DEN170029 · Longhorn Vaccines and Diagnostics, LLC · QBD · Mar 19, 2018 · Microbiology
Device Facts
Record ID
DEN170029
Device Name
PrimeStore MTM
Applicant
Longhorn Vaccines and Diagnostics, LLC
Product Code
QBD · Microbiology
Decision Date
Mar 19, 2018
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 866.2950
Device Class
Class 2
Indications for Use
PrimeStore MTM is intended for the stabilization, transportation and inactivation of infectious unprocessed nasal washes suspected of containing Influenza A virus RNA. PrimeStore MTM is also intended for the stabilization, transportation and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis DNA from human samples.
Device Story
PrimeStore MTM is a transport device consisting of a storage tube with O-ring/lip seal containing 1.5 mL stabilization solution. Solution contains guanidine thiocyanate, TCEP, N-Lauroylsarcosine sodium, and other reagents. Device inactivates Influenza A and M. tuberculosis, lyses cells, disrupts lipid membranes, denatures proteins, and stabilizes nucleic acids. Used in clinical settings for specimen collection; operated by healthcare professionals. Specimens are stored at 2-25 °C. Inactivation protects laboratory personnel from occupational exposure during transport and processing. Stabilized nucleic acids are subsequently extracted and analyzed via molecular assays (e.g., real-time PCR). Device facilitates safe transport and preserves sample integrity for downstream diagnostic testing.
Clinical Evidence
Bench testing only. No clinical data. Studies evaluated nucleic acid stability and pathogen inactivation. Stability: MTB DNA (1x10^3 CFU/mL) stable in sputum for 36 days at 4°C and 27°C; Influenza A RNA (1x10^2 TCID50/mL) stable in nasal wash for 29 days at 4°C and 8 days at 27°C. Inactivation: MTB (1.5x10^6 CFU/mL) inactivated in sputum at 1:3 ratio after 60 minutes; Influenza A inactivated at 1:3 ratio after 10 seconds.
Technological Characteristics
Storage tube with O-ring/lip seal. Chemical stabilization solution: guanidine thiocyanate, TCEP, sodium citrate, N-Lauroylsarcosine sodium, Antifoam A, TRIS, EDTA, ethanol, HCl, nuclease-free water. Non-electronic, standalone device. No software or energy source.
Indications for Use
Indicated for stabilization, transport, and inactivation of infectious unprocessed nasal washes containing Influenza A virus RNA and sputum samples containing Mycobacterium tuberculosis DNA from human patients.
Regulatory Classification
Identification
A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device is not intended for preserving morphology or viability of microorganisms.
Special Controls
The special controls for this device are: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including all device components. (ii) Performance characteristics from applicable analytical studies, including but not limited to, nucleic acid stability and microorganism inactivation. (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing. (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design including all device components and all control elements incorporated into the analytical validation procedures. (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, but not limited to, extraction platforms and assays used for the detection of preserved nucleic acids. (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of microorganisms and types of human specimens intended to be preserved.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A detailed device description, including all device components;
(ii) Performance characteristics from applicable analytical studies, including nucleic acid stability and microorganism inactivation;
(iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing; and
(iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies.
(3) Design verification and validation must include the following:
(i) Overall device design, including all device components and all control elements incorporated into the analytical validation procedures;
(ii) Thorough description of the microorganisms and methodology used in the validation of the device including, extraction platforms and assays used for the detection of preserved nucleic acids; and
(iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.
Submission Summary (Full Text)
{0}------------------------------------------------
### EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR PrimeStore MTM DECISION SUMMARY
### A. DEN Number:
DEN170029
### B. Purpose for Submission:
De Novo request for evaluation of automatic class III designation for the PrimeStore MTM
### C. Measurands:
Storage and stability of nucleic acids from Mycobacterium Tuberculosis and Influenza A virus.
### D. Type of Device:
Transport device for the stabilization of microbial nucleic acids
### E. Applicant:
Longhorn Vaccines and Diagnostics, LLC
### F. Proprietary and Established Names:
PrimeStore MTM
### G. Regulatory Information:
- 1. Regulation section:
21 CFR 866.2950
- 2. Classification:
Class II (Special Controls)
- 3. Product code(s):
QBD
- 4. Panel:
83- Microbiology
{1}------------------------------------------------
### H. Indications for Use:
### 1. Indications for use:
PrimeStore MTM is intended for the stabilization, transportation and inactivation of infectious unprocessed nasal washes suspected of containing Influenza A virus RNA. PrimeStore MTM is also intended for the stabilization, transportation and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis DNA from human samples.
### 2. Special conditions for use statement(s):
For in vitro diagnostic use only For prescription use only
- 3. Special instrument requirements:
None
### I. Device Description:
The PrimeStore MTM device consists of a storage tube with an O-ring and lip seal containing 1.5 mL of the stabilization solution. These components are intended to inactivate Influenza A and Mycoplasma tuberculosis, lyse cells, disrupt/lyse lipid membranes, denatures proteins, inactivates enzymes, and stabilize Influenza A RNA and M. tuberculosis DNA. The transport device is designed for storage of specimens between 36-77 °F (2-25 °C).
The media contains the following reagents:
- o Guanidine thiocyanate
- o TCEP
- Sodium Citrate ●
- N-Lauroylsarcosine sodium (NLS) ●
- Antifoam A, TRIS ●
- EDTA ●
- Ethanol (molecular grade)
- HCl ●
- Nuclease-free water ●
### J. Standard/Guidance Document Referenced (if applicable):
Not applicable
### K. Test Principle:
Not applicable
{2}------------------------------------------------
### L. Performance Characteristics:
#### 1. Analytical performance:
#### a. Limit of detection
LoD testing was conducted to determine the lowest concentration of organisms that contains measurable nucleic acids that can be repeatedly recovered from the transport media with a greater than 95% accuracy. The LoD studies for M. tuberculosis and Influenza A were designed using a specific extraction platforms and amplification instrument to establish a concentration of organisms which will form the basis of additional testing. Testing near the LoD for the additional studies will challenge PrimeStore MTM ability to preserver nucleic acids from degradation under a variety of test conditions.
#### Mycobacterium Tuberculosis Limit of Detection:
LoD testing was initially performed by spiking seven concentrations of MTB diluted 1:10 into pooled sputum which was known to be negative for MTB. The spiked sputum was then added at a concentration of 1:3 into the PrimeStore MTM for recovery using the PrimeXtract extraction kit and amplification using the ABI 7500 realtime PCR instrument. The study was conducted in quadruplicate to determine a recoverable concentration of MTB after being spiked into sputum and subsequently added to PrimeStore MTM. No detection (Ct = 40) was observed for one of the replicates at the 10% CFU/mL concentration. All other concentrations demonstrated recovery of MTB DNA from all the replicates. Table 1 shows the results of MTB organisms diluted 1:10 at multiple concentrations, extraction of DNA and amplification at each concentration. Additional LoD testing was provided at a concentration of 101 CFU/mL with 25 replicates. An acceptance criteria with a range of 3 C; was used to determine the concentration that yielded at least a 95% of the replicates were recoverable within this range. At a concentration of 101 CFU/mL 25 of 25 replicates had recoverable concentrations and all replicates fell within the range of 3 Ct values, Table 2. The MTB DNA extracted from PrimeStore MTM had an average CT = 34.0; S.D. = 0.98 at a concentration of 101 CFU/mL.
The M. tuberculosis strain used for LoD studies was strain HN878 (BEI Resources, Catalog No. NR-13647)
{3}------------------------------------------------
| MTB<br>Concentration<br>(CFU/mL) | Rep 1<br>(Ct) | Rep 2<br>(Ct) | Rep 3<br>(Ct) | Rep 4<br>(Ct) | Average<br>(Ct) | SD<br>(Ct) |
|----------------------------------|---------------|---------------|---------------|---------------|-----------------|------------|
| 106 | 17.8 | 17.7 | 17.6 | 17.6 | 17.7 | 0.1 |
| 105 | 22.1 | 22.6 | 21.9 | 22.1 | 22.2 | 0.4 |
| 104 | 23.5 | 23.4 | 23.2 | 23.5 | 23.4 | 0.2 |
| 103 | 26.3 | 26.7 | 26.6 | 26.8 | 26.6 | 0.2 |
| 102 | 29.4 | 29.3 | 29.6 | 28.7 | 29.3 | 0.2 |
| 101 | 32.8 | 32.6 | 33.4 | 32.2 | 32.8 | 0.4 |
| 10 | 34.4 | 40.0 | 34.6 | 35.6 | 36.2 | 3.2 |
Table 1. MTB preliminary Limit of Detection
Table 2. MTB LoD
| 10¹ CFU/mL | |
|------------|----------|
| Replicates | Ct Value |
| 1 | 33.3 |
| 2 | 33.2 |
| 3 | 33.0 |
| 4 | 32.7 |
| 5 | 33.0 |
| 6 | 34.1 |
| 7 | 33.6 |
| 8 | 33.4 |
| 9 | 35.4 |
| 10 | 35.1 |
| 11 | 35.3 |
| 12 | 35.4 |
| 13 | 36.0 |
| 14 | 35.1 |
| 15 | 34.1 |
| 16 | 33.4 |
| 17 | 33.2 |
| 18 | 32.3 |
| 19 | 34.3 |
| 20 | 34.3 |
| 21 | 33.3 |
| 22 | 33.1 |
| 23 | 33.9 |
| 24 | 34.0 |
| 25 | 34.6 |
| AVG: | 34.0 |
| SD: | 1.0 |
LoD testing at both 104 CFU/mL and 102 CFU/mL resulted in all 25 replicates for each concentration to meet the pre-defined acceptance criteria.
{4}------------------------------------------------
#### Influenza A Limit of Detection:
LoD testing was initially performed by spiking multiple concentrations of influenza A diluted 1:10 into pooled, clinically negative, nasal washes. The final concentration of each spiked nasal was then added to achieve a final ratio of 1:3 specimen to PrimeStore MTM for recovery using the PrimeXtract extraction kit and amplification in conjunction with the ABI 7500 realtime PCR instrument. The study objective was to determine the detectable concentration of influenza A (A/Texas/78209/2008 (H3N2)) after spiking into nasal wash and added to PrimeStore MTM. No detection (C: = 40) was observed at the 101 TCID50/mL concentration. All other concentrations demonstrated recovery of Influenza A. Table 3 shows the results of the 1:10 dilutions of Influenza A. Additional LoD testing was provided at a concentration of 102 TCID50/mL. 102 TCID50/mL was replicated 25 times. An acceptance criteria with a range of 3 Ct was used to determine the concentration that yielded at least a 95% of the replicates were recoverable within this range. At a concentration of 102 TCID50/mL 25 of 25 replicates had recoverable concentrations; however, one replicate had a high C value and fell outside the range of 3 Ct values, Table 4. The Influenza A RNA extracted from PrimeStore MTM had the following performance average CT = 34.5; S.D. = 0.88 at a concentration of 102 TCIDso/mL.
| FLUA<br>Concentration<br>(TCID50/mL) | Rep 1<br>(Ct) | Rep 2<br>(Ct) | Rep 3<br>(Ct) | Average<br>(Ct) | SD<br>(Ct) |
|--------------------------------------|---------------|---------------|---------------|-----------------|------------|
| 105 | 25.5 | 25 | 25.1 | 25.2 | 0.26 |
| 104 | 28.1 | 28 | 27.9 | 28 | 0.10 |
| 103 | 31.4 | 31.3 | 32.1 | 31.6 | 0.44 |
| 102 | 35.3 | 35.5 | 34.9 | 35.2 | 0.31 |
| 101 | 40 | 40 | 40 | 40 | 0 |
Table 3. Influenza A Preliminary Limit of Detection
{5}------------------------------------------------
| 102 TCID50/mL | |
|---------------|----------|
| Replicates | CT Value |
| 1 | 35.8 |
| 2 | 33.7 |
| 3 | 34.6 |
| 4 | 33.3 |
| 5 | 34.2 |
| 6 | 33.9 |
| 7 | 34.1 |
| 8 | 36.3 |
| 9 | 34.9 |
| 10 | 33.9 |
| 11 | 34.6 |
| 12 | 34 |
| 13 | 34.2 |
| 14 | 33.8 |
| 15 | 34.7 |
| 16 | 34.4 |
| 17 | 33.6 |
| 18 | 34.4 |
| 19 | 34.5 |
| 20 | 34.4 |
| 21 | 34.9 |
| 22 | 36.1 |
| 23 | 35.5 |
| 24 | 32.6 |
| 25 | 35.6 |
| AVG: | 34.5 |
| SD: | 0.88 |
Table 4. Influenza A LoD
LoD testing at 102 TCID50 resulted in all 25 replicates for the concentration to meet the pre-defined acceptance criteria.
### b. Stability
#### Mycobacterium Tuberculosis stability:
Stability studies evaluated the stability of DNA from whole organism Mycobacterium tuberculosis (MTB; 1x103 CFU/mL) spiked into clinically negative sputum samples incubated in PrimeStore MTM at: refrigerated temperature (4°C, 39.2°F) for 36 days, Table 5, and ambient temperature (27°C, 80.6°F) for 36 days, Table 6. DNA from MTB (Strain HN878) in PrimeStore MTM was extracted using PrimeXtract spin columns. Nucleic acid stability was measured and analyzed by real-time PCR using the ABI 7500 (Thermo Fisher Scientific) instrument. The study analyzed 25 replicates extracted at each time point and at each temperature range. An initial time point designated as Day 0 for each temperature was included as the initial Cr average from
{6}------------------------------------------------
which all other time points were compared to. Time points for sample extractions were performed at Day 0, 1, 8, 15, 22, 29, 36, with an embedded Internal Negative Control (INC) included at both temperatures. The INC consisted of a test that used negative clinical samples in PrimeStore MTM, i.e., sputum without MTB. Additionally, a control that contained clinical specimen plus target, i.e., sputum plus MTB, was incubated in PBS, without PrimeStore MTM ((POS-(PS)), and included in the evaluation. A pre-defined acceptance criteria of (+/-) 3.0 C; from time zero was used to establish stability and preservation of nucleic acids (DNA from MTB) as determined by real-time PCR, for 4 and 27°C without loss of detection signal using statistical analysis.
| Day (4°C) | 0 | 1 | 8 | 15 | 22 | 29 | 36 |
|-----------------|------|------|------|------|------|------|------|
| AVG (Ct): | 29.8 | 29.5 | 29.4 | 29.3 | 30.8 | 29.5 | 29.3 |
| SD (Ct): | 0.8 | 0.4 | 0.4 | 0.3 | 0.3 | 0.5 | 0.7 |
| INC: | NEG | NEG | NEG | NEG | NEG | NEG | NEG |
| POS (-PS) (Ct): | 28.9 | 29.7 | 29.8 | 29.9 | 32.4 | 32.2 | 30.8 |
Table 5 MTB (1 x 103 CFU/mI ) stability at 4°C
Table 6 MTR (1 x 103 CFU/mI ) stability at 27°C
| Day (27°C) | 0 | 1 | 8 | 15 | 22 | 29 | 36 |
|-----------------|------|------|------|------|------|------|------|
| AVG (Ct): | 29.4 | 29.4 | 29.4 | 29.4 | 31.0 | 29.8 | 29.4 |
| SD (Ct): | 0.2 | 0.2 | 0.3 | 0.3 | 0.3 | 0.4 | 0.3 |
| INC: | NEG | NEG | NEG | NEG | NEG | NEG | NEG |
| POS (-PS) (Ct): | 29.3 | 34.6 | 34.9 | 34.7 | 34.3 | 34.8 | 34.7 |
Stability testing of DNA from M. tuberculosis whole organism spiked into sputum and stored in PrimeStore MTM resulted in a variation of 1.6 C; or less over 36 days at both 4°C and 27°C. The Positive samples not stored in Primestore MTM demonstrated degraded DNA as time and temperature increased.
#### Influenza A stability:
Stability studies evaluated the stability of Influenza A virus (Flu A: 1 x 102 TCID50/mL) RNA from clinical samples incubated in PrimeStore MTM at: refrigerated temperature (4°C, 39.2°F) for 29 days, Table 7, and ambient temperature (27°C, 80.6°F) for 8 days, Table 8. The clinical matrix for the flu A stability study was nasal washes. The RNA was stabilized under the predefined conditions and was then extracted using PrimeXtract spin columns. Nucleic acid stability was measured and analyzed with the ABI 7500 (Thermo Fisher Scientific) instrument. The experiment analyzed 25 replicates extracted at each time point and for each temperature range. An initial time point designated as Day 0 and was included as the initial Ct average for each of the two temperature ranges tested. Time points for sample extractions were performed at Day 0, 1, 8, 15, 22 and 29 for refrigerated temperature (4°C, 39.2°F) and Day 0, 1 and 8 for ambient temperature (27°C, 80.6°F). An embedded Internal Negative Control (INC) was included in the study at both temperatures. The INC consisted of a test using negative clinical sample in PrimeStore MTM, i.e., nasal washing without Flu A virus. Additionally, a control that contained clinical specimen plus target, i.e., nasal washing plus Flu A was incubated
{7}------------------------------------------------
in PBS (POS (-PS)) and included in the evaluation. Pre-defined acceptance criteria of (+/-) 3.0 C; from time zero was used to establish stability and preservation of nucleic acids (RNA from Flu A virus) as determined by real-time PCR, for 4°C and 27°C without loss of detection signal using statistical analysis.
| Day (4°C) | 0 | 1 | 8 | 15 | 22 | 29 |
|-----------------|------|------|------|------|------|------|
| AVG (Ct): | 31.9 | 32.3 | 34.4 | 34.3 | 34.1 | 34.6 |
| SD (Ct): | 0.6 | 0.8 | 1.3 | 0.9 | 0.8 | 0.8 |
| INC: | NEG | NEG | NEG | NEG | NEG | NEG |
| POS (-PS) (Ct): | 34.4 | 33.6 | 35.8 | NEG | 36.5 | NEG |
Table 7. Flu A (1 x 103 TCID50 / mL) stability at 4°C
Table 8. Flu A (1 x 103 TCID50 / mL) stability at 27°C
| Day (27°C) | 0 | 1 | 8 |
|-----------------|------|------|------|
| AVG (Ct): | 32.3 | 33.0 | 34.3 |
| SD (Ct): | 0.4 | 0.9 | 1.6 |
| INC: | NEG | NEG | NEG |
| POS (-PS) (Ct): | 34.4 | 34.9 | 37.6 |
Stability testing of RNA from Influenza A whole virus spiked into nasal wash and stored in PrimeStore MTM resulted in a variation of 2.7 Cr over 29 days at 4°C and a variation of 2.0 Ct over 8 days at 27℃. The Positive samples not stored in Primestore MTM demonstrated degraded RNA as time and temperature increased.
- c. Inactivation
#### Mycobacterium tuberculosis inactivation:
Sputum samples not submitted for MTB investigation were obtained from the diagnostic laboratory at the University of Pretoria (Pretoria, South Africa) and assessed for the presence of acid-fast bacilli by smear microscopy, cultured by MGIT 960 system to confirm the absence of MTB followed by quality assessment using the Bartlett Scoring System. Good quality purulent sputum specimens (Bartlett test score of 2+) were included for use in the spiking matrix experiments. These sputa were split and spiked with MTB H37Rv strain with concentrations of 1.5 x106 and 1.5 x 108 CFU/mL followed by inoculation into PrimeStore MTM (without decontamination or other pre-culture steps). A matrix assessment to determine effect of concentration was performed by adding to 1 mL of spiked sputum to 3, 2, and 1 mL of PrimeStore MTM. Samples were incubated in triplicate at ambient temperature for 1, 5, 10, 30, 60 and 180 minutes, including two positive and negative controls, and analyzed using the MGIT 960 system. Effective inactivation at each concentration and time point is defined as no growth in all samples after 42 days.
Transport media inactivation results:
No growth of MTB was observed when the concertation of MTB in sputum was 1.5 x 100 CFU/mL and incubated for a time of greater than 5 minutes with PrimeStore
{8}------------------------------------------------
MTM. Intermittent growth was observed at each time point when MTB concentration was1.5 x 10° CFU/ml and added at a ratio of 1:1, 1:2 and 1:3 sample to PrimeStore MTM. Growth was determined after 42 days of incubation (Table 9. below).
| Concentration<br>of<br>$M. tuberculosis$<br>spiked in<br>sputum | Ratio of Sputum to<br>PrimeStore MTM | Exposure to PrimeStore MTM Time<br>(Minutes)<br>Number of replicates which exhibited<br>growth after 42 days | | | | | | |
|-----------------------------------------------------------------|--------------------------------------|--------------------------------------------------------------------------------------------------------------|-----|-----|-----|-----|-----|-----|
| | | 1 | 5 | 10 | 15 | 30 | 60 | 180 |
| 1.5 x 106<br>CFU/mL | 1-3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | 1-2 | 1/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| | 1-1 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 |
| 1.5 x 108<br>CFU/mL | 1-3 | 0/3 | 1/3 | 0/3 | 1/3 | 0/3 | 0/3 | 0/3 |
| | 1-2 | 3/3 | 0/3 | 0/3 | 0/3 | 2/3 | 0/3 | 0/3 |
| | 1-1 | 3/3 | 1/3 | 2/3 | 0/3 | 0/3 | 1/3 | 2/3 |
Table 9. MTB Growth at 42 days after exposure to PrimeStore MTM at different concentrations and ratios
The data shows that PrimeStore MTM must be used at a ratio of at least 1:3 sputum to PrimeStore MTM and a minimum of 60 minutes exposure time to demonstrate MTB inactivation.
#### Influenza A inactivation:
Influenza/A/Wuhan/359/95 (10° TCID50/ml) were incubated with PrimeStore MTM for 10, 30 and 60 seconds. Influenza only and PrimeStore MTM only were also incubated accordingly to serve as internal controls. Four days after inoculation, the cells were fixed and stained with 0.06% crystal violet in 1% glutaraldehyde. Wells that were not stained demonstrated the cytopathic effect (CPE) of the virus. The titer of the virus was recorded as the TCID50.
### Inactivation rate:
PrimeStore MTM showed cytotoxicity on MDCK cells when diluted 1:100 but not cytotoxicity at 1:1.000. The mixture of PrimeStore MTM and Influenza had similar profiles as PrimeStore MTM alone with 10, 30 and 60 second incubations, while Influenza only samples had viral loads of 1 x 107.75, 1 x 10 7.5 and 1 x 107.75 TCIDs0 respectively.
### Transport media inactivation:
PrimeStore MTM rapidly inactivated Influenza virus with a >4.0 log reduction in concentration at 10 seconds. Viral CPE could not be observed at < 3.0 logs due to cellular destruction by PrimeStore MTM, Table 10 below.
{9}------------------------------------------------
| | 10s incubation<br>1 X 10x TCID50 | 30s incubation<br>1 X 10x TCID50 | 60s incubation<br>1 X 10x TCID50 |
|--------------------------------|----------------------------------|----------------------------------|----------------------------------|
| Flu A only | 7.75 | 7.5 | 7.75 |
| Flu A and<br>PrimeStore<br>MTM | < 3.0 | < 3.0 | < 3.0 |
| PrimeStore<br>MTM only* | < 3.0 | < 3.0 | < 3.0 |
Table 10 MTB inactivation in PrimeStore MTM
*PrimeStore shows cytotoxicity on MDCK cells when diluted to 1:1,000.
The PrimeStore MTM must be used at a ratio of at least 1:3 nasal to PrimeStore MTM and a minimum of 10 seconds exposure time to demonstrate inactivation of Influenza A. Measuring Influenza inactivation below 1 X 103 was not possible because of the cytotoxic affects PrimeStore MTM has on the cell culture based testing.
### 2. Comparison Studies:
#### Clinical Comparison: a.
Method comparison is not applicable for a nucleic acid transport device. The device itself does not provide a result that can be used in making a clinical decision. Bench testing studies were done to determine the ability of PrimeStore MTM to stabilize the nucleic acids from MTB and Flu A samples. Nucleic Acid stability was performed using clinically negative matrix pooled and a minimum of 50 distinct matrices were used for each analyte for the stability study.
- b. Extraction platform and amplification instrument compatibility
### Mycobacterium Tuberculosis:
A study to establish that the PrimeStore MTM media is compatible with commercially available nucleic acid isolation reagents was performed using MTB spiked into sputum with final concentrations ranging from 3 to 250,000 CFU/mL. Aliquots of sputum in PrimeStore MTM were prepared and processed using the following nucleic acid extraction systems as per manufacturers' instructions:
- The QIAamp DNA mini kit (Qiagen, Hilden, Germany)
- MagNA Pure 96 System (Roche Diagnostics, Mannheim, Germany) using the DNA Bacterial/Viral small volume kit
- NucliSENS easyMAG (bioMerieux, Marcy I'Etoile, France) and using the generic protocol.
- An FDA cleared assay was performed as a control. ●
{10}------------------------------------------------
An aliquot of a sputum sample was tested using the control assay prior to inoculating in PrimeStore MTM, as instructed by the manufacturer. An input volume of 200 µL PrimeStore MTM sample was used along with the manufacturer recommended output volume of 50 uL (QIAamp and NucliSENS) and 100 uL (MagNA Pure 96) for further analysis. Real-time PCR targeting the MTB specific insertion sequence element 6110 (IS6110) was performed on the LightCycler 480II platform (Roche Diagnostics, Mannheim, Germany) to identify MTB in the DNA extracts. For each extraction method, the limit of detection for MTB DNA in PrimeStore MTM was similar to the PBS control as noted in Table 11.
| | MagNA Pure<br>(PrimeStore<br>MTM)1,2 | EasyMAG<br>(PrimeStore<br>MTM)1,3 | QIAamp<br>(PrimeStore<br>MTM)1,4 | FDA Cleared<br>control (PBS<br>Control) |
|----------------------------------|-------------------------------------------------|-----------------------------------|----------------------------------|-----------------------------------------|
| MTB<br>Concentration<br>(CFU/mL) | Number of positive replicates out of 20 samples | | | |
| 2.5 X 105 | 20 | 20 | 20 | 20 |
| 2.5 X 104 | 20 | 20 | 20 | 20 |
| 2.5 X 103 | 20 | 20 | 16 | 13 |
| 2.5 X 102 | 20 | 20 | 18 | 5 |
| 2.5 X 101 | 13 | 10 | 7 | 3 |
Table 11. MTB extraction comparison
1 LightCycler 480II platform (Roche Diagnostics, Mannheim, Germany).
2 MagNA Pure LoD, 2.5 X 102 CFU/mL
3 EasyMAG LoD, 2.5 X 102 CFU/mL
4 QIAamp LoD, 2.5 X 104 CFU/mL
A study was performed to establish the performance of PrimeStore MTM with the same amplification assay performed using two PCR amplification instruments. The study evaluated the detection of three 10-fold concentrations (1 X 10, 1 X 104, and 1x103 CFU/mL) of MTB (Strain HN878, 1x108 CFU/mL) extracted from sputum stored in PrimeStore MTM. Amplification was performed using two commercial real-time PCR platforms: 1) The ABI 7500 (Thermo Fisher Scientific) and 2) LightCycler 2.0 (Roche Diagnostics). The results using the two platforms are similar as shown in Table 12.
{11}------------------------------------------------
| | | | Average CT Value |
|---------------------|----------------|----------------|----------------------|
| ABI 7500 Instrument | CT Value Rep 1 | CT Value Rep 2 | (High Concentration) |
| 105 Extraction 1 | 19.2 | 18.8 | 19.0 |
| 105 Extraction 2 | 19.4 | 18.9 | 19.1 |
| 104 Extraction 1 | 22.0 | 22.1 | 22 |
| 104 Extraction 2 | 21.6 | 22.0 | 21.8 |
| 103 Extraction 1 | 24.4 | 24.7 | 24.5 |
| 103 Extraction 2 | 24.8 | 24.7 | 24.7 |
| NTC | NEG | NEG | |
| NTC | NEG | NEG | |
| Positive Control | POS | POS | |
| | | | |
| | | | Average CT Value |
| Roche LightCycler | CT Value Rep 1 | CT Value Rep 2 | (Low Concentration) |
| 105 Extraction 1 | 19.9 | 20.3 | 20.1 |
| 105 Extraction 2 | 20.3 | 20.0 | 20.1 |
| 104 Extraction 1 | 23.4 | 22.8 | 23.1 |
| 104 Extraction 2 | 22.6 | 23.1 | 22.8 |
| 103 Extraction 1 | 25.6 | 25.9 | 25.7 |
| 103 Extraction 2 | 26.9 | 26.7 | 26.8 |
| NTC | NEG | NEG | |
| NTC | NEG | NEG | |
Table 12: Amplification comparison for two amplification instruments
The test results indicate that the extraction method used may affect the LoD of the downstream assay. The extraction methods listed above all indicate they are compatible with PrimeStore MTM but the QIAamp increases the LoD by two logs. A similar result is observed when the ABI 7500 and Roche LightCycler are compared to each other. The variation within each instrument is on average within 1 C. The variation between instruments is between 1 to 1.7 Cc depending on the concentration of M. tuberculosis used for testing. The results indicate that the extraction platform and amplification instrument may need additional validation to ensure robust results are provided.
- c. Matrix Equivalence Study
Not applicable
- 3. Clinical Studies:
Not applicable
- 4. Clinical cut-off:
Not applicable
{12}------------------------------------------------
- 5. Expected values/Reference range:
Not applicable.
### M. Instrument Name
Not applicable
### N. System Descriptions:
- 1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ___________ or No ________________________________________________________________________________________________________________________________________________________
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ___________ or No _______X___
- 2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ___________ or No _____X_____
Hazard Analysis and software development are not applicable to this class of device:
- 3. Specimen Identification:
Transport device are not intended to identify specimens. The device itself does not provide a result that can be used in making a clinical decision. Transport devices are intended to preserve and stabilize nucleic acids.
- 4. Specimen Sampling and Handling: See section L.1.b regarding specimen stability.
- 5. Calibration: Not applicable.
- 6. Quality Control: Not applicable.
### O. Other Supportive Instrument Performance Characteristics Data Not Covered in the "Performance Characteristics" Section above:
Not applicable
{13}------------------------------------------------
## P. Proposed Labeling:
The labeling is sufficient and satisfies the requirements of 21 CFR parts 801 and 809 as well as the Special Controls for this type of device.
{14}------------------------------------------------
# Q. Identified Risks to Health and Mitigation Measures:
| Identified Risks to Health | Mitigation Measures |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Failure to stabilize pathogen nucleic acid<br>resulting in a false negative result | General Controls and Special Controls (1),<br>(2), and (3) |
| Failure to inactivate the specimen | General Controls and Special Controls (1),<br>(2)(i), (2)(ii), (3)(i), (3)(ii) and (3)(iv) |
## R. Benefit/Risk Analysis:
| Summary | | |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Summary of<br>the Benefit(s) | 0<br>PrimeStore MTM is a transport media that inactivates and stabilizes sputum specimens<br>from patients suspected of influenza A or M. tuberculosis infection. | |
| | PrimeStore MTM facilitates downstream diagnostic testing by stabilizing pathogen<br>●<br>nucleic acids to allow storage and transport of specimens between facilities or from<br>patient care areas to the laboratory. | |
| | PrimeStore MTM may help prevent occupational transmission of infection to<br>●<br>laboratory personnel by inactivating infectious agents. | |
| Summary of<br>the Risk(s) | · False negative results or failure to inactivate infectious pathogens are the primary risks<br>associated with use of the PrimeStore MTM transport media. | |
| | · If PrimeStore MTM does not stabilize pathogen nucleic acids, a false negative result<br>could occur and may result in a delay of antimicrobial therapy, with subsequent<br>worsening of infection and associated increase in morbidity or mortality. | |
| | · Failure to inactivate pathogen nucleic acids may increase risk of transmission to<br>laboratory personnel. | |
| Summary of<br>Other<br>Factors | None | |
| Conclusions<br>Do the<br>probable<br>benefits<br>outweigh the<br>probable<br>risks? | The probable benefits of the PrimeStore MTM device outweigh the potential risks in light<br>of the listed special controls and applicable general controls. The PrimeStore MTM<br>inactivates and stabilizes influenza A and M. tuberculosis nucleic acids, which may protect<br>laboratory workers from occupational exposures and assist downstream molecular assays<br>to detect Influenza A and M. tuberculosis. The proposed special controls will ensure that<br>errors will be uncommon, and potential errors are further mitigated by current laboratory<br>practices, which include universal precautions to protect laboratory technicians. The<br>PrimeStore MTM performance characteristics suggest that the device will be safe and<br>effective, if used as directed by the package insert, and could provide potential benefits to<br>patients and laboratory personnel by facilitating the safe transport of clinical specimens. | |
{15}------------------------------------------------
### S. Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
### T. Conclusion:
The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.2950. FDA believes that the stated special controls, and applicable general controls provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:
| Product Code: | QBD |
|---------------|---------------------------------------------------------|
| Device Type: | Microbial nucleic acid storage and stabilization device |
| Class: | II (special controls) |
| Regulation: | 21 CFR 866.2950 |
- (a) Identification. A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device in not intended for preserving morphology or viability of microorganisms.
- (b) Classification. Class II (special controls). the special controls for this device are:
- The intended use for the 21 CFR 809.10 labeling must include a detailed description (1) of microorganisms and types of human specimens intended to be preserved.
- (2) The 21 CFR 809.10(b) labeling must include:
- (i) A detailed device description, including all device components.
- (ii) Performance characteristics from applicable analytical studies, including but not limited to, nucleic acid stability and microorganism inactivation.
- (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing.
- (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies.
- (3) Design verification and validation must include the following:
- (i) Overall device design including all device components and all control elements incorporated into the analytical validation procedures.
- (ii) Thorough description of the microorganisms and methodology used in the
{16}------------------------------------------------
validation of the device including, but not limited to, extraction platforms and assays used for the detection of preserved nucleic acids.
- (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.