- (iii) A summary of the clinical testing with the device.
## BENEFIT-RISK DETERMINATION
The known probable risks of the device are based on the non-clinical data and the data collected in the clinical study described above. Namely, the study identified adverse events such as nausea. motion sickness, and discomfort due to the headset. The device exhibited an acceptable safety profile in the clinical study, and any adverse events that occurred were temporary and had complete resolution by discontinuing use or adjusting the device. No device-related serious adverse events were observed. The results of the nonclinical testing demonstrated that the device performed as per specifications and the results did not raise concerns regarding risks to the patients.
The probable benefits of the device are also based on data collected in the clinical studies as described above. There is evidence of a clinically meaningful (>30% change) and statistically significant improvement in pain and pain interference DVPRS scores. Results show that 65% of patients in the EaseVRx treatment group achieved a 30% reduction in pain at end of treatment, which was sustained throughout the 1-, 2-, and 3-month follow-up period, whereas only 41% of the sham group achieved a 30% or greater reduction in pain and this improvement was not sustained through follow-up. Additionally, an average of 73% of EaseVRx patients achieved a 30% or greater reduction in pain interference on activity, mood, sleep, and stress, whereas an average of 58% to sham patients achieved a 30% or greater reduction in pain interference on activity, mood, sleep, and stress respectively. Therefore, the clinical study demonstrates a greater improvement in pain and pain interference in the EaseVRx treatment group compared to the sham group.
Sources of uncertainty in the benefits include the effect of concomitant use of prescription and over the counter medications for pain during treatment and imperfect comparator method. specifically that the sham device could have been administering a treatment effect. The use of concomitant medications was addressed by specifying the Ease VRx is intended for adjunctive treatment in the indications for use. Additionally, the clinical data showed a greater improvement in the EaseVRx treatment group compared to the sham group. Even if the sham group did have a treatment effect, the EaseVRx group still demonstrated a greater improvement. This addresses the uncertainty introduced by a potential imperfect comparator method.
# PATIENT PERSPECTIVES
{10}------------------------------------------------
The primary and secondary outcome measures in the supportive clinical study were collected using patient reported outcomes. Five patient reported primary effectiveness endpoints were defined: average pain intensity and pain interference on activity, mood, sleep, and stress at 8 weekly time points across the 8-week treatment phase. Each outcome was measured using the Defense and Veterans Pain Rating Scale (DVPRS).
Patient reported secondary outcomes included: Patient Global Impression of Change Scale (PGIC), PROMIS physical function, PROMIS sleep disturbance, Pain Self-Efficacy Questionnaire (PSEQ-2), Pain Catastrophizing Scale (PSCS-4), Chronic Pain Acceptance Questionnaire (CPAQ-8), Patient satisfaction, Positive and Negative Affect Scale (PANAS).
## BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
EaseVRx is a prescription-use immersive virtual reality system intended to provide adjunctive treatment based on cognitive behavioral therapy skills and other evidencebased behavioral methods for patients (age 18 and older) with a diagnosis of chronic lower back-pain (defined as moderate to severe pain lasting longer than three months). The device is intended for in-home use for the reduction of pain and pain interference associated with chronic lower back pain.
The probable benefits outweigh the probable risks for Ease VRx. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
## CONCLUSION
The De Novo request for EaseVRx is granted and the device is classified as follows:
Product Code: ORA Device Type: Virtual reality behavioral therapy device for pain relief Regulation Number: 21 CFR 882.5800 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.