Clinical samples from patients receiving clozapine therapy
Clinical samples collected from patients at six U.S. sites were used in a method comparison study to evaluate the accuracy of the MyCare Psychiatry Clozapine Assay Kit against a validated LC-MS/MS reference method.
Method comparison; Clinical samples; Accuracy
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Method comparison study using clinical samples
Patients receiving clozapine therapy; Sample Size: (b)(4) samples; Number of Sites: 6
Validated clozapine LC-MS/MS method
Accuracy/Method comparison (Deming regression)
Indications for Use
The MyCare Psychiatry Clozapine Assay Kit is intended for the in vitro quantitative measurement of clozapine in adult human serum using automated clinical chemistry analyzers. Measurements obtained can be used to aid in the management of individuals prescribed clozapine for treatment-resistant schizophrenia. This assay should be used in conjunction with other clinical and laboratory findings and results from this test alone should not be used to make treatment decisions.
Device Story
MyCare Psychiatry Clozapine Assay Kit is a homogeneous two-reagent nanoparticle agglutination immunoassay. Input: human serum sample. Principle: competition between clozapine in sample and drug-conjugates for binding to clozapine-specific monoclonal antibodies on nanoparticles. Aggregation measured spectrophotometrically at ~ (b)(4) nm on automated clinical chemistry analyzers (validated on Beckman Coulter AU480). High clozapine levels result in less aggregation and lower light scattering/absorption; low levels result in more aggregation and higher absorption. Used in clinical laboratories by technicians/pathologists. Output: quantitative clozapine concentration (ng/mL). Healthcare providers use results to aid management of clozapine therapy, such as during dose titration or monitoring for toxicity, in conjunction with clinical assessment. Benefits: provides objective blood level data to assist in optimizing clozapine dosing, potentially reducing risks of sub-therapeutic efficacy or toxic adverse events.
Clinical Evidence
No clinical data provided in the document. Device classification supported by required special controls including precision studies, method comparison, and interference/cross-reactivity testing.
Technological Characteristics
In vitro diagnostic test system for automated clinical chemistry analyzers. Measures clozapine in human serum. Subject to special controls regarding precision, accuracy, and interference from co-administered medications and metabolites.
Indications for Use
Indicated for adult patients with treatment-resistant schizophrenia prescribed clozapine. For prescription use only. For In Vitro Diagnostic Use Only.
Regulatory Classification
Identification
A clozapine test system is a device intended to measure clozapine in human specimens. Measurements obtained by this device are used in monitoring levels of clozapine to ensure appropriate therapy in patients with treatment-resistant schizophrenia.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Precision study data that demonstrates precision that is clinically appropriate, as determined by FDA, for the clozapine test system. Precision studies must include a minimum of three samples containing different concentrations of clozapine including near medical decision points and throughout the expected therapeutic range of clozapine. Samples near the medical decision points must be clinical specimens collected from patients taking clozapine;
(ii) Method comparison data that demonstrates accuracy that is clinically acceptable, as determined by FDA, for the clozapine test system;
(iii) Data from studies that demonstrate that the device is free from clinically significant interference, as determined by FDA, from commonly co-administered medications that are used in patients with treatment-resistant schizophrenia; and
(iv) Data from studies that demonstrate that the device is free from clinically significant cross-reactivity, as determined by FDA, from major circulating metabolites found in the intended use population.
(2) The labeling required under § 809.10 of this chapter must include a limiting statement conveying that the assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures and that results from the assay alone should not be used in making treatment decisions.
Submission Summary (Full Text)
{0}------------------------------------------------
## EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR MyCare Psychiatry Clozapine Assay Kit DECISION SUMMARY
## A. DEN Number:
DEN190028
## B. Purpose for Submission:
De Novo request for evaluation of class III designation of the MyCare Psychiatry Clozapine Assay Kit
## C. Measurand:
Clozapine (8-chloro-11-(4-methyl-1-piperazinyl)-5H-dibenzo [b,e] [1,4] diazepine)
## D. Type of Test:
Homogeneous nanoparticle agglutination immunoassay
## E. Applicant:
Saladax Biomedical, Inc.
## F. Proprietary and Established Names:
MyCare Psychiatry Clozapine Assay Kit
## G. Regulatory Information:
- 1. Regulation section:
21 CFR 862.3245
- 2. Classification:
Class II (Special Controls)
- 3. Product code:
QKT
{1}------------------------------------------------
- 4. Panel:
- 91- Toxicology
## H. Intended Use:
- 1. Intended use(s):
See indications for use
- 2. Indication(s) for use:
The MyCare Psychiatry Clozapine Assay Kit is intended for the in vitro quantitative measurement of clozapine in adult human serum using automated clinical chemistry analyzers. Measurements obtained can be used to aid in the management of individuals prescribed clozapine for treatment-resistant schizophrenia. This assay should be used in conjunction with other clinical and laboratory findings and results from this test alone should not be used to make treatment decisions.
- 3. Special conditions for use statement(s):
For prescription use only For In Vitro Diagnostic Use Only
- 4. Special instrument requirements:
The assay was validated on the Beckman Coulter AU480 Clinical Analyzer
## I. Device Description:
The MyCare Psychiatry Clozapine Assay Kit is a homogenous two reagent nanoparticle agglutination assay used for detection of clozapine in human serum. It is based on competition between drug and drug-conjugates for binding to drug-specific antibodies covalently bound to nanoparticles. The extent of particle aggregation can be followed spectrophotometrically on clinical chemistry analyzers. This aggregation is measured at a wavelength of around (b)(4) by automated clinical chemistry analyzers.
The assay contains:
- Reagent 1 (R1) reaction buffer that contains drug-conjugate in a buffered protein . solution
- . Reagent 2 (R2) - nanoparticle reagent that contains clozapine-specific monoclonal antibody bound to nanoparticles in a buffered solution
{2}------------------------------------------------
## J. Standard/Guidance Document Referenced (if applicable):
- . CLSI EP05-A3 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition, October 2014.
- CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A ● Statistical Approach; Approved Guideline, April 2003.
- CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement ● Procedures; Approved Guideline - second edition, June 2012.
- CLSI EP25-A Evaluation of Stability of In vitro Diagnostic Reagents; Approved ● Guideline, September 2009.
## K. Test Principle:
Drug-conjugates serve as binding partners to antibodies selective for clozapine which are covalently attached to the surface of nanoparticles. In the absence of free clozapine, this reaction creates large aggregates, resulting in a solution that scatters incident light and leads to an increase in the observed absorption of the solution. When incubated with a sample containing clozapine, the free clozapine in the sample binds to the reagent antibodies, and those antibody-nanoparticle conjugates are no longer available to aggregate, resulting in less scattering of incident light and lower observed absorption of the solution. Thus, maximum absorption occurs with low levels of drug, and minimum absorption occurs with high levels of drug.
## L. Performance Characteristics (if/when applicable):
- 1. Analytical performance:
- a. Precision/Reproducibility:
A precision study was conducted according to recommendations in CLSI EP05-A3 using 4 human serum sample pools, which were prepared from clinical samples from patients taking clozapine therapy, serum pools spiked with clozapine, and 3 levels of controls. Each sample was assessed in replicates of (b)(4) a day, for (b)(4) days. Samples were assessed using three lots of reagents, calibrators, and controls on two Beckman Coulter AU480 Clinical Analyzers. The results are summarized below.
# Sample Results Across All Reagent Lots and Analyzers
| Sample | N | Mean<br>(ng/mL) | Within - Run | Total |
|--------------|-----|-----------------|--------------|-------|
| | | | CV | CV |
| Serum pool 1 | 480 | 148 | 4.6% | 5.6% |
| Serum pool 2 | 480 | 338 | 3.3% | 4.0% |
{3}------------------------------------------------
| Sample | N | Mean | Within - Run | Total |
|----------------|-----|---------|--------------|-------|
| | | (ng/mL) | CV | CV |
| Serum pool 3 | 480 | 577 | 3.3% | 4.0% |
| Serum pool 4 | 480 | 926 | 3.5% | 4.8% |
| Spiked Serum 1 | 480 | 98 | 6.0% | 7.8% |
| Spiked Serum 2 | 480 | 1,094 | 4.4% | 6.6% |
| Control 1 | 480 | 156 | 4.5% | 5.4% |
| Control 2 | 480 | 474 | 3.7% | 4.4% |
| Control 3 | 480 | 945 | 3.8% | 4.5% |
## b. Linearity/assay reportable range:
A study was conducted to evaluate linearity across the measuring range of the assay following the CLSI EP06-A guideline. Samples with 11 different clozapine levels were evaluated: 55, 132, 169, 352, 491, 588, 821, 1003, 1214, 1467, and 1554 ng/mL. Two samples were prepared by pooling human serum specimens that were then spiked with clozapine to low and high clozapine concentrations, which were then mixed to produce samples with the concentrations tested. Each sample was tested in five replicates using three lots of reagents, calibrators, and controls and two Beckman Coulter AU480 clinical analyzers. The observed values were plotted against the expected values and linear regression analysis was performed. The linear regression results from the worst performing lot of three lots are shown below.
| Claimed<br>Range | Slope | Intercept | R |
|------------------|------------------------------|---------------------|--------|
| 68-1500 ng/mL | 0.920<br>(0.910 to<br>0.930) | 4.1<br>(1.4 to 6.8) | 0.9985 |
The 95% confidence intervals are shown for slope and intercept.
These results support the claimed measuring range of 68 to 1500 ng/mL for clozapine.
- Traceability, Stability, Expected values (controls, calibrators, or methods): C.
#### Traceability
The MyCare Psychiatry Calibrator Kit is traceable to a certified USP clozapine reference standard. Master calibrators are prepared by (b) (4) of this
{4}------------------------------------------------
certified reference material in buffer and verified by LC-MS/MS.
- d. Detection limit:
Detection limits were assessed according to recommendations in CLSI EP17-A2.
The limit of blank (LOB) was determined using the non-parametric "classical approach" using "individual normal(b) (4) serum samples containing no clozapine. Each of the serum samples were tested in duplicate on three different days on two different analyzers for n="", using" lots of reagents, calibrators and controls. The
95th percentile of each n = " data set was calculated, and the highest 95th percentile value from among the three n = data sets was determined. The sponsor determined the LOB to be (b) (4) (4) (4) (4) (4) (4) 1) 1) (4) 1) 1) (4) 1) 1) 1 (4) 1) 1 (4) 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1 1 (4) 1 1 1 (4) 1 1 1 (
The limit of detection (LOD) was determined using the variant (non-parametric analysis) "classical approach" with a beta of (b) (4) individual normal human serum samples spiked with clozapine at 15, 20, 25, 30, and 35 ng/mL. Each of the serum samples were tested in duplicate on three different days on (0K4) different analyzers for n=""", using " lots of reagents, calibrators and controls. The LOD was determined using the highest median value observed at the lowest spiking level for which each of the three n = "" data sets associated with that spiking level had(b) (4) of results below the LOB. The sponsor determined the LOD to be 39 ng/mL.
The limit of quantitation (LOO) was determined by 4 individual normal human serum samples spiked with clozapine at " targeted concentrations: 40, 50, 60, 70 and 8 ng/mL. Each of the serum sample was tested n = "on(b) (4) different days on analyzer, using lots of reagents/calibrators/controls, for n=000 (results combined across all lots). The LOQ was defined as the specimen concentration where the total allowable error was ≤ 35% when calculated according to the Westgard model. The sponsor determined the LOO to be 68 ng/mL.
- e. Analytical specificity:
## Potentially Cross-Reacting Substances
Cross-reactivity was evaluated by adding each potentially cross-reacting clozapine metabolite into a human serum pool prepared from multiple individual human serum samples. The sponsor tested the three metabolites in the presence of clozapine at 350 ng/mL. Testing was conducted using one lot of reagents, one lot of calibrators, and two analyzers. Samples were tested in (0) (4) replicates. The formula used for calculating percent cross reactivity was:
(b) (4)
The results are summarized below. The sponsor did not observe cross-reactivity
{5}------------------------------------------------
exceeding ±10%:
| Compound | Tested at (ng/mL) | % Cross-reactivity |
|---------------------------------|-------------------|--------------------|
| Clozapine-N-oxide | 250 | 3% |
| 8-Hydroxy-8-deschloro-clozapine | 100 | 9% |
| Norclozapine | 2,700 | 1% |
Endogenous and Exogenous Interfering Substances:
Interference studies were conducted based on the CLSI guideline EP07-A2. Potentially interfering substances (test sample) or solvent (control sample) were spiked into human serum sample pools with clozapine at 350 and 600 ng/mL. Each sample was tested at a minimum of 5 runs using one lot of reagent on two analyzers. The percentage difference between the measurement of clozapine in the test sample and the control sample was calculated. The sponsor considered a percent difference between test samples and control samples of > ±10% to be clinically significant interference. None of the substances in the following tables (endogenous or exogenous substances) were found to lead to clinically significant interference (> ±10%) for the device.
| Interfering Substance | Concentration Tested |
|------------------------------|----------------------|
| Bilirubin | 18.18 mg/dL |
| Hemolysate | 1,050 mg/dL |
| Human IgG (HIgG) | 12.5 g/dL |
| Human Serum<br>Albumin (HSA) | 10.9 g/dL |
| Rheumatoid Factor (RF) | 508 IU/mL |
| Triglycerides | 2,585.87 mg/dL |
#### Endogenous Substances
## Exogenous Interferences
| Compound | Tested at<br>(ng/mL) | Compound | Tested at<br>(ng/mL) |
|--------------------------------|----------------------|-------------------------------------|----------------------|
| Acetaminophen | 200,000 | Acetazolamide | 60,000 |
| Acetylsalicylic acid | 500,000 | Albuterol | 1,000 |
| Alendronate sodium | 1,000 | Alpha - tocopherol | 130,000 |
| Alprazolam | 2,000 | Amantadine<br>Hydrochloride | 10,000 |
| Compound | Tested at<br>(ng/mL) | Compound | Tested at<br>(ng/mL) |
| Amikacin sulfate | 144,000 | Amiloride HCl<br>dihydrate | 500 |
| Amisulpride | 1,200 | Amitriptyline | 1,000 |
| Amlodipine besylate | 100 | S (+)-<br>amphetamine | 1,000 |
| Amoxapine | 2,900 | Amoxicillin | 80,000 |
| Aripiprazole | 1,400 | L-ascorbic acid | 60,000 |
| Asenapine | 500 | Atomoxetine | 7,900 |
| Atorvastatin calcium | 800 | Baclofen | 3,000 |
| Benztropine | 600 | Betamethasone | 400 |
| Biotin | 3,600 | Biperiden | 300 |
| Blonanserin | 100 | Brexpiprazole | 1,000 |
| Bromperidol | 100 | Budesonide | 50 |
| Bupropion | 3,000 | Buspirone | 200 |
| Caffeine | 108,000 | Calcium carbonate | 315,000 |
| Cannabidiol | 100 | Cannabinol | 100 |
| Carbamazepine | 45,000 | Cariprazine | 50 |
| L-Carnosine | 100,000 | Cefalexin | 200,000 |
| Celecoxib | 8,800 | Cetirizine<br>dihydrochloride | 4,400 |
| 8-chlorotheophylline | 3,000 | Chlorpromazine<br>HCl | 3,300 |
| Cimetidine | 30,000 | Ciprofloxacin | 12,000 |
| Citalopram HBr | 5,500 | Clindamycin | 51,000 |
| Clonazepam | 300 | Clotiapine | 500 |
| Clotrimazole | 50 | Codeine | 2,000 |
| Cortisol | 300 | (-)-Cotinine | 2,000 |
| Cyclosporin A | 9,000 | Desloratadine | 600 |
| Desvenlafaxine | 800 | Dextro-<br>methorphan | 1,000 |
| Diazepam | 30,000 | Diphenhydramine<br>HCl | 6,000 |
| Divalproex Sodium | 400,000 | Docosahexaenoic<br>acid ethyl ester | 150,000 |
| Donepezil | 50,000 | Doxycycline HCl | 35,000 |
| Droperidol | 200 | D-Serine | 100,000 |
| Duloxetine | 200 | Erythromycin | 138,000 |
| Escitalopram | 200 | Estradiol | 10 |
| Compound | Tested at (ng/mL) | Compound | Tested at (ng/mL) |
| Eszopiclone | 300 | Ethanol | 10,000,000 |
| Famotidine | 2,500 | Fenofibrate | 50,000 |
| Fentanyl | 600 | Fluoxetine HCl | 4,000 |
| Fluticasone propionate | 50 | Fluvoxamine | 2,000 |
| Folic acid | 15 | Furosemide | 60,000 |
| Galantamine | 200 | Gentamicin sulfate | 30,000 |
| Glyburide | 2,000 | Haloperidol | 1,000 |
| Heparin sodium salt | 50 U/mL | Hydrochlorothiazide | 6,000 |
| Hyoscine (Scopolamine HBr) | 100 | Hyperforin (St. John's Wort) | 200 |
| Hypericin (St. John's Wort) | 100 | Ibuprofen | 500,000 |
| Iloperidone | 100 | Imipramine | 700 |
| Indinavir sulfate | 400 | Lactulose | 10,000 |
| Lamivudine | 10,500 | Lamotrigine | 42,000 |
| Lansoprazole | 9,400 | Levonorgestrel | 100 |
| Lisinopril dihydrate | 350 | Lithium carbonate | 250,000 |
| Lorazepam | 1,000 | Lovastatin | 500 |
| Loxapine | 300 | Lurasidone | 400 |
| Meclizine dihydrochloride | 500 | Metformin | 40,000 |
| Methotrimeprazine | 600 | Methylphenidate HCl | 350 |
| Metoclopramide HCl | 500 | Metoprolol tartrate | 5,000 |
| Metronidazole | 123,000 | Midazolam | 3,800 |
| Milnacipran | 10,000 | Mirtazapine | 900 |
| Mometasone furoate | 50 | Morphine | 7,800 |
| Naltrexone | 200 | Naproxen sodium | 500,000 |
| Nateglinide | 30,000 | Nefazodone HCl | 6,000 |
| Nicotine | 1,000 | Nicotinic acid | 27,900 |
| Nordiazepam | 5,000 | Nortriptyline | 1,200 |
| Olanzapine | 400 | Omeprazole | 8,400 |
| Oxazepam | 5,000 | Oxcarbazepine | 105,000 |
| Oxycodone | 500 | Paliperidone | 60 |
| Pantothenic acid | 1,800 | Paroxetine | 1,200 |
| Penicillin V | 42,000 | Perazine | 1,400 |
| Compound | Tested at (ng/mL) | Compound | Tested at (ng/mL) |
| Perlapine | 150 | Perphenazine | 100 |
| Phenobarbital | 690,000 | Phentermine | 500 |
| Phenytoin | 60,000 | Pimozide | 100 |
| Pipamperone<br>dihydrochloride | 1,200 | Potassium EDTA | 1,000 |
| Pravastatin sodium | 300 | Prednisolone | 3,000 |
| Pregabalin | 22,500 | Procyclidine | 1,900 |
| Promethazine | 1,200 | R,R-(-)-<br>pseudoephedrine | 10,000 |
| S,S-(+)-<br>pseudoephedrine | 10,000 | Pyridoxine HCl | 100 |
| Quetiapine | 2,800 | Quinidine | 15,000 |
| Raloxifene | 50 | Ranitidine | 10,500 |
| Retinol | 4,000 | Riboflavin | 200 |
| Rifampicin | 65,000 | Risperidone | 200 |
| Rosuvastatin calcium | 200 | Salicylic acid | 500,000 |
| Sarcosine | 1,500 | Sertindole | 300 |
| Sertraline<br>hydrochloride | 1,000 | Simvastatin | 1,700 |
| Sodium benzoate | 400,000 | Sodium fluoride | 900 |
| Spironolactone | 600 | Sulfamethoxazole | 400,000 |
| Sulpiride | 50,000 | Temazepam | 5,000 |
| Terbinafine | 9,000 | Theophylline | 60,000 |
| Thiamine HCl | 500 | Topiramate | 75,000 |
| Trazodone HCl | 14,700 | Triamcinolone<br>acetonide | 300 |
| Triamterene | 9,000 | Triazolam | 40 |
| Valproic acid | 500,000 | Vancomycin HCl | 120,000 |
| Varenicline | 50 | Venlafaxine HCl | 700 |
| Vitamin B12 | 50 | Vitamin D2 | 200 |
| Vitamin K1 | 50 | Warfarin | 75,000 |
| Ziprasidone | 600 | Zolpidem<br>hemitartrate | 5,000 |
| Zonisamide | 120,000 | Zopiclone | 200 |
| Zuclopenthixol | 300 | | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
- f. Assay cut-off:
Not applicable
- 2. Comparison studies:
- a. Method comparison study:
A method comparison study was conducted using the MyCare Psychiatry Clozapine Assay Kit on the Beckman Coulter AU480 by testing clinical samples that were collected at six different U.S. clinical sites from patients receiving clozapine therapy. Results from the MyCare Psychiatry Clozapine Assay Kit were compared to results from a validated clozapine LC-MS/MS method. Deming regression analysis was performed with (b)(4) samples, resulting in the following line equation (with 99% confidence intervals shown) from a representative lot:
y = 1.037 (0.939 to 1.135) x - 27.8 (-78.3 to +22.7); R = 0.9269
A summary of the device accuracy (compared to the LC-MS/MS comparator method) at different concentration ranges throughout the claimed measuring interval is shown below.
| | Difference Range Between the Serum Clozapine Level<br>by LC/MS-MS and the MyCare Psychiatry Clozapine<br>Assay | | | | |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------|------------------------|------------------------|------------------------|------------------------|
| Concentration of<br>Clozapine<br>(ng/mL) | Within<br>10%<br>(n/N) | Within<br>15%<br>(n/N) | Within<br>20%<br>(n/N) | Within<br>30%<br>(n/N) | Within<br>40%<br>(n/N) |
| 68-350 | 30%<br>(15/50) | 42%<br>(21/50) | 60%<br>(30/50) | 78%<br>(39/50) | 84%<br>(42/50) |
| 350-1000 | 34%<br>(24/71) | 56%<br>(40/71) | 75%<br>(53/71) | 93%<br>(66/71) | 96%<br>(68/71) |
| 1000-1500 | 50% (1/2) | 50% (1/2) | 100% (2/2) | N/A | N/A |
- b. Matrix comparison:
Not applicable
- 3. Clinical studies:
- a. Clinical Sensitivity:
Not applicable
{10}------------------------------------------------
- b. Clinical specificity:
Not applicable
- c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor provided a discussion of the clinical use of therapeutic drug monitoring for clozapine in patients with treatment-resistant schizophrenia, as provided in the U.S. clinical practice guideline, "The American Psychiatric Association Practice Guideline for the Treatment of Patients with Schizophrenia" (2020).
- 4. Clinical cut-off:
Not applicable.
- 5. Expected values/Reference range:
There is no definitive therapeutic range for clozapine, as this range may be different for each patient.
## N. Proposed Labeling:
The labeling supports the decision to grant the De Novo request for this device.
#### O. Identified Risks to Health and Mitigation Measures
| Identified Risks to Health | Mitigation Measures |
|------------------------------------------|------------------------------------------------------------------------------------------|
| Incorrect test results | Certain design verification and validation<br>activities<br>Certain labeling information |
| Incorrect interpretation of test results | Certain design verification and validation<br>activities<br>Certain labeling information |
## P. Benefit/Risk Determination
#### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
#### Summary of the Assessment of Benefit
There is currently no available FDA cleared or approved device for determining Clozapine
{11}------------------------------------------------
blood concentrations. Toxic side effects from Clozapine include seizures, myocarditis, agranulocytosis, and other severe adverse effects, and Clozapine currently requires a REMS (Risk Evaluation and Mitigation Strategy) to ensure the benefits of the drug outweigh the risk of severe neutropenia. The availability of the test may aid in the management of patients with treatment-resistant schizophrenia under various clinical scenarios as described in the American clinical practice guideline, "The American Psychiatric Association Practice Guideline for the Treatment of Patients with Schizophrenia" (2020) (e.g. "while the dose of clozapine is being titrated, it is useful to obtain blood levels of clozapine").
#### Summary of the Assessment of Risk
In the setting of poor or incomplete drug response, a falsely low clozapine level may increase the risk of a prescriber falsely concluding that the outcome is due to sub-therapeutic levels which may prompt an unnecessary increase in drug dose and increase the risk of adverse events such as seizures. In the setting of potential toxicity, a falsely low level may delay clozapine dose de-escalation, increasing the risk of persistent adverse events and severe adverse events. In the setting of poor or incomplete drug response, a falsely high level may increase the risk of a prescriber incorrectly assuming that the patient was receiving an adequate dose and continue a suboptimal dose or conclude that the patient was not responding to clozapine and prescribe alternate therapies that may be less effective for the patient, increasing the risk of continued psychological distress from persistent symptoms, poorer social and occupational functioning, hospitalization, or transfer to residential settings. Additionally, if a patient's clozapine level is falsely elevated, a prescriber may not fully investigate the possibility of suboptimal dosing, which could result in a missed opportunity to identify and rectify conditions that could lead to suboptimal dosing or to consider treatment alternatives that are more acceptable to the patient. Falsely elevated levels may lead to a prescriber deciding to decrease the dose or discontinue medication, even if the patient were tolerating the medication and had experienced benefit, which could increase the risk of relapse, functional impairment, or transition to higher levels of care.
#### Summary of the Assessment of Benefit-Risk
General controls are insufficient to mitigate the risks of the device. However, the probable clinical benefits outweigh the probable risks for the assay, considering the mitigation of the risks provided for in the special controls. Device design verification and validation, including precision, method comparison, interference, and cross-reactivity studies will help ensure that the device functions as intended and mitigate the risk of falsely high test results. A limitation statement conveying that results from the assay alone should not be used in making treatment decisions will be included in the labeling, as a mitigation against the risk of incorrect interpretation of results. Overall, the probable benefits outweigh the probable risks of incorrect test results or incorrect interpretation of test results for the proposed indications for use, in light of the special controls and general controls.
{12}------------------------------------------------
# Q. Conclusion
The De Novo request for the MyCare device is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:
Product Code: QKT Device Type: Clozapine test system Class: II (special controls) Regulation: 21 CFR 862.3245
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.