Study of Women's Health Across the Nation (SWAN) retrospective serum samples
Retrospective serum samples from the SWAN study were used to evaluate the clinical performance of the picoAMH ELISA in determining menopausal status categories.
Retrospective cohort; Clinical validation; Menopausal status; SWAN study
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Study of Women's Health Across the Nation (SWAN) validation cohort; Retrospective cohort study
690 apparently healthy women ages 42.9 to 62.4; Sample Size: 690
Not applicable for this study
Correlation of AMH levels with menopausal status (time to final menstrual period)
Indications for Use
The picoAMH ELISA is an enzyme-linked immunosorbent assay (ELISA) for the in vitro quantitative measurement of anti-Müllerian hormone (AMH), also known as Müllerian Inhibiting Substance (MIS), concentrations in human serum. It is intended to be used as an aid in the determination of menopausal status in women between 42 and 62 years of age. This assay should only be used in conjunction with other clinical and laboratory findings and results from this test alone should not be used to make diagnostic or treatment decisions. It is intended for in vitro diagnostic use and for prescription use only.
Device Story
picoAMH ELISA is an in vitro diagnostic assay measuring anti-Müllerian hormone (AMH) concentrations in human serum samples. The device utilizes enzyme-linked immunosorbent assay (ELISA) technology to quantify AMH levels. It is intended for use by laboratory professionals to assist clinicians in determining menopausal status in women aged 42 to 62. The output is a quantitative hormone concentration value. Clinicians interpret this result alongside other clinical and laboratory findings to assess menopausal status. The device is not intended for standalone diagnostic use or for making treatment decisions, such as discontinuing contraception or evaluating endometrial cancer risk. By providing an objective measurement of AMH, the device aids in the clinical assessment of the menopausal transition.
Clinical Evidence
The device requires clinical study documentation to demonstrate performance relative to a well-accepted comparator and accuracy across the measuring range for the intended population. Specific clinical performance metrics are not detailed in the provided document.
Indicated for women aged 42-62 to aid in determining menopausal status. Not for fertility assessment, monitoring ovarian response, or predicting fertility. Contraindicated for use as a sole diagnostic tool; results must be assessed with clinical history and examination. Results <10 pg/mL require caution regarding contraceptive discontinuation; results >10 pg/mL require caution regarding potential endometrial cancer.
Regulatory Classification
Identification
A menopause test system is an in vitro diagnostic device intended to measure hormones or other analytes in human clinical specimens as an aid in the determination of menopausal status in women.
Special Controls
A menopause test system must comply with the following special controls: (1) Design verification and validation must include the following: (i) An appropriate traceability plan to minimize the risk of drift in the menopause test system results over time. (ii) Detailed documentation of a clinical study to demonstrate clinical performance or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information: (A) Results must demonstrate appropriate clinical performance relative to a well-accepted and appropriate comparator. (B) Data must demonstrate accuracy of device output for each indicated specimen type throughout the device measuring range as appropriate for the intended use population. (2) The 21 CFR 809.10 labeling must include the following: (i) A statement in the intended use that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions. (ii) A limiting statement that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions. (iii) A limiting statement appropriately describing the risks of false test results and that test results should not be relied upon in clinical decision making (e.g., to discontinue contraceptive use and/or to evaluate patients for the presence of endometrial cancer) without other clinical and laboratory findings.
*Classification.* Class II (special controls). A menopause test system must comply with the following special controls:(1) Design verification and validation must include the following:
(i) An appropriate traceability plan to minimize the risk of drift in the menopause test system results over time.
(ii) Detailed documentation of a clinical study to demonstrate clinical performance or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(A) Results must demonstrate appropriate clinical performance relative to a well-accepted and appropriate comparator.
(B) Data must demonstrate accuracy of device output for each indicated specimen type throughout the device measuring range as appropriate for the intended use population.
(2) The labeling required under § 809.10 of this chapter must include the following:
(i) A statement in the indications for use that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions.
(ii) A limiting statement that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions.
(iii) A limiting statement appropriately describing the risks of false test results and that test results should not be relied upon in clinical decision making (
*e.g.,* to discontinue contraceptive use and/or to evaluate patients for the presence of endometrial cancer) without other clinical and laboratory findings.
Submission Summary (Full Text)
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR picoAMH ELISA
## DECISION SUMMARY
### A. DEN Number:
DEN180004
### B. Purpose for Submission:
New device
### C. Measurand:
Anti-Müllerian Hormone (AMH)
### D. Type of Test:
Quantitative enzyme-linked immunosorbent assay (ELISA)
### E. Applicant:
Ansh Labs
### F. Proprietary and Established Names:
picoAMH ELISA
### G. Regulatory Information:
| Regulation | Name | Product Code | Panel |
|-----------------|-----------------------|--------------|----------------|
| 21 CFR 862.1093 | Menopause test system | QDH | Chemistry (75) |
### H. Indications for Use:
- 1. Indication(s) for use:
The picoAMH ELISA is an enzyme-linked immunosorbent assay (ELISA) for the in vitro quantitative measurement of anti-Müllerian hormone (AMH), also known as Müllerian Inhibiting Substance (MIS), concentrations in human serum. It is intended to be used as an aid in the determination of menopausal status in women between 42 and 62 years of age. This assay should only be used in conjunction with other clinical and laboratory findings and results from this test alone should not be used to make diagnostic or treatment decisions. It is intended for in vitro diagnostic use and for prescription use only.
{1}------------------------------------------------
- 2. Special conditions for use statement(s):
For prescription use only.
picoAMH results should always be assessed in conjunction with the patient's medical history, clinical examination, and other findings when being interpreted for diagnostic purposes.
picoAMH test results < 10 pg/mL should be carefully evaluated in the context of a full clinical work up to ensure that the use of contraceptives is not discontinued in women who have not yet reached menopause.
picoAMH test results > 10 pg/mL should be carefully evaluated in the context of a full clinical work up to ensure that uterine bleeding due to endometrial cancer is not dismissed as a potential diagnosis.
This test should not be used to assess a woman's fertility status or for use in monitoring or predicting the ovarian response in women undergoing or planning to undergo fertility treatments.
# 3. Special instrument requirements:
Microplate reader capable for absorbance measurement at 450 nm, 405 nm, and 630 nm.
# I. Device Description:
The picoAMH ELISA device is supplied as a reagent kit containing the following components in buffer with preservatives:
- AMH/MIS (Müllerian Inhibiting Substance) Coated Microtitration strips: one strip-● holder, containing 12 strips and 96 microtitration wells with mouse monoclonal AMH antibody immobilized to the inside wall of each well.
- AMH/MIS Assay Buffer: one 12 mL bottle containing protein-based buffer with preservative.
- picoAMH Biotin Conjugate Ready-To-Use: one 12 mL bottle containing biotinylated ● mouse anti-AMH antibody in protein-based buffer with preservative.
- picoAMH Streptavidin-Enzyme Conjugate Ready-To-Use: one 12 mL bottle containing streptavidin-HRP (horseradish peroxidase) in a protein-based buffer and preservative.
- . TMB Chromogen Solution: one 12 mL bottle containing a solution of tetramethylbenzidine (TMB) in buffer with hydrogen peroxide.
- Stopping Solution: one 12 mL bottle containing 0.2 M sulfuric acid. ●
- Wash Concentrate A: one 60 mL bottle containing buffered saline with a nonionic ● detergent that requires a 25-fold dilution with deionized water prior to use.
{2}------------------------------------------------
## J. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Methods, 3rd Edition
CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures; A Statistical Approach, 2nd Edition
CLSI EP07-A2: Interference Testing in Clinical Chemistry, 2nd Edition
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, 2nd Edition
CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents, 1st Edition
CLSI CP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, Third Edition
## K. Test Principle:
The picoAMH ELISA is a quantitative three-step sandwich type immunoassay that is designed to measure human AMH. In the first step, calibrators, controls, and unknown samples are added to AMH antibody-coated microtiter wells. After an incubation and washing, biotinylated AMH antibody solution is added. After a second incubation and washing, the wells are incubated with streptavidin horseradish peroxidase conjugate (SHRP) solution. Finally, after a third incubation and washing step, the wells are incubated with substrate solution (TMB) followed by an acidic stopping solution.
The AMH antibody-biotin conjugate binds to the solid phase antibody-antigen complex, which in turn binds to the streptavidin-enzyme conjugate. The antibody-antigen-biotin conjugate-SHRP (streptavidin horseradish peroxidase conjugate solution) complex bound to the well is detected by enzyme-substrate reaction. The degree of enzymatic turnover of the substrate is determined by dual wavelength absorbance measurement at 450 nm as the primary test filter and 630 nm as the reference filter. The absorbance measured is directly proportional to the concentration of AMH in the calibrators, controls, and specimens tested.
### L. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
### a. Precision/Reproducibility:
Precision was evaluated using four human serum specimens containing AMH concentrations ranging from about 15 pg/mL to about 930 pg/mL. Samples were tested at one site in replicates of four, in two runs per day, for 20 days with three reagent lots (n = 160 per lot).
{3}------------------------------------------------
| Sample | Lot | N | Mean<br>pg/mL | Repeatability<br>SD (pg/mL) | Repeatability<br>% CV | Intermediate<br>Precision<br>SD (pg/mL) | Intermediate<br>Precision<br>%CV |
|---------|-----|-----|---------------|-----------------------------|-----------------------|-----------------------------------------|----------------------------------|
| Serum 1 | 1 | 160 | 14.6 | 0.8 | 5.5% | 1.2 | 8.1% |
| | 2 | 160 | 14.2 | 0.6 | 4.2% | 0.8 | 5.9% |
| | 3 | 160 | 15.5 | 0.7 | 4.5% | 1.0 | 6.7% |
| Serum 2 | 1 | 160 | 80.1 | 2.2 | 2.8% | 3.6 | 4.5% |
| | 2 | 160 | 80.0 | 2.0 | 2.5% | 3.4 | 4.2% |
| | 3 | 160 | 80.8 | 2.8 | 3.5% | 3.4 | 4.2% |
| Serum 3 | 1 | 160 | 620.3 | 17.4 | 2.8% | 22.9 | 3.7% |
| | 2 | 160 | 609.6 | 16.8 | 2.8% | 24.2 | 4.0% |
| | 3 | 160 | 643.2 | 19.2 | 3.0% | 24.0 | 3.7% |
| Serum 5 | 1 | 160 | 942.8 | 28.9 | 3.1% | 51.3 | 5.4% |
| | 2 | 160 | 924.1 | 23.7 | 2.6% | 48.5 | 5.3% |
| | 3 | 160 | 935.2 | 36.8 | 3.9% | 47.4 | 5.1% |
Results for within-lab imprecision are as follows:
# b. Linearity/assay reportable range:
## Linearity
Linearity was evaluated based on CLSI EP06-A. A high analyte (1200 pg/mL) concentration specimen ("High Pool") was prepared by adding recombinant human AMH to low-analyte pooled human serum. 14 intermediate levels were prepared by mixing low-analyte serum with the High Pool for a total of 16 level (b) (4) Each level was measured in triplicate with three lots and the allowable nonlinearity was calculated for each lot. Representative data for one reagent lot is shown below:
# y =(b) (4)
The data supports the sponsor's claimed measuring range of 6.0 to 1150 pg/mL.
# Dilution
A dilution study was performed using nine human serum samples spiked with recombinant AMH (ranging from (b) (4) pg/mL). Samples were manually diluted 20-fold with picoAMH Calibrator A/Sample diluent and tested in triplicate using four reagent lots. The data support the sponsor's instructions for use stating that samples containing AMH concentrations above the claimed measuring range (6 to 1150 pg/mL) and up to 23,000 pg/mL can be measured with the picoAMH ELISA assay when diluted up to 20-fold.
- c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The picoAMH ELISA assay is traceable to an internal recombinant AMH standard
{4}------------------------------------------------
manufactured at Ansh Labs. The applicant submitted a detailed traceability assurance plan which was reviewed and found to be acceptable.
The picoAMH Calibrators consist of a picoAMH Calibrator A/Sample Diluent containing AMH in equine serum with preservative. The target values for the calibrators are shown below:
| Calibrator | AMH Concentration<br>(pg/mL) |
|------------|------------------------------|
| CAL A | 0 |
| CAL B | 8 |
| CAL C | 31 |
| CAL D | 105 |
| CAL E | 360 |
| CAL F | 1091 |
The picoAMH Controls consist of two vials, labeled Levels I and II, containing low and high AMH concentrations in serum with preservative. Controls are reconstituted with 1 mL of deionized water.
- d. Detection limit:
Detection limits were evaluated based on CLSI EP17-A2.
The Limit of Blank (LoB) was evaluated using 18 independent human serum pools containing no detectable AMH. Samples were tested with four reagent lots for a total of 73 replicates per specimen. The overall LoB (n = 324) was calculated using a nonparametric analysis.
The Limit of Detection (LoD) and Limit of Quantitation (LoQ) were evaluated using 12 low-level serum pools (containing between 2.76 and 9.95 pg/mL AMH) measured over four reagent lots (n = 144 per lot). The LoD was calculated as the Mean LoB + 1.645*SDL (SDL = standard deviation ofpooled low level specimens). The sponsor defined the LoQ as the lowest amount of analyte in a sample that can be quantified reliably with an intra-assay coefficient of variance (CV) of 20% based on precision profiling.
The data supports the following claims:
LoB = 0.5 pg/mL
LoD = 1.3 pg/mL
LoQ = 3.2 pg/mL
{5}------------------------------------------------
# e. Analytical specificity:
# Endogenous and exogenous interference
Endogenous and exogenous interference was evaluated according to CLSI EP07-A2 using low (27 pg/mL) and high (300 pg/mL) serum samples spiked with possible interferents. High samples were tested in replicates of six and low samples were tested in replicates of four. Test samples were compared to control samples without interferent. The sponsor considered a percent difference between test samples and control samples of >10% to be significant interference.
The highest concentrations of endogenous and exogenous substances tested that show non-significant interference are summarized in the table below:
| Substance | Highest Concentration Tested<br>Without Significant Interference |
|-------------------------------------------|------------------------------------------------------------------|
| Acetaminophen | 0.2 mg/mL |
| Acetylcysteine | 0.15 mg/mL |
| Acetylsalicylic acid | 1 mg/mL |
| Ampicillin-Na | 1 mg/mL |
| Ascorbic acid | 0.3 mg/mL |
| Bazedoxifene | 1 µg/mL |
| Bilirubin | 0.66 mg/mL |
| Biotin* | 10,000 ng/mL |
| Bisphosphonate<br>(Alendronate (Fosamax)) | 0.02 mg/mL |
| Cefoxitin Na | 2.5 mg/mL |
| Cholesterol | 5 mg/mL |
| Citalopram | 0.01 mg/mL |
| Cyclosporine | 5 µg/mL |
| Cyproterone acetate | 0.3 mg/mL |
| Doxycycline Hyclate | 50 µg/mL |
| Escitalopram Oxalate | 0.1 mg/mL |
| Estradiol (beta) | 1 ng/mL |
| Estrone sulfate | 1 ng/mL |
| Estropipate | 0.015 mg/mL |
| Fluoxetine HCl | 3.5 µg/mL |
| Folic Acid | 0.4 µg/mL |
| Gabapentin | 0.09 mg/mL |
| Hemoglobin (Human) | 10 mg/mL |
| Heparin | 30 U/mg |
| Ibuprofen | 0.5 mg/mL |
| Levodopa | 30 µg/mL |
| Levonorgestrel | 3 µg/mL |
| Levothyroxine | 0.2 µg/mL |
{6}------------------------------------------------
| Substance | Highest Concentration Tested<br>Without Significant Interference |
|-----------------------------|------------------------------------------------------------------|
| Medroxyprogesterone acetate | 1 µg/mL |
| Metformin | 2 mg/mL |
| Methyldopa | 0.02 mg/mL |
| Metronidazole | 0.2 mg/mL |
| Norethindrone | 0.03 mg/mL |
| Paroxetine HCl | 1 µg/mL |
| Phenylbutazone | 0.1 mg/mL |
| Pregabalin | 0.01 mg/mL |
| Progesterone | 0.4 mg/mL |
| Raloxifene HCl | 0.12 mg/mL |
| Rifampicin | 60 µg/mL |
| Theophylline | 0.1 mg/mL |
| Intralipid 20% | 20 mg/mL |
| Triptorelin acetate | 15 µg/mL |
| Venlafaxine HCl | 15 µg/mL |
* Though this assay uses biotin-streptavidin binding technology, no interference is observed in specimens containing up to 10,000 ng/mL biotin.
### Cross-reactivity
The cross reactivity of the picoAMH ELISA assay was evaluated using pooled serum samples containing undetectable concentrations of AMH spiked with potential crossreacting compounds. Pro+Mature recombinant human AMH was included as a positive control. All samples were tested in duplicate with a single reagent lot. The AMH measured as pg/mL after addition was used to calculate a percent cross reactivity based on the mass of potential cross reactant tested as follows:
% cross reactivity = (2.5 pg/mL divided by pg/mL cross-reactant tested) x 100
For cross-reactants yielding detectable AMH concentrations, the formula for calculating percent cross reactivity was:
% cross-reactivity = ((measured value - true value)/concentration of cross-reactant tested)*100)
| Cross-Reactant | Concentration Tested<br>(pg/mL) | AMH Value<br>(pg/mL) | % Cross-<br>reactivity |
|----------------|---------------------------------|----------------------|------------------------|
| Activin B | 50,000 | < 1.3 | 0.003% |
| Inhibin A | 100,000 | < 1.3 | 0.001% |
| Inhibin B | 100,000 | < 1.3 | 0.001% |
None of the potential cross-reactants showed > 0.5% cross-reactivity.
{7}------------------------------------------------
| Cross-Reactant | Concentration Tested<br>(pg/mL) | AMH Value<br>(pg/mL) | % Cross-<br>reactivity |
|------------------|---------------------------------|----------------------|------------------------|
| alpha-2 | 65.000 | < 1.3 | 0.002% |
| Follistatin-288 | 50,000 | < 1.3 | 0.003% |
| Follistatin-315 | 50,000 | < 1.3 | 0.003% |
| hAMH, Pro+Mature | 600 | 627.8 | 104.641% |
| hAMH, Mature | 600 | < 1.3 | 0.217% |
| Myostatin | 50,000 | < 1.3 | 0.003% |
| FSH | 39.683 | < 1.3 | 0.003% |
| TSH | 869,565 | < 1.3 | 0.000% |
| LH | 9,312 | < 1.3 | 0.014% |
| Prolactin | 211,000 | < 1.3 | 0.001% |
| Testosterone | 100,000 | < 1.3 | 0.001% |
| Estrone Sulphate | 100.000 | < 1.3 | 0.001% |
| DHEA | 100,000 | < 1.3 | 0.001% |
| Progesterone | 100,000 | < 1.3 | 0.001% |
| Estradiol | 50,000 | < 1.3 | 0.003% |
## Hook effect
In order to assess hook effect, recombinant human AMH was added to pooled human serum AMH to generate a specimen containing known AMH levels at least (b) (4) the upper limit of the measuring range. The specimen was tested neat and at a series of dilutions in each of three reagent lots. The sponsor concludes that there is no hook effect on the picoAMH ELISA by AMH concentrations in human serum up to 256,000 pg/mL.
- f. Assay cut-off:
Please see section 4 (clinical cut-off) below.
- 2. Comparison studies:
- a. Method comparison with predicate device:
Not applicable.
- b. Matrix comparison:
Not applicable. Serum is the only claimed sample matrix.
{8}------------------------------------------------
### 3. Clinical studies:
A clinical study was performed to evaluate the clinical performance of the device. 690 retrospective serum samples from 690 apparently healthy women ages 42.9 to 62.4 (no history of ovarian surgery, no diagnosis of PCOS, no oral contraceptive intake) participating in the Study of Women's Health Across the Nation (SWAN) study were randomly chosen and assigned to the clinical validation cohort. Each woman contributed one sample and one outcome to the study. The sponsor provided data supporting the stability of the frozen serum specimens used in the clinical validation study for the length and conditions of storage of SWAN study specimens. Menopausal status for each woman was determined based on time to final menstrual period (FMP). The sponsor defined three menopausal status categories: "> 5 years from FMP", "< 5 years from FMP", and "at FMP or later". The samples were evenly distributed across all three menopausal categories (n = 230) based on outcome and each category contained women with a range of ages.
Image /page/8/Figure/2 description: This image shows a plot with the y-axis labeled "Ansh picoAMH (pg/mL)". The x-axis is labeled with categories "> 5 years", "< 5 years", and "0 or later". The plot is labeled with "(b) (4)" in the top left corner. The plot area is filled with a solid gray color.
Time to Final Menstrual Period (FMP)
| | Menopausal category | | | |
|-----------|---------------------|--------------------|-----------------|-------|
| Age | > 5 years from FMP | < 5 years from FMP | At FMP or later | Total |
| 42.9 - 45 | 50 | 12 | 2 | 64 |
| 45 - 49.9 | 164 | 108 | 47 | 319 |
| 50 - 54.9 | 16 | 103 | 138 | 257 |
| 55 - 62.4 | 0 | 7 | 43 | 50 |
| Total | 230 | 230 | 230 | 690 |
Age distribution in each menopausal category:
{9}------------------------------------------------
The following table describes the distribution of picoAMH ELISA test results across the three menopausal categories. Percentages indicate the agreement of the picoAMH ELISA test result with the corresponding menopausal category.
| picoAMH | Menopausal category | | | |
|-----------------|---------------------|--------------------|--------------------|-----------------|
| | | > 5 years from FMP | < 5 years from FMP | At FMP or later |
| < 10 pg/mL | 19 (8.2%) | 89 (38.7%) | 198 (86.1%) | 306 |
| 10 - 99.9 pg/mL | 22 (9.6%) | 80 (34.8%) | 28 (12.2%) | 130 |
| ≥ 100 pg/mL | 189 (82.2%) | 61 (26.5%) | 4 (1.7%) | 254 |
| Total | 230 (100%) | 230 (100%) | 230 (100%) | 690 |
Clinical performance was estimated for each menopausal category. The results are summarized below:
| picoAMH | Menopausal Category | Detection Rate (%)<br>(95% CI) | picoAMH | Classified<br>Menopausal<br>Category | True Menopausal<br>Status | False Positive Rate<br>(%)<br>(95% CI) |
|---------------|---------------------|--------------------------------|---------------------|--------------------------------------|---------------------------|----------------------------------------|
| ≥100 pg/mL | >5 years from FMP | 82.2<br>(76.6–86.9) | ≥100 pg/mL | >5 years from FMP | < 5 years from FMP | 26.5<br>(20.9-32.7) |
| <10 pg/mL | at FMP or later | 86.1<br>(80.4–89.9) | ≥100 pg/mL | >5 years from FMP | at FMP or later | 1.7<br>(0.4-4.4) |
| 10-99.9 pg/mL | < 5 years from FMP | 34.8<br>(28.6–41.3) | <10 pg/mL | at FMP or later | >5 years from FMP | 8.2<br>(5.0-12.6) |
| <10 pg/mL | at FMP or later | < 5 years from FMP | 38.7<br>(32.4-45.3) | | | |
| 10-99.9 pg/mL | < 5 years from FMP | >5 years from FMP | 9.6<br>(6.1-14.1) | | | |
| 10-99.9 pg/mL | < 5 years from FMP | at FMP or later | 12.2<br>(8.2, 17.1) | | | |
The labeling contains the following statement:
"This analysis shows that the picoAMH ELISA performs reasonably well in distinguishing women at FMP or later and women > 5 years from FMP. Specifically, 86.1% of women who were at FMP or later also had a picoAMH level <10 pg/mL and 82.2% of women who were > 5 years from FMP also had a picoAMH level > 100 pg/mL.
{10}------------------------------------------------
However, the picoAMH ELISA test result has limited deterministic value when AMH values fall between 10 and 99.9 pg/mL, and test results falling into this range should be interpreted with caution."
To aid in the interpretation of picoAMH ELISA test results obtained for women in the menopausal transition (i.e., < 5 years away from their FMP), the following table provides the likelihood ratios for the comparison of women < 5 years away from their FMP vs. adjacent menopausal categories given the AMH concentrations obtained in the clinical studv.
| Menopausal Category<br>Comparison | Positive<br>Likelihood Ratio<br>(95% CI) | Negative<br>Likelihood Ratio<br>(95% CI) |
|----------------------------------------------|------------------------------------------|------------------------------------------|
| < 5 years from FMP vs.<br>at FMP or later | 2.86<br>(1.94, 4.22) | 0.74<br>(0.67, 0.83) |
| < 5 years from FMP vs.<br>> 5 years from FMP | 3.64<br>(2.35, 5.62) | 0.72<br>(0.65, 0.80) |
The results above show that, comparing to any woman with an unknown AMH level who has reached her FMP or is < 5 years from her FMP, a woman with an AMH test result between 10 and 99.9 pg/mL is 2.86 times more likely to be < 5 years away from her FMP than to be at FMP or later. Likewise, comparing to any woman with an unknown AMH level who is < 5 years away from her FMP or > 5 years away from her FMP, a woman with an AMH test result between 10 and 99.9 pg/mL is 3.64 times more likely to be < 5 years away from her FMP than to be > 5 years away from her FMP.
- a. Clinical Sensitivity:
Not applicable.
- b. Clinical specificity:
Not applicable.
- c. Other clinical supportive data (when a. and b are not applicable):
Not applicable.
- 4. Clinical cut-off:
The sponsor specified two clinical cut-offs for the pico AMH ELISA assay: < 10 pg/mL to distinguish women at or later than their final menstrual period (FMP) and > 100 pg/mL to distinguish women > 5 years from FMP.
{11}------------------------------------------------
# 5. Expected values/Reference range:
A reference range was determined by analyzing 644 serum samples from 644 apparently healthy women enrolled in the SWAN study. The picoAMH test results for these samples was then stratified into four different age groups and percentiles calculated. The results are as follows:
| Age (years) | N | picoAMH ELISA test result centile | | | | | CI for 95th centile |
|--------------|-----|-----------------------------------|------|------|-------|-------|---------------------|
| | | 5th | 25th | 50th | 75th | 95th | |
| 42.9 to 44.9 | 159 | <6 | 100 | 390 | 1,200 | 2,900 | 2,200-3,500 |
| 45.0 to 49.9 | 175 | <6 | 3.1 | 75 | 280 | 640 | 700 - 1,500 |
| 50.0 to 54.9 | 175 | <6 | <6 | <6 | 16 | 98 | 160-350 |
| 55.0 to 62.4 | 135 | <6 | <6 | <6 | <6 | 28 | 6-110 |
# M. Proposed Labeling:
The labeling supports the decision to grant the De Novo request for this device.
## N. Identified Risks and Identified Mitigations:
| Identified Risks to Health | Identified Mitigations |
|------------------------------------------|---------------------------------------------------|
| Incorrect test results | General controls and special controls (1) and (2) |
| Incorrect interpretation of test results | General controls and special controls (1) and (2) |
# O. Benefit/Risk Analysis:
## Summary of the Assessment of Benefit For the Proposed Indications for Use
Natural menopause is determined after a women has experienced 12 months of amenorthea without any other obvious pathological or physiological cause. Knowing in what timeframe a woman may become menopausal may prompt preventive approaches for loss in bone mineral density and cardiovascular disease, which are known to increase after menopause.
## Summary of the Assessment of Risk For the Proposed Indications for Use
The main risks associated with the test are a falsely high AMH level. A falsely low value may incorrectly indicate that a woman was post-menopausal or closer to her final menstrual period (FMP). In this case, a clinician may have a lower threshold to discontinue contraception, and thus may increase the risk of pregnancy. However, this risk is only as high
{12}------------------------------------------------
as the differential likelihood of pregnancy, considering age and other factors contributing to pregnancy risk for the individual. A falsely low value may also increase the risk of prescribing hormone replacement therapy in a woman who is not yet menopausal. However, hormone replacement therapy may be prescribed based on menopausal symptoms which may occur during the menopausal transition and would be prompted based on the severity of the symptoms, more so than post-menopausal status. According to the American College of Obstetricians and Gynecologists Practice Bulletin on the management of Menopausal Symptoms, "the decision to continue [systemic hormone therapy] should be individualized and be based on a woman's symptoms and the risk-benefit ratio, regardless of age."
A falsely high AMH value may incorrectly indicate that a woman was not post-menopausal or farther from her final menstrual period. In a woman with irregular bleeding which may be confused with menopause transition bleeding, a falsely high AMH value may delay the evaluation of endometrial cancer. However, the character of bleeding (e.g., increased bleeding or more frequent bleeding) is the main determination for suspecting endometrial cancer. If the bleeding described by a patient is concerning for features of endometrial cancer, the AMH test result would not likely impact the standard of care for the work up of endometrial cancer. Additionally, continued symptoms of irregular bleeding would prompt a clinician to perform endometrial cancer evaluation without further delay.
There is a risk for incorrect interpretation of test results. The clinical study described above was designed to detect menopause rather than predict menopause since the study analysis categorized women by the outcome (FMP) first, then correlated an AMH value, instead of predicting a clinical outcome based on a measured AMH value. Therefore, the use of the test is limited to aiding in detecting a women's menopausal category (more than 5 years from her expected FMP, within 5 years from FMP, or at FMP or later) and cannot be used to predict an individual woman's FMP or fertile lifespan.
# Summary of the Assessment of Benefit-Risk For the Proposed Indications for Use
Given that there are possible risks associated with an incorrect test result and the incorrect interpretation of test results, the benefit-risk balance of this device is undetermined and requires additional mitigations in the form of limitations and special controls, beyond general controls.
## Summary of the Assessment of Benefit-Risk, considering risk mitigation strategies For the Proposed Indications for Use
Overall, the likelihood of benefit of the picoAMH test to detect postmenopausal status and prompt preventive strategies for conditions associated with menopause outweighs the likelihood of pregnancy due to a falselv low AMH level or the likelihood of delaying an opportunity for endometrial cancer evaluation due to a falsely high AMH level, when considering the mitigations provided by the limitations and special controls, beyond general controls.
{13}------------------------------------------------
# P. Conclusion:
The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 862.1093. FDA believes that the special controls, in combination with the general controls provide a reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:
Product Code: QDH Device Type: Menopause test system Class: II (special controls) Regulation: 21 CFR 862.1093
- a) Identification: A menopause test system is an in vitro diagnostic device intended to measure hormones or other analytes in human clinical specimens as an aid in the determination of menopausal status in women.
- b) Classification: Class II (special controls). A menopause test system must comply with the following special controls:
- (1) Design verification and validation must include the following:
- (i) An appropriate traceability plan to minimize the risk of drift in the menopause test system results over time.
- (ii) Detailed documentation of a clinical study to demonstrate clinical performance or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
- (A) Results must demonstrate appropriate clinical performance relative to a well-accepted and appropriate comparator.
- (B) Data must demonstrate accuracy of device output for each indicated specimen type throughout the device measuring range as appropriate for the intended use population.
- (2) The 21 CFR 809.10 labeling must include the following:
- (i) A statement in the intended use that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions.
- (ii) A limiting statement that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions.
- (iii) A limiting statement appropriately describing the risks of false test results and that test results should not be relied upon in clinical decision making
{14}------------------------------------------------
(e.g., to discontinue contraceptive use and/or to evaluate patients for the presence of endometrial cancer) without other clinical and laboratory findings.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.