cobas EBV, cobas EBV/BKV Control Kit, cobas Buffer Negative Control Kit
Applicant
Roche Molecular Systems, Inc.
Product Code
QLX · Microbiology
Decision Date
Jul 30, 2020
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 866.3183
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN200015 · Jul 30, 2020
cobas EBV, cobas EBV/BKV Control Kit, cobas Buffer Negative Control Kit
Roche Molecular Systems, Inc.
Longitudinal clinical samples from transplant patients; Laboratory-developed test (LDT) results from a major transplant center
The clinical performance of the cobas EBV test was evaluated by comparing its results against a validated, well-established comparator nucleic acid test currently used in routine clinical practice at a major transplant center.
Concordance analysis between cobas EBV and the comparator EBV; Retrospective clinical concordance study; Follow-up/Duration: Longitudinal
Transplant subjects (SOT and HSCT); Sample Size: 439 clinical samples from 72 transplant subjects
Validated, well-established laboratory-developed test (LDT) used in clinical practice
Clinical concordance (percent agreement) between cobas EBV and comparator LDT
Indications for Use
cobas EBV is an in vitro nucleic acid amplification test for the quantitation of Epstein-Barr virus (EBV) DNA in human EDTA plasma on the cobas 6800/8800 Systems. cobas EBV is intended for use as an aid in the management of EBV in transplant patients. In patients undergoing monitoring of EBV, serial DNA measurements can be used to indicate the need for potential treatment changes and to assess response to treatment. The results from cobas EBV are intended to be read and analyzed by a qualified licensed healthcare professional in conjunction with clinical signs and symptoms and relevant laboratory findings. Negative test results do not preclude EBV infection or EBV disease. Test results must not be the sole basis for patient management decisions. cobas EBV is not intended for use as a screening test for donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps).
Device Story
The cobas EBV is a quantitative PCR assay performed on the automated cobas 6800/8800 Systems. It processes human EDTA plasma samples from transplant patients. The system performs automated nucleic acid extraction and purification using magnetic glass particles, followed by PCR amplification of EBV DNA (EBNA-1 and BMRF-2 genes) and a non-EBV DNA Quantitation Standard (DNA-QS). The DNA-QS serves as an internal control for extraction and amplification. The system detects target-specific probes labeled with fluorescent dyes. The software calculates EBV viral load in IU/mL using lot-specific calibration constants and Ct values. Results are reviewed by healthcare professionals to monitor transplant patients for EBV-related conditions, such as post-transplant lymphoproliferative disorders (PTLD), and to guide clinical management decisions. The device provides standardized viral load measurements to assist in assessing treatment response.
Clinical Evidence
No clinical data provided in the summary document. Device requires clinical performance testing documentation, including method comparison studies with an appropriate comparator or prospective clinical studies demonstrating clinical validity, including patient samples near the lower limit of quantitation.
Technological Characteristics
Quantitative nucleic acid amplification test; utilizes cobas 6800/8800 Systems instrumentation; includes cobas EBV Control Kit and cobas Buffer Negative Control Kit; requires specific primer/probe sequences and gene targets; calibrated to recognized consensus standards; software-driven automated processing and analysis.
Indications for Use
Indicated for transplant patients requiring monitoring of Epstein-Barr virus (EBV) DNA levels in EDTA plasma to aid in management, treatment adjustment, and response assessment. Not for use as a donor screening test.
Regulatory Classification
Identification
A quantitative viral nucleic acid test for transplant patient management is identified as a device intended for prescription use in the detection of viral pathogens by measurement of viral DNA or RNA using specified specimen processing, amplification, and detection instrumentation. The test is intended for use as an aid in the management of transplant patients with active viral infection or at risk for developing viral infections. The test results are intended to be interpreted by qualified healthcare professionals in conjunction with other relevant clinical and laboratory findings.
Special Controls
The De Novo request is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request: Product Code(s): QLX Device Type: Quantitative Viral Nucleic Acid Test for Transplant Patient Management Class: II (special controls) Regulation: 21 CFR 866.3183
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the device is not intended for use as a donor screening test for the presence of viral nucleic acid in blood or blood products.
(ii) Limitations which must be updated to reflect current clinical practice. These limitations must include, but are not limited to, statements that indicate:
(A) Test results are to be interpreted by qualified licensed healthcare professionals in conjunction with clinical signs and symptoms and other relevant laboratory results; and
(B) Negative test results do not preclude viral infection or tissue invasive viral disease and that test results must not be the sole basis for patient management decisions.
(iii) A detailed explanation of the interpretation of results and acceptance criteria must be provided and include specific warnings regarding the potential for variability in viral load measurement when samples are measured by different devices. Warnings must include the following statement, where applicable: “Due to the potential for variability in [analyte] measurements across different [analyte] assays, it is recommended that the same device be used for the quantitation of [analyte] when managing individual patients.”
(iv) A detailed explanation of the principles of operation and procedures for assay performance.
(2) Design verification and validation must include the following:
(i) Detailed documentation of the device description, including all parts that make up the device, ancillary reagents required for use with the assay but not provided, an explanation of the methodology, design of the primer/probe sequences, rationale for the selected gene target, and specifications for amplicon size, guanine-cytosine content, and degree of nucleic acid sequence conservation. The design and nature of all primary, secondary and tertiary quantitation standards used for calibration must also be described.
(ii) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions;
(iii) Documentation and characterization (
*e.g.,* determination of the identity, supplier, purity, and stability) of all critical reagents and protocols for maintaining product integrity throughout its labeled shelf-life.(iv) Stability data for reagents provided with the device and indicated specimen types, in addition to the basis for the stability acceptance criteria at all time points chosen across the spectrum of the device's indicated life cycle, which must include a time point at the end of shelf life.
(v) All stability protocols, including acceptance criteria.
(vi) Final lot release criteria along with documentation of an appropriate justification that lots released at the extremes of the specifications will meet the claimed analytical and clinical performance characteristics as well as the stability claims.
(vii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Mode Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel viral stains (
*e.g.,* regular review of published literature and annual in silico analysis of target sequences to detect possible primer or probe mismatches). All results of this protocol, including any findings, must be documented.(viii) Analytical performance testing that includes:
(A) Detailed documentation of the following analytical performance studies: limit of detection, upper and lower limits of quantitation, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, quality control, specimen stability studies, and additional studies as applicable to specimen type and intended use for the device;
(B) Identification of the viral strains selected for use in analytical studies, which must be representative of clinically relevant circulating strains;
(C) Inclusivity study results obtained with a variety of viral genotypes as applicable to the specific assay target and supplemented by in silico analysis;
(D) Reproducibility studies that include the testing of three independent production lots;
(E) Documentation of calibration to a reference standard that FDA has determined is appropriate for the quantification of viral DNA or RNA (
*e.g.,* a recognized consensus standard); and(F) Documentation of traceability performed each time a new lot of the standardized reference material to which the device is traceable is released, or when the field transitions to a new standardized reference material.
(ix) Clinical performance testing that includes:
(A) Detailed documentation from either a method comparison study with a comparator that FDA has determined is appropriate, or results from a prospective clinical study demonstrating clinical validity of the device;
(B) Data from patient samples, with an acceptable number of the virus-positive samples containing an analyte concentration near the lower limit of quantitation and any clinically relevant decision points. If an acceptable number of virus-positive samples containing an analyte concentration near the lower limit of quantitation and any clinically relevant decision cannot be obtained, contrived samples may be used to supplement sample numbers when appropriate, as determined by FDA;
(C) The method comparison study must include predefined maximum acceptable differences between the test and comparator method across all primary outcome measures in the clinical study protocol; and
(D) The final release test results for each lot used in the clinical study.
Submission Summary (Full Text)
{0}------------------------------------------------
## EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR cobas EBV DECISION SUMMARY
## A. De Novo Number
DEN200015
## B. Purpose for Submission
De Novo request for evaluation of automatic class III designation for the cobas EBV test.
#### C. Measurand
EBV DNA
## D. Type of Test
Quantitative Polymerase Chain Reaction (PCR)
## E. Applicant
Roche Molecular Systems, Inc.
## F. Proprietary and Established Names
cobas EBV
## G. Regulatory Information
- 1. Regulation section
- 21 CFR 866.3183
- 2. Classification
Class II
- 3. Product code(s):
QLX
- 4. Panel
{1}------------------------------------------------
Microbiology (83)
## H. Indications For Use
- 1. Indications for use:
cobas EBV is an in vitro nucleic acid amplification test for the quantitation of Epstein-Barr virus (EBV) DNA in human EDTA plasma on the cobas 6800/8800 Systems.
cobas EBV is intended for use as an aid in the management of EBV in transplant patients. In patients undergoing monitoring of EBV, serial DNA measurements can be used to indicate the need for potential treatment changes and to assess response to treatment.
The results from cobas EBV are intended to be read and analyzed by a qualified licensed healthcare professional in conjunction with clinical signs and symptoms and relevant laboratory findings. Negative test results do not preclude EBV infection or EBV disease. Test results must not be the sole basis for patient management decisions.
cobas EBV is not intended for use as a screening test for donors of blood or blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps).
Special conditions for use statement(s)
For prescription use only.
For in vitro diagnostic use only.
- 2. Special instrument requirements
The test is run on the cobas 6800/8800 instrument systems.
# I. Device Description
cobas EBV is a quantitative test performed on the cobas 6800 System and cobas 8800 System. cobas EBV enables the detection of EBV DNA in plasma specimens. The cobas EBV assay is a dual target assay, with both targets using the same dye. The DNA Internal Control, used to monitor the entire sample preparation and PCR amplification process, is introduced into each specimen during sample processing. cobas EBV enables the detection and quantitation of EBV DNA in EDTA plasma from solid organ transplant patients (SOT) and from hematopoietic stem cell transplant (HSCT) patients. The test is intended for use as an aid in the management of SOT patients and HSCT patients.
The cobas EBV consists of:
- Proteinase Solution ●
- DNA Quantitation Standard (DNA QS) ●
- Elution Buffer ●
{2}------------------------------------------------
- Master Mix Reagent 1
- . EBV Master Mix Reagent 2
The EBV viral load is quantified against a non-EBV DNA quantitation standard (DNA-OS), which is introduced into each specimen during sample preparation. The DNA-QS also functions as an internal control for sample preparation and the PCR amplification process.
In addition, the test utilizes the following separately packed and sold control materials:
- 1. cobas EBV Positive Control Kit:
- . EBV Low Positive Control (EBV L(+)C)
- EBV High Positive Control (EBV H(+)C) ●
The positive control contains phage packaged EBV DNA in normal human plasma and serves as a control for the cobas EBV test.
- 2. cobas Negative Control Kit:
- cobas Buffer Negative Control (BUF (-) C) ●
Testing with the cobas EBV test requires the following materials that are not provided:
- cobas OMNI Reagents: Including the following reagents used for specimen ● processing, PCR and detection:
- cobas EBV Assay Specific Analysis Package (ASAP) software .
The cobas EBV test uses sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection, all steps are fully automated by the cobas 6800/8800 platform.
#### Instrumentation and Software
The cobas 6800/8800 platform consists of two instrument versions: the cobas 6800 System, and the cobas 8800 System. Each system is comprised of a cobas 6800 or cobas 8800 instrument, system software, Assay Specific Analysis Packages (ASAP), and a sample source unit, which can be connected to a conveyor system for automated transport of samples to and from the system. The test kits consist of assay-specific reagents and omni reagents (or common reagents) which can be used with any of the cobas assays, and on either the cobas 6800 or the cobas 8800 system.
In addition, the cobas omni (common) reagents and consumables, such as the P-plates, racks, AD-plates, waste bags, pipette tips, and secondary tubes, can be used with any of the cobas assays, and can be used for both the cobas 6800 and the cobas 8800 systems.
#### J. Standard/Guidance Document Referenced
EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition
{3}------------------------------------------------
EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline.
EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition
EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline -Second Edition
# K. Test Principle
The cobas EBV test is a quantitative PCR test performed on the fully automated cobas 6800/ 8800 Systems that detects and quantifies EBV DNA from (EDTA) plasma specimens of transplant patients as follows:
# Target Selection
Selective amplification of EBV target nucleic acid from the sample is achieved by the use of specific forward and reverse primers which are selected to amplify highly-conserved regions of the EBV DNA EBNA-1 and BMRF-2 gene. Specific probes for each amplicon are used to detect and quantify the EBV targets. Selective amplification of DNA-OS is achieved by the use of DNA-QS specific forward and reverse primers, selected to have no homology with the EBV genome, detected through a DNA-QS specific probe.
## Sample Preparation (Nucleic Acid Extraction and Purification)
Nucleic acid from patient samples and external controls are extracted upon addition of a DNA Quantitation standard (DNA-QS). The DNA-QS molecules are extracted simultaneously with the samples/controls serving as an extraction control. Viral nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid, along with the added DNA-QS binds to magnetic glass particles. Unbound substances and impurities are removed with subsequent wash reagent steps and purified nucleic acid is then eluted from the magnetic glass particles with elution buffer.
#### Nucleic Acid Amplification and Target Detection
The cobas EBV master mix contains detection probes which are specific for the two EBV target sequences and the DNA-QS nucleic acid, respectively. The two EBV specific detection probes are labeled with the same fluorescent dye while the DNA-OS detection probe is labeled with a second fluorescent dye both acting as reporter dyes. Each probe also has a second dye which acts as a quencher that suppresses the fluorescent signals of the intact probes when they are not bound to their respective target sequence. Target bound probes however, emit fluorescence of the two reporter dyes. This fluorescence is measured at defined wavelengths, thus permitting simultaneous detection and discrimination of the EBV targets and the DNA-QS amplification products generated by a thermostable DNA polymerase enzyme.
#### EBV DNA Quantitation
During the extension phase of the PCR process fluorescence readings are processed to generate Ct values for the EBV DNA target and the EBV QS DNA. The lot-specific
{4}------------------------------------------------
calibration constants provided with the cobas EBV test are used to calculate the titer value for the specimens and controls based on both the EBV DNA target and the EBV QS DNA Ct values. EBV viral load results are reported in International Units/mL (IU/mL).
#### L. Performance Characteristics
## 1. Analytical performance
#### a. Precision
The Precision was assessed for the predominant genotype (EBV Genotype 1) with seven panel members ranging from (b) (4) Panel members were prepared by spiking a high titer EBV lambda phagemid into EBV negative EDTA plasma.
Precision was calculated on results generated over twelve days using three kit lots and three cobas 6800 systems by three operators. Per test day, two (2) runs were performed containing three (3) within-run replicates per panel member. The study design accounts for a total of 72 replicates per panel member.
For data analysis only samples with titers above the Lower Limit of Ouantification (LLoO). (i.e., (b) (4)), were used. Precision was determined according to the CLSI guideline EP05-A3 as a multivariance analysis accounting for reagent lots, operators/instruments, days, runs and within-run replicates.
| Nominal<br>Concentration | Assigned<br>concentration<br>(IU/mL) | Standard Deviation | | | |
|--------------------------|--------------------------------------|--------------------|-------|-------|----------|
| | | Lot 1 | Lot 2 | Lot 3 | All Lots |
| 5.00E+07 | 5.40E+07 | 0.03 | 0.04 | 0.04 | 0.04 |
| 1.00E+06 | 1.08E+06 | 0.02 | 0.03 | 0.02 | 0.02 |
| 1.00E+05 | 1.08E+05 | 0.02 | 0.02 | 0.03 | 0.02 |
| 1.00E+04 | 1.08E+04 | 0.04 | 0.02 | 0.03 | 0.03 |
| 1.00E+03 | 1.08E+03 | 0.05 | 0.05 | 0.05 | 0.05 |
| 1.00E+02 | 1.08E+02 | 0.17 | 0.18 | 0.15 | 0.17 |
| 6.00E+01 | 6.48E+01 | 0.17 | 0.17 | 0.13 | 0.16 |
#### Table 1: Precision - Standard Deviation
* Titer data are considered to be log-normally distributed and are analyzed following logio transformation. Standard deviations (SD) columns present the total of the log-transformed titer for each of the three reagent lots.
The results for cobas EBV Precision are shown in Table 1. The Variance Component Analysis demonstrated the contribution of components of variance to the total precision
{5}------------------------------------------------
variance (Table 2). Overall, the total Precision as SD of log10 titer is comparable across all kits.
| Panel | Assigned<br>Concentration<br>(IU/mL) | | N | Instrument/<br>Operator | Kit<br>Lot | Day | Run | Within<br>Run | Total |
|--------|--------------------------------------|---------------------------|----|-------------------------|------------|-----|-----|---------------|-------|
| Member | Titer<br>(IU/mL) | Log10<br>Titer<br>(IU/mL) | | %CV | %CV | %CV | %CV | %CV | %CV |
| PM01 | 1.08E+06 | 7.73 | 72 | 5 | 1 | 2 | 2 | 8 | 10 |
| PM02 | 1.08E+05 | 6.03 | 72 | 1 | 4 | 1 | 3 | 4 | 7 |
| PM03 | 5.94E+03 | 5.03 | 72 | 2 | 2 | 3 | 2 | 5 | 7 |
| PM04 | 1.08E+04 | 4.03 | 72 | 2 | 1 | 3 | 3 | 7 | 8 |
| PM05 | 1.08E+03 | 3.03 | 72 | 4 | 1 | 4 | 4 | 10 | 12 |
| PM06 | 1.08E+02 | 2.03 | 72 | 3 | 5 | 8 | 14 | 42 | 43 |
| PM07 | 6.48E+01 | 1.81 | 68 | 7 | 3 | 6 | 12 | 39 | 40 |
Table 2: Lognormal Percent Coefficient of Variation (% CV) *
* Titer data are considered to be log-normally distributed and the %CV values are analyzed as Lognormal CV (%) = sqrt (10^ [SD^2 * ln (10)] - 1) * 100%
#### b. Reproducibility
Reproducibility of cobas EBV was evaluated in EDTA plasma across three Reagent Lots, three Test Sites, three Instruments (two cobas 6800 Systems and one cobas 8800 System). Two runs were performed per operator per day (1 run = 1 batch; 1 batch = 1 panel + 3 controls) over five days per reagent lot and each run had three replicates per panel member. The total number of tests (not including controls) was as follows: 3 lots × 3 sites × 5 days/lot × 2 runs × 3 replicates/concentration = 270 test results/concentration.
Test panel members were prepared from EBV-VCA IgG sero-negative and RNA negative EDTA plasma spiked with EBV genotype 1 genomic material in the form of EBV cell culture supernatant. EBV phagemid was used for preparing the high-positive panel member (5x107 IU/mL) due to the lack of adequate volumes of high concentration sample. Test panel members had the following concentrations: Negative, 105 IU/mL, 5x103 IU/mL, 5x104 IU/mL, 5x105 IU/mL, and 5x107 IU/mL. Two invalid results were excluded. The results are summarized in Table 3 below:
{6}------------------------------------------------
Table 3: Reproducibility Study
| EBV DNA<br>Concentration<br>(log10 IU/mL) | | Number<br>of<br>Testsb | Percent of Total Variance<br>[Log-Normal CV (%)] | | | | | Total<br>Precision | |
|-------------------------------------------|-------------------|------------------------|--------------------------------------------------|----------------|------------------|----------------|------------------|--------------------|------------------------------|
| Expected | Observed<br>Meana | | Lot | Site | Day/<br>Operator | Batch | Within<br>-Batch | SDc | Log-<br>Normal<br>CV<br>(%)d |
| 2.02 | 2.09 | 270 | 11%<br>(11.97) | 2%<br>(5.30) | 0%<br>(0.00) | 3%<br>(6.34) | 84%<br>(34.25) | 0.158 | 37.56 |
| 3.70 | 3.68 | 270 | 43%<br>(10.07) | 15%<br>(5.92) | 0%<br>(0.00) | 16%<br>(6.23) | 26%<br>(7.81) | 0.067 | 15.43 |
| 4.70 | 4.68 | 270 | 39%<br>(8.54) | 10%<br>(4.24) | 0%<br>(0.00) | 24%<br>(6.63) | 28%<br>(7.18) | 0.059 | 13.70 |
| 5.70 | 5.50 | 268 | 7%<br>(11.39) | 58%<br>(34.36) | 0%<br>(0.00) | 21%<br>(20.18) | 15%<br>(17.08) | 0.191 | 46.16 |
| 7.70 | 7.76 | 270 | 27%<br>(8.63) | 15%<br>(6.52) | 0%<br>(0.88) | 13%<br>(6.01) | 45%<br>(11.26) | 0.073 | 16.83 |
a Calculated using SAS MIXED procedure.
b Number of valid tests with detectable DNA level.
C Calculated using the total variability from the SAS MIXED procedure.
d Lognormal CV (%) = sqrt (10^ [SD^2 * In (10)] - 1) * 100
Note: The table only includes results with detectable DNA level. CV (%) = percent coefficient of variation; SD = standard deviation.
Analysis of variance and a mixed model that included lot, site, day/operator, batch and within-batch (random error) as random effects was performed. The variance contribution of each component to the total variance was estimated. The range of the total lognormal coefficient of variation, among positive panel members, was from 13.7% to 46.16%. The largest total lognormal coefficient of variation was observed in the expected 5x10 -panel member and most of that variability (58% of the total variance) was attributed to site. The largest total lognormal coefficient of variation observed in the lowest panel member (37.56%) was explained by the within-batch component.
{7}------------------------------------------------
#### c. Linearity
#### Linearity of the cobas EBV with Genotype 1
Linearity of the cobas EBV test was evaluated for the predominant EBV genotype (GT 1) in EDTA-plasma using a 17-member test panel. Eleven panel members were generated using an EBV phagemid DNA and covered the entire linear range. Six panel members were generated using an EBV GT1-positive clinical specimen. Due to the lack of sufficient volume of high positive EBV-positive transplant patient samples the clinical sample panel was designed to cover the range from 13.6 IU/mL to 4 log IU/mL (i.e., only the low and intermediate part of the measuring range) to overlap with the phagemid-based higher concentration panel members.
Panel member concentrations spanned the range of 1.5E+01 IU/mL to 2.00E+08 IU/mL (nominal concentration). Each panel member was tested in 36 replicates across three lots of cobas EBV test reagents (12 replicates/lot) and the results of the study are presented in Table 4 and Figure 1 below. The Linearity panel was tested across three instruments by three operators. Resulting data were analyzed to identify the linear range according to CLSI guideline EP6-A and the best fitting polynomial regression fit.
| Lot | Equation 1st Order<br>$Y=b_0+b_1x$ | Equation 2nd Order<br>$Y=b_0+b_1x+b_2x^2$ | Maximum<br>Difference<br>(log10 IU/mL) |
|---------------------------------------|------------------------------------|-------------------------------------------|----------------------------------------|
| Clinical Sample | | | |
| 1 | (b) (4) | | |
| 2 | (b) (4) | | N/A |
| 3 | (b) (4) | | N/A |
| Phagemid | | | |
| 1* | (b) (4) | | |
| 2 | (b) (4) | | N/A |
| 3 | (b) (4) | | N/A |
| Clinical Sample and Phagemid Combined | | | |
| 1 | (b) (4) | | N/A |
| 2 | (b) (4) | | N/A |
| 3 | (b) (4) | | N/A |
| ALL | (b) (4) | | |
#### Table 4: cobas EBV Linearity with EBV Genotype 1
bo-Intercept; bj=slope; all coefficients are provided in log10 IU/mL; best fitting model is bolded N/A=not applicable; if linearity is described best by the 1st order model, there is no deviation to be shown between the 1st order model and any higher order model
* For this lot the 3d order was the best fitting model, however, the difference to the 1st order regression model was so minor that it not shown here.
{8}------------------------------------------------
Image /page/8/Figure/0 description: The image shows the title of a figure. The title is "Figure 1: Linearity Across Both Panel Types (all lots)". The title is written in a bold, sans-serif font. The figure number is 1.
Image /page/8/Figure/1 description: The image is a scatter plot that shows the relationship between AssignedLogTiter and ObservedLogTiter. The x-axis represents AssignedLogTiter, ranging from 1 to 8, while the y-axis represents ObservedLogTiter, ranging from 0 to 9. The plot shows a cluster of data points between 1 and 2 on the x-axis, with a linear trend between the two variables as x increases. A dashed line is plotted through the data, showing a positive correlation between the two variables.
Except for Lot 1 (clinical sample), the 1st order model is the best fitting model for all lots and all panel members (clinical sample and phagemid) when analyzed separately. However, for all panel member and lots combined the 2nd order model (62) is significant at a 5% level. Therefore, the 2nd order polynomial was chosen as the best-fitting regression model for the analysis. However, the absolute difference between the 1st order and the better fitting 2nd order regression is minimal (i.e., equal or less than ± 0.01 log10 IU/mL). Across the linear range, the accuracy of the test was within ± 0.15 log10 IU/mL (Mean Square Error).
Based on the LLoQ (35 IU/mL) and the determined linear range, the claimed linear measurement range of the cobas EBV test is from 35 IU/mL (LLoQ) to 1.0E+08 IU/mL (ULoQ).
#### Verification of Linearity of cobas EBV with Genotype 2
Linearity was assessed with a cell culture derived EBV Genotype 2 panel (EBV strain Jiyoye) spanning the expected linear range from the 1 x 10° IU/mL (at or near expected ULoQ) to 30 IU/mL (at or near expected LLoQ). The following 8 panel members were prepared in EBV negative EDTA-plasma: 30 IU/mL, 150 IU/mL, 3, 4, 5, 6, 7, and 8 log IU/mL. Twelve replicates across three test-specific reagent lots were tested per concentration level (i.e., 4 replicates per kit lot). Data were analyzed according to CLSI guideline EP6-A and the 3td order polynomial regression was the best fitting modes using a 5% significance level. The following equations were obtained.
| | 1st order: | y = 0.2124 + 0.9713x |
|--|------------|---------------------------------------------|
| | 2nd order: | y = 0.2800 + 0.9346x + 0.0039x2 |
| | 3rd order: | y = 0.7323 + 0.5420x + 0.0980x2 + -0.0066x3 |
{9}------------------------------------------------
The differences in logio values calculated from the 1st order and the 3th order regression were minor, ranging from -0.06 log IU/mL to +0.08 log IU/mL. Linearity for EBV genotype 2 is shown in
Figure 2. Across the linear range, the accuracy of the test was within ± 0.12 log10 IU/mL (Mean Square Error).
Image /page/9/Figure/2 description: The image shows the label (b) (4) in the upper left corner. The rest of the image is a gray rectangle. The gray rectangle takes up most of the image.
Figure 2: Verification of Linearity for EBV Genotype 2
Figure 1 Regression plot of EBV Genotype 2 - including outliers
#### d. Traceability, Stability, Expected values (controls, calibrators, or methods)
#### i) Traceability
Several standards and controls were used during development of the cobas EBV test to provide traceability to the WHO EBV Standard [1st WHO International Standard for Epstein-Barr Virus for Nucleic Acid Amplification Techniques (NIBSC 09/260)]. The standards used during development of the test include the WHO EBV Standard, the Roche Molecular Systems (RMS) EBV Secondary Standard, and the RMS EBV Calibration Panel.
Traceability of the calibration panel and the RMS EBV Secondary Standard to the 1st EBV WHO Standard was verified as shown in Figure 3.
| The concentration range tested for the EBV WHO Standard was from (b) (4) | 10 | |
|--------------------------------------------------------------------------|---------------------------------------------------|--|
| (b) (4) | . the RMS EBV Secondary Standard was tested at | |
| (b) (4) | and the RMS EBV Calibration Panel was tested from | |
| (b) (4) | | |
#### Figure 3: Traceability to the 1st WHO International Standard for EBV (NIBSC 09/260) using cobas EBV
{10}------------------------------------------------
Image /page/10/Figure/0 description: The image is a scatter plot comparing observed EBV DNA concentration to expected EBV DNA concentration. The x-axis represents the expected EBV DNA concentration (Log 10 IU/mL) ranging from 1 to 9, while the y-axis represents the observed EBV DNA concentration (Log 10 IU/mL) ranging from 1 to 9. The plot includes data points for EBV Calibration Panel, EBV Secondary Standard, and EBV WHO Standard, with trend lines showing a positive correlation between expected and observed concentrations.
Figure 3 shows the bivariate fit of observed EBV DNA concentration (log10 IU/mL) by expected EBV DNA concentration (log10 IU/mL). Observed quantitation values are similar to the expected values with deviation of not more than 0.15 log10 IU/mL. All materials demonstrated co-linear dilution performance across the linear range of cobas EBV (Figure 2) The maximum deviation was observed at 200 IU/mL (approximately 6x LLoQ). The following linear regression equations were obtained:
EBV Calibration Panel: y = 1.000x - 0.002; R2 = 1.000 EBV 1st WHO Standard: y = 0.975x + 0.159; R2 = 0.983
Based on these results, the calibration and standardization process of cobas EBV provides quantitation values for the cobas EBV calibration panel and the RMS EBV Secondary Standard provide traceability to the 1st WHO international Standard for EBV.
#### ii) Expected values
To monitor the assay performance, reagent performance, and procedural errors, positive and negative external controls must be run in accordance with the guidelines or requirements of local, state, and/or federal regulations or accrediting organizations.
External controls are provided separately from the cobas EBV test kit in the cobas EBV Positive Control Kit and the cobas Negative Control. The cobas EBV Positive Control includes an EBV High Positive control, and a Low Positive control containing EBV phagemid-DNA diluted into negative human plasma. The cobas Negative Control includes Tris buffer. The cobas Negative Control, the EBV Low Positive Control, and the EBV High Positive Control must be included in each run. The validity of the results for the controls as well as for the DNA-QS is determined by the assay specific analysis software package used by the cobas 6800/8800 instrument. The amount of EBV DNA for EBV high and low positive controls must fall within their acceptable titer ranges.
{11}------------------------------------------------
The failure rates of the kit controls and samples were assessed by evaluating outcomes of the nine analytical studies (LoD, LoD verification for Genotype, Linearity, Linearity verification for Genotype, Precision, Cross contamination, Lot interchangeability, Accuracy, and LoD in Plasma vs Buffer). Results demonstrated an overall invalid rate of 0% for QS and RMC and a sample failure rate of 0.02%.
## iii) Stability
#### iii.a. Clinical Specimen Stability
The clinical specimen stability of EBV viral target in whole blood specimens collected in EDTA-plasma preparation tubes and/or plasma samples after various storage conditions with or without freeze-thawing cycles was evaluated using cobas EBV for use on the cobas 6800/8800 Systems.
Freshly drawn whole blood from ten unique individual EBV-negative donors collected in EDTA-plasma tubes (five donor samples collected in PPT and five in lavender top tubes) were spiked with EBV to a concentration of 150 IU/mL (approximately 5xLLoQ). The reference time point (T0) was processed directly after spiking of the target by separating the blood into EDTA-plasma; all other samples were stored at indicated conditions until further processing. All samples used for this study in each PPT/ lavender top tube type were tested unspiked at time point 0 (T0).
All EBV-positive samples tested generated positive results for EBV and the mean log10 titers including the two-sided 95 confidence interval of each of the tested time points/conditions and tube types was within ±0.5 logio of the mean logio titer of the respective reference condition (T0 = reference).
The results support the following storage conditions for whole blood collected in BD Vacutainer PPT or BD Vacutainer EDTA tubes (lavender top) and the respective separated plasma:
- Whole blood collected in EDTA-plasma tubes (lavender top and PPT) may be ● stored or transported for up to 24 hours at 2℃ to 25℃ before further processing and plasma separation.
- . Afterwards, whole blood samples should be centrifuged, and the resulting plasma samples are additionally stable for:
- o 24 hours at 2°C to 30°C in primary or secondary tubes, followed by
- o up to 6 days at 2℃ to 8℃ in primary or secondary tubes, or
- o up to 6 months at -15°C to -80°C in secondary tubes.
- Plasma samples are stable in secondary tubes for up to four freeze/thaw cycles . when frozen between -15°C to -80°C.
{12}------------------------------------------------
#### iii.b. Open kit and On-board Stability
The data submitted support the claim that, once opened, the cobas EBV 192T test-specific reagent cassettes are stable for up to 90 days at 2-8℃ (Open Kit Stability) and remain stable for up to 40 hours at 37°C (On-Board Stability). Furthermore. 192T test-specific cassettes once opened are re-usable for up to 40 runs.
#### iii.c. Reagent Stability
Three lots of the cobas EBV were tested to demonstrate stability of the test-specific reagents of cobas EBV and cobas EBV Control Kit when stored at stressed temperature conditions (accelerated stability) and at the targeted storage temperature of 2°C to 8°C (real-time stability).
Real-time stability: The data submitted supports a shelf-life stability claim of 12 months. when stored at 2-8°C.
#### e. Detection limit
#### i) Limit of Detection (LoD) using the 1st WHO International Standard for EBV
The LoD of the cobas EBV test for the 1st WHO EBV Standard (Genotype 1) was determined by analysis of serial dilutions of the Standard diluted into a pooled EDTA-plasma derived from EBV IgG/IgM negative individuals following the recommendations in CLSI Guideline EP17-A2. Panels of six concentration levels plus a blank were tested with three lots of cobas EBV test reagents and three instruments with multiple runs and operators over a period of three days. Each dilution was determined in " replicates per lot and day (n=10)(4) total replicates per day). The results from testing the WHO EBV Standard in EDTA plasma as well as the calculated LoD values are shown in Table 5. The LoD values in Table 5 were determined by Probit analysis and by 95% hit rate.
| Kit Lot | Nominal<br>Concentration<br>(IU/mL) | Number of<br>Positive<br>Replicates | Number of<br>Valid<br>Replicates | Hit Rate<br>[%] | LoD by<br>Probit<br>[95% CI] |
|---------|-------------------------------------|-------------------------------------|----------------------------------|-----------------|------------------------------|
| Lot 1 | (b) (4) | | | | |
#### Table 5: LoD with EBV DNA 1st WHO International Standard in EDTA Plasma
{13}------------------------------------------------
| Kit Lot | Nominal<br>Concentration<br>(IU/mL) | Number of<br>Positive<br>Replicates | Number of<br>Valid<br>Replicates | Hit Rate<br>[%] | LoD by<br>Probit<br>[95% CI] |
|----------------------|-------------------------------------|-------------------------------------|----------------------------------|-----------------|------------------------------|
| Lot 2 | (b) (4) | | | | |
| Lot 3 | | | | | |
| All lots<br>combined | | | | | |
When determined by Probit analysis, the different lots have similar LoD for all tested lots; the highest LoD of 18.8 IU/mL was obtained with lot 1, which is only slightly lower than the LoD determined by 95% hit rate. The LoD by 95% hit rate was (b) (4) and was the same for all tested lots. The claimed LoD value is (b) (4) determined by the least sensitive kit lot and this concentration was used in studies for confirmation of the LoD.
#### ii) Limit of Detection (LoD) Confirmation with EBV Genotype 2
The Limit of Detection (18.8 IU/mL) was verified for the cobas EBV test with EBV genotype 2 following the CLSI Guideline EP17-A2. EBV cell culture supernatants for genotype 2 strain Jiyoye (GT2-J) were diluted to three different concentration levels in EBV negative EDTA plasma. The hit rate determination was performed with 63 replicates for each level. Testing was conducted with three lots of cobas EBV reagents across three days of testing.
The results are shown in Table 6 and verify that a hit rate of 95% or higher was observed above 18.8 IU/mL for EBV genotype 2. Thus, the observed hit rates verify the LoD for EBV Genotype 2 at 18.8 IU/mL.
{14}------------------------------------------------
| Lot | Nominal<br>Concentration<br>(IU/mL) | Number of<br>Positive<br>Replicates | Number of<br>Valid<br>Replicates | Hit Rate<br>[%] | LOD by Hit Rate |
|----------|-------------------------------------|-------------------------------------|----------------------------------|-----------------|-----------------|
| Lot 1 | (b) (4) | | | | |
| Lot 2 | | | | | |
| Lot 3 | | | | | |
| All Lots | | | | | |
## Table 6: Verification of the LOD for EBV Genotype 2
# iii) Limit of Detection (LoD) in Plasma vs Buffer
This study evaluated whether the Limit of Detection in Generic Specimen Diluent (GSD) is equivalent to the LoD in EDTA-plasma in the cobas EBV assay, so that GSD can be used as a negative control for cobas EBV. Three independent dilution series (0.5x, 1.0x, and 1.5x LoD of cobas EBV) were prepared on three consecutive days using EBV WHO International Standard in GSD and tested.
The results (Table 7) demonstrated that the hit rate was 98.4% and 92.1% at 1.5xLoD and 1.0xLoD, respectively. The study demonstrated a comparable LoD performance of cobas EBV in Plasma and GSD.
#### Table 7: Hit Rates for all Dilution Series combined
| Conc. Level | Conc. IU/mL | Hit Rate % | Two-sided 95% CI |
|-------------|-------------|------------|------------------|
| (b) (4) | | | |
#### Lower Limit of Quantitation (LLoQ) f.
# LLoQ using the 1st WHO International Standard for EBV
The LLoQ for cobas EBV was calculated using data generated from the Limit of Detection study using EBV WHO International Standard. The LLoQ was determined, per CLSI document EP17-A2, as the lowest titer within the linear range with a hit rate of at least 95% and at which the total analytical error (TAE) meets both of the following two criteria:
- The TAE, when calculated as (b) (4) and
{15}------------------------------------------------
- The TAE has to be such that the standard deviation for the difference between two measurements calculated as (b) (4)
The LLoQ was determined for each kit lot; all calculations are based on non-rounded values but the results shown in here are rounded.
| Lot | Nominal<br>Concentrati<br>on (IU/mL) | Log10<br>titer<br>Nominal | Mean log10<br>titer<br>Observed | SD<br>(log10) | Absolute<br>Bias | TAE<br>([Bias] +<br>2SD) | Difference<br>Between<br>Measurements<br>(SD) |
|-------------|--------------------------------------|---------------------------|---------------------------------|---------------|------------------|--------------------------|-----------------------------------------------|
| 1 | 20 | 1.30 | 1.34 | 0.38 | 0.04 | 0.80 | 1.07 |
| 1 | 35 | 1.54 | 1.60 | 0.25 | 0.06 | 0.56 | 0.71 |
| 1 | 50 | 1.70 | 1.74 | 0.22 | 0.04 | 0.49 | 0.63 |
| 2 | 20 | 1.30 | 1.29 | 0.37 | 0.01 | 0.75 | 1.05 |
| 2 | 35 | 1.54 | 1.58 | 0.27 | 0.04 | 0.58 | 0.76 |
| 2 | 50 | 1.70 | 1.77 | 0.23 | 0.07 | 0.53 | 0.65 |
| 3 | 20 | 1.30 | 1.33 | 0.31 | 0.03 | 0.65 | 0.88 |
| 3 | 35 | 1.54 | 1.58 | 0.32 | 0.04 | 0.67 | 0.89 |
| 3 | 50 | 1.70 | 1.76 | 0.21 | 0.06 | 0.48 | 0.60 |
| All<br>lots | 20 | 1.30 | 1.32 | 0.35 | 0.02 | 0.73 | 1.00 |
| | 35 | 1.54 | 1.59 | 0.28 | 0.05 | 0.60 | 0.79 |
| | 50 | 1.70 | 1.76 | 0.22 | 0.06 | 0.50 | 0.63 |
Table 8: LLoQ - TAE and Difference between Measurements
The LLoQ was determined to be 35 IU/mL for lots 1 and 20 IU/mL for lot 3, calculated based on the calculation of the Total Analytical Error (TAE) and the difference between two measurements. The LLoQ for the cobas EBV test is 35 IU/mL.
#### g. Analytical specificity
#### i) Cross reactivity
For potential cross reactants 35 microorganisms, including 17 viral isolates, 15 bacterial strains and three fungal isolates were used and divided into seven pools with 4 to 5 microorganisms per cross reactant pool and HCV as single interferent. Potential cross reactants in EBV-seronegative EDTA plasma were tested in the absence and presence of EBV DNA at a concentration of 5xLLoQ. Potential cross reactants were tested at bl(4)
Results are shown in Table 9. For EBV-negative samples the negativity rate was determined. For EBV-positive samples the positivity rate was determined together with the correct
{16}------------------------------------------------
quantitation of EBV DNA by computing the Mean concentration detected across the replicates, the SD, and the difference between the control condition (no cross reactant) and the test condition containing the potential cross reactant organism. The mean log10 titer of each of the positive EBV samples containing potentially cross-reacting organisms was within ± 0.5 log10 of the mean log10 titer of the respective positive spike control.
{17}------------------------------------------------
| | Organisms | Test<br>Concentrat<br>ion | No<br>EBV | | EBV | | |
|------|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------------------|------------------------|-----------------|---------------|-----------------------------------------|
| Pool | | | Negati<br>vity<br>Rate | Positi<br>vity<br>Rate | Mean<br>[Log10] | SD<br>[Log10] | Mean<br>Difference<br>in log10<br>Titer |
| 1 | HSV 1<br>HSV 2<br>HSV 6<br>HSV 7<br>HSV 8 | (b) (4) | | | | | |
| 2 | Adenovirus<br>Type 5<br>Candida<br>albicans<br>Chlamydia<br>trachomatis<br>Clostridium<br>perfringens<br>CMV | | | | | | |
| 3 | Enterococcus<br>faecalis<br>Escherichia coli<br>HBV<br>HIV-1<br>HIV-2 | | | | | | |
| 4 | Klebsiella<br>pneumoniae<br>Listeria<br>monocytogenes<br>Mycobacterium<br>avium<br>Mycoplasma<br>pneumoniae<br>Neisseria<br>gonorrhoeae | | | | | | |
## Table 9: Cross Reactivity
{18}------------------------------------------------
| | | (b) (4) |
|---|---------------------------------------------------------|---------|
| 5 | Parvovirus B19 | |
| | Propionibacterium acnes | |
| | Salmonella<br>enterica | |
| | Simian Virus 40 | |
| | Staphylococcus<br>aureus | |
| 6 | Staphylococcus<br>epidermis | |
| | Streptococcus<br>pyogenes | |
| | Streptococcus<br>pneumoniae | |
| | VZV | |
| 7 | Aspergillus<br>niger | |
| | Cryptococcus<br>neoformans | |
| | Human<br>Papilloma Virus<br>(HPV) | |
| | JC Virus | |
| | BK Virus | |
| | Hepatitis C<br>Virus (HCV) | |
| | Control (EBV<br>negative) | |
| | Control (EBV<br>positive) – for<br>Pools 1-6 and<br>HCV | |
| | Control (EBV<br>positive) – for<br>Pool 7 | |
*cp/mL= copies /mL
{19}------------------------------------------------
## ii) Endogenous Interference
The effect of potentially interfering endogenous substances on the sensitivity/quantitation of cobas EBV was determined by testing 20 individual clinical EBV-seronegative samples spiked with selected endogenous substances and EBV target at 150 IU/mL (5xLLoQ). The negative sample spiked solely with EBV target was used as a Positive Spike Control (PSC). To analyze specificity, the same 20 individual clinical negative samples were individually spiked with potentially interfering endogenous substances and tested in the absence of EBV target DNA. The un-spiked samples were used as Negative Spiked Controls (NSC). Interferent concentrations were used as recommended by the CLSI guideline EP7-A2. Human DNA levels were tested at 2mg/mL. Control conditions were tested with one replicate per specimen, and test conditions were tested with 3 replicates per specimen. Results are summarized in Table 10.
| Interferent | C1 | No EBV | | EBV [150 IU/mL] | | | |
|------------------------|---------|--------------------|--------------------|-----------------|------------------|----------------|-----------------------------------------|
| | | Negativity<br>Rate | Positivity<br>Rate | Mean Ct | Mean<br>[Log 10] | SD<br>[Log 10] | Mean<br>Difference<br>in log10<br>Titer |
| Control | (b) (4) | | | | | | |
| NaOH2 | | | | | | | |
| Albumin | | | | | | | |
| Bilirubin<br>(conj.) | | | | | | | |
| Bilirubin<br>(unconj.) | | | | | | | |
| Human DNA | | | | | | | |
| Hemoglobin | | | | | | | |
| Triglycerides | | | | | | | |
Table 10: Endogenous Interference
1C= Test Concentration; 2 solvent control; 3 0.2 g/L = 342 umol/L
EBV-negative samples with endogenous interferents all produced valid negative results (target not detected) in the presence of endogenous interferents.
For EBV-positive samples with endogenous interferents the mean log10 titer of each of the positive EBV samples containing endogenous interferents was within ± 0.05 logio of the mean log10 titer of the spike control.
{20}------------------------------------------------
#### iii) Exogenous Interference
The effect of potentially interfering exogenous substances on the sensitivity/quantitation of cobas EBV was determined by testing 10 individual clinical EBV-negative samples spiked with pools of 24 commercially available drugs at three times the plasma peak level per CLSI EP7-A2. The same samples were also tested in the presence of EBV target at 150 IU/mL (5x LLoQ). The negative sample spiked solely with EBV target was used as a Positive Spike Control (PSC). The un-spiked samples were used as Negative Spiked Controls (NSC). Conditions were tested with 3 replicates per specimen. The following drugs were tested, and the results are summarized in Table 11 below.
- Pool 1: Azathioprine, Sulfamethoxzole, Trimethoprim. Cefotan, Cidofovir .
- Pool 2: Foscarnet, Piperacillin, Tazobactam, Prednisone, Vancomycin .
- Pool 3: Cyclosporine, Everolimus, Fluconazole, Ganciclovir .
- Pool 4: Mycophenolate mofetil, Mycophenolic acid, Valganciclovir .
- Pool 5: Sirolimus, Tacrolimus .
- Pool 6: Letermovir, Micafungin, Acyclovir Clavulanate potassium .
- Pool 7: Ticarcillin disodium .
| | No EBV | EBV [150 IU/mL] | | | | |
|------------|--------------------|--------------------|------------|------------------|----------------|-----------------------------------------|
| Pool | Negativity<br>Rate | Positivity<br>Rate | Mean<br>Ct | Mean<br>[Log 10] | SD<br>[Log 10] | Mean<br>Difference<br>in log10<br>Titer |
| Pool 11 | 10/10<br>100% | 30/30<br>100% | 34.5 | 2.22 | 0.18 | -0.18 |
| Pool 21 | 10/10<br>100% | 30/30<br>100% | 34.7 | 2.19 | 0.15 | -0.21 |
| Pool 32 | 10/10<br>100% | 30/30<br>100% | 34.7 | 2.23 | 0.18 | -0.17 |
| Pool 42 | 10/10<br>100% | 30/30<br>100% | 34.7 | 2.23 | 0.15 | -0.17 |
| Pool 53 | 10/10<br>100% | 30/30<br>100% | 34.6 | 2.24 | 0.19 | -0.16 |
| Pool 61 | 10/10<br>100% | 30/30<br>100% | 34.7 | 2.20 | 0.16 | -0.20 |
| Pool 71 | 10/10<br>100% | 30/30<br>100% | 34.6 | 2.21 | 0.17 | -0.19 |
| PBS SC | 10/10<br>100% | 30/30<br>100% | 34.4 | 2.30 | 0.19 | -0.10 |
| DMSO SC | 10/10<br>100% | 30/30<br>100% | 34.5 | 2.26 | 0.15 | -0.13 |
| Ethanol SC | 10/10<br>100% | 30/30<br>100% | 34.4 | 2.37 | 0.23 | -0.03 |
#### Table 11: Exogenous Interference
{21}------------------------------------------------
| Negative<br>Control | 10/10<br>100% | - | N/A | N/A | N/A | N/A |
|---------------------|---------------|---------------|------|-----|------|-----|
| Positive<br>Control | - | 30/30<br>100% | 34.2 | 2.4 | 0.22 | N/A |
The superscripts in Pools 1-7 indicate the solvent that was used for constituting the interferents (i.e., 1 = PBS; 2 = DMSO; and 3 = Ethanol); SC = solvent control.
- h. Cross Contamination
The cross-contamination rate for cobas EBV was determined by testing 240 replicates of an EBV-negative matrix sample and 225 replicates of a high titer EBV sample at approximately 2.00E+07 IU/mL. In total, five runs were performed with positive and negative samples in a checkerboard configuration.
All 240 replicates of the negative sample were negative, resulting in a cross-contamination rate of 0% (upper one-sided 95% confidence interval: 1.24%).
- Assay cut-off i.
Not applicable
- 2. Comparison studies
- a. Method comparison with predicate device
Not applicable
- b. Matrix comparison
Not applicable
- 3. Clinical studies
#### Concordance of cobas EBV with a Comparator EBV Test
The clinical performance of cobas EBV was compared to a validated well-established comparator nucleic acid test(comparator EBV) by measuring EBV DNA levels in longitudinal clinical samples (neat and diluted) of EBV-infected and non-infected patients. Contrived EDTA plasma samples spiked with cultured EBV virus were used to cover the linear range.
The comparator EBV is well described, currently used in clinical practice at a major transplant center in the United States, is traceable to the WHO standard and its use is acceptable. Due to different methods of measuring EBV viral load at each institution, EBV viral load quantitation may vary between laboratories and hence should not be compared to make clinical management decisions.
A total of 464 samples (439 neat or diluted clinical samples of 72 transplant subjects and 25 contrived samples) were valid on both assays and evaluable for the clinical concordance
{22}------------------------------------------------
analysis. Results presented in Table 12 demonstrate a column percent agreement between the cobas EBV and comparator EBV ranging from between 82.5% to 100% depending on the analyte concentration in the samples. DNA sequencing on representative samples from subjects with results consistently offset by more than 1 log10 IU/mL DNA level did not reveal any sequence mismatches for any primer or probe targets for the cobas EBV assay.
| | Comparator EBV (log10 IU/mL) | | | | | | |
|-----------------------------|------------------------------|------------------|-------------------|-------------------|-------------------|-------------------|-------|
| cobas® EBV<br>(log10 IU/mL) | Target<br>Not<br>Detected | <LLOQ<br>(<2) | 2 to <2.6 | 2.6 to<br><3.2 | 3.2 to 3.8 | > 3.8 | Total |
| Target Not<br>Detected | 95 | 17 | 17 | 0 | 0 | 0 | 129 |
| < LLoQ (< 2) | 39 | 46 | 75 | 11 | 0 | 0 | 171 |
| 2 to < 2.6 | 1 | 2 | 16 | 37 | 6 | 0 | 62 |
| 2.6 to < 3.2 | 1 | 0 | 5 | 15 | 30 | 1 | 52 |
| 3.2 to 3.8 | 0 | 0 | 0 | 0 | 9 | 11 | 20 |
| > 3.8 | 0 | 0 | 0 | 0 | 1 | 29 | 30 |
| Total | 136 | 65 | 113 | 63 | 46 | 41 | 464 |
| Column | (134/136) | (65/65) | (96/113) | (52/63) | (40/46) | (40/41) | |
| Agreement (%) | 98.5% | 100% | 85.0% | 82.5% | 87.0% | 97.6% | |
| (95% Score CI)a | (94.8,<br>99.6%) | (94.4%,<br>100%) | (77.2%,<br>90.4%) | (71.4%,<br>90.0%) | (74.3%,<br>93.9%) | (87.4%,<br>99.6%) | |
Table 12: Concordance analysis between cobas EBV and the comparator EBV DNA level results for all samples
Note: LLoO = lower limit of quantitation of comparator EBV (100 IU/mL). Standard Deviation of comparator EBV estimated at 0.3 logio IU/mL (comparator EBV analytical precision study).
Paired samples evaluable for clinical concordance analysis were included in this table.
a Assumed independence between all samples: CI =Confidence Interval by Score Method.
Discordant results were defined as those that are more than one box away from the diagonal (indicated by shading). For Target Not Detected by comparator EBV Column Agreement the cobas EBV Target Not Detected and < LLoQ (< 2) cells were combined. The rationale for adding the adjacent < LLoQ and TND cells for the TND column is that the difference between a TND and < LLoQ is not clinically meaningful and that these are analytically at the lower end of the measuring range, which may be impacted by random error.
Forty one of the 43 comparator EBV negative samples collected for the estimation of the Negative Percent Agreement (NPA) with the cobas EBV were negative by cobas EBV: therefore, the NPA was 95.4% (95% Exact CI: 84.2%-99.4%). The two comparator EBV negative samples were positive (<LLoQ) by cobas EBV and were seropositive for EBV VCA IgG and EBNA-1 IgG by supplemental serology testing. Concordance between cobas EBV and the comparator EBV was also evaluated using different clinical thresholds as in Table 13.
{23}------------------------------------------------
| | Percent Agreement<br>< Threshold (n/N)<br>95% CI | Percent Agreement<br>≥ Threshold (n/N)<br>95% CI |
|------------------------|--------------------------------------------------|--------------------------------------------------|
| Target Not Detected | 98.5% (134/136)<br>(94.8, 99.6%) | 89.6% (294/328)<br>(85.9%, 92.5%) |
| LLOQ (2.0 Log10 IU/mL) | 98.0% (197/201)<br>(95.0%, 99.2%) | 60.8% (160/263)<br>(54.8%, 66.5%) |
| 3.0 Log10 IU/mL | 100.0% (363/363)<br>(99.0%, 100.0%) | 64.4% (65/101)<br>(54.6%, 73.0%) |
| 4.0 Log10 IU/mL | 100.0% (431/431)<br>(99.1%, 100.0%) | 84.8% (28/33)<br>(69.1%, 93.3%) |
Table 13: Concordance summary of cobas EBV and comparator EBV using different thresholds
From all samples tested with cobas EBV that were EBV-positive with the comparator EBV there were a total of 158 (139 neat or diluted clinical samples of 28 transplant subjects and 19 contrived samples) that were evaluable for the correlation analysis at the three testing sites.
Analysis revealed no bias between neat clinical and contrived samples between the sites. Similarly, for the neat clinical and diluted samples the corresponding column percent agreements were poolable based on their 95% CIs, except for the "Target Not Detected" category where the 95%CI of column percent agreement did not overlap between neat clinical samples [90.6% (79.7%-95.9%)] and diluted clinical samples [57.3% (46.5%-67.5%)].
{24}------------------------------------------------
Figure 4: Correlation between cobas EBV and comparator EBV for all samples: Deming linear regression plot of DNA levels (log10 IU/mL)
Image /page/24/Figure/1 description: The image is a scatter plot comparing two EBV measurement methods. The x-axis represents the Comparator EBV LDT (log10 IU/mL), and the y-axis represents the cobas EBV (log10 IU/mL). The plot includes a Deming regression line with the equation Y = -0.45 + 1.025X, an R-squared value of 0.91, and 95% confidence intervals for the intercept (-0.637, -0.263) and slope (0.974, 1.076).
Additional bias plot analysis of DNA level differences indicated a systematic difference between both assays that is constant across the overlapping linear range. The 95% CI of the intercept of the fitted line in the bias plots was (-0.456 to -0.104), which is within ±0.6 log10 IU/mL (± 2 times analytical precision standard deviation of comparator EBV).
#### External Controls
During the conduct of the clinical trial protocols, external control testing was performed according to the Instructions for Use.
- Clinical specificity C.
Enrollment Inclusion/Exclusion Criteria
Study Demographics
- d. Other clinical supportive data (when a. and b. are not applicable):
- 4. Clinical cut-off:
Not applicable. Clinical thresholds for anti-EBV treatment are institution specific where they should follow professional guidelines (if available)
{25}------------------------------------------------
#### 5. Expected values/Reference range:
Not applicable
#### M. Instrument Name
cobas 6800/8800
#### N. System Descriptions
- 1. Modes of Operation
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes __________________________________________________________________________________________________________________________________________________________________________
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ____________or No ________________________________________________________________________________________________________________________________________________________
#### 2. Software
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes
The device does not contain any software or instrument components.
#### 3. Specimen Identification
#### 4. Specimen Sampling and Handling
See Instructions for Use.
- 5. Calibration
- 6. Quality Control
#### P. Labeling
The labeling supports the decision to grant the De Novo request for this device.
{26}------------------------------------------------
#### Q. Identified Risks to Health and Mitigation Measures
| Identified Risk | Mitigation Measures |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Risk of false results | Certain warnings, limitations, results<br>interpretation information, and explanation of<br>procedures in labeling. |
| | Certain device descriptions and specifications,<br>analytical studies, clinical studies, and risk<br>analysis in design verification and validation. |
| Failure to correctly interpret test results | Certain warnings, limitations, results<br>interpretation information, and explanation of<br>procedures in labeling. |
| Failure to correctly operate the device | Certain warnings, limitations, results<br>interpretation information, and explanation of<br>procedures in labeling. |
#### R. Benefit/Risk Analysis:
#### Summary of the Assessment of Benefit
The benefits of the assay are the quantitation of Epstein-Barr virus (EBV) DNA in plasma samples from transplant patients. The assay will be used to monitor and treat patients after transplant for post-transplant lymphoproliferative disorders (PTLD) that are associated with EBV. PTLD are considered among the most serious and fatal complications of transplantation. Known sequelae of PTLD include continued symptoms and increased morbidity and mortality due to organ dysfunction. Monitoring and treatment of PTLD can result in improved patient outcomes by mitigating the sequelae of untreated disease, namely by preserving transplanted organs and eradicating PTLD, thus decreasing morbidity and mortality in these patients.
An additional benefit of the device is standardization of EBV viral load measurement in transplant patients as there are currently no FDA-approved alternatives marketed in the United States. Standardization of EBV viral load measurement will benefit patients by facilitating medical research about the clinical significance of specific viral load measurements with the device.
#### Summary of the Assessment of Risk
The risks associated with the device, when used as intended, are those related to the risk of false test results, failure to correctly interpret the test results and failure to correctly operate the instrument.
Risks of incorrectly high test results are improper patient management, including imaging, biopsy, and/or implementation of treatment for PTLD with its associated side effects and risks such as rejection of a transplanted organ or graft. Risks of incorrectly low test results are improper patient management, including missing the opportunity to mitigate the known sequelae of PTLD, such as continued symptoms and increased morbidity and mortality due to organ dysfunction, preserving transplanted organs, and eradicating PTLD.
{27}------------------------------------------------
While the performance of the device in the clinical and analytical studies suggests that patients will benefit from the assay, expected and acceptable sources of uncertainty are the wide confidence intervals around point estimates during subgroup analysis. Another source of uncertainty of the benefits of the assay are that the physiological or clinically meaningful range of the diagnostic output are unknown. The special controls, including the interpretation of results and the limiting statements in device labeling will help to ensure that errors will be uncommon and will facilitate accurate assay implementation and interpretation of results.
A third source of uncertainty of the benefits of the assay is an imperfect comparator as there is currently no gold standard to measure EBV DNA levels. This uncertainty is acceptable given that the comparator in the clinical study is laboratory-developed test that is currently used in clinical practice at a major transplant center in the United States and is traceable to the World Health Organization (WHO) standard
#### Summary of the Assessment of Benefits-Risks:
The clinical benefits outweigh the risks for the proposed assay when considering the mitigations of the risks provided in the special controls as well as general controls. The special controls, including performance and total product life cycle commitments, an explanation of procedures, and the warnings, limiting statements, and results interpretation in device labeling will help to ensure that errors will be rare and will facilitate accurate assay implementation and interpretation of results. The device's performance observed in the clinical study suggests that errors will be infrequent and that the assay will provide substantial benefits to patients, along with other clinical information, to measure EBV DNA to monitor and manage PTLD in appropriate populations.
#### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
#### S. Conclusion
The De Novo request is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request: Product Code(s): QLX Device Type: Quantitative Viral Nucleic Acid Test for Transplant Patient Management Class: II (special controls) Regulation: 21 CFR 866.3183
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.