Retrospective remnant clinical samples were used to supplement the prospective clinical trial data to ensure a sufficient number of positive specimens (at least 100 total) for performance evaluation.
Retrospective analysis of remnant clinical samples
Patients undergoing revision surgery for total knee or hip arthroplasty; Sample Size: 65 MSIS positive samples
MSIS diagnostic criteria (neutrophil %, positive culture, and WBC count)
Positive percent agreement with MSIS diagnosis
Indications for Use
The Synovasure Alpha Defensin Lateral Flow Test Kit is a qualitative visually read immunochromatographic assay for the detection of human host response proteins, Alpha Defensins 1-3, in the synovial fluid of adults with a total joint replacement who are being evaluated for revision surgery. The Synovasure Alpha Defensin Lateral Flow Test Kit results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of periprosthetic joint infection (PJI). The Synovasure Alpha Defensin Lateral Flow Test Kit is not intended to identify the etiology or severity of a PJI. The Synovasure Alpha Defensin Control Kit is used in the Synovasure Alpha Defensin Lateral Flow Test Kit as assayed quality control samples to monitor performance and reliability of the Synovasure Alpha Defensin Lateral Flow Test Kit.
Device Story
Synovasure Alpha Defensin Lateral Flow Test Kit is a qualitative, visually read immunochromatographic assay. It detects human Alpha Defensins 1-3 in synovial fluid. Input: synovial fluid sample; processed via dilution buffer and sample prep assembly. Principle: antibodies specific to alpha defensin bind host proteins; immobilized on test strip; detected via colloidal gold reporter. Output: visual colored line on test cassette (positive/negative). Used in clinical settings by healthcare providers. Results used as an adjunct to clinical/diagnostic findings to aid PJI diagnosis. Benefits: assists in identifying PJI, potentially guiding treatment (antibiotics/revision surgery) or hardware retention decisions.
Lateral flow immunoassay. Components: test cassette with reagent strip, sample prep assembly (dilution buffer, sample cup, Microsafe tubes), and control kit (positive/negative controls). Detection principle: colloidal gold-labeled anti-alpha defensin antibody. Form factor: manual, visually read cassette. No instrumentation required. Storage: 2-30°C. Stability: 24-month shelf life supported by Arrhenius extrapolation.
Indications for Use
Indicated for adults with total joint replacement undergoing evaluation for revision surgery to aid in the diagnosis of periprosthetic joint infection (PJI).
Regulatory Classification
Identification
A device to detect and measure non-microbial analytes to aid in the detection and identification of localized human infections is identified as an in vitro diagnostic device intended for the detection and qualitative measurement, quantitative measurement, or both of one or more non-microbial analytes in human clinical specimens to aid in the assessment, identification, or both of a localized microbial infection when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.
Special Controls
The special controls for this device are:
*Classification.* Class II (special controls). The special controls for this device are:(1) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of human specimens; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) An intended use with a detailed description of what the device detects and measures, the type of results provided to the user, the sample type, whether the measure is qualitative and/or quantitative, the clinical indications for the test use, and the specific population(s) for which the device is intended.
(ii) A detailed description of the performance characteristics of the device for all intended specimen types from the analytical and clinical studies (as applicable) required under paragraphs (b)(3)(ii) and (iii) of this section.
(iii) A detailed explanation of the interpretation of results, including acceptance criteria for evaluating the validity of individual runs (
*e.g.,* assessment of internal and/or external quality controls, as applicable).(iv) The following limiting statements:
(A) A statement that a negative test result does not preclude the possibility of infection;
(B) A statement that the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) A statement that consistent device performance is dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) A statement that details any limitations associated with the samples, as appropriate (
*e.g.,* collected on the day of admission to the intensive care unit).(3) Design verification and validation must include the following:
(i) A detailed device description, including as appropriate, all device parts; control elements incorporated into the test procedure; instrument requirements; reagents required but not provided; and the principle of device operation and test methodology, including all preanalytical methods for the processing of specimens and the methodology from obtaining a sample to the result; design of primer/probe sequences; rationale for target analyte selection; and computational path from collected raw data to reported result (
*e.g.,* how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies including analytical sensitivity (Limit of Detection, Limit of Quantitation, and Limit of Blank), inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross-contamination, specimen stability, within-lab precision, reproducibility, and linearity, as applicable.
(iii) Detailed documentation and results either from a clinical study, that includes prospective (sequentially collected) samples for each intended specimen type that are representative of the intended use populations and, when determined to be acceptable by FDA, additional characterized clinical samples; or, when determined to be acceptable by FDA, an equivalent sample set. The clinical study must compare the device performance to results obtained from an FDA-accepted reference method and/or FDA-accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol (
*e.g.,* the predefined statistical analysis plan), clinical study report, testing results, and results of all statistical analyses.(iv) An evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
*e.g.,* age, racial, ethnic, and sex distribution) similar to the intended use population of the device.(v) Documentation of an appropriate end user device training program that will be offered as part of efforts to mitigate the risks of false results, failure to operate the device correctly, and failure to interpret test results correctly.
(vi) An appropriate risk mitigation strategy to ensure that the device does not prevent any other device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use (
*e.g.,* safety and effectiveness of the functions of the indicated device(s) remain unaffected).(vii) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.
Submission Summary (Full Text)
{0}------------------------------------------------
### EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR SYNOVASURE ALPHA DEFENSIN LATERAL FLOW TEST KIT DECISION SUMMARY
### A. De Novo Number:
DEN180032
#### B. Purpose for Submission:
De Novo request for evaluation of automatic class III designation for the Synovasure Alpha Defensin Lateral Flow Test Kit.
#### C. Measurand:
Human Alpha Defensins 1-3
### D. Type of Test:
Lateral flow immunoassay
#### E. Applicant:
CD Diagnostics, a Zimmer Biomet Subsidiary
#### F. Proprietary and Established Names:
Synovasure Alpha Defensin Lateral Flow Test Kit; Synovasure Alpha Defensin Lateral Flow Test Kit (5 Test); Synovasure Alpha Defensin Lateral Flow Test Kit (10 Test); Synovasure Alpha Defensin Lateral Flow Test Kit (30 Test); Synovasure Alpha Defensin Control Kit;
### G. Regulatory Information:
- 1. Regulation section:
21 CFR 866.3230
- 2. Classification:
Class II
{1}------------------------------------------------
### 3. Product code(s):
QGN
- 4. Panel:
Microbiology (83)
## H. Intended Use:
- 1. Intended use(s):
The Synovasure Alpha Defensin Lateral Flow Test Kit is a qualitative visually read immunochromatographic assay for the detection of human host response proteins, Alpha Defensins 1-3, in the synovial fluid of adults with a total joint replacement who are being evaluated for revision surgery. The Synovasure Alpha Defensin Lateral Flow Test Kit results are intended to be used in conjunction with other clinical and diagnostic findings as an aid in the diagnosis of periprosthetic joint infection (PJI). The Synovasure Alpha Defensin Lateral Flow Test Kit is not intended to identify the etiology or severity of a PJI.
The Synovasure Alpha Defensin Control Kit is used in the Synovasure Alpha Defensin Lateral Flow Test Kit as assayed quality control samples to monitor performance and reliability of the Synovasure Alpha Defensin Lateral Flow Test Kit.
- 2. Indication(s) for use:
Same as intended use.
- 3. Special conditions for use statement(s):
For prescription use only.
For in vitro diagnostic use only.
- 4. Special instrument requirements:
Not applicable.
# I. Device Description:
The Synovasure Alpha Defensin Lateral Flow Test Kit (Synovasure LFT) is an immunoassay for the detection of alpha defensin levels in the synovial fluid of patients with a potential PJI. Antibodies specific to alpha defensin bind host alpha defensin in the synovial fluid, become immobilized on the lateral flow test strip, and are detected as a colored line due to the use of a colloidal gold reporter.
{2}------------------------------------------------
Synovasure Alpha Defensin Lateral Flow Test Kit contains two sub components:
- 1. Synovasure Alpha Defensin Lateral Flow Test Device
- 2. Synovasure Lateral Flow Sample Prep Assembly
The Synovasure Lateral Flow Sample Prep Assembly further contains
- 1. One Synovasure Dilution Buffer Bottle
- 2. One Sample Cup
- 3. Two Microsafe Tubes
The Synovasure Alpha Defensin Lateral Flow Test Device is a cassette that includes a reagent strip. Each cassette contains a reagent strip with all the critical components for the assay.
The Synovasure Alpha Defensin Lateral Flow Test Kit is accompanied by the Synovasure Alpha Defensin Control Kit. The Synovasure Alpha Defensin Control Kit further contains
- 1. Synovasure Alpha Defensin Positive Control
- 2. Synovasure Alpha Defensin Negative Control
- 3. Synovasure Control Reconstitution Bottle
The positive control contains 0.25 mL of 16 µg/mL alpha defensin in synthetic synovial fluid and the negative control contains 0.25 mL of synthetic synovial fluid.
Additional materials required but not provided include
1. Timer
### J. Standard/Guidance Document Referenced (if applicable):
EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition
EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition
EP12-A2, User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline-Second Edition
{3}------------------------------------------------
EP14-A3, Evaluation of Commutability of Processed Samples; Approved Guideline-Third Edition
EP25-A , Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline-Third Edition
# K. Test Principle:
Alpha defensins are antimicrobial peptides secreted by neutrophils in response to infection. The Synovasure Alpha Defensin Lateral Flow Test detects the presence of alpha defensin in synovial fluid specimens utilizing lateral flow technology. A sample of the synovial fluid is diluted and added to the lateral flow test device. The first pad in the device filters out the cellular material and the filtered sample contacts the buffering pad that contains the components for blocking and pH control. The sample then migrates to the pad containing the anti-alpha defensin antibody which is labeled with a colloidal gold conjugate. Finally, the sample mixture migrates to the which has immobilized anti-alpha defensin antibody and then to the control line that captures unbound antibody and verifies that the device flowed properly. A line present in the test zone indicates a positive result and the absence of a line in the test zone indicates a negative result.
# L. Performance Characteristics (if/when applicable):
- 1. Analytical performance:
- a. Precision/Reproducibility:
The precision study was performed at three external laboratories over a minimum of 5 days with 3 operators per site. 3 runs per day, 3 reagent lots, and 18 blinded samples per run consisting of 2-4 blinded replicates of each sample. Combined results are shown in Table 1.
{4}------------------------------------------------
| Panel Member | % Positive | 95%<br>Confidence<br>Interval | % Negative | 95%<br>Confidence<br>Interval |
|---------------|--------------------|-------------------------------|--------------------|-------------------------------|
| Negative | 1.0% (4/403) | 0.3-2.5% | 99.0%<br>(399/403) | 97.5-99.7% |
| High Negative | 9.9% (40/404) | 7.2-13.2% | 90.1%<br>(364/404) | 86.8-92.8% |
| Cutoff | 49.9%<br>(202/405) | 44.9-54.9% | 50.1%<br>(203/405) | 45.1-55.1% |
| Low Positive | 79.9%<br>(321/403) | 75.4-83.5% | 20.3% (82/403) | 16.5-24.6% |
| Positive | 96.0%<br>(388/404) | 93.6-97.7% | 4.0% (16/404) | 2.3-6.4% |
| High Positive | 98.5%<br>(396/402) | 96.8-99.5% | 1.5% (6/402) | 0.5-3.2% |
Table 1: Synovasure LFT Precision of Alpha Defensin Spiked into Synthetic Synovial Fluid
The low positive sample results failed to meet an acceptance criterion of 90% reactivity. To identify a root cause for this discrepancy, a sub-analysis was performed on data stratified form each site and obtained from each operator. The sponsor noted that data obtained from Operators 2 and 3 from Site 1 exhibited significantly lower reactivities in the low positive samples (Table 2). Furthermore, it was noted that the the actual concentration of some panel members was lower than expected due to an error in the preparation of the panel.
| Panel Member | % Positive<br>Site 1 | 95%<br>Confidence<br>Interval | % Positive<br>Site 2 | 95%<br>Confidence<br>Interval | % Positive<br>Site 3 | 95%<br>Confidence<br>Interval |
|---------------|----------------------|-------------------------------|----------------------|-------------------------------|----------------------|-------------------------------|
| Negative | 0.7% (1/134) | 0-4.1% | 1.5%<br>(2/135) | 0.2-5.2% | 0.7%<br>(1/134) | 0.0-4.1% |
| High Negative | 0.0% (0/135) | 0-2.7% | 3.7%<br>(5/134) | 1.2-8.5% | 25.9%<br>(35/135) | 18.8-34.2% |
| Cutoff | 25.2%<br>(34/135) | 18.1-33.4% | 51.9%<br>(70/135) | 43.1-60.5% | 72.6%<br>(98/135) | 64.3-79.9% |
| Low Positive | 66.4%<br>(89/134) | 57.8-74.3% | 80.0%<br>(108/135) | 72.3-86.4% | 92.5%<br>(124/134) | 86.7-96.4% |
| Positive | 93.3%<br>(126/135) | 87.7-96.9% | 97.0%<br>(131/135) | 92.6-99.2% | 97.8%<br>(131/134) | 93.6-99.5% |
| High Positive | 97.8%<br>(132/135) | 93.6-99.5% | 99.2%<br>(132/133) | 95.9-100% | 98.5%<br>(132/134) | 94.7-99.8% |
Table 2: Precision Performance Stratified by Site
An additional analysis was performed excluding the low positive data from Operators 2 and 3 at Site 1 and the low positive sample reactivity becomes 87% (271/315). According to the probit regression obtained in the C5-C95 study, this is in close agreement with the expected reactivity (88.14%) for the actual alpha defensin concentration of the low positive panel member (10.3 ug/mL). Furthermore, the
{5}------------------------------------------------
positive and negative control kit includes an alpha defensin concentration near the LoD and could provide an adequate control for user test interpretation.
- b. Linearity/assay reportable range:
Not applicable
- c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Stability
#### External Controls
To monitor the assay performance, reagent performance, and procedural errors, positive and negative external controls must be run in accordance with the guidelines or requirements of local, state, and/or federal regulations or accrediting organizations. The positive and negative controls are packaged lyophilized. Prior to use, the end user rehydrates the lyophilized pellets using the reconstitution fluid provided in the Synovasure Control Reconstitution Bottle following the instructions in the Synovasure Alpha Defensin Control Kit package insert.
#### Reagent Stability:
Reagent Integral Stability:
The sponsor evaluated the effect of freeze-thaw cycles on Synovasure LFT devices over time. While not reflective of the recommended storage temperatures (2-30°C) for assay components, the product may experience freezing and thawing during transport. A total of 30 Synovasure LFT kits (including the device and buffer bottle with sample cup and Microsafe tube) were stored at -20°C for one week and thawed at room temperature to evaluate performance of spiked synthetic synovial fluid positive and negative controls (N = 10). Additionally, another set of 30 Synovasure LFT kits were stored at -20°C for 24 h. thawed, and returned to -20°C for at least another 24 h three times. Performance of the Synovasure LFT against three synthetic synovial fluid quality control (QC) samples is reported below in Table 3.
{6}------------------------------------------------
| 3X Freeze-Thaw<br>Percent Positive (N) | | |
|----------------------------------------|---------------|----------------|
| QC1 (0 ug/mL) | QC2 (4 ug/mL) | QC3 (12 ug/mL) |
| 0% (0/10) | 0% (0/10) | 90% (9/10) |
| RT Controls | | |
| 0% (0/3) | 0% (0/3) | 100% (3/3) |
| 1X Freeze-Thaw | | |
| Neg. Control | Pos. Control | |
| 0% (0/10) | 100% (10/10) | |
| RT Controls | | |
| 0% (0/3) | 100% (3/3) | |
## Table 3: Synovasure LFT Kit Freeze-Thaw Stability
Three lots of Synovasure LFT assay kits were subjected to real-time stability testing at both 2-8°C and 30°C over a period of either 1021. 990. or 772 days. Positive (16 ug/mL alpha defensin) and negative controls from the associated control kit were evaluated at each time point. Assay performance was evaluated by comparison to a color card unit scale with scores <1 being negative. While assay output does decrease at higher temperatures, all 3 lots met the acceptance criteria over the entire course of the study at both 2-8°C and 30°C. An additional lot was subjected to an accelerated stability study at both 37℃ and 50℃. The assay kit stored at 37℃ met acceptance criteria for 224 days. Extrapolating these results utilizing the Arrhenius equation did not support the desired 24-month assay stability claim. However, the assay kit stored at 50°C met acceptance criteria for 190 days. The Arthenius equation extrapolation of these stability data to 30℃ do support a 24-month stability claim.
#### Calibrator Stability:
To confirm stability of rehydrated control reagents, 6 vials each of positive and negative control were reconstituted and tested so that reactivity was assessed on day 0, 5, 7, 12, 17, and 20 after resuspension. All controls met predetermined acceptance criteria with positive specimens maintaining assay reactivity during the 20 day length of the study.
Three lots of Synovasure LFT Control Kits were subjected to real-time stability testing. Positive control specimens were stored at both 2-8°C and evaluated on the Synovasure LFT over time to support a 36 month expiration date. The positive and negative controls, along with the Reconstitution Buffer, were shown to give correct test results with the lateral flow tests at 28 months for both storage temperatures.
#### Sample Stability:
Five negative synovial fluid specimens were each halved and divided into two independent sample groups with 1 group receiving a spike of alpha defensin at 23.2 ug/mL and the other group receiving a vehicle control (water). Each sample was then
{7}------------------------------------------------
aliquoted and stored at either 2-8°C. 13-16°C, 20-25°C, or 28-32°C and analyzed on days 0, 1, 3, 5, and 7. Aliquots were evaluated for correct identification by the Synovasure LFT. Results are listed below in Table 4.
| | % Correctly Identified at Indicated Storage Temperature | | | |
|-------------|---------------------------------------------------------|--------------|--------------|--------------|
| Time (Days) | 4-8°C | 13-16°C | 20-25°C | 28-32°C |
| T0 | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) |
| T1 | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) |
| T3 | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) |
| T5 | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) |
| T7 | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) |
Table 4: Specimen Stability Study Results
All samples agreed with the expected results.
#### Freeze-Thaw Stability Study
All clinical synovial fluid specimens evaluated by the sponsor were tested fresh without freezing. The only specimens that were frozen prior to testing were the alpha defensin spiked synthetic synovial fluid panels prepared for the precision study. In order to determine the stability of alpha defensin spiked synthetic synovial fluid panel members, the sponsor evaluated the effects of multiple freeze-thaw cycles on samples containing 2, 4, 8, 12 ug/mL alpha defensin. Specifically, samples were assayed on an alpha defensin ELISA after 1 freeze-thaw cvcle and after 3 separate freeze-thaw cycles. Generally, samples after a freeze-thaw cycle exhibited no loss of analyte signal due to the freezing process (Table 5).
| Conc.: | 2 $ \mu $ g/mL | | | 4 $ \mu $ g/mL | | | 8 $ \mu $ g/mL | | | 12 $ \mu $ g/mL | | |
|-------------|------------------|--------|--------|------------------|-------|-------|------------------|-------|-------|-------------------|-------|-------|
| Repeat | Wet | 1X FT | 3X FT | Wet | 1X FT | 3X FT | Wet | 1X FT | 3X FT | Wet | 1X FT | 3X FT |
| 1 | 0.26 | 0.24 | 0.21 | 0.44 | 0.48 | 0.58 | 0.90 | 1.24 | 0.86 | 1.04 | 1.14 | 1.06 |
| 2 | 0.26 | 0.24 | 0.25 | 0.47 | 0.47 | 0.55 | 0.90 | 0.95 | 0.87 | 1.18 | 1.18 | 1.11 |
| 3 | 0.26 | 0.24 | 0.25 | 0.47 | 0.59 | 0.57 | 0.96 | 0.99 | 1.00 | 1.19 | 1.33 | 1.10 |
| 4 | 0.25 | 0.24 | 0.25 | 0.46 | | 0.50 | 0.88 | | 1.08 | 1.30 | 1.19 | 1.31 |
| 5 | 0.25 | | 0.23 | 0.50 | | | 0.82 | | 1.24 | 1.13 | 1.21 | 1.07 |
| Average: | 0.25 | 0.24 | 0.24 | 0.47 | 0.52 | 0.55 | 0.89 | 1.06 | 1.01 | 1.17 | 1.21 | 1.13 |
| Difference: | | -0.012 | -0.014 | | 0.05 | 0.081 | | 0.17 | 0.12 | | 0.041 | -0.04 |
Table 5: Freeze-Thaw Stability of C5-C95 Study Synthetic Synovial Fluid Specimens
Samples containing a range of alpha defensin concentrations were also compared to a standard curve of serially diluted alpha defensin to estimate any difference from target concentrations after a single freeze-thaw cycle (Table 6). No significant differences were observed.
{8}------------------------------------------------
| Target<br>(µg/mL) | Fresh<br>(µg/mL) | Difference<br>(%) | 1X FT<br>(µg/mL) | Difference<br>(%) | % Positive<br>After 1X FT |
|-------------------|------------------|-------------------|------------------|-------------------|---------------------------|
| 0 | 0.1 | - | 0.4 | - | 0 |
| 2.0 | 1.7 | 15.0 | 1.8 | 9.0 | 0 |
| 4.0 | 3.8 | 5.0 | 4.3 | 7.0 | 0 |
| 4.8 | 4 | 16.7 | 4.8 | 0 | 0 |
| 5.6 | 5 | 10.7 | 5.5 | 1.0 | 0 |
| 6.4 | 6.1 | 4.7 | 6.2 | 3.0 | 0 |
| 8.0 | 7.9 | 1.3 | 8.9 | 12.0 | 33.3% (1/3) |
| 9.6 | 8.7 | 9.4 | 10.7 | 11.0 | 100% (3/3) |
| 10.4 | 10.2 | 1.9 | 11.8 | 13.0 | 100% (3/3) |
| 11.2 | 10.5 | 6.2 | 12.1 | 8.0 | 100% (3/3) |
| 12.0 | 11.1 | 7.5 | 12.9 | 8.0 | 100% (3/3) |
Table 6: Freeze-Thaw Stability of Spiked Synthetic Synovial Fluid Samples
#### d. Detection limit:
A study to establish the detection limit of the Synovasure LFT was conducted over 5 days utilizing 3 reagent lots. A panel of 88 samples was tested by 3 operators. This design results in a total of 120 replicates per sample. The results of this study were analyzed by probit regression to identify C5 and C95 concentrations. Results are listed below in Table 7.
Table 7: Establishment of C5/C95 Concentrations for Synovasure LFT
| Concentration<br>(µg/mL) | % Positive | 95% Confidence<br>Interval | % Negative | 95% Confidence<br>Interval |
|--------------------------|-----------------|----------------------------|------------|----------------------------|
| 0 | 0% (0/118) | 0-3.1% | 100% | 97.0-100% |
| 2.2 | 0% (0/120) | 0-3% | 100% | 97.0-100% |
| 4.3 | 5% (6/120) | 1.9-10.6% | 95.0% | 89.4-98.1% |
| 4.8 | 8.3% (10/120) | 4.1-14.8% | 91.7% | 85.2-95.9% |
| 5.4 | 11.7% (14/120) | 6.5-18.8% | 88.3% | 81.2-93.5% |
| 6.1 | 49.2% (59/120) | 39.9-58.5% | 50.8% | 41.6-60.1% |
| 8.9 | 77.5% (93/120) | 69.0-84.6% | 22.5% | 15.4-31.0% |
| 10.7 | 93.3% (112/120) | 87.3-97.1% | 6.7% | 2.9-12.7% |
| 11.9 | 93.3% (112/120) | 87.3-97.1% | 6.7% | 2.9-12.7% |
| 12.2 | 92.5% (111/120) | 86.2-96.5% | 7.5% | 3.5-13.8% |
| 13 | 96.7% (116/120) | 91.7-99.1% | 3.3% | 0.9-8.31% |
The probit model identified C5, C50, and C95 alpha defensin concentrations of 4.17, 7.05, and 11.95 ug/mL, respectively.
The verification of the analytical sensitivity was performed over 1 day using 3 reagent lots. 20 replicates of 11 panel members that span the expected C5-C95 range were evaluated for a total of 220 samples per lot. Results are below in Table 8.
{9}------------------------------------------------
| Concentration<br>(ug/mL) | % Positive | 95% Confidence<br>Interval | % Negative | 95% Confidence<br>Interval |
|--------------------------|---------------|----------------------------|------------|----------------------------|
| 0 | 0% (0/60) | 0-6.0% | 100% | 94.0-100% |
| 1.385 | 0% (0/60) | 0-6.0% | 100% | 94.0-100% |
| 2.505 | 0% (0/60) | 0-6.0% | 100% | 94.0-100% |
| 4.675 | 21.7% (13/60) | 12.1-34.2% | 78.3% | 65.8-87.9% |
| 5.58 | 42.4% (25/59) | 29.6-55.9% | 57.6% | 44.1-70.4% |
| 6.875 | 53.3% (32/60) | 40.0-66.3% | 46.7% | 33.7-60.0% |
| 8.01 | 73.3% (44/60) | 60.3-83.9% | 26.7% | 16.1-39.7% |
| 10.46 | 93.3% (56/60) | 83.8-98.2% | 6.7% | 1.9-16.2% |
| 11.035 | 91.7% (55/60) | 81.6-97.2% | 8.3% | 2.8-18.4% |
| 11.87 | 98.3% (59/60) | 91.1-100% | 1.7% | 0.04-8.9% |
| 13.88 | 96.7% (58/60) | 88.5-99.6% | 3.3% | 0.4-11.5% |
Table 8: Verification of C5/C95 Concentrations for Synovasure LFT
The results of the verification study establish C5 and C95 alpha defensin concentrations of 3.44 and 11.66 ug/mL, respectively.
The sponsor evaluated Synovasure LFT performance at temperatures including room temperature (22-25°C), 12-18°C, and 27-32°C. Results included in Table 9 demonstrate that assay sensitivity deteriorates at temperatures ≥27°C. The assay package insert specifically states that the assay should be performed at room temperature only.
Table 9: Lateral Flow Assay In-Use Temperature Effects
| | | | % Correctly Classified According to QC<br>Criteria | | | |
|-----------------|-----------------|----|----------------------------------------------------|---------|----------|----------|
| Target<br>Temp. | Actual<br>Temp. | N | 0 µg/mL | 4 µg/mL | 12 µg/mL | 16 µg/mL |
| 27-32 | 30.3-30.8 | 20 | 100 | 100 | 15 | 85 |
| 27 | 26.8 | 20 | 100 | 100 | 80 | 100 |
| 22-25 | 23.3-23.7 | 20 | 100 | 100 | 100 | 100 |
| 12-18 | 12.8-17.9 | 20 | 100 | 100 | 100 | 100 |
- e. Analytical specificity:
Cross reactivity:
Not applicable
Interference:
Exogenous and endogenous interfering substances were evaluated for interference by spiking into negative clinical synovial fluid specimens. Half of samples tested contained 17 ug/mL alpha defensin and the other half a water vehicle control. Both alpha defensin spiked and unspiked specimens were further subdivided to included
{10}------------------------------------------------
interferent or carrier solvent controls. Since synthetic synovial fluid used by the sponsor primarily contains physiological levels of Hyaluronic acid (HA), the sponsor also wanted to evaluate the effects of differing HA concentrations on assay performance.
| | | Low Positive | | Negative | |
|----------------------------------------------|---------------------------|--------------|-----------|----------|-----------|
| Substance | Test<br>Concentratio<br>n | N | % Correct | N | % Correct |
| Rheumatoid Factor | 300 IU | 25 | 100% | 20 | 100% |
| Whole Blood<br>Hemoglobin | 12.1 g/dL | 15 | 100% | 20 | 100% |
| Lysed Blood<br>Hemoglobin | 8.7 g/dL | 9 | 100% | 3 | 100% |
| Bilirubin<br>(unconjugated) | 20 mg/dL | 25 | 100% | 20 | 100% |
| Bilirubin (conjugated) | 29 mg/dL | 25 | 100% | 20 | 100% |
| Triglyceride | 418 mg/dL | 25 | 96% | 20 | 100% |
| Hyaluronic Acid | 0 mg/mL | 25 | 100% | 25 | 100% |
| Hyaluronic Acid | 2 mg/mL | 25 | 100% | 25 | 100% |
| Hyaluronic Acid | 4 mg/mL | 25 | 100% | 25 | 100% |
| Hyaluronic Acid | 8 mg/mL | 25 | 100% | 25 | 100% |
| Metal Ion Cobalt | 150 mg/L | 25 | 96% | 20 | 100% |
| Metal Ion Chromium | 150 mg/L | 25 | 100% | 20 | 100% |
| Metal Ion Titanium | 150 mg/L | 25 | 100% | 20 | 100% |
| Bone Cement | 10 mg/mL | 25 | 100% | 20 | 100% |
| Ultra-high-molecular-<br>weight polyethylene | 10 mg/mL | 25 | 92% | 20 | 100% |
Table 10: Interfering Substances Percent Agreement with Expected Results
An interference effect was observed when lyophilized hemoglobin was evaluated on the Synovasure LFT with an effect that disappeared at 25 mg/dL concentrations. Hemoglobin interference was further assessed by testing synovial fluid samples containing both lysed and unlysed whole blood. Since no noticeable interference was observed with these specimens, the interefence from lyophilized hemoglobin was likely an artifact. Importantly, while whole blood does not appear to negatively interfere with assay performance, too much whole blood in the synovial fluid sample could cause false negative results by diluting any alpha defensin present. This point is noted in a limitation described on the package insert and evidenced in the clinical studies.
### f. High Dose Hook Effect Study
Performance of the Synovasure LFT was evaluated utilizing high concentrations of alpha defensin to demonstrate the absence of a negative hook effect. A single operator utilizing a single lot of reagents tested specimens corresponding to physiological
{11}------------------------------------------------
alpha defensin concentrations of 0. 50. 100. 500. 1.000. 5.000 and 10.000 us/mL. There was no visible decline in band intensity at these higher analyte concentrations.
- f. Assay cut-off:
Not applicable
- 2. Comparison studies:
- a. Method comparison with predicate device:
Not applicable
- b. Matrix comparison:
Some of the assay validation studies utilized synthetic synovial fluid (SSF) as a sample matrix while other testing utilized samples prepared from clinical synovial fluid specimens. Clinical matrix was specifically utilized for interference testing. sampling method comparison studies, stability studies, and in-use stability testing. A study performed to assess in-use temperature as well as the precision and C5-C95 studies were conducted utilizing SSF as a sample matrix due to limited availability of clinical specimen volumes to meet experimental designs. Preliminary comparisons of these two distinct sample matrices utilized only an alpha defensin concentration of 16.4 ug/mL. An additional matrix comparison study was therefore performed to demonstrate equivalent performance of the device at analyte levels below the assay LoD for both matrices. Alpha defensin was spiked into negative clinical pool (NP) of synovial fluid and synthetic synovial fluid at concentrations of 0, 6.7, ad 8 ug/mL and both matrices were evaluated on the Synovasure LFT. No significant difference between synthetic synovial fluid and clinical synovial fluid matrix was observed.
| Matrix | Alpha<br>defensin<br>(ug/mL) | Percent<br>Positive | 95% CI |
|--------|------------------------------|---------------------|-------------|
| SSF | 0 | 0% (0/40) | 0 - 8.81% |
| NP | 0 | 0% (0/40) | 0 - 8.81% |
| SSF | 6.7 | 5.0% (2/40) | 0.6 - 16.9% |
| NP | 6.7 | 5.0% (2/40) | 0.6 - 16.9% |
| SSF | 8.0 | 55.0% (22/40) | 38.5-70.7% |
| NP | 8.0 | 67.5% (27/40) | 50.9-81.4% |
| SSF | 16.4 | 100% (25/25) | 86.3-100% |
| NP | 16.4 | 100% (25/25) | 86.3-100% |
Table 11 : Performance Comparison of Clinical Matrix and Synthetic Synovial Fluid
{12}------------------------------------------------
### 3. Clinical studies:
- a. Clinical Sensitivity:
## Prospective Study
Reference method
The reference method utilized by the sponsor entails the identification of Musculoskeletal Infection Society (MSIS) criteria as listed below. Patients are diagnosed with periprosthetic joint infection if they meet one major criterion or any 3 of the 5 minor criteria. Final status determination was adjudicated by a two-physician panel, with discrepant opinions being resolved by consultation of a third physician.
### Table 12: MSIS PJI Diagnostic Criteria
| Major Criteria |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| 1. A sinus tract communicating with the prosthesis |
| 2. A pathogen is isolated by culture from two separate tissue or fluid samples obtained from the affected prosthetic joint |
| Minor Criteria |
| 1. Elevated Erythrocyte Sedimentation rate (≥30mm/hr) and C-reactive protein levels (≥10 mg/L) |
| 2. Elevated synovial fluid WBC count (≥3000) |
| 3. Elevated synovial fluid neutrophil percentage (≥80%) |
| 4. Isolation of a microorganism in one periprosthetic tissue or fluid culture |
| 5. ≥5 neutrophils per high-powered field in 5 high-power fields from histological analysis of periprosthetic tissue at 400X magnification |
#### Acceptance Criteria
The sponsor employed a sensitivity criterion of >90% with lower confidence bound of >85% and a specificity criterion of >90% with a lower confidence bound of >90% for combined prospective and retrospective data.
#### Study Demographics
A total of 386 subjects were evaluated for prospective study eligibility. 81 subjects were excluded because there was insufficient information for an MSIS diagnosis, the subject did not have a total knee or hip arthroplasty, the subject did not have revision surgery, the subject had a joint injection or synovial fluid collection recently, or there was insufficient sample volume for testing. The clinical study therefore analyzed a total of 305 prospective synovial fluid samples and 65 retrospective fresh remnant samples. The sponsor provided the demographic breakdown for prospective clinical study samples as identified in Table 13.
{13}------------------------------------------------
| Variable | Value |
|----------------------------------------------------------|-------------|
| Age (years) | |
| Mean | 64.8 |
| STDEV | 9.76 |
| Min. | 38 |
| Max. | 92 |
| Gender (n, %) | |
| Male | 131 (43.0%) |
| Female | 174 (57.0%) |
| Race (n, %) | |
| White | 266 (87.2%) |
| Asian | 1 (0.3%) |
| African American | 26 (8.5%) |
| Native American | 2 (0.7%) |
| Other | 10 (3.3%) |
| Infection History (n, %) | |
| Yes | 87 (28.5%) |
| No | 218 (71.5%) |
| Antibiotic History (n, %) | |
| Yes | 40 (13.1%) |
| No | 77 (25.2%) |
| Inflammatory Medications<br>within the last month (n, %) | |
| Yes | 43 (14.1%) |
| No | 262 (85.9%) |
| Immunocompromised<br>Conditions (n, %) | |
| Yes | 19 (6.2%) |
| No | 286 (93.8%) |
| Medical History (n, %) | |
| Yes | 219 (71.8%) |
| No | 86 (28.2%) |
Table 13: Prospective Clinical Study Demographics and Baseline Characteristics
### External Controls
During the conduct of the clinical trial protocol, external control testing was performed once per week per operator performing lateral flow testing. The tests were conducted per kit instructions and no control failures were observed over the 74 weeks of clinical testing across all sites.
{14}------------------------------------------------
## Enrollment Inclusion/Exclusion Criteria
The prospective clinical study had the following subject inclusion criteria:
- 1. Subject had a total knee and/or hip joint arthroplasty
- 2. Subject was evaluated for revision surgery
- a. Operative samples were required for full MSIS classification
- 3. Subject was >22 years of age
- 4. Subject had no recent injections or surgeries of the joint (within past 6 weeks)
- 5. Subject had all the medical tests required to allow MSIS classification
- 6. Subject signed informed consent form.
The prospective clinical study had the following subject exclusion criteria:
- 1. Subject did not have a total knee and/or hip joint arthroplasty
- 2. Healthy subject without medical need for aspiration
- 3. Subject did not have a revision surgery
- 4. Subject had a diagnostic synovial fluid specimen collection within the past 14 days
- 5. Subject was <22 years of age
- 6. Subject had an injection, lavage, or surgery of the joint within the past 6 weeks
- 7. Subject did not have all the medical tests required for MSIS classification
- 8. Subject did not sign informed consent form
# Analysis
The sponsor conducted a prospective clinical trial at 3 US medical centers with high volumes of revision surgery. Enough specimens were collected to ensure at least 50 positives were obtained as determined by Musculoskeletal Infection Society (MSIS) criteria. After adjudication, 57 prospective samples were identified as positive and the Synovasure LFT achieved a sensitivity of 89.5% (95% CI:78.5-96.1%). (Table 14).
### Table 14: Clinical Performance of Synovasure LFT in Prospective Clinical Specimens
| | | MSIS PJI Diagnosis | | |
|------------------------------------------|-------|--------------------|-----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 51 | 13 | 64 |
| | N | 6 | 235 | 241 |
| | Total | 57 | 248 | 305 |
| Sensitivity (95% CI): 89.5% (78.5-96.0%) | | | | |
| Specificity (95% CI): 94.8% (91.2-97.2%) | | | | |
Whole blood contamination potentially reduces the alpha defensin concentration in the synovial fluid sample. Therefore, the sponsor performed an additional analysis excluding those samples with higher levels of whole blood contamination (>20%
{15}------------------------------------------------
dilution by volume) (Table 15). Out of 17 samples with greater than 20% whole blood contamination, 1 was a true positive, 3 were false negatives, and 13 were true negatives. To address this risk of generating false results, a specific limitation noting the potential for false negative results in samples containing a significant amount of blood is listed in the package insert.
Table 15: Clinical Performance of Synovasure LFT in Prospective Clinical Specimens (RBC>1x106/uL excluded)
| | | MSIS Diagnosis | | |
|------------------------------------|-------|----------------|-----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 50 | 13 | 63 |
| | N | 3 | 222 | 225 |
| | Total | 53 | 235 | 288 |
Two false negative Synovasure LFT results from the clinical trial were identified in patients with a draining sinus tract connecting to the joint. One of these was also identified as contaminated with greater than 20% whole blood. A draining sinus should be readily observable and is also a major diagnostic criterion for PJI: therefore. the impact of this on the patient's diagnosis would be minimal.
### Retrospective Remnant Clinical Samples
To supplement the testing of fresh prospectively collected specimens. the sponsor also obtained fresh retrospective synovial fluid collections to bring the total number of positive specimens tested to at least 100. The residual samples were anonymized. assigned new case ID numbers and tested. Synovasure LFT results were compared to a status determination based positive confirmation of three minor MSIS criteria (neutrophil %, positive culture, and WBC count). A total of 65 MSIS positive retrospective samples were evaluated interspersed with sufficient negative specimens to maintain relative disease prevalence consistent with what was observed in the prospective study population. The Synovasure LFT exhibited a 98.5% (95% CI: 91.7-100%) positive percent agreement with MSIS diagnosis for these retrospective specimens.
#### Synovasure LFT Analysis of Covariates
Performance of the Synovasure LFT in the prospective clinical study was further assessed among several covariates which could potentially impact device performance: age, gender, race, history of infection, history of inflammatory disease, use of anti-inflammatory medication. gram positive or gram negative culture isolates. type of prosthetic joint, and individual MSIS criteria. Insignificant differences in assay performance were observed based upon these variables (Table 16 - Table 32).
{16}------------------------------------------------
| | | MSIS PJI Diagnosis | | |
|------------------------------------------|-------|--------------------|-----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 45 | 13 | 58 |
| | N | 6 | 202 | 208 |
| | Total | 51 | 215 | 266 |
| Sensitivity (95% CI): 88.2% (76.1-95.6%) | | | | |
| Specificity (95% CI): 94.0% (89.9-96.7%) | | | | |
Table 16: Prospective Study Performance of Synovasure LFT for White Subjects
Table 17: Prospective Study Performance of Synovasure LFT for African American Subjects
| | | MSIS PJI Diagnosis | | | |
|------------------------------------|-----------------------------------------|--------------------|----|-------|--|
| | | P | N | Total | |
| Synovasure<br>Lateral Flow<br>Test | P | 5 | 0 | 5 | |
| | N | 0 | 21 | 21 | |
| | Total | 5 | 21 | 26 | |
| | Sensitivity (95% CI): 100% (47.8-100%) | | | | |
| | Specificity (95% CI): 94.0% (83.9-100%) | | | | |
| | | Table 18: Prospective Study Performance of Synovasure LFT for Other Races | | | |
|------------------------------------|-------|---------------------------------------------------------------------------|----|-------|--|
| | | MSIS PJI Diagnosis | | | |
| | | P | N | Total | |
| Synovasure<br>Lateral Flow<br>Test | P | 1 | 0 | 1 | |
| | N | 0 | 12 | 12 | |
| | Total | 1 | 12 | 13 | |
| | | Sensitivity (95% CI): 100% (2.5-100%) | | | |
| | | Specificity (95% CI): 100% (73.5-100%) | | | |
1. Other races include Asian, American Indian or Alaskan Native, and races identified as "Other"
Table 19: Prospective Study Performance of Synovasure LFT for Male Subjects (excluding RBC>1 x106 /uL samples)
| | | MSIS PJI Diagnosis | | | |
|------------------------------------|------------------------------------------|--------------------|----|-------|--|
| | | P | N | Total | |
| Synovasure<br>Lateral Flow<br>Test | P | 24 | 9 | 33 | |
| | N | 1 | 90 | 91 | |
| | Total | 25 | 99 | 124 | |
| | Sensitivity (95% CI): 96.0% (79.6-99.9%) | | | | |
| | Specificity (95% CI): 90.9% (83.4-95.8%) | | | | |
Table 20: Prospective Study Performance of Synovasure LFT for Female Subjects (excluding RBC>1 x106/μL samples)
| | MSIS PJI Diagnosis | | | |
|--------------|--------------------|----|-------|-----|
| | P | N | Total | |
| Synovasure | P | 26 | 4 | 30 |
| Lateral Flow | N | 2 | 132 | 134 |
{17}------------------------------------------------
| Test | Total | 28 | 136 | 164 |
|------|------------------------------------------|----|-----|-----|
| | Sensitivity (95% CI): 92.9% (76.5-99.1%) | | | |
| | Specificity (95% CI): 97.1% (92.6-99.2%) | | | |
To evaluate whether prior infections impact device performance, assay results were tabulated among patients with any prior history of infection or infectious disease (e.g., sinusitis, pneumonia or other respiratory infections, urinary tract infections, the common cold, HIV, lyme disease, viral hepatitis, etc.) within the last 6 months (Table 21) and for for those without such history (Table 22).
Table 21: Prospective Study Performance of Synovasure LFT for Subjects with Infection History (excluding RBC>1 x109/μL samples)
| | | MSIS PJI Diagnosis | | |
|------------------------------------|-------|------------------------------------------|----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 30 | 6 | 36 |
| | N | 1 | 48 | 49 |
| | Total | 31 | 54 | 85 |
| | | Sensitivity (95% CI): 96.8% (83.3-99.9%) | | |
| | | Specificity (95% CI): 88.9% (77.4-95.8%) | | |
Table 22: Prospective Study Performance of Synovasure LFT for Subjects without Infection History (excluding RBC>1 x106 /uL samples)
| | | MSIS PJI Diagnosis | | |
|------------------------------------|-------|------------------------------------------|-----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 20 | 7 | 27 |
| | N | 2 | 174 | 176 |
| | Total | 22 | 181 | 203 |
| | | Sensitivity (95% CI): 90.9% (70.8-98.9%) | | |
| | | Specificity (95% CI): 96.1% (92.2-98.4%) | | |
Ongoing use of antibiotics might reduce infection severity and affect the performance of the Synovasure Lateral Flow test. Assay sensitivity and specificity were evaluated among study subjects who were actively on an antibiotic regimen upon study enrollment (Table 23) and those not currently taking antibiotic medications (Table 24).
{18}------------------------------------------------
| | | MSIS PJI Diagnosis | | |
|------------------------------------------|-------|--------------------|----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 25 | 1 | 26 |
| Synovasure<br>Lateral Flow<br>Test | N | 1 | 12 | 13 |
| Synovasure<br>Lateral Flow<br>Test | Total | 26 | 13 | 39 |
| Sensitivity (95% CI): 96.2% (83.3-99.9%) | | | | |
| Specificity (95% CI): 92.3% (64.0-99.8%) | | | | |
Table 23: Prospective Study Performance of Synovasure LFT for Subjects with Ongoing Antibiotic Use (excluding RBC>1 x106/uL samples)
| Table 24: Prospective Study Performance of Synovasure LFT for Subjects without |
|--------------------------------------------------------------------------------|
| Ongoing Antibiotic Use (excluding RBC>1 x106/µL Samples) |
| | | MSIS PJI Diagnosis | | |
|--------------|-------|------------------------------------------|----|-------|
| | | P | N | Total |
| Synovasure | P | 8 | 5 | 13 |
| Lateral Flow | N | 1 | 60 | 61 |
| Test | Total | 9 | 65 | 74 |
| | | Sensitivity (95% CI): 88.9% (51.8-99.7%) | | |
| | | Specificity (95% CI): 92.3% (83.0-97.5%) | | |
The presence of an underlying inflammatory or autoimmune disease (e.g., rheumatoid arthritis, lupus, diabetes, multiple sclerosis, autoimmune hepatitis, psoriasis, scleroderma, sarcoidosis, etc.) was also evaluated for potential impact on Synovasure LFT performance. Study subjects with a history of any inflammatory disease at least 6 months before enrollment were analysed (Table 25) and compared to those with a medical history of no inflammatory or autoimmune disorders (Table 26).
| Table 25: Prospective Study Performance of Synovasure LFT for Subjects with | |
|-----------------------------------------------------------------------------|--|
| Inflammatory Disease History (excluding RBC>1 x106/μL samples) | |
| | | MSIS PJI Diagnosis | | |
|------------------------------------------|-------|--------------------|----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 12 | 2 | 14 |
| | N | 1 | 52 | 53 |
| | Total | 13 | 54 | 67 |
| Sensitivity (95% CI): 92.3% (64.0-99.8%) | | | | |
| Specificity (95% CI): 96.3% (87.3-99.5%) | | | | |
Table 26: Prospective Study Performance of Synovasure LFT for Subjects without Inflammatory Disease History (excluding RBC>1 x109/uL samples)
| | | MSIS PJI Diagnosis | | |
|------------------------------------|-------|------------------------------------------|-----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 38 | 11 | 49 |
| | N | 2 | 170 | 172 |
| | Total | 40 | 181 | 221 |
| | | Sensitivity (95% CI): 95.0% (83.1-99.4%) | | |
{19}------------------------------------------------
### Specificity (95% CI): 93.9% (89.4-96.9%)
The performance of the Synovasure LFT was also evaluated in subjects who are currently on anti-inflammatory or autoimmune medications (e.g., steroids or immunosuppressants) or have taken such medications in the past 6 months (Table 27) and those subjects who have not taken such medications (Table 28).
### Table 27: Prospective Study Performance of Synovasure LFT for Subjects Using Anti-Inflammatory Medication (excluding RBC>1 x106/µL samples)
| | | MSIS PJI Diagnosis | | | |
|------------------------------------|------------------------------------------|--------------------|----|-------|--|
| | | P | N | Total | |
| Synovasure<br>Lateral Flow<br>Test | P | 10 | 0 | 10 | |
| | N | 1 | 30 | 31 | |
| | Total | 11 | 30 | 41 | |
| | Sensitivity (95% CI): 90.9% (58.7-99.8%) | | | | |
| | Specificity (95% CI): 100% (88.4-100%) | | | | |
| Table 28: Prospective Study Performance of Synovasure LFT for Subjects without |
|--------------------------------------------------------------------------------|
| Current Anti-Inflammatory Medications (excluding RBC>1 x106/μL samples) |
| | | MSIS PJI Diagnosis | | Total |
|------------------------------------|------------------------------------------|--------------------|-----|-------|
| | | P | N | |
| Synovasure<br>Lateral Flow<br>Test | P | 40 | 13 | 53 |
| | N | 2 | 192 | 194 |
| | Total | 42 | 205 | 247 |
| | Sensitivity (95% CI): 95.2% (83.8-99.4%) | | | |
| | Specificity (95% CI): 93.7% (89.4-96.6%) | | | |
Since alpha defensins are host response proteins, it is possible that their expression level varies depending on the type of organism responsible for the joint infection. Performance of the Synovasure LFT was compared in study subjects who had a gram positive pathogen isolated by culture (Table 29) versus those study subjects with a gram negative pathogen isolated by culture (Table 30).
### Table 29: Prospective Study Performance of Synovasure LFT for Subjects with Gram Positive Culture (excluding RBC>1 x106/uL samples)
| | | MSIS PJI Diagnosis | | |
|------------------------------------------|-------|--------------------|----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 29 | 2 | 31 |
| | N | 3 | 21 | 24 |
| | Total | 32 | 23 | 55 |
| Sensitivity (95% CI): 90.6% (75.0-98.0%) | | | | |
| Specificity (95% CI): 91.3% (72.0-98.9%) | | | | |
{20}------------------------------------------------
| | | MSIS PJI Diagnosis | | |
|------------------------------------|-------|------------------------------------------|---|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 6 | 0 | 6 |
| | N | 1 | 0 | 1 |
| | Total | 7 | 0 | 7 |
| | | Sensitivity (95% CI): 85.7% (42.1-99.6%) | | |
| | | Specificity (95% CI): Non-estimable | | |
Table 30: Prospective Study Performance of Synovasure LFT for Subjects with Gram Negative Culture (excluding RBC>1 x106/uL samples)
All of the data from the prospective study (including specimens with greater than 1 x106/uL RBC contamination) were also analyzed to evaluate Synovasure LFT Performance in the presence of a positive culture (Table 31), a key component of MSIS diagnostic criteria, and in samples for which no culture was positive (Table 32).
Table 31: Prospective Study Performance of Synovasure LFT for Subjects with at least one Positive Culture
| | | MSIS PJI Diagnosis | | |
|------------------------------------|-------|--------------------|----|-------|
| | | P | N | Total |
| Synovasure<br>Lateral Flow<br>Test | P | 36 | 2 | 38 |
| | N | 5 | 23 | 28 |
| | Total | 41 | 25 | 66 |
Table 32: Prospective Study Performance of Synovasure LFT for Subjects with all Negative Cultures
| | MSIS PJI Diagnosis | | Total | |
|------------------------------------------|--------------------|----|-------|-----|
| | P | N | | |
| Synovasure<br>Lateral Flow<br>Test | P | 15 | 11 | 26 |
| | N | 1 | 210 | 211 |
| | Total | 16 | 221 | 237 |
| Sensitivity (95% CI): 93.8% (69.8-99.8%) | | | | |
| Specificity (95% CI): 92.0% (91.3-97.5%) | | | | |
{21}------------------------------------------------
### Prevalence Analysis
A detailed description of the type of infections observed during the clinical study is provided in Table 33.
### Table 33: PJI Prevalence Overall and by Type of Infection for All Prospective Subjects with a Synovial Fluid and/or Tissue Culture Performed
| | Positive PJI<br>Diagnosis<br>(N, [%, 95% CI]) | Negative PJI<br>Diagnosis<br>(N, [%, 95% CI]) | Total<br>(N, [%, 95% CI]) |
|------------------------------------------|-----------------------------------------------|-----------------------------------------------|---------------------------|
| Subjects with Gram positive<br>culture | 35 (61%, 48-74%) | 24 (10%, 6-14%) | 59 (19%, 15-24) |
| Subjects with a Gram Negative<br>Culture | 7 (12%, 5-24%) | 1 (0%, 0-2%) | 8 (3%, 1-5%) |
| Subjects with a Fungal Culture | 1 (2%, 0-9%) | 1 (0%, 0-2%) | 2 (1%, 0-2%) |
| Subjects with any positive<br>culture | 41 (72%, 58-83%) | 25 (10%, 7-15%) | 66 (22%, 17-27) |
| Subjects with all negative<br>cultures | 16 (28%, 17-42) | 221 (90%, 85-93%) | 237 (78%, 73-83%) |
| Total subjects with culture<br>performed | 57 | 246 | 303 |
- b. Clinical specificity:
See section M.3a above.
- c. Other clinical supportive data (when a. and b. are not applicable):
See section M.3a above.
- 4. Clinical cut-off:
Not applicable.
- 5. Expected values/Reference range:
Not Applicable
## M. Instrument Name:
Not applicable. The device does not utilize an instrument for result generation.
{22}------------------------------------------------
### N. System Descriptions:
### 1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ___________ or No ________________________________________________________________________________________________________________________________________________________
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ___________ or No ____X_________
- 2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ____________ or No _______________________________________________________________________________________________________________________________________________________
The device does not contain any software or instrument components.
- 3. Specimen Identification:
Not applicable.
- 4. Specimen Sampling and Handling:
Not applicable.
- 5. Calibration:
Not applicable.
- 6. Quality Control:
Not applicable.
## O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None.
- P. Proposed Labeling:
{23}------------------------------------------------
The labeling supports the decision to grant the De Novo request for this device.
| | | | O. Identified Risks to Health and Identified Mitigations |
|--|--|--|----------------------------------------------------------|
| | | | |
| | | | |
| | | | |
| Identified Risks to Health | Mitigation Measures |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Risk of false test results | Certain device descriptions, performance<br>characteristics, results interpretation<br>information, limitations, and study details in<br>labeling.<br>Certain device description information,<br>demographic analysis, validation procedures,<br>risk mitigation strategies and end user trainings,<br>and studies.<br>Collection device specification. |
| Failure to correctly interpret test results | Certain device descriptions, performance<br>characteristics, results interpretation<br>information, limitations, and study details in<br>labeling.<br>Certain demographic analysis, validation<br>procedures, risk mitigation strategies and end<br>user trainings, and studies. |
| Failure to correctly operate the device | Certain device descriptions, performance<br>characteristics, results interpretation<br>information, limitations, and study details in<br>labeling.<br>Certain demographic analysis, validation<br>procedures, risk mitigation strategies and end<br>user trainings, and studies.<br>Collection device specification. |
# R. Benefit/Risk Analysis:
### Summary of the Assessment of Benefit
The benefit of the assay is aiding the accurate diagnosis of prosthetic joint infection (PI). Accurate diagnosis of prosthetic joint infection can be helpful to initiate appropriate treatment for prosthetic joint infection, including, but not limited to, antibiotics and revision surgery. Appropriate treatment of PJI can lead to alleviation of symptoms associated with infection and restoration of function. Additionally, appropriate exclusion of a prosthetic joint infection will aid clinicians in deciding to retain existing hardware.
### Summary of the Assessment of Risk
The risks associated with the device, when used as intended, are those related to the risk of false test results. failure to correctly interpret the test results, and failure to correctly operate the device.
The risk of a false positive test result is improper patient management, including inappropriate administration of prolonged courses of antibiotics and inappropriate
{24}------------------------------------------------
explanting of hardware. Inappropriate administration of prolonged courses of antibiotics is associated with toxicity, allergic reactions, and adverse outcomes, including secondary infections such as C. difficile colitis. Inappropriate explanting of hardware may further obscure the anatomy of a joint, decrease function, worsen symptoms, and complicate further surgical manipulation of the joint. The risk of a false positive test is mitigated by the intended use clearly stating that the assay is intended as an adjunct to aid in the diagnosis of infection, in conjunction with other clinical and diagnostic findings.
The risk of a false negative test result is improper patient management, including inappropriate discontinuation of antibiotics or failure to treat a prosthetic joint infection with antibiotics and explanting of infected hardware. Failure to treat a prosthetic joint infection could lead to decreased function of the joint and worsen symptoms of infection. If not treated, a prosthetic joint infection could develop into a more complicated or a systemic infection, including skin and soft tissue infection, acute or chronic osteomyelitis, bacteremia, and/or sepsis. A patient who is symptomatic from a prosthetic joint infection will likely return to care, most likely delaying treatment instead of resulting in a complete failure to diagnose and treat. The risk of a false negative test is mitigated by the fact that this assay is intended as an adjunct to aid in the diagnosis of infection in conjunction with other clinical and diagnostic findings.
Failure to correctly operate the device can lead to false test results. Failure to correctly interpret test results can lead to treatment of a clinically positive patient in the same manner as a false negative test result and a clinically negative patient in the same manner as a false positive test result with the corresponding implications discussed above.
#### Summary of the Assessment of Benefit-Risk
General controls are insufficient to mitigate the risks associated with the device. However, the probable clinical benefits outweigh the potential risks for the proposed assay, considering the mitigations of the risks provided for in the listed special controls established for this device as well as general controls. The required special controls will help ensure that errors will be uncommon and will facilitate accurate assay implementation and interpretation of results. The clinical performance observed in the clinical trial suggests that errors will be uncommon and that the assay will provide substantial benefits to patients in the diagnosis of prosthetic joint infections and when used in conjunction with other clinical and diagnostic findings.
#### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
#### S. Conclusion
The information provided in this De Novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.3230. FDA believes that the stated special controls, in combination with the applicable general controls, provide a reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:
{25}------------------------------------------------
Product Code(s): QGN Device Type: Device to detect and measure non-microbial analytes to aid in the detection and identification of localized human infections Class: II (special controls) Regulation: 21 CFR 866.3230
- (a) Identification: A device to detect and measure non-microbial analytes to aid in the detection and identification of localized human infections is identified as an in vitro device intended for the detection and qualitative measurement, quantitative measurement, or both of one or more non-microbial analytes in human clinical specimens to aid in the assessment, identification, or both of a localized microbial infection when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.
- (b) Classification: Class II (special controls). The special controls for this device are:
- 1. Any sample collection device used must be FDA-cleared. -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of human specimens; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
- 2. The labeling required under 21 CFR 809.10(b) must include:
- i. An intended use with a detailed description of what the device detects and measures, the type of results provided to the user, the sample type, whether the measure is qualitative and/or quantitative, the clinical indications for the test use, and the specific population(s) for which the device is intended.
- A detailed description of the performance characteristics of the device for all ii. intended specimen types from the analytical and clinical studies (as applicable) required under paragraphs 3(ii) and 3(iii).
- iii. A detailed explanation of the interpretation of results, including acceptance criteria for evaluating the validity of individual runs (e.g., assessment of internal and/or external quality controls, as applicable).
- iv. The following limiting statements:
- A statement that a negative test result does not preclude the possibility (A) of infection:
- A statement that the test results should be interpreted in conjunction with (B) other clinical and laboratory data available to the clinician;
- A statement that consistent device performance is dependent on (C) adequate specimen collection, transport, storage, and processing. Failure
{26}------------------------------------------------
to observe proper procedures in any one of these steps can lead to incorrect results; and
- A statement that details any limitations associated with the samples, as (D) appropriate (e.g., collected on the day of admission to the ICU).
- 3. Design verification and validation must include the following:
- i. A detailed device description, including as appropriate, all device parts; control elements incorporated into the test procedure; instrument requirements; reagents required but not provided; and the principle of device operation and test methodology, including all pre-analytical methods for the processing of specimens and the methodology from obtaining a sample to the result; design of primer/probe sequences; rationale for target analyte selection; and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported result).
- ii. Detailed documentation of analytical studies including analytical sensitivity (Limit of Detection, Limit of Quantitation, and Limit of Blank), inclusivity, crossreactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross-contamination, specimen stability, within-lab precision, reproducibility, and linearity, as applicable.
- iii. Detailed documentation and results either from: a clinical study, that includes prospective (sequentially collected) samples for each intended specimen type that are representative of the intended use populations and, when determined to be acceptable by FDA, additional characterized clinical samples; or, when determined to be acceptable by FDA, an equivalent sample set. The clinical study must compare the device performance to results obtained from an FDA accepted reference method and/or FDA accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol (e.g., the predefined statistical analysis plan), clinical study report, testing results, and results of all statistical analyses.
- iv. An evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (e.g., age, racial, ethnic, and gender distribution) similar to the intended use population of the device.
- Documentation of an appropriate end user device training program that will be V. offered as part of efforts to mitigate the risks of false results, failure to correctly operate the device, and failure to correctly interpret test results.
- An appropriate risk mitigation strategy to ensure that the device does not prevent vi. any other device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use (e.g., safety and effectiveness of the functions of the indicated device(s) remain unaffected).
{27}------------------------------------------------
- vii. A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.