EchoTip Insight Portosystemic Pressure Gradient Measuring System
Applicant
Cook Ireland, Ltd.
Product Code
QIJ · Gastroenterology, Urology
Decision Date
Nov 20, 2019
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 876.1050
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN180062 · Nov 20, 2019
EchoTip Insight Portosystemic Pressure Gradient Measuring System
Cook Ireland, Ltd.
Published clinical literature (human pilot study)
The study was provided as confirmatory evidence to demonstrate the device's ability to safely measure portal pressure gradient in a human clinical setting.
Human pilot study; Portal pressure gradient; Clinical cohort
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
EUS Guided portal pressure gradient measurement with a simple novel device: a human pilot study; Single-center human pilot study
28 patients, including 19 with clinical signs or symptoms highly suspicious of cirrhosis; Sample Size: 28; Number of Sites: 1
Not applicable for this study
Safety and ability to measure portal pressure gradient (PPG)
Indications for Use
The EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System is indicated to directly measure pressures in the hepatic and portal venous vasculatures and is used in conjunction with an ultrasound endoscope.
Device Story
System provides endoscopic access to portal and hepatic veins for pressure measurement; used with ultrasound endoscope. Components: 25-gauge stainless steel needle, PEEK sheath, connecting tube, syringe, stopcock, and Compass CT pressure transducer. Physician advances needle through endoscope accessory channel into liver parenchyma under ultrasound guidance. System primed with heparinized saline; pressure transducer (disposable manometer) measures local blood pressure. Output displayed on Compass CT liquid crystal display. Facilitates direct portal pressure gradient (PPG) measurement; avoids risks of trans-jugular approach (radiation, contrast, jugular puncture complications).
Clinical Evidence
Evidence includes a human pilot study (n=28) and a comparative animal study (n=3 swine). Human study demonstrated safe PPG measurement in patients with suspected cirrhosis; no intraprocedural or postprocedural adverse events (bleeding, perforation, infection) reported. Animal study compared subject device to standard trans-jugular balloon catheter approach; Pearson's correlation coefficient of 0.988 for PV and WHVP measurements, indicating excellent correlation.
Technological Characteristics
Needle: 25-gauge stainless steel, beveled tip, echogenic. Sheath: 5 Fr. PEEK, 142 cm, adjustable length. Connectivity: Luer lock connections. Energy: Passive (mechanical pressure transmission). Sterilization: Ethylene Oxide (ISO 11135-1). Shelf life: 3 years (ASTM F1980-07). Software: Compass CT transducer includes embedded pressure sensor and pre-programmed diagnostic computer chip.
Indications for Use
Indicated for patients requiring direct pressure measurement in hepatic and portal venous vasculatures via endoscopic ultrasound guidance.
Regulatory Classification
Identification
An endoscopic transhepatic venous access needle is inserted through the liver into the patient's portal/hepatic venous system under endoscopic ultrasound guidance. It is connected to a separate device intended to measure a physiological parameter.
Special Controls
In combination with the general controls of the FD&C Act, the endoscopic transhepatic venous access needle is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(3) The patient-contacting components of the device must be demonstrated to be non-pyrogenic.
(4) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be tested:
(i) Needle crumple testing;
(ii) Tensile testing;
(iii) Dimensional verification for all components; and
(iv) Simulated use testing.
(6) Labeling must include the following:
(i) Instructions for use, including specific instructions regarding device preparation;
(ii) The recommended training for safe use of the device; and
(iii) A shelf life for any sterile components.
Submission Summary (Full Text)
{0}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR ECHOTIP® INSIGHT™ PORTOSYSTEMIC PRESSURE GRADIENT MEASUREMENT SYSTEM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Endoscopic transhepatic venous access needle. An endoscopic transhepatic venous access needle is inserted through the liver into the patient's portal/hepatic venous system under endoscopic ultrasound guidance. It is connected to a separate device intended to measure a physiological parameter.
NEW REGULATION NUMBER: 21 CFR 876.1050
CLASSIFICATION: Class II
PRODUCT CODE: OIJ
#### BACKGROUND
DEVICE NAME: EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System
SUBMISSION NUMBER: DEN180062
DATE DE NOVO RECEIVED: November 19, 2018
- Cook Ireland, Ltd CONTACT: O'Halloran Road, National Technology Park Limerick Ireland
#### INDICATIONS FOR USE
The EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System is indicated to directly measure pressures in the hepatic and portal venous vasculatures and is used in conjunction with an ultrasound endoscope.
#### LIMITATIONS
The sale, distribution, and use of the EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System are restricted to prescription use in accordance with 21 CFR 801.109.
The device is not intended for uses other than that described in the labeling.
{1}------------------------------------------------
## PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System is a system used to provide endoscopic access to the portal and hepatic veins to measure local blood pressure.
As shown in Figure 1, the system is composed of:
- 1. EchoTip Insight Endoscopic Ultrasound Needle
- 2. Connecting Tube
- 3. 10 mL syringe
- 4. Stopcock
- 5. Compass CT Pressure Transducer (K133624).
Image /page/1/Figure/9 description: The image shows a medical device setup with labeled components. The components include an EchoTip Insight Needle (1), a connecting tube (2), a syringe (3), a stopcock (4), and a Compass CT Pressure Transducer (5). The needle and syringe are connected by the tube, with the stopcock and transducer placed in between.
### Figure 1. EchoTip Insight needle, with Connecting Tube and Compass CT transducer
The EchoTip® Insight™ Needle functions to provide access to the venous vasculature related to the portal circulation and hepatic outflow to facilitate the measurement of physiological pressure by the Compass® CT pressure transducer, a disposable manometer (cleared under K133624).
The needle is enclosed within a 5.2 Fr. outer sheath for protection. The sheath has an adjustable length (0-5 cm) to allow the user to adjust for the working length of the endoscope. The sheath length extension lock ring allows the sheath to be secured in place by tightening the thumbscrew.
The needle extension is also adjustable and ranges from 0-8 cm. The "zero" reference ensures complete needle retraction within the sheath. The handle has a safety ring that slides and locks using the thumbscrew at the desired needle extension.
The Connecting Tube is used for the transfer of liquids (heparinized saline) between the
{2}------------------------------------------------
EchoTip® Insight™ Needle and the Compass® CT pressure transducer. The Connecting Tube uses a female Luer lock to connect to the Compass CT pressure transducer, and a male Luer lock to connect to the stopcock. The Compass CT is a disposable, point-of-use blood pressure measurement and monitoring device that incorporates an embedded pressure sensor and an integrated, pre-programmed diagnostic computer chip with a liquid crystal display.
Once the system is connected and primed with heparinized saline, the EchoTipe Insight™ Needle is passed through the accessory channel of an ultrasound endoscope and the needle is advanced to the desired vasculature (portal or hepatic vein) through the liver parenchyma. Using ultrasound guidance, the needle is placed at the desired location, allowed to stabilize, and a measurement reading is taken.
Table 1 provides a technological description of the device.
| Device Component | Attribute |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Needle | 25 gauge |
| | 0-8 cm length |
| | Beveled tip |
| | Stainless steel |
| | Echogenic texturing |
| Sheath | 5 Fr. |
| | 142 cm |
| | 0-5 cm adjustable length |
| | Polyetheretherketone (PEEK) Material |
| Handle | Proximal female luer |
| | Sheath and needle adjustment controls |
| Connectors | Luer connection |
| Access/visualization<br>technique | Manually advanced through the working<br>channel (2.8 mm) of an ultrasound<br>endoscope under ultrasound visualization |
# Table 1. Description of Technology for the EchoTip Insight Portosystemic Pressure Gradient Measurement System.
# SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS
The Connecting Tube, stopcock, and syringe are externally communicating devices with indirect contact with blood for a limited (≤24 hrs) duration. As such, the following tests are expected: Cytotoxicity, Sensitization, Irritation, Acute systemic toxicity, material mediated pyrogenicity, hemocompatibility. The EchoTip Insight needle is categorized as an externally communicating device in direct contact with circulating blood for a limited (≤24 hrs) duration. As such, the same tests were expected, with the addition of genotoxicity. The device system therefore, has components that have direct and indirect
{3}------------------------------------------------
patient contact. All device components were evaluated according to the FDA guidance (2016). "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," and the ISO 10993-4:2017, "Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood." From the evaluations and supporting information, the components of the device were found to be biocompatible for its use.
## SHELF LIFE/STERILITY
The EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System is a sterile, single use system. All device components are single use devices provided sterile to the end user. Device components in contact with blood were also tested for pyrogenicity using the LAL (Limulus Amebocyte Lysate test).
Sterilization methods for the EchoTip Insight needle and Connecting Tube have been validated in accordance with ISO 11135-1:2007 "Sterilization of Health Care Products-Ethylene Oxide – Part 1: Requirements for Development. Validation and Routine Control of a Sterilization Process for Medical Devices," to ensure a sterility assurance level of 10-6 before the device is marketed.
Accelerated aging to support a 3-year shelf life was performed for the EO sterilized EchoTip Insight needle and Connecting Tube per ASTM F1980-07, Standard Guidance for Accelerated Aging of Sterile Medical Device Packages. The expiration date of 3 years was verified by demonstrating package integrity through dye penetration and burst testing on the stored products.
### PERFORMANCE TESTING - BENCH
Non-clinical performance data was generated to mitigate the risk associated with the failure of the device components and/or materials. The following tests were conducted and passed: needle crumple testing, tensile strength testing, product integrity testing for Connecting Tube, dimensional verification for needle and Connecting Tube, package integrity testing on all device components post distribution, and design verification/validation testing.
A simulated use test was performed to measure pressures of known values. The simulated use testing consisted of design verification and validation of the EchoTip Insight Portosystemic Pressure Gradient Measuring System. After passing the device through an echoendoscope, the device was tested for its ability to measure three known pressures, which were created by filling a column of liquid at three different points. A statistically significant number of distinct needle devices, connecting tubes, and Compass CT transducers were tested, and the ability of an end user to accomplish each of the tasks conducted to obtain a pressure measurement under simulated use conditions was verified.
{4}------------------------------------------------
## PERFORMANCE TESTING - ANIMAL TESTING
The EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System was compared versus the standard trans-jugular approach, in which an indirect portal/hepatic pressure is obtained with the use of a balloon catheter . In this study, 17 pressure measurements were carried out in three swine at baseline, and then following induction of portal hypertension (PH) by rapid peripheral infusion of Dextran-40. Manometric data were obtained using an interventional radiologic (IR) trans-jugular balloon catheter tip from the right hepatic vein (both free and wedged pressures) and then using the subject device to obtain pressure gradients of the portal and hepatic veins. The correlation between the 2 methods of measurement was charted in scatter plots, and the Pearson's correlation coefficient (R) was calculated. Among other measurements, the authors calculated the correlation of 17 paired manometric data points from Wedge Hepatic Venous Pressure (WHVP) using the transjugular method, and Portal Venous Pressure (PV) using the subject device. There was excellent correlation between the subject device and transjugular approach at all pressure ranges. The authors calculated a Pearson's correlation coefficient of 0.999 for all vessels, 0.985 for all veins, 0.988 for PV and WHVP, and 0.986 for free HV pressure.
1 Huang, J., Samarasena, J., Tsujino, T., Chang, K. EUS Guided portal pressure gradient measurement with a novel 25-gauge needle device versus standard transjugular approach: A comparison animal study. Gastrointestinal Endoscopy, 2016 Aug; 84(2), p. 358-362.
### SUMMARY OF CLINICAL INFORMATION
A published clinical study2 was provided to FDA as confirmatory evidence that the device could safely measure portal pressure gradient (PPG) (i.e the difference between hepatic and portal vein pressures). In this single center study, PPG was determined in 28 patients. Of these patients, 19 patients had clinical signs or symptoms that were highly suspicious of cirrhosis. The study excluded pregnant women, patients at risk of bleeding (International Normalized Ratio > 1.5, platelet count < 50 X 10° /L), patients with active gastrointestinal bleeding, and patients with post sinusoidal portal hypertension.
No patient experienced intraprocedural or postprocedural adverse events. Specifically, no patient developed bleeding, perforation, infection, or complained of pain.
2Huang, J., Samarasena, J., Tsujino, T., Lee, J., Hu, K., McLaren, C., Chen, W., Chang, K. EUS Guided portal pressure gradient measurement with a simple novel device: a human pilot study. Gastrointestinal Endoscopy, 2017 May; 85(5), p. 996-1001.
### Pediatric Extrapolation
In this De Novo request, existing clinical information was not leveraged to support the use of the device in a pediatric patient population.
{5}------------------------------------------------
# LABELING
Labeling has been provided that includes instructions for use and an appropriate prescription statement as required by 21 CFR 801.109. The labeling includes:
- Instructions for Use Manual: The manual is the primary labeling material for the device. It provides information about the device and its components, indications, contraindications, precautions, warnings, possible adverse reactions, device functions, and guidelines for use. The manual includes instructions and diagrams that explain the steps to prepare the device for use prior to use and explains the steps to measure hepatic and portal venous pressures.
- Package Label: This provides sizing information, manufacturer information, a shelf life ● and product summary.
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the endoscopic transhepatic venous access needle and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measures |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Adverse tissue reaction | Biocompatibility testing<br>Pyrogenicity testing |
| Infection | Sterilization validation<br>Pyrogenicity testing<br>Shelf life testing<br>Package integrity testing |
| Use error leading to:<br>Access site hemorrhage/thrombosis Portal vein penetration leading to<br>intrahepatic bleeding | Labeling |
| Improper patient management due to<br>inaccurate measurement | Non-clinical performance testing<br>Labeling |
Table 2: Identified Risks to Health and Mitigation Measures
### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the endoscopic transhepatic venous access needle is subject to the following special controls:
- 1. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 2. Performance data must demonstrate the sterility of the patient-contacting components of the device.
{6}------------------------------------------------
- 3. The patient-contacting components of the device must be demonstrated to be nonpyrogenic.
- 4. Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
- 5. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be tested:
- a. Needle crumple testing;
- b. Tensile testing:
- c. Dimensional verification for all components; and
- d. Simulated use testing.
- 6. Labeling must include the following:
- a. Instructions for use, including specific instructions regarding device preparation;
- b. The recommended training for safe use of the device; and
- c. A shelf life for any sterile components.
#### BENEFIT-RISK DETERMINATION
The probable risks of the device include the risks associated with an endoscopic ultrasound procedure, adverse tissue reactions, infection, use error such as those leading to hemorrhage/thrombosis, portal vein penetration leading to intrahepatic bleeding and improper patient management due to inaccurate pressure measurement.
Based on the available performance data, the probability of such harmful events is low, and the incidence is reduced with the mitigation measures and special controls identified above.
The probable benefits of the device include direct measurement of both portal and hepatic venous pressures, without risks associated with trans jugular measurements, which include bleeding from the jugular puncture site, cardiac arrhythmias, and exposure to radiation and contrast agents.
#### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
#### Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System is indicated to directly measure pressures in the hepatic and portal venous vasculatures and is used in conjunction with an ultrasound endoscope.
The probable benefits outweigh the probable risks for this device. The device provides benefits
{7}------------------------------------------------
and the risks can be mitigated using general controls and the identified special controls.
# CONCLUSION
The De Novo request for the EchoTip® Insight™ Portosystemic Pressure Gradient Measurement System is granted and the device is classified as follows:
Product Code: QIJ Device Type: Endoscopic transhepatic venous access needle Regulation Number: 21 CFR 876.1050 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.