DEN200070 · Letsgetchecked Inc. (Formerly Privapath Diagnostics Inc.) · QYA · Nov 15, 2023 · Microbiology
Device Facts
Record ID
DEN200070
Device Name
Simple 2 Test
Applicant
Letsgetchecked Inc. (Formerly Privapath Diagnostics Inc.)
Product Code
QYA · Microbiology
Decision Date
Nov 15, 2023
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 866.3385
Device Class
Class 2
Indications for Use
The Simple 2 Test is intended for in vitro detection and identification of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) in home-collected specimens which are shipped to a clinical laboratory for testing using the Aptima Combo 2 Assay on the Panther System. This product is available over-the-counter (OTC) to consumers 18 years of age and older. The Simple 2 Test contains all the necessary components to collect urine from male patients (Simple 2 Urine Home Collection Kit (Penile)) or vaginal swabs from female patients (Simple 2 Swab Home Collection Kit (Vaginal)) in their home, or in similar environments, without supervision from a healthcare provider. The Simple 2 Test Collection Kits may also be used to self-collect specimens in a clinic. The testing is performed, as determined to be appropriate, based on the results of LetsGetChecked Suitability Questionnaire. This test system is not a substitute for visits to a healthcare provider. The information provided by this product should not be used to start, stop, or change any course of treatment unless advised by your healthcare provider. Testing is limited to the manufacturer, Priva Path laboratories (d.b.a. LetsGetChecked. Inc.).
Device Story
Simple 2 Test is an OTC home-collection kit for Chlamydia trachomatis and Neisseria gonorrhoeae; includes components for male urine or female vaginal swab collection. User completes a suitability questionnaire; collects specimen at home; ships to Priva Path laboratories. Laboratory processes samples using Aptima Combo 2 Assay on Panther System. Results provided to user; intended to aid diagnosis. Not a substitute for healthcare provider visits; users advised not to change treatment without professional consultation. Benefits include increased access to STI testing; facilitates early detection and management.
Clinical Evidence
Requires clinical study documentation including prospective self-collected samples for each claimed specimen type. Study must compare device performance to an FDA-appropriate comparator. Documentation must include protocol, statistical analysis plan, and results. Specific metrics (sensitivity, specificity, etc.) not provided in the summary letter.
Technological Characteristics
Nucleic acid amplification assay. Components: urine collection cup, graduated transfer pipette, Aptima Urine Specimen Transport Tube, vaginal swab, Sample Transport Tube. Processing: Aptima Combo 2 Assay (target capture, transcription-mediated amplification, hybridization protection, dual kinetic assay) on Panther System. Connectivity: online portal for activation and result delivery.
Indications for Use
Indicated for in vitro detection and identification of Chlamydia trachomatis and/or Neisseria gonorrhoeae in home-collected urine (males) or vaginal swabs (females) for consumers 18 years and older. Not a substitute for healthcare provider visits.
Regulatory Classification
Identification
The Simple 2 Test is an over-the-counter (OTC) in vitro diagnostic test intended for the detection and identification of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) in home-collected specimens. The device consists of the Simple 2 Urine Home Collection Kit (Penile) and the Simple 2 Swab Home Collection Kit (Vaginal). Collected specimens are shipped to a CLIA-certified laboratory for testing using the Aptima Combo 2 Assay on the Panther System. The test is intended for consumers 18 years of age and older and includes a suitability questionnaire to determine if the test is appropriate for the individual.
Reference Devices
Aptima Combo 2 Assay
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine. The text is written in a clear, sans-serif font, with "FDA" in a blue box.
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Simple 2 Test DECISION SUMMARY
#### I Background Information:
- A De Novo Number
DEN200070
# B Applicant
LetsGetChecked Inc. (formerly PrivaPath Diagnostics Inc.)
# C Proprietary and Established Names
Simple 2 Test
# D Regulatory Information
| Product<br>Code(s) | Classification | Regulation<br>Section | Panel |
|--------------------|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| QYA | Class II | 21 CFR 866.3385 - System for detection of<br>nucleic acid from non-viral microorganism(s)<br>causing sexually transmitted infections using<br>home-collected specimens | MI - Microbiology |
#### Submission/Device Overview: II
## A Purpose for Submission:
De Novo request for evaluation of automatic class III designation for the Simple 2 Test.
# B Measurand:
Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) ribosomal RNA (rRNA)
# C Type of Test:
Nucleic acid amplification assay
#### III Indications for Use:
# A Intended Use(s):
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov
{1}------------------------------------------------
See Indications for Use below.
#### B Indication(s) for Use:
The Simple 2 Test is intended for in vitro detection and identification of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) in home-collected specimens which are shipped to a clinical laboratory for testing using the Aptima Combo 2 Assay on the Panther System. This product is available over-the-counter (OTC) to consumers 18 years of age and older.
The Simple 2 Test contains all the necessary components to collect urine from male patients (Simple 2 Urine Home Collection Kit (Penile)) or vaginal swabs from female patients (Simple 2 Swab Home Collection Kit (Vaginal)) in their home, or in similar environments, without supervision from a healthcare provider.
The Simple 2 Test Collection Kits may also be used to self-collect specimens in a clinic.
The testing is performed, as determined to be appropriate, based on the results of LetsGetChecked Suitability Questionnaire.
This test system is not a substitute for visits to a healthcare provider. The information provided by this product should not be used to start, stop, or change any course of treatment unless advised by your healthcare provider.
Testing is limited to the manufacturer, Priva Path laboratories (d.b.a. LetsGetChecked. Inc.).
# C Special Conditions for Use Statement(s):
OTC - Over the Counter
# D Special Instrument Requirements:
Panther System
#### IV Device/System Characteristics:
## A Device Description:
The LetsGetChecked Simple 2 Test is composed of the Simple 2 Urine Home Collection Kit (Penile) and the Simple 2 Swab Home Collection Kit (Vaginal) and the samples are shipped to LetsGetChecked's Clinical Laboratory Improvement Amendments (CLIA)-certified laboratory for processing using the Aptima Combo 2 Assay on the Panther System. The Aptima Combo 2 Assay was previously cleared for the qualitative in vitro detection of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) when used with male urine and self-collected vaginal swabs collected in clinical settings.
The home collection kits are intended to be distributed for purchase online without the need for prescription from a physician or healthcare professional (over-the-counter use). After purchase, the user activates the kit on LetsGetChecked's website and is requested to fill out the Suitability Questionnaire to determine whether this test is appropriate for this individual. The user follows detailed directions for self-collecting a urogenital specimen (male urine or female vaginal swab,
{2}------------------------------------------------
as applicable), packaging and shipping it to the LetsGetChecked CLIA certified laboratory for testing (PrivaPath lab). The test results are delivered to the individual via LetsGetChecked online portal. Positive and invalid/inconclusive results are followed up with a call from the LetsGetChecked Care Team. If the patient received a positive chlamydia result, treatment options are discussed. Patients receiving a positive gonorrhea test result, will be advised to contact their healthcare provider to receive appropriate treatment. The LetsGetChecked Care Team will also contact all patients who test negative for chlamydia and gonorrhea, but do report symptoms, or concerns, or where further investigation is indicated. These patients are internally flagged for follow-up.
## Simple 2 Urine Home Collection Kit (Penile)
Each Simple 2 Urine Home Collection Kit (Penile) includes a urine collection cup with indicated urine fill area, one Aptima Urine Specimen Transport Tube, a graduated transfer pipette, a biohazard bag, home specimen collection and shipping instructions, and Frequently Asked Questions (FAQs) sheets with information about CT/GC infections and test results. All components are included in the box which is also used for shipping of the sample to the laboratory for testing.
# Simple 2 Swab Home Collection Kit (Vaginal)
Each Simple 2 Swab Home Collection Kit (Vaginal) includes Aptima Multitest Swab Specimen Collection Kit, which includes sterile vaginal swab and Sample Transport Tube. The kit also includes a biohazard bag, home specimen collection and shipping instructions, and Frequently Asked Questions (FAQs) sheets with information about CT/GC infections and test results. All components are included in the box which is also used for shipping of the sample to the laboratory for testing.
## B Principle of Operation
The Aptima Combo 2 Assay utilizes target capture, transcription-mediated amplification, hybridization protection assay, and dual kinetic assay for specimen processing to amplify target rRNA and detect the amplicon, respectively. Please refer to the decision summaries of K111409, K132251, K180681, and K200866.
# C Instrument Description Information
- 1. Instrument Name:
Panther System
- 2. Specimen Identification:
Refer to K111409, K132251, K180681, and K200866
- 3. Specimen Sampling and Handling:
Refer to K111409, K132251, K180681, and K200866
- 4. Calibration:
{3}------------------------------------------------
Refer to K111409, K132251, K180681, and K200866
- 5. Quality Control:
Refer to K111409, K132251, K180681, and K200866
#### V Standards/Guidance Documents Referenced:
Not applicable.
#### VI Performance Characteristics:
## A Analytical Performance:
- 1. Precision/Reproducibility:
Reproducibility and within laboratory precision studies were previously reviewed and described in K111409, K132251, K180681, and K200866.
- 2. Linearity:
Not Applicable; this is a qualitative assay.
- 3. Analytical Specificity/Interference:
Analytical Specificity/Interference studies were previously reviewed and described in K111409, K132251, K180681, and K200866.
- 4. Assay Reportable Range:
Not Applicable; this is a qualitative assay.
- 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability, Stability, Expected Values (Controls, Calibrators, or Methods) were previously reviewed and described in K111409, K132251, K180681, and K200866.
- 6. Detection Limit:
The Detection Limit was previously reviewed and described in K111409, K132251, K180681, and K200866.
- 7. Assay Cut-Off:
Assay cut-off was previously reviewed and described in K111409, K132251, K180681, and K200866.
{4}------------------------------------------------
#### 8. Accuracy (Instrument):
Instrument accuracy was previously reviewed and described in K111409, K132251, K180681, and K200866.
- 9. Carry-Over:
Carry-Over study was previously reviewed and described in K111409, K132251, K180681, and K200866.
#### B Comparison Studies:
- 1. Method Comparison:
Not applicable.
- 2. Matrix Comparison:
Not applicable.
#### C Clinical Studies:
- 1. Clinical Sensitivity:
Clinical Sensitivity was previously reviewed and described in K111409, K132251, K180681, and K200866.
- 2. Clinical Specificity:
Clinical Specificity was previously reviewed and described in K111409, K132251, K180681, and K200866.
- 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
## D Clinical Cut-Off:
Not applicable.
#### Expected Values/Reference Range: E .
Expected Values/Reference Range was previously reviewed and described in K111409, K132251, K180681, and K200866.
## F Other Supportive Performance Characteristics Data:
## 1. USABILITY AND COMPREHENSION
{5}------------------------------------------------
# A. Simple 2 Urine Home Collection Kit (Penile)
Two usability and comprehension studies with lay users were conducted to evaluate the ease of use of the collection kit, ability to interpret the test results and comprehension of the educational materials included in the Simple 2 Urine Home Collection Kit (Penile). The first study was conducted with 89 male participants. The summary of demographics of participants included in this study is presented in Table 1.
| Participants | N | % |
|-----------------------------|----|------|
| Male | 89 | 100% |
| Age Group | | |
| 18-25 | 7 | 8% |
| 26-35 | 25 | 28% |
| 36-45 | 17 | 19% |
| 46-55 | 20 | 22% |
| 56-65 | 14 | 16% |
| > 65 | 6 | 7% |
| Education | | |
| < High school | 0 | 0% |
| High school/Some College | 42 | 47% |
| Bachelor's degree or higher | 47 | 53% |
| Race/Ethnicity | | |
| Caucasian | 55 | 62% |
| Asian | 7 | 8% |
| Hispanic/Latin American | 9 | 10% |
| Black/African American | 11 | 12% |
| Mixed/Unknown/Other | 7 | 8% |
# Table 1. Summary of Demographics of Participants in the Initial Usability and Comprehension Study for the Simple 2 Urine Home Collection Kit (Penile).
The results from this study showed that 9% (8/89) of lay users had difficulty transferring correct volume of urine to the transport tube. Therefore, modifications to the home collection kit and instructions were necessary and to mitigate the risk of users transferring incorrect volume of urine to the transfer tubes, the collection kit was modified to include a graduated transfer pipette and revisions were made to the instructions on how to transfer the proper volume of urine. In addition, FAQs sheets were added to the kit to help the lay user with understating of test results and to provide educational information on CT/GC infections. To evaluate the modifications introduced to Simple 2 Urine Home Collection Kit (Penile), the second usability and comprehension study was conducted. The summary of demographics of participants in the final usability and comprehension study for the Simple 2 Urine Home Collection Kit (Penile) is presented in Table 2.
Similar to the first study, the second usability and comprehension study was conducted remotely from lay user's home via online video conferencing. The study participants were asked to follow the home collection Instructions for Use provided with the kit to complete tasks required for use of the Simple 2 Urine Home Collection Kit (Penile), including kit
{6}------------------------------------------------
activation, sample preparation and packaging, and sample shipment. Participants had an option to watch the instructional video. The lay users did not have to collect a urine sample. In this study, the collection cup was filled with water to simulate urine and evaluate users' ability to transfer the appropriate volume of liquid to the transport tube. After packaging of the sample, the study participants answered study staff questions to evaluate comprehension of the instructions/safety warnings, possible test results, and general information about the diseases. In addition, the samples were evaluated at the testing laboratory for critical errors that could lead to inaccurate test results or sample rejection upon receipt in the laboratory. The results from the study are summarized in the Table 3 and Table 4.
| Participants | N | % |
|-----------------------------------------------------|----|-----|
| Male | 32 | 100 |
| Age Group | | |
| 18-25 | 5 | 16% |
| 26-35 | 10 | 31% |
| 36-45 | 6 | 19% |
| 46-55 | 3 | 9% |
| 56-65 | 5 | 16% |
| > 65 | 3 | 9% |
| Education | | |
| < High school | 1 | 3% |
| High school diploma | 4 | 13% |
| Some college/Associates<br>Degree/Vocational Degree | 11 | 34% |
| Bachelor's degree | 12 | 38% |
| Graduate degree | 4 | 13% |
| Literacy (SAHL)a) | | |
| Low (<14) | 0 | 0% |
| Average (15-16) | 2 | 6% |
| High (17-18) | 30 | 94% |
| Race/Ethnicity | | |
| Caucasian | 15 | 47% |
| Asian | 2 | 6% |
| Hispanic/Latin American | 8 | 25% |
| Black/African American | 3 | 9% |
| Mixed/Unknown/Other | 3 | 9% |
Table 2. Summary of Demographics of Participants in the Final Usability and Comprehension Study for the Simple 2 Urine Home Collection Kit (Penile).
a) Short Assessment of Health Literacy
#### Table 3. Summary of Results of Usability and Comprehension Study for the Simple 2 Urine Home Collection Kit (Penile).
| Instructions Step/Question | Correct<br>Action/Response | Success Rate |
|----------------------------|----------------------------|--------------|
|----------------------------|----------------------------|--------------|
{7}------------------------------------------------
| Sample Collection Assessment | | |
|--------------------------------|--------|--|
| | (D)(4) | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| IFU and Warnings Comprehension | (b)(4) | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
{8}------------------------------------------------
| | (D(4) | |
|--------------------------------------------------|-------|--|
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| General Information about Diseases Comprehension | | |
| | | |
| | 10(4) | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | Success Rate |
|--------------------|--------|--------------|
| | (b)(4) | |
| Rejection Reason: | | |
| | (b)(4) | |
| N/A Not applicable | | |
N/A Not applicable.
# B. Simple 2 Swab Home Collection Kit (Vaginal)
Two usability and comprehension studies with lay users were conducted to evaluate the ease of use, results comprehension, and understanding of the educational materials about
{9}------------------------------------------------
gonorrhea and chlamydia included in the Simple 2 Swab Home Collection Kit (Vaginal). First study was conducted with 85 female participants. The summary of demographics of participants included in this study is presented in Table 5.
| Participants | N | % |
|-----------------------------|----|------|
| Female | 85 | 100% |
| Age Group | | |
| 18-25 | 17 | 20% |
| 26-35 | 34 | 40% |
| 36-45 | 14 | 16% |
| 46-55 | 15 | 18% |
| 56-65 | 4 | 5% |
| > 65 | 1 | 1% |
| Education | | |
| < High school | 2 | 2% |
| High school/Some College | 49 | 58% |
| Bachelor's degree or higher | 34 | 40% |
| Race/Ethnicity | | |
| Caucasian | 44 | 52% |
| Asian | 7 | 8% |
| Hispanic/Latin American | 14 | 16% |
| Black/African American | 14 | 16% |
| Mixed/Unknown/Other | 6 | 7% |
Table 5. Summary of Demographics of Participants in the Initial Usability and Comprehension Study for the Simple 2 Swab Home Collection Kit (Vaginal).
Upon completion of the study the home collection kit instructions were modified to ensure closer alignment with the instructions from the Hologic's Aptima Multitest Swab Specimen Collection Kit for Patient Collected Specimen. In addition, FAQs sheets were added to the home collection kit to help the lay user with understating of test results and provide education on CT/GC infections. The second usability and comprehension study was conducted to evaluate the modifications introduced to the Simple 2 Swab Home Collection Kit (Vaginal). The summary of demographics of participants in the final usability and comprehension study is presented in Table 6.
Similar to the first study, the second usability and comprehension study was conducted remotely from lay user's home via online video conferencing. The study participants were asked to follow the Instructions for Use provided with the kit to complete tasks required for use of the Simple 2 Swab Home Collection Kit (Vaginal), including kit activation, sample preparation and packaging, and sample shipment. Participants had an option to watch the instructional video. After packaging of the sample, the study participants answered study staff questions to evaluate comprehension of the instructions/safety warnings, possible test results, and general information about the diseases. In addition, the samples were evaluated at the testing laboratory for critical errors that could lead to inaccurate test results or sample
{10}------------------------------------------------
rejection upon shipping. The results from the study are summarized in the Table 7 and Table 8.
| Participants | N | % |
|--------------------------------------------------|----|-----|
| Female | 33 | 97% |
| Transgender Male* | 1 | 3% |
| Age Group | | |
| 18-25 | 5 | 15% |
| 26-35 | 8 | 24% |
| 36-45 | 8 | 24% |
| 46-55 | 9 | 26% |
| 56-65 | 3 | 9% |
| > 65 | 1 | 3% |
| Education | | |
| < High school | 1 | 3% |
| High school diploma | 6 | 18% |
| Some college/Associates Degree/Vocational Degree | 15 | 44% |
| Bachelor's degree | 10 | 29% |
| Graduate degree | 2 | 6% |
| Literacy (SAHL)a) | | |
| Low (<14) | 2 | 6% |
| Average (15-16) | 6 | 18% |
| High (17-18) | 26 | 77% |
| Race/Ethnicity | | |
| Caucasian | 15 | 44% |
| Asian | 3 | 9% |
| Hispanic/Latin American | 5 | 15% |
| Black/African American | 9 | 26% |
| Mixed/Unknown/Other | 2 | 6% |
# Table 6. Summary of Demographics of Participants in the Final Usability and Comprehension Study for the Simple 2 Swab Home Collection Kit (Vaginal).
Short Assessment of Health Literacy a)
# Table 7. Summary of Results of Usability and Comprehension Study for the Simple 2 Swab Home Collection Kit (Vaginal).
| Instructions Step/Question | Correct<br>Action/Response | Success Rate |
|------------------------------|----------------------------|--------------|
| Sample Collection Assessment | | |
| (b)(4) | | |
| | | |
| | | |
| | | |
{11}------------------------------------------------
| | (DG) | |
|--------------------------------|--------|--|
| | | |
| | | |
| | | |
| | | |
| | | |
| IFU and Warnings Comprehension | | |
| | (0)143 | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
{12}------------------------------------------------
Table 8. Errors Noted at the Testing Laboratory During Accessioning for the Simple 2 Swab Home Collection Kit (Vaginal).
| Success Rate |
|--------------------|
| (b)(4) |
| Rejection Reasons: |
| (b)(4) |
N/A Not applicable.
The results from the usability studies for both kits demonstrated that lay users are able to properly collect their own specimens and correctly perform the tasks associated with the packaging and shipping the samples. Further, results from the questionnaire support the conclusion that lay users demonstrated comprehension of the critical elements in the labeling,
{13}------------------------------------------------
including warnings, implications of the test results and the need to contact a healthcare provider as needed.
In addition, the second study for the urine kit showed that replacing generic transfer pipette with the graduated transfer pipette allows the users to transfer an appropriate amount of urine, which was a challenge with the previous version of the pipette.
# 2. FLEX STUDIES
To conduct analytical studies at the challenging concentrations (low positive) of CT and NG, the empirical LoD of the assay was established. In urine, the LoD was established at | @101 for NG and | | | | for CT. In negative vaginal swab specimens, the LoD was established at 10145 for NG and (000) for CT. All Flex Studies, Interference with Hand Contaminant, and Sample Shipping Stability studies were performed in reference to the empiric LoD.
# A. Incorrect Volume of Urine Transferred to the Aptima Urine Transport Tube (Minimum Volume)
The manufacturer's instructions for the Aptima Combo 2 assay recommend a urine volume of 1000 to be transferred into the Aptima Urine Transport Tube for a final sample volume (urine + Aptima transport media) of (01/01/11 The package insert indicates that the final sample volume must be within the black fill lines indicated on the urine specimen transport tube. The study was performed to determine the effects of sample volumes below the maximum fill line, i.e. [b](4) of added urine (underfilled tubes) and above the maximum fill line (overfilled tubes). Negative male urine matrix was spiked with CT or GC organisms at the target concentration of LoD and the test samples were prepared by adding the spiked urine (for positive samples) and unspiked urine (for negative samples) into the | | | | of the transport medium at the volumes listed in Tables 9 and 10. 1014 positive and 1010 megative replicates of each sample were tested. The results for underfilled samples are presented in Table 9 below. Correct results were obtained for all negative samples tested and positive samples when at least | |0 | of urine was transferred into the Urine Transport Tube. For positive samples, urine volume below [014] failed to generate 100% positivity. The results for overfilled samples are presented in Table 10 below. Correct results were obtained for all negative and positive samples tested.
| Urine<br>Volume<br>(mL) | (b)(4) LoD<br>CT<br>#positives/<br>#tested | CT<br>Agreement<br>with<br>expected<br>result | (b)(4) LoD<br>GC<br>#positives/<br>#tested | GC<br>Agreement<br>with<br>expected<br>result | Negative<br>Samples | Negative Samples<br>Agreement with<br>Expected Results<br>(for CT and NG) |
|-------------------------|--------------------------------------------|-----------------------------------------------|--------------------------------------------|-----------------------------------------------|---------------------|---------------------------------------------------------------------------|
| | (b)(4) | 100% | (b)(4) | 100% | (b)(4) | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 100% | | 100% |
| | | 80% | | 60% | | 100% |
| | | 80% | | 40% | | 100% |
| | | 0% | | 20% | | 100% |
## Table 9. Summary of Testing Underfilled Urine Transport Tubes.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows a gray rectangle with a red border. Inside the rectangle, there is the text '(b)(4)' in a small, light-colored font. The text is centered horizontally and vertically within the rectangle.
| Urine<br>Volume<br>(mL) | (b)(4) LoD<br>CT<br>#positives/<br>#tested | CT<br>Agreement<br>with<br>expected<br>result | (b)(4) LoD<br>GC<br>#positives/<br>#tested | GC<br>Agreement<br>with<br>expected<br>result | Negative<br>Samples | Negative Samples<br>Agreement with<br>Expected Results<br>(for CT and NG) |
|-------------------------|--------------------------------------------|-----------------------------------------------|--------------------------------------------|-----------------------------------------------|---------------------|---------------------------------------------------------------------------|
| (b)(4) | (b)(4) | 100% | (b)(4) | 100% | (b)(4) | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 100% | | 100% |
# Table 10. Summary of Testing Overfilled Urine Transport Tubes.
The study results showed that underfilling the Transport Tube with urine will generate accurate results if at least (0)(4) of the specified volume) of urine is transferred. The results also showed that the test is not sensitive to overfilling the Transport Tube. To mitigate the risk of user errors, the transfer pipette is marked to indicate the proper volume to be transferred and the instructions guide the user to ensure that the transferred urine volume is appropriate.
# B. Incorrect Amount of Transport Media in the Aptima Urine Transport Tubes.
Aptima urine transport tubes are prepared by the manufacturer with a fixed amount of the transport media | | The transport media is responsible for cell lysis and subsequent stabilization of the resulting rRNA. Spilling or removing of the transport media from the sample tube could lead to a reduced transport media-to-urine volume ratio (intended ratio (0)4) which could lead to erroneous results. To determine if lower than the recommended volume of the transport media leads to erroneous results, the tubes were filled with several volumes, followed by addition of of urine samples; one set of test samples used low positive LoD) CT or GC urine samples, and the second set of samples used negative urine samples. [ 1 ] positive and [ 14 negative replicates were tested for each volume of the transport media. The results are presented in Table 11 below. Correct results were obtained for all negative and all positive samples tested.
## Table 11. Summary of Testing Various Transport Media Volumes in the Urine Transport Tube.
| Transport<br>Media<br>Volume<br>(mL) | Urine<br>Volume<br>(mL) | (b)(4) LoD<br>CT<br>#positives/<br>#tested | CT<br>Agreement<br>with<br>expected<br>result | (b)(4) LoD<br>GC<br>#positives<br>/#tested | GC<br>Agreement<br>with expected<br>result | Negative<br>Samples | Negative<br>Samples<br>Agreement with<br>Expected Results<br>(for CT and NG) |
|--------------------------------------|-------------------------|--------------------------------------------|-----------------------------------------------|--------------------------------------------|--------------------------------------------|---------------------|------------------------------------------------------------------------------|
| (b)(4) | (b)(4) | | 100% | (b)(4) | 100% | (b)(4) | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 0% | | 100% |
{15}------------------------------------------------
# C. Delay of Urine Transfer to the Aptima Urine Transport Tubes.
The collection instructions direct the user to transfer the Urine Transport Tube after collection. The study was performed to determine the effect of delay of urine transfer from the collection cup to the Aptima Urine Transport Tubes. Negative male urine spiked with LoD CT or GC at time/day 0, was placed into a urine collection cup/box and stored on the bench top at room temperature ( 10%(4) | At each time point, as presented in the Table 12 below, a | b() | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Aptima Urine Specimen Transport Tube, and tested for the presence of GC/CT using the Aptima Combo 2 assay. [1019] positive and [0] negative replicates were tested at each time point. The results are presented in Table 12 below. Expected results were obtained for all negative samples. False negative results for CT were obtained when the urine was transferred to the Urine Transport Tube three weeks after placing in the collection cup/box. False negative results were obtained for GC when the transfer of urine to the Urine Transport Tube was delayed more than one day. Therefore, the IFU directs the user to transfer the urine right after the collection to mitigate the risk of erroneous results due to the delay in urine transfer to the Urine Transport Tube. Further, the directions specify that the sample should be sent to the lab on the same day as it is collected.
| Time | (b)(4) LoD CT<br>#positives/<br>#tested | CT<br>Agreement<br>with<br>expected<br>result | (b)(4) LoD GC<br>#positives/<br>#tested | GC<br>Agreement<br>with<br>expected<br>result | Negative Samples | Negative Samples<br>Agreement with<br>Expected Result<br>(for CT and NG) |
|--------|-----------------------------------------|-----------------------------------------------|-----------------------------------------|-----------------------------------------------|------------------|--------------------------------------------------------------------------|
| (b)(4) | | 100% | (b)(4) | 100% | (b)(4) | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 100% | | 100% |
| | | 100% | | 0% | | 100% |
| | | 100% | | 0% | | 100% |
| | | 100% | | 0% | | 100% |
| | | 100% | | 0% | | 100% |
| | | 60% | | 0% | | 100% |
Table 12. Summary of Testing the Delay of Urine Transfer to the Urine Transport Tube.
negative and equivocal result were obtained. a) (10014)
b) negative results were obtained.
c) negative results were obtained.
## 3. INTERFERING SUBSTANCES
## A. Simple 2 Urine Home Collection Kit (Penile)
Studies were performed to evaluate the effect of inadvertent introduction of common hand contaminants to urine samples collected with Aptima urine transport tubes and tested using. the Aptima Combo 2 assay. Pooled male urine samples containing common hand contaminants (i.e., water, soap, hand sanitizer, lotion, and sunscreen) were spiked with GC or CT at PLOD to determine if accidental introduction of exogenous substances that may be introduced during specimen collection present a risk for false results. Multiple brands of each substance were tested at [ [0] positive and [0] negative replicates were tested for each
{16}------------------------------------------------
substance. The results are presented in Table 13 below. Expected results were obtained for negative and all CT positive samples. However, false negative results were obtained for GC in the presence of certain brands of hand soap and hand samitizer (Table 13). The limitations have been introduced in the labeling to caution the user about the interference. Additionally, the instructions direct the user to wash their hands in warm soapy water and dry thoroughly which serves to lower the risk of sample contamination.
| Substance<br>Tested | Brand | (b)(4) LoD<br>CT<br>#positives/<br>#tested | CT<br>Agreement<br>with<br>expected<br>result | (b)(4) LoD<br>GC<br>#positives/<br>#tested | GC<br>Agreement<br>with<br>expected<br>result | Negative<br>Samples | Negative Samples<br>Agreement with<br>Expected Results<br>(for CT and NG) |
|---------------------|--------|--------------------------------------------|-----------------------------------------------|--------------------------------------------|-----------------------------------------------|---------------------|---------------------------------------------------------------------------|
| | (b)(4) | | 100% | (b)(4) | 100% | (b)(4) | 100% |
| | | | 100% | | 0% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 67% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 50% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
# Table 13. Summary of Testing Interfering Substances in the Urine Transport Tube.
[[b](4) negative and [b] equivocal results were obtained. a)
b) [co) equivocal results were obtained.
c) negative and be equivocal result were obtained.
# B. Simple 2 Swab Home Collection Kit (Vaginal)
Studies were performed to evaluate the effect of inadvertent introduction of common hand contaminants to vaginal swab samples collected with the Simple 2 Swab Home Collection Kit (Vaginal) and tested using the Aptima Combo 2 assay. Negative vaginal swab STM samples containing common hand contaminants (i.e., water, soap, hand sanitizer, lotion, and sunscreen) were spiked with GC or CT at (0) LoD to determine if inadvertent introduction of exogenous substances that may be introduced during specimen collection present a risk for false positive or false negative results. Multiple brands of each substance were tested at | [1] 1010 DI positive and prosive replicates were tested for each substance. The results are presented in Table 14 below. All samples generated expected results.
# Table 14. Summary of Testing Interfering Substances in the Simple 2 Swab Home Collection Kit (Vaginal).
{17}------------------------------------------------
| Substance<br>Tested | Brand | 3x LoD<br>CT<br>#positives/<br>#tested | CT<br>Agreement<br>with<br>expected<br>result | 3x LoD<br>GC<br>#positives/<br>#tested | GC<br>Agreement<br>with<br>expected<br>result | Negative<br>Samples | Negative<br>Samples<br>Agreement<br>with<br>Expected<br>Results |
|---------------------|--------|----------------------------------------|-----------------------------------------------|----------------------------------------|-----------------------------------------------|---------------------|-----------------------------------------------------------------|
| | (b)(4) | (b)(4) | 100% | (b)(4) | 100% | (b)(4) | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
| | | | 100% | | 100% | | 100% |
## 4. SAMPLE SHIPPING STABILITY
## A. Simple 2 Urine Home Collection Kit (Penile)
Shipping stability studies were performed to evaluate the stability of samples under extreme shipping conditions that samples may encounter when shipped via commercial carriers. Pooled negative male urine was spiked with live CT or GC at 01 (low positive) or 1084 LoD (moderate positive). Now positive. Pla moderate positive and Por negative samples were evaluated according to the Summer and Winter Shipping Temperature Cycles as indicated in Tables 15 and 17, respectively. The results from the study are presented in Tables 16 and 18. The expected results were generated for all samples tested under both Summer and Winter Shipping Temperature Cycles except DX() == Jsample tested at 1014 LoD under Winter Shipping Temperature Cycle. These data are acceptable sincel (014) (>95%) positive results were generated at [0] LoD. The data demonstrate that the urine sample integrity is maintained even when exposed to extremes of temperature during shipping.
| Temperature | Cycle Period | Cycle Period (Hours) | Total Time (Hours) |
|-------------|--------------|----------------------|--------------------|
| (b)(4) | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
{18}------------------------------------------------
| | (b)(4) | |
|--|--------|--|
|--|--------|--|
# Table 16. Summary Results for the Summer Shipping Temperature Cycle.
| Concentration/<br>Organism | CT<br># positive/#tested | GC<br># positive/#tested | CT/GC<br>#negatives/#tested |
|----------------------------|--------------------------|--------------------------|-----------------------------|
| (b)(4) | | | |
# Table 17. Winter Shipping Temperature Cycle.
| Temperature | Cycle Period | Cycle Period (Hours) | Total Time (Hours) |
|-------------|--------------|----------------------|--------------------|
| | | (b)(4) | |
Table 18. Summary Results for the Winter Shipping Temperature Cycle.
| Concentration/<br>Organism | CT<br># positive/#tested | GC<br># positive/#tested | CT/GC<br>#negatives/#tested |
|----------------------------|--------------------------|--------------------------|-----------------------------|
| | (b)(4) | | |
# B. Simple 2 Swab Home Collection Kit (Vaginal)
The stability of the collected vaginal specimens under shipping conditions was evaluated by testing negative vaginal swabs in transport medium and positive vaginal swabs in transport medium (spiked with live CT or GC at (1) LoD (low positive) and 10/4) LoD (moderate positive)) after exposure to the Summer and Winter Shipping Temperature Cycles, as indicated in Tables 15 and 17, respectively. 10(4) |low positive, " | |high positive, and 10 negative samples were subjected to the extended Summer and Winter shipping profiles, as indicated in Tables 15 and 17 above. The results are presented in Tables 19 and 20 below. The expected results were generated for all samples tested under both Summer and Winter Shipping Temperature Cycles. The study results demonstrate that the vaginal swab sample integrity is maintained even when exposed to extremes of temperature during shipping.
# Table 19. Summary Results for the Summer Shipping Temperature Cycle.
| Concentration/<br>Organism | CT<br># positive/#tested | GC<br># positive/#tested | CT/GC<br>#negatives/#tested |
|----------------------------|--------------------------|--------------------------|-----------------------------|
| (b)(4) | | | |
{19}------------------------------------------------
(D)(4)
# Table 20. Summary Results for the Winter Shipping Temperature Cycle.
#### VII Proposed Labeling:
The labeling supports the decision to grant the De Novo request for this device.
#### VIII Identified Risks and Mitigations:
| Identified Risks to Health | Mitigation Measures |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Risk of false results | Certain labeling information including limitations,<br>device descriptions, performance information, and<br>explanations of procedures.<br>Use of certain specimen collection devices.<br>Certain design verification and validation including<br>documentation of device descriptions, certain analytical<br>studies and clinical studies, risk analysis strategies. |
| Failure to correctly interpret test<br>results | Certain labeling information including limitations,<br>device descriptions, performance information, and<br>explanations of procedures.<br>Certain design verification and validation including<br>documentation of device descriptions, certain analytical<br>studies and clinical studies, risk analysis strategies. |
| Failure to correctly operate the<br>device | Certain labeling information including limitations,<br>device descriptions, performance information, and<br>explanations of procedures.<br>Certain design verification and validation including<br>documentation of device descriptions, certain analytical<br>studies and clinical studies, risk analysis strategies. |
#### IX Benefit/Risk Assessment:
# A Summary of the Assessment of Benefit:
The Simple 2 Test is the first over-the counter device for detection of chlamydia and gonorhea that allows for self-collection of specimens in the privacy of one's home (home collection). It is probable that the wide availability of this test without prescription will increase testing for STIs in patients who cannot or chose not to visit a healthcare provider, and who would not be tested otherwise. Because infections with Chlamydia trachomatis and Neisseria gonorrhoeae are often
{20}------------------------------------------------
asymptomatic. opening the access to testing at home will likely help curb the spread of these infections. Further, home collection of specimens for testing for sexually transmitted infections is desirable to patients, facilitating access to information about their sexual health without the need for a visit to a physician. Additionally, identification of these infections early in the course of disease will lead to more effective or more timely treatments and a reduction in the morbidity in sexually active individuals. The data submitted in this De Novo application demonstrates that lay users can effectively collect samples at home without compromising the performance of the test results provided by the Aptima Combo 2 assay.
#### B Summary of the Assessment of Risk:
The risks associated with the test are mainly the possibility of false positive and false negative test results. False negative test results can cause delays to effective treatment, progression to disseminated disease, and spread of infection to other persons throughout the community. According to the CDC guidelines, patients with known exposure to either gonorrhea or chlamydia (e.g., a sexual partner with positive diagnosis), should be empirically treated with antibiotics, regardless of the test result. Thus, a false negative result obtained by persons with known exposures to gonorrhea or chlamydia using this test may lead to delays or lost opportunities for treatment. A false negative result may also be obtained if the test is used within a short time after exposure. False positive results when using this test could lead to an incorrect diagnosis and unnecessary treatment for chlamvdia and gonorrhea. Additionally, false positive results may lead to risk of side effects from unnecessary treatments, as well as psychological distress. Further, the risks associated with the device are failure to correctly interpret test results and failure to correctly operate the device. These risks are mitigated by the special controls, mainly required analytical and clinical performance and required information in the labeling and educational information about CT and GC infections contained in the enclosed FAQ pamphlet.
#### C Summary of the Assessment of Benefit-Risk:
In conclusion, given the available information above, for the following indication statement:
The Simple 2 Test is intended for in vitro detection and identification of Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) in home-collected specimens which are shipped to a CLIA compliant laboratory for testing using the Aptima Combo 2 Assay on the Panther System. This product is available over-the-counter (OTC) to consumers 18 years of age and older.
The Simple 2 Test contains all the necessary components to collect urine from male patients (Simple 2 Urine Home Collection Kit (Penile)) or vaginal swabs from female patients (Simple 2 Swab Home Collection Kit (Vaginal)) in their home, or in similar environments, without supervision from a healthcare provider.
The Simple 2 Test Collection Kits may also be used to self-collect specimens in a clinic.
The testing is performed, as determined to be appropriate, based on the results of LetsGetChecked Suitability Questionnaire.
This test system is not a substitute for visits to a healthcare provider. The information provided by this product should not be used to start, stop, or change any course of treatment unless advised by your healthcare provider.
{21}------------------------------------------------
Testing is limited to the manufacturer, Priva Path laboratories (d.b.a. LetsGetChecked, Inc.).
The probable benefits outweigh the probable risks for the Simple 2 Test. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.
#### X Conclusion:
The De Novo request is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:
Product Code: QYA Device Type: System for detection of nucleic acid from non-viral microorganism(s) causing sexually transmitted infections using home-collected specimens.
Class: II Regulation: 21 CFR 866.3385
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.