← Product Code [PQK](/productcode/PQK) · DEN170004

# Ellipsys Vascular Access System (DEN170004)

_Avenu Medical, Inc. · PQK · Jun 22, 2018 · Cardiovascular · DENG_

**Canonical URL:** https://fda-staging.innolitics.com/device/DEN170004

## Device Facts

- **Applicant:** Avenu Medical, Inc.
- **Product Code:** [PQK](/productcode/PQK.md)
- **Decision Date:** Jun 22, 2018
- **Decision:** DENG
- **Submission Type:** Direct
- **Regulation:** 21 CFR 870.1252
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Ellipsys® System is indicated for the creation of a proximal radial artery to perforating vein anastomosis via a retrograde venous access approach in patients with a minimum vessel diameter of 2.0mm and less than 1.5mm of separation between the artery and vein at the fistula creation site who have chronic kidnev disease requiring dialysis.

## Device Story

Single-use percutaneous catheter system; creates arteriovenous fistula (AVF) in arm for hemodialysis access. Retrograde venous access approach; connects proximal radial artery to perforating vein. Used in clinical setting; operated by physician. Device delivers, deploys, and creates anastomosis; enables blood flow suitable for dialysis. Benefits patients by providing vascular access for hemodialysis via minimally invasive percutaneous procedure rather than traditional surgery.

## Clinical Evidence

Prospective, non-randomized, multi-center study (N=103 ITT). Primary safety endpoint (device-related SAEs) met (0% rate). Primary effectiveness endpoint (90-day maturation success: venous diameter >4mm, flow >500 mL/min) met with 89.3% success (p<0.0001). 12-month assisted primary patency was 81%. 96.1% of subjects required maturation/maintenance procedures (MAPs), primarily PTA and coil embolization. 94.5% of MAPs had no associated AEs.

## Technological Characteristics

Percutaneous catheter system; single-use. Includes electrical components requiring safety and EMC testing. Software-controlled. Biocompatible patient-contacting components. Requires sterilization validation and shelf-life testing.

## Regulatory Identification

This device is a single use percutaneous catheter system that creates an arteriovenous fistula in the arm of patients with chronic kidney disease who need hemodialysis.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate:
(i) The ability to safely deliver, deploy, and remove the device;
(ii) The ability of the device to create an arteriovenous fistula;
(iii) The ability of the arteriovenous fistula to attain a blood flow rate and diameter suitable for hemodialysis;
(iv) The ability of the fistula to be used for vascular access for hemodialysis;
(v) The patency of the fistula; and
(vi) The rates and types of all adverse events.
(2) Animal testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be assessed:
(i) Delivery, deployment, and retrieval of the device;
(ii) Compatibility with other devices labeled for use with the device;
(iii) Patency of the fistula;
(iv) Characterization of blood flow at the time of the fistula creation procedure and at chronic followup; and
(v) Gross pathology and histopathology assessing vascular injury and downstream embolization.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Simulated-use testing in a clinically relevant bench anatomic model to assess the delivery, deployment, activation, and retrieval of the device;
(ii) Tensile strengths of joints and components;
(iii) Accurate positioning and alignment of the device to achieve fistula creation; and
(iv) Characterization and verification of all dimensions.
(4) Electrical performance, electrical safety, and electromagnetic compatibility (EMC) testing must be performed for devices with electrical components.
(5) Software verification, validation, and hazard analysis must be performed for devices that use software.
(6) All patient-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance data must demonstrate the sterility of the device components intended to be provided sterile.
(8) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(9) Labeling for the device must include:
(i) Instructions for use;
(ii) Identification of system components and compatible devices;
(iii) Expertise needed for the safe use of the device;
(iv) A detailed summary of the clinical testing conducted and the patient population studied; and
(v) A shelf life and storage conditions.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a De Novo](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### DE NOVO CLASSIFICATION REQUEST FOR ELLIPSYS VASCULAR ACCESS SYSTEM

#### REGULATORY INFORMATION

FDA identifies this generic type of device as:

Percutaneous catheter for creation of an arteriovenous fistula for hemodialysis access. This device is a single use percutaneous catheter system that creates an arteriovenous fistula (AVF) in the arm of patients with chronic kidney disease who need hemodialysis.

NEW REGULATION NUMBER: 21 CFR 870.1252

CLASSIFICATION: II

PRODUCT CODE: POK

#### BACKGROUND

DEVICE NAME: Ellipsys Vascular Access System (Ellipsys® System)

SUBMISSION NUMBER: DEN170004

DATE OF DE NOVO: January 10, 2017

| CONTACT: | Avenu Medical, Inc.<br>27123 Calle Arroyo, Suite 2101<br>San Juan Capistrano, CA 92675 |
|----------|----------------------------------------------------------------------------------------|
|----------|----------------------------------------------------------------------------------------|

#### INDICATIONS FOR USE

The Ellipsys® System is indicated for the creation of a proximal radial artery to perforating vein anastomosis via a retrograde venous access approach in patients with a minimum vessel diameter of 2.0mm and less than 1.5mm of separation between the artery and vein at the fistula creation site who have chronic kidnev disease requiring dialysis.

#### LIMITATIONS

The sale, distribution, and use of the Ellipsys® System are restricted to prescription use in accordance with 21 CFR 801.109.

The Avenu Medical Ellipsys® System is intended to only be used by physicians trained in ultrasound guided percutaneous endovascular interventional techniques using appropriate clinical standards of care for fistula maintenance and maturation including balloon dilatation and coil embolization.

{1}------------------------------------------------

The Ellipsys® System is contraindicated for use in patients with target vessels that are < 2mm in diameter. The Ellipsys® System is contraindicated for use in patients who have a distance between the target artery and vein > 1.5mm.

The Ellipsys® System has only been studied for the creation of an AV fistula using the proximal radial artery and the adjacent perforating vein. It has not been studied in subjects who are candidates for surgical fistula creation at other locations, including sites distal to this location.

Additional procedures are expected to be required to increase and direct blood flow into the AVF target outflow vein and to maintain patency of the AVF. Care should be taken to proactively plan for any fistula maturation procedures when using the device.

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

{2}------------------------------------------------

## DEVICE DESCRIPTION

The Ellipsys® Vascular Access System is a catheter based system that is used to percutaneously create a vascular anastomosis between adjacent blood vessels using direct current (DC) thermal heating. The system consists of several components:

- Ellipsys Catheter ●
- Ellipsys Crossing Needle
- Ellipsys Power Controller ●

## Ellipsys Catheter

The Ellipsys Catheter is a single-use, sterile, flexible catheter with a heating element on the distal tip. The catheter is tracked over a guidewire in the target anastomosis site. Then the distal tip components are extended, with part of the tip placed into the target artery, and the tip is retracted to approximate the vessels (i.e., pull the vessels together). Thermal energy is then delivered from the heating element on the tip to seal the vessels together and create an arteriovenous fistula (AVF). The figures below illustrate the catheter and its mechanism of action.

Image /page/2/Figure/7 description: The image shows a medical device with a long, thin probe extending from a white handle. The handle has the word "ELLIPSYS" printed on it in red letters. A blue and white connector is attached to the handle by a white cable, which is coiled up next to the device.

Figure 1: Ellipsys Catheter

Image /page/2/Picture/9 description: The image shows a medical illustration of a surgical procedure. Two surgical tools with blue handles and gold tips are positioned near red and blue vessels. The tools appear to be puncturing or accessing the vessels, possibly for a procedure like angioplasty or catheterization. The background includes a faint outline of skeletal structures, suggesting the anatomical context of the procedure.

Figure 2: Ellipsys Catheter Position in Vasculature Prior to Activation

{3}------------------------------------------------

Image /page/3/Picture/0 description: In the image, a medical illustration depicts a surgical procedure involving blood vessels. A golden-colored instrument with a blue handle is positioned between a red and a blue vessel, possibly indicating arteries and veins. A circular, transparent overlay with concentric hexagonal patterns surrounds the area of intervention, suggesting a focus on the surgical site or a technology being used.

Figure 3: Creation of a Fistula with the Ellipsys Catheter

# Ellipsys Crossing Needle

The Ellipsys Crossing Needle is a single use, sterile, telescoping needle that is used to cross from inside the vein into the artery at the target anastomosis site. When the inner needle is retracted, it has a round, atraumatic tip that can be positioned against the wall of the adjacent vessel prior to advancing the needle. The sharp needle tip housed within the cannula can be manually deployed a distance of 10mm by pushing forward on the inner hub of the handle. The Ellipsys Crossing Needle is compatible with a non-coated .014" guidewire and a 6F introducer sheath.

Image /page/3/Figure/4 description: The image shows a diagram of a needle in both retracted and deployed positions. When retracted, the needle measures 13 cm. A detailed view labeled "DETAIL A" shows the needle tip measuring 10 mm in length. The image also includes a "FLAT SIDE" view of the needle hub.

Figure 4: Ellipsys Crossing Needle

# Ellipsys Power Controller

Unlike the other components, the Ellipsys Power Controller is a reusable, non-sterile device. This portable electronic console is powered with AC inputs, has internal control electronics, and a user interface display. The catheter can connect to the controller which provides the catheter with DC power (converted from AC power) and a closed loop temperature control of the distal heating element.

{4}------------------------------------------------

The Power Controller performs the following functions:

- Performs a start-up sequence to verify catheter function prior to insertion into the patient. .
- Contains a visual display screen so the user can verify catheter functionality and start the . thermal cycle.
- Controls the thermal cycle. .
	- o Applies DC energy to heat the catheter
	- o Monitors the tip temperature of the catheter
	- o Monitors the tip position of the catheter
	- o Provides closed loop temperature control algorithm to ensure the temperature is maintained at the programmed parameter
- Stores patient ID, the process parameters and control calculations by the main controller . board.
- Has a USB port to allow the user to externally export procedural data to a USB flash drive. .

Image /page/4/Picture/10 description: The image shows a gray electronic device with a large rectangular screen in the center. The screen is black, suggesting it is either off or displaying a dark image. There is a red button with a yellow border in the upper left corner and a power button symbol in the upper right corner. A small circular button is located in the bottom left corner.

Figure 5: Ellipsys Power Controller

Please refer to the Instructions for Use for additional details.

### SUMMARY OF NONCLINICAL/BENCH STUDIES

### BIOCOMPATIBILITY/MATERIALS

The Catheter and the Crossing Needle are externally communicating components that are in direct contact with circulating blood for a limited duration (b) (4) The biocompatibility testing summarized below was performed to demonstrate that these components are biocompatible for their intended uses. All tests passed.

The Power Controller is a non-patient-contacting component and was not tested.

| Test          | Description (Method)                                                    |
|---------------|-------------------------------------------------------------------------|
| Cytotoxicity  | MEM Elution Assay with L-929 Mouse Fibroblast<br>Cells<br>(ISO 10993-5) |
| Sensitization | Guinea Pig Maximization<br>(ISO 10993-10)                               |

Table 1: Biocompatibility Testing Summary

{5}------------------------------------------------

| Irritation            | Intracutaneous Reactivity<br>(ISO 10993-10)                                                                                                                                                 |  |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Acute System Toxicity | Acute Systemic Injection<br>(ISO 10993-11)                                                                                                                                                  |  |
| Hemocompatibility     | ASTM Hemolysis Assay - Direct Contact &<br>Extract Methods<br>(ASTM Method F756-08)                                                                                                         |  |
|                       | Complement Activation C3a and SC5b-9<br>Determination of SC5b-9 Terminal Chain Complex<br>(TCC) and C3a-desArg Present in Normal Human<br>Serum Through Enzyme Immunoassay<br>(ISO 10993-4) |  |
|                       | In-vivo Thromboresistance<br>(evaluated as part of the animal studies)                                                                                                                      |  |
| Pyrogenicity          | Material-mediated Rabbit Pyrogen<br>(USP Rabbit Pyrogen Test Procedure, Section 151)<br>Endotoxin-mediated (LAL)<br>(AAMI ST72)                                                             |  |

# SHELF LIFE/STERILITY

The shelf-life of the Ellipsys System has been established at 2 years based on accelerated aging testing equivalent to 2 years in accordance with ASTM F1980 - 07 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Following 2X sterilization, environmental conditioning per ISTA 2A, distribution simulation per ASTM D4169, and accelerated aging, the devices were visually inspected for damage per ASTM F1886. bubble leak tested per ASTM F2096. and package seals were tested per ASTM F88. Aged devices also underwent repeat engineering bench testing to confirm acceptable performance.

The Ellipys Catheter and Crossing Needle are labeled as sterile and have a validated sterility assurance level (SAL) of 106. The Ellipsys Catheter and Crossing Needle have been validated to be sterilized via ethylene oxide (EO). The sterilization cycle was validated using the (b) (4) , and the EO and ECH residuals were shown to meet the limits specified by ISO 10993-7:2008.

The Ellipsys Power Controller is provided non-sterile and is labeled as such. The Power Controller user manual includes instructions for cleaning.

### ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY

The electromagnetic compatibility and electrical safety of the Ellipsys System was evaluated by demonstrating compliance to the following standards:

| Standard                                                   | Name                                                                                                         |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005 +<br>CORR. 1 (2006) +<br>CORR. 2 (2007) + | Medical Electrical Equipment - Part 1: General<br>Requirements For Basic Safety And Essential<br>Performance |

Table 2: Electrical Performance Testing Summary

{6}------------------------------------------------

| AM12012            |                                                                                                                                                                                                   |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2007 | Medical Electrical Equipment - Part 1-2:<br>General Requirements For Basic Safety And<br>Essential Performance - Collateral Standard:<br>Electromagnetic Disturbances - Requirements<br>and Tests |

#### MAGNETIC RESONANCE (MR) COMPATIBILITY

The Ellipsys System is intended as a temporary use device and has not been tested for MR compatibility.

#### SOFTWARE

The Ellipsys Power Controller software controls the catheter temperature. The controller also receives, analyzes, and adjusts the power output of the power control module to maintain the desired temperature. The controller reports the data via the user interface. archives recent process data for later analysis, and directs the user through most of the Ellipsys system, automating most functions. The software monitors the catheter for errors and provides alerts and instructions to the user if it cannot resolve the error automatically.

The Ellipsys Power Controller has a Moderate Level of Concern (LOC). Appropriate documentation was provided to support the validation of the Power Controller software for a Moderate LOC in accordance with FDA's 2005 guidance titled, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

### PERFORMANCE TESTING - BENCH

The Ellipsys System was subjected to a series of bench tests to assess its functional performance. These tests were performed on final, sterilized product. The engineering bench testing summarized in the table below was performed to demonstrate acceptable mechanical performance of the device for its intended use.

| Test                                   | Description/Acceptance Criteria                |                                                                                                                                                                                                                                                                                                                                                                           |
|----------------------------------------|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ellipsys Catheter                      |                                                |                                                                                                                                                                                                                                                                                                                                                                           |
| Dimensional<br>Verification/Validation | Catheter Working<br>Length                     | The working length of the catheter must be > 15cm                                                                                                                                                                                                                                                                                                                         |
|                                        | 6F Sheath<br>Compatibility                     | The device must pass through a 6F sheath without<br>excessive friction.                                                                                                                                                                                                                                                                                                   |
|                                        | 0.014" Guidewire<br>Compatibility              | 0.014" guidewire must pass through device lumen with<br>ease.                                                                                                                                                                                                                                                                                                             |
|                                        | Hub/Tip<br>Alignment                           | Hub must correctly identify the tip orientation; i.e., taper of<br>heater element must face away from thumb tab.                                                                                                                                                                                                                                                          |
| Electrical<br>Performance              | Electrical<br>Connection to<br>Generator       | Catheter connector shall only be connected to the generator<br>in one orientation.                                                                                                                                                                                                                                                                                        |
|                                        | Control<br>Thermocouple<br>Accuracy            | Ambient Temperature: $700 \pm 10$ °F                                                                                                                                                                                                                                                                                                                                      |
|                                        | Base<br>Thermocouple<br>Accuracy               | Ambient Temperature: $700 \pm 50°F$                                                                                                                                                                                                                                                                                                                                       |
|                                        | Position Sensor<br>Function                    | The voltage range for the position sensor corresponding to<br>opening and closing of device shall have a 1-1.3 V voltage<br>differential.                                                                                                                                                                                                                                 |
| Tensile Strength                       |                                                | Bonds/joints shall maintain mechanical integrity during use Each bond/joint was tested against specifications based on the clinical use of the device                                                                                                                                                                                                                     |
| Simulated Use                          |                                                | Catheter shall properly connect to the Power Controller. Power Controller software shall run without errors or interruption when connected to the catheter. Track device samples through vein/artery samples and activate device to create a fistula. After simulated use, the pull force required to separate the vessels must be > 20g. Device shall be free of damage. |
| Corrosion Resistance                   |                                                | Metallic components of the catheter intended for fluid path contact shall show no signs of corrosion.                                                                                                                                                                                                                                                                     |
| Coating Integrity Testing              |                                                | After simulated use in a clinically-relevant model, and<br>subjecting the device to 2 heat cycles, the device coating<br>shall not exhibit any damage or peeling when inspected<br>under magnification.                                                                                                                                                                   |
| Ellipsys Crossing Needle               |                                                |                                                                                                                                                                                                                                                                                                                                                                           |
|                                        | Crossing Needle<br>Working Length              | The working length of the needle must be $13.0 \pm 0.5$ cm.                                                                                                                                                                                                                                                                                                               |
|                                        | 6F Sheath<br>Compatibility                     | The device must pass through a 6F sheath without<br>excessive friction.                                                                                                                                                                                                                                                                                                   |
|                                        | 0.014" Guidewire<br>Compatibility              | 0.014" guidewire must pass through device lumen with<br>ease.                                                                                                                                                                                                                                                                                                             |
|                                        | Needle Extension<br>Length                     | Needle shall extend between 8-11mm.                                                                                                                                                                                                                                                                                                                                       |
| Dimensional<br>Verification/Validation | Needle Tip<br>Orientation and<br>Grind Profile | Needle hub shall correctly indicate the position of the needle. Needle tip shall be sharp, not bent or flattened, when inspected under magnification.                                                                                                                                                                                                                     |
|                                        | Distal End of<br>Outer Cannula<br>Protrusion   | Distal end of cannula shall conform to drawing dimensions and needle shall not protrude from cannula when retracted. Tip of cannula shall be free of debris or sharp edges when inspected under magnification.                                                                                                                                                            |
| Tensile Strength                       |                                                | Bonds/joints shall maintain mechanical integrity during use. Each bond/joint was tested against specifications based on the clinical use of the device.                                                                                                                                                                                                                   |
| Corrosion Resistance                   |                                                | The crossing needle shall show no signs of corrosion.                                                                                                                                                                                                                                                                                                                     |
| Ellipsys Power Controller              |                                                |                                                                                                                                                                                                                                                                                                                                                                           |
| Electrical Performance                 | Position Sensor<br>Function                    | Position sensor shall have an accuracy of $3.0 \pm 0.5$ mm.                                                                                                                                                                                                                                                                                                               |
|                                        | Electrical<br>Connection to<br>Catheter        | The Power Controller shall connect to the catheter in only<br>one orientation.                                                                                                                                                                                                                                                                                            |

Table 3: Performance Testing (Bench) Summary

{7}------------------------------------------------

{8}------------------------------------------------

# PERFORMANCE TESTING - ANIMAL &/OR CADAVER

Two main in vivo animal studies were conducted to demonstrate the feasibility and safety of the Ellipsys System. The studies are summarized in the table below.

| Purpose                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Methods                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Results                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Evaluate the acute and<br>chronic safety, fistula<br>blood flow<br>characteristics, fistula<br>patency, handling<br>characteristics, and<br>histopathological<br>response at the<br>treatment site and off-<br>target tissue sites using<br>the final device<br>design. Perform gross<br>comparison of fistulae<br>created with the final<br>device design<br>compared to a<br>previous device design<br>to demonstrate<br>equivalent fistula<br>creation. Perform<br>study according to<br>Good Laboratory<br>Practices. | • GLP<br>• N=6 ovine models<br>• Insert, track, deploy, and<br>activate device to create<br>bilateral arteriovenous fistulae<br>in each animal. The original<br>device design was used to<br>create a fistula in one femoral<br>artery and the final device<br>design was used to create a<br>fistula in the other femoral<br>artery.<br>• Acute and post-treatment<br>assessments of device<br>performance included needle<br>and catheter access/removal,<br>and anastomosis and sheath<br>appearance.<br>• Immediately after fistula<br>creation, fistulae were evaluated<br>by ultrasound to determine<br>fistula diameter, patency, and<br>fistula flow.<br>• At 1, 7, and 30 days post-<br>procedure, animals were<br>evaluated for safety events<br>(swelling, hematoma,<br>thrombosis, etc.) and ultrasound<br>was used to evaluate fistula<br>diameter, patency, and fistula<br>flow.<br>• Gross necropsy, including<br>external and internal<br>examination of the fistulae, was<br>performed at 30 days.<br>• Histopathological examination<br>of the fistulae and downstream<br>organs such as the brain, heart,<br>lungs, liver, spleen, and kidneys<br>was conducted at 30 days to<br>evaluate chronic safety of the<br>fistula and any thromboembolic<br>events.<br>• All results from fistulae created<br>with the final device design<br>were compared to the fistulae | • All fistulae were successfully created.<br>• Needle access/removal, catheter<br>insertion/removal, anastomosis<br>appearance, and sheath appearance were<br>acceptable for all devices.<br>• No procedural or handling differences<br>were observed between the two device<br>designs.<br>• Intra-operative ultrasound demonstrated<br>a focal seal zone around the<br>circumference of all anastomoses.<br>• All animals appeared normal at routine<br>health assessments.<br>• One unexpected death occurred due to<br>anesthetic complications that occurred<br>immediately post-sedation.<br>• 7/10 fistulae (4/5 created using final<br>device design) were patent at 7-day<br>follow-up. Loss of patency was due to<br>flow reduction due to thrombosis at<br>fistula, though the adjacent vein and<br>artery were patent in all cases.<br>• There was no evidence of excessive<br>submural bleeding, aneurysm, or loss of<br>vessel approximation.<br>• The fistula sizes and flow rates for both<br>device designs were acceptable at 30<br>days.<br>• Gross pathology supported the<br>ultrasound findings.<br>• Thermal necrosis, granulation tissue,<br>mineralization and neointimal<br>hyperplasia were localized to the tissue<br>site as expected.<br>• Vascular healing appeared to be normal<br>for this type of device.<br>• No evidence of safety concerns, such as<br>thrombo-emboli, was observed in<br>downstream organs. |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | created with the previous device<br>design to determine<br>equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Evaluate the ability to<br>create an endovascular<br>fistula, acute<br>performance, and<br>chronic safety in an<br>animal model using an<br>earlier design version<br>of the device<br>compared to surgical<br>fistula creation.<br>Following the study,<br>no major changes were<br>made to the electro-<br>mechanical system<br>involved in vessel<br>access or energy<br>delivery.                                                                                                                                     | • Non-GLP<br>• N=6 ovine models<br>• Insert, track, deploy, and<br>activate device to create a<br>fistula in one of the femoral<br>arteries of each animal. A<br>surgical fistula was<br>subsequently created in the<br>other femoral artery of each<br>animal.<br>• Immediately after fistula<br>creation, fistulae were evaluated<br>by ultrasound to determine<br>fistula diameter, patency, and<br>fistula flow.<br>• At 15 and 30 days post-<br>procedure, ultrasound was used<br>to evaluate fistula diameter,<br>patency, and fistula flow.<br>• Gross pathological examination<br>and histopathological<br>examination performed for N=3<br>animals at 15 days and N=3<br>animals at 30 days to evaluate<br>safety and healing response.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | • Procedural performance was acceptable<br>for all devices.<br>• 5/6 fistulae were successfully created.<br>One fistula separated five minutes after<br>the procedure, but no intervention was<br>required and no clinical sequelae were<br>observed.<br>• All successfully created fistulae were<br>patent at all follow-up evaluations, with<br>acceptable fistula flow.<br>• No adverse clinical events were observed<br>in any animals.<br>• No evidence of chronic tissue damage<br>was observed.<br>• No excessive inflammatory response,<br>delayed healing or stenosis was<br>observed.<br>• Fistula performance appeared to be<br>acceptable and comparable to the<br>surgical control fistulae.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

Table 4: Summary of Preclinical Studies

{9}------------------------------------------------

# SUMMARY OF CLINICAL INFORMATION

The Ellipsys System was primarily supported by a pivotal study entitled the "Ellipsys Vascular Access System Clinical Trial." Details of the study design and selected clinical results are provided below.

Purpose: To demonstrate the safety and effectiveness of the Ellipsys Vascular Access System in creating a native AV fistula via percutaneous access to support future hemodialysis.

Design: The study was a non-randomized, prospective multi-center study conducted in the United States, in which a total of 117 subjects from 5 sites were enrolled to undergo creation of an AV fistula using the Ellipsys Vascular Access System. Patients who met the study criteria and had suitable vascular anatomy, as confirmed by pre-procedure ultrasound vascular mapping, were enrolled in the study.

Follow-up visits were scheduled at 24 hours, 1 weeks, 90 days, 6 months and 1 year using the following standard of care assessments: vital signs, upper extremity vessel dimensions by ultrasound, post-procedure patency and flow rates, and adverse events.

{10}------------------------------------------------

All study data were monitored by an independent study monitor, and adverse events were adjudicated by an independent medical monitor.

# Primary Endpoints:

- 1) Safety: The primary safety endpoint was the percent of subjects showing at least one of the following device-related serious adverse events (SAEs): vessel perforation, vessel dissection, and electrical shock during index; and embolization in a previously uninvolved arterial territory within 90 days (+/- 14 days) of the index procedure.
The following definitions were used for the safety analyses in the study:

- Adverse event (AE): any undesirable experience associated with the use of a medical product in a patient.
- Serious adverse event (SAE): an AE or any untoward medical consequence that meets ● any of the following criteria:
	- o Results in death.
	- Is life-threatening (i.e., an AE that, in the opinion of the investigator, places o the subject at immediate risk of death).
	- Requires in-patient hospitalization or prolongs an existing hospitalization O (hospitalizations for scheduled medical or surgical procedures to conduct scheduled treatments or routine examinations do not meet this criterion).
	- Results in persistent or significant disability or incapacity. о
	- Is a congenital anomaly or birth defect in the infant/newborn of a mother that O has been exposed to the research device.
	- It is considered by the investigator to be a significant medical event based on o medical criteria (e.g., puts the subject at risk or may require medical or surgical intervention to prevent one of the aforementioned results).
- . Device-related adverse event: AE that is directly attributable to the use of the device.
- . Procedure-related adverse event: AE that is directly related to accessing the vessel prior to introducing the study device and creating the AV fistula.
- 2) Effectiveness: The primary effectiveness endpoint was the maturation success rate at the 90day follow-up visit, with maturation defined as an access site intended for dialysis needle cannulation with venous diameter > 4 mm, blood flow > 500 ml/min.

Secondary Endpoints: Secondary endpoints consisted of the following:

- . Device Success Rate: Percent of Ellipsys Vascular Access Systems that successfully created an AVF upon deployment.
- . Technical Success Rate: Percent of access sites that demonstrated physical exam patency through clinic discharge:

{11}------------------------------------------------

- Rapid Maturation Rate: Percent of access sites that could sustain three 2-needle . cannulations at the prescribed needle gauge and blood flow rate (Ob) between the 4 week and 90 day follow-up visits after initial AV fistula creation.
- Assisted Maturation Rate: Percent of access sites which achieved or maintained . maturation following intervening manipulations (surgical or endovascular) designed to promote or reestablish maturation.
- . Unassisted Maturation Rate: Percent of access sites that achieved and maintained maturation without any surgical or endovascular intervention designed to promote or reestablish maturation.
- . Assisted Patency Rate: Percent of access sites which were patent after intervening manipulations (surgical or endovascular) intended to promote or reestablish patency.
- . Unassisted Patency Rate: Percent of access sites that maintained patency without any surgical or endovascular intervention designed to maintain or reestablish blood flow in the access site.
- . Intervention Rate: Percent of subjects who had one or more surgical or endovascular interventions to maintain or reestablish blood flow in the access site
- . Transposition Rate: Percent of subjects who required one or more surgical transpositions performed to facilitate needle access.

Other Outcome Measures: Other outcome measures included the following:

- Time to First Cannulation: Elapsed time to first use of access site.
- . Ultrasound Flow: Flow rate of blood through outflow vein(s).
- Time to Access Site Abandonment: Elapsed time to abandonment of the access site. ●
- Catheter Utilization: Total number of days a catheter was used before access site maturation per subject.
- 6 Month Patency and Flow: Patency and flow at 6 months following AVF creation. .
- 1 Year (12 Month) Patency and Flow: Patency and flow at 12 months following AVF . creation.

In addition to the secondary endpoints and other outcome measures, additional analyses that were not pre-specified in the clinical protocol were performed and are discussed in the results section below.

Eligibility Criteria Summary: The study population consisted of male and female patients from the United States, at least 18 years of age and no more than 80 years of age.

Kev inclusion criteria included the following:

{12}------------------------------------------------

- . Diagnosed with end-stage renal disease (ESRD) or chronic kidney disease requiring dialysis or anticipated start of dialysis within 6 months of enrollment.
- . Patients deemed medically eligible for upper extremity autogenous AV fistula creation, per institutional guidelines and/or clinical judgment.
- Arterial lumen diameter of ≥ 2.0mm and adjacent vein diameter of ≥ 2.0mm at the target anastomosis site.
- . Radial artery-adjacent vein proximity ≤ 1.5 mm measured lumen edge-to-lumen edge as determined by pre-procedural ultrasound and confirmed pre-procedure.
- Adequate collateral arterial perfusion and vein quality based on pre-operative assessment. . without evidence of subclavian artery stenosis on the ipsilateral side.
- Negative Allen's Test for ulnar artery insufficiency. .

Key exclusion criteria included the following:

- Patient required creation of an arteriovenous fistula distal to the wrist. ●
- . Documented or suspected central venous stenosis including upper extremity arterial stenosis (≥ 50%).
- Prior vascular surgery at or proximal (central) to the AVF target site.
- Prior axillary dissection or mastectomy on the ipsilateral side as intended AV fistula site. ●
- Evidence of vascular disease at the radial artery/adjacent vein site on the ipsilateral side. ●
- History of steal syndrome from a previous hemodialysis vascular access on the ipsilateral . side which required intervention or abandonment.

Accountability: Figure 6 below shows the disposition of all subjects through the end of the study. In the figure, long term follow-up (LTFU) refers to the 6- and 12-month study visits.

Note: Study subjects were not consented for LTFU until after a large percentage of the study subjects had completed their 90-day study visit and some had already passed the window for their 6-month visit. As a result, 6 subjects declined to re-consent to rejoin the study for LTFU and 34 subjects completed the 12-month study visit but not the 6-month study visit.

Patients were not included in the ITT population for the following reasons:

- Roll-ins the first two patients at each clinical site who were included to facilitate in . vivo training on the device and the specific vascular anatomy.
- . Access failures - the device could not be introduced into the vessel or vessel spasm precluded continuation of the procedure.

Patients in the ITT population exited the study for the following reasons:

- . No AVF created – an AVF was attempted during the index procedure but could not be successfully created.
- Total occlusions three subjects whose access sites were abandoned within the first 30 . days due to total occlusion of the anastomosis which could not be re-opened.

{13}------------------------------------------------

- . Death - one death was reported at the 90-day visit, two deaths were reported at the 6month visit, and five deaths were reported at the 12-month visit. Among the seven deaths that occurred after 90 days, five subjects had not consented to LTFU while two subjects had consented for LTFU.
- Lost to follow-up the patient failed to return for follow-up visits, and documented . attempts to locate and contact the subject were unsuccessful. A patient who missed a study visit but attended a subsequent visit was no longer considered lost to follow-up. A missed visit was considered a protocol deviation and the deviation was documented and reported.
- . Other - subjects who terminated the study early for various other reasons such as declining to re-join the study after completing the 90-day visit, subjects who were using other means of vascular access, or experienced an SAE unrelated to the study device or procedure.

Image /page/13/Figure/3 description: This image is a flowchart showing the enrollment and follow-up of subjects in a study. The flowchart starts with 117 subjects enrolled, then 10 roll-ins and 4 access failures. The ITT population is 103, and the potential LTFU is 98. 84 subjects consented to LTFU, and 77 completed the 12-month visit.

Figure 6: Subject Disposition Through End of Study

{14}------------------------------------------------

Demographics: The total Intent-to-Treat (ITT) population consisted of 103 patients. Information on the patient demographics is provided in Table 5 below.

| Attribute                                             |                                                       | ITT Patients<br>(N=103) |
|-------------------------------------------------------|-------------------------------------------------------|-------------------------|
| Age (mean, yrs)                                       |                                                       | 56.6                    |
| BMI (mean, kg/m²)                                     |                                                       | 31.2                    |
| Sex (%)                                               | Male                                                  | 73.8                    |
|                                                       | Female                                                | 26.2                    |
| Race (%)                                              | White                                                 | 74.8                    |
|                                                       | Black                                                 | 20.4                    |
|                                                       | Asian                                                 | 1.9                     |
|                                                       | Other                                                 | 2.9                     |
| Ethnicity (%)                                         | Hispanic/Latino                                       | 35.9                    |
|                                                       | Not Hispanic or Latino                                | 63.1                    |
| Diabetes (%)                                          | Type I                                                | 4.9                     |
|                                                       | Type II                                               | 61.2                    |
| Hypertension (%)                                      |                                                       | 98.1                    |
| Vascular Disease other than Target Site (%)           |                                                       | 34.0                    |
| Relevant Previous<br>Surgical Procedures<br>(%)       | Prior Failed Surgical<br>AVF on Contralateral<br>Side | 6.8                     |
|                                                       | Prior Nephrectomy                                     | 4.9                     |
|                                                       | Other Relevant Surgical<br>Procedure                  | 21.4                    |
| Catheter Present for Hemodialysis at Screening<br>(%) |                                                       | 62.1                    |

Table 5: Demographic Characteristics of ITT Patients

Table 6 below presents the pre-procedure vessel mapping results by ultrasound in the ITT population, and Table 7 shows the vessel locations used in the study. Table 8 presents the patients' dialysis status at screening and end of study.

Table 6: Pre-Procedure Ultrasound Vessel Characteristics, ITT Population

| Attribute                                       | # of ITT Patients (N=103) |
|-------------------------------------------------|---------------------------|
| Proximal Radial Artery Diameter (mean ± SD, mm) | $3.08 \pm 0.62$           |
| Perforating Vein Diameter (mean ± SD, mm)       | $3.48 \pm 0.88$           |
| Artery-Vein Proximity (mean ± SD, mm)           | $0.65 \pm 0.48$           |

Table 7: Target Vessel Locations, Completed Cases Population

| Target Vessel | # of CC Patients (N=90) |
|---------------|-------------------------|
| Cephalic Vein | 68 (75.6%)              |
| Basilic Vein  | 20 (22.2%)              |

{15}------------------------------------------------

| Brachial Vein 1 | 1 (1.1%) |
|-----------------|----------|
| Brachial Vein 2 | 1 (1.1%) |

Table 8: Dialysis Status at Screening and End of Study

|                                              | Dialysis Status           |                         |
|----------------------------------------------|---------------------------|-------------------------|
| Time Period (Analysis Population)            | Pre-<br>Dialysis<br>N (%) | On<br>Dialysis<br>N (%) |
| Screening (ITT Population, N=103)            | 39 (37.9)                 | 64 (62.1)               |
| Screening (Long Term Visit Population, N=81) | 28 (34.6)                 | 53 (65.4)               |
| 12 Months (Long Term Visit Population, N=81) | 14 (17.3)                 | 67 (82.7)               |

Results: The principal safety and effectiveness results from patients in the study are provided below. The primary endpoints were evaluated for the Intent-to-Treat (ITT) population.

Primary Safety Endpoint: The primary safety endpoint of the study was met as none of the subjects experienced one or more of the following device-related SAEs: vessel perforation, vessel dissection, electrical shock during index, and significant embolization in a previously uninvolved arterial territory within 90 days of the procedure. A hypothesis test was not used for the primary safety endpoint.

| Safety Endpoint                                             | # of Subjects | % of Subjects | 95%<br>Confidence<br>Interval |
|-------------------------------------------------------------|---------------|---------------|-------------------------------|
| Vessel perforation during index<br>procedure                | 0             | 0.0%          | 0.0%, 3.5%                    |
| Vessel dissection during index<br>procedure                 | 0             | 0.0%          | 0.0%, 3.5%                    |
| Electrical shock during index<br>procedure                  | 0             | 0.0%          | 0.0%, 3.5%                    |
| Embolization in previously<br>uninvolved arterial territory | 0             | 0.0%          | 0.0%, 3.5%                    |
| Composite Safety Endpoint                                   | 0             | 0.0%          | 0.0%, 3.5%                    |

Table 9: Primary Safety Endpoint Results - 90-Day Device Related SAE Rate

Primary Effectiveness Endpoint: The primary effectiveness endpoint was the maturation success rate at the 90 day follow-up visit, with maturation defined as an access site intended for dialysis needle cannulation with venous diameter > 4 mm and blood flow > 500 ml/min, as measured by duplex ultrasound.

{16}------------------------------------------------

The primary effectiveness endpoint was tested for the ITT population using a one-sample hypothesis comparing the Ellipsys Vascular Access System (Test) to a Performance Goal of 49%. The point estimate for the historical performance goal was derived from a meta-analysis of eight previous studies of open surgical AVF procedures. The null and alternative statistical hypotheses were:

Ho: PTest < 49% VS. HA: PTest > 49%,

where Press was the maturation success rate in the Test group. This hypothesis was tested with a one-sided binomial test. The null hypothesis was tested using a one-sided significance level of 0.025. A supportive analysis of the primary endpoint was calculated as the lower one-sided 97.5% exact binomial confidence limit based on Clopper-Pearson methods for the ITT population. Additionally, supportive analyses for the primary effectiveness endpoint were done using the same methodology for the As Treated (AT) and Completed Cases (CC) populations.

The primary endpoint was met with a 90-day maturation success rate of 89.3%.

The statistical analysis populations used in the study were defined as follows:

- . Intent-to-Treat (IT): The ITT population included all subjects enrolled following the roll-in stage in whom the procedure had been attempted and the study device deployed, whether successful or not.
- As-Treated (AT): The AT population included all subjects in the ITT population where . the study device was successfully deployed in the target vasculature and an AVF was created.
- Completed Cases (CC): The CC population included all subjects in the ITT population . with a patent AVF completing the 90-day evaluation.

| Analysis<br>Population | Successes (N) | Total (N) | Percent (%) | Lower<br>97.5%<br>Confidence<br>Interval | p-value |
|------------------------|---------------|-----------|-------------|------------------------------------------|---------|
| ITT                    | 92            | 103       | 89.3%       | 81.7%                                    | <0.0001 |
| AT                     | 92            | 102       | 90.2%       | 82.7%                                    | <0.0001 |
| CC                     | 86            | 90        | 95.6%       | 89.0%                                    | <0.0001 |

Table 10: Primary Effectiveness Endpoint Results - 90-Day Maturation Success Rate

Secondary Endpoints: A summary of the secondary endpoint analyses are provided below. All analyses presented below are based on the ITT population.

{17}------------------------------------------------

| Adverse<br>Events<br>Relatedness              | SAEs Through 90 Days    |                           |                    | SAEs Through 12 Months  |                           |       |
|-----------------------------------------------|-------------------------|---------------------------|--------------------|-------------------------|---------------------------|-------|
| # of<br>Events (N)                            | # of<br>Subjects<br>(N) | Percent<br>(N=103)<br>(%) | # of<br>Events (N) | # of<br>Subjects<br>(N) | Percent<br>(N=103)<br>(%) |       |
| Related to<br>Device Only                     | 0                       | 0                         | 0.0%               | 0                       | 0                         | 0.0%  |
| Related to<br>Procedure<br>Only               | 1                       | 1                         | 1.0%               | 2¹                      | 2                         | 1.9%  |
| Related to<br>Both Device<br>and<br>Procedure | 0                       | 0                         | 0.0%               | 0                       | 0                         | 0.0%  |
| Not Related<br>to Device or<br>Procedure      | 18                      | 17                        | 16.5%              | 76                      | 40                        | 38.8% |

Table 11: Secondary Safety Endpoint Results - SAEs Through 12 Month Follow-Up

The two SAEs adjudicated as related or possibly related to the procedure included a reaction 1. to pre-procedure medication and a dialysis catheter infection.

| Secondary<br>Endpoint            | Successes (N) | Total (N) | Percent (%) | 95%<br>Confidence<br>Interval |
|----------------------------------|---------------|-----------|-------------|-------------------------------|
| Device<br>Success                | 102           | 103       | 99.0%       | [94.7, 100.0]                 |
| Technical<br>Success             | 98            | 103       | 95.1%       | [89.0, 98.4]                  |
| Intervention<br>Rate             | 98            | 103       | 95.1%       | [89.0, 98.4]                  |
| Transposition<br>Rate            | 19            | 103       | 18.4%       | [11.5, 27.3]                  |
| Rapid<br>Maturation<br>Rate      | 29            | 103       | 28.2%       | [19.7, 37.9]                  |
| Assisted<br>Maturation<br>Rate   | 85            | 103       | 82.5%       | [73.8, 89.3]                  |
| Unassisted<br>Maturation<br>Rate | 1             | 103       | 1.0%        | [0.0, 5.3]                    |
| Assisted<br>Patency Rate         | 89            | 103       | 86.4%       | [78.2, 92.4]                  |
| Unassisted<br>Patency Rate       | 1             | 103       | 1.0%        | [0.0, 5.3]                    |

Table 12: Secondary Effectiveness Endpoint Results - 90 Day Follow-Up

{18}------------------------------------------------

Adverse Events: The adverse events (AEs) presented in Table 13 were observed in the study through 90 days and 12 months. The data represent all AEs that were observed during the study, regardless of seriousness or relation to the device and/or procedure. All events were adjudicated by an independent Medical Monitor.

A total of 25 occlusion events were observed in 19 subjects (17.5%) throughout the study. For 15 of the 19 subjects (78.9%), the occlusion event occurred within 24 hours of the study procedure. There was no propagation of thrombus to adjacent vessels. In all cases, this was not a lifethreatening or serious event that required hospitalization, and there was no incapacity or disability. In all but two of the acute thrombosis cases, flow was restored quickly the same day or the next day using standard percutaneous techniques and overall 20/25 (80%) of the occlusion events were successfully resolved. Though 5/25 (20%) of the AVFs that remained occluded after treatment had minor associated thrombosis, the primary mechanism for occlusion was attributed to vessel spasm.

One death was reported at the 90-day visit, two deaths were reported at the 6-month visit, and five deaths were reported at the 12-month visit.

| Table 13: Adverse Events Summary (111 Population, N=103) |              |                                                      |              |                                                        |
|----------------------------------------------------------|--------------|------------------------------------------------------|--------------|--------------------------------------------------------|
| Event Type                                               | Total Events | Cumulative Through 90 Days<br># of Subjects w/ Event | Total Events | Cumulative Through 12 Months<br># of Subjects w/ Event |
| Occlusion of anastomosis1                                | 18           | 18 (17.5%)                                           | 25           | 19 (18.4%)                                             |
| Total occlusion                                          | 12           | 12 (11.7%)                                           | 15           | 13 (12.6%)                                             |
| Partial occlusion                                        | 6            | 6 (5.8%)                                             | 10           | 8 (7.8%)                                               |
| Stenosis at the AV anastomosis                           | 8            | 7 (6.8%)                                             | 9            | 8 (7.8%)                                               |
| Death2                                                   | 1            | 1 (0.9%)                                             | 8            | 8 (7.8%)                                               |
| Hematoma                                                 | 6            | 6 (5.8%)                                             | 8            | 8 (7.8%)                                               |
| Infection of vascular access                             | 2            | 2 (1.9%)                                             | 8            | 5 (4.9%)                                               |
| Infection unrelated to vascular access<br>site3          | 2            | 2 (1.9%)                                             | 8            | 6 (5.8%)                                               |
| Infiltration                                             | 3            | 2 (1.9%)                                             | 7            | 6 (5.8%)                                               |
| Volume overload                                          | 0            | 0 (0.0%)                                             | 7            | 5 (4.9%)                                               |
| Venous stenosis                                          | 5            | 4 (3.9%)                                             | 7            | 5 (4.9%)                                               |
| Coronary artery disease                                  | 0            | 0 (0.0%)                                             | 5            | 3 (2.9%)                                               |
| Anemia                                                   | 0            | 0 (0.0%)                                             | 5            | 5 (4.9%)                                               |
| Bleeding unrelated to access site                        | 3            | 3 (2.9%)                                             | 4            | 4 (3.9%)                                               |
| Bleeding at access site                                  | 3            | 2 (1.9%)                                             | 3            | 2 (1.9%)                                               |
| End-stage kidney disease                                 | 2            | 2 (1.9%)                                             | 3            | 3 (2.9%)                                               |
| Myocardial infarction                                    | 0            | 0 (0.0%)                                             | 2            | 2 (1.9%)                                               |
| Congestive heart failure                                 | 1            | 1 (0.9%)                                             | 2            | 2 (1.9%)                                               |
| Pulmonary embolism                                       | 1            | 1 (0.9%)                                             | 2            | 2 (1.9%)                                               |
| Stroke                                                   | 0            | 0 (0.0%)                                             | 1            | 1 (0.9%)                                               |
| Steal syndrome4                                          | 0            | 0 (0.0%)                                             | 1            | 1 (0.9%)                                               |
| Embolization coil migration5                             | 1            | 1 (0.9%)                                             | 1            | 1 (0.9%)                                               |
| Arterial extravasation                                   | 1            | 1 (0.9%)                                             | 1            | 1 (0.9%)                                               |

Table 13: Adverse Events Summary (ITT Population, N=103)

{19}------------------------------------------------

| Deep vein pseudo-aneurysm | 1  | 1 (0.9%)   | 1  | 1 (0.9%)   |
|---------------------------|----|------------|----|------------|
| Failure to create an AVF  | 1  | 1 (0.9%)   | 1  | 1 (0.9%)   |
| Seroma                    | 0  | 0 (0.0%)   | 1  | 1 (0.9%)   |
| Mild paresthesia          | 1  | 1 (0.9%)   | 1  | 1 (0.9%)   |
| Other6                    | 19 | 12 (11.7%) | 66 | 32 (31.1%) |

1. 15/18 (83.3%) of occlusions at 90 days were successfully resolved. 20/25 (80.0%) of occlusions at 12 months were successfully resolved. Two subjects experienced a partial occlusion and a total occlusion, and are counted in the number of subjects for both types.

- 2. No deaths were adjudicated as related to the study device or procedure.
- 3. Infections unrelated to the access site included a toe infection (1), ischemic infection (1), AICD lead infection (1), hernia repair infection (1), infection of surgical wound (1), infected peritoneal dialysis catheter (1), and unspecified infections (2).
- 4. Steal syndrome related to surgical transposition error was successfully resolved.
- 5. Coil migrated to lungs during a maturation assistance procedure 28 days post-index procedure. AE was resolved and was asymptomatic.
- 6. Types and numbers of Other adverse events included: osteomyelitis (6), cellulitis (6), hyperkalemia (5), fall resulting in fracture/hematoma (4), hypoglycemia (3), pain (3), atrial flutter (2), COPD/asthma (2), clostridium difficile (2), incident hypertension, atrial thrombus, atrial fibrillation, cardiac A/V block, hypotension, cancer, aortic dissection, respiratory failure, pneumonia, bacteremia, tooth abscess, discitis, necrotic cholecystitis, peritonitis, gallstone pancreatitis, DIC, allergic reaction to contrast, hematoma at lower extremity, neurological reaction to under-dialysis, epistaxis, musculoskeletal injury, cephalic vein occlusion, hypotension, DKA, PAD ischemia, apnea, intubations, elevated INR, diarrhea, herpes, low back pain, and nausea/vomiting.

Other Outcome Measures: Other selected outcome measures are reported below.

### Study AVF Utilization

Table 14 below presents the total number of subjects who received 2-needle cannulation of their study AVF, the time to first 2-needle cannulation for hemodialysis, and estimated duration of AVF use in ITT subjects who were eligible for hemodialysis. Results are also stratified by dialysis status at the time of study enrollment (i.e. pre-dialysis or already on dialysis). Subjects were considered eligible for hemodialysis if they were recorded as being on hemodialysis, regardless of access site. Three (3) subjects who were on peritoneal dialysis but not on hemodialysis at the 6- and 12-month visits were not considered eligible for hemodialysis.

There were 71 subjects who achieved 2-needle cannulation out of a total 81 ITT subjects (87.7%) eligible for hemodialysis at any time during the study.

AVF time used was estimated using the first recorded cannulation date and last recorded dialysis date using the study AVF for each subject at their final visit. AVF time used was not estimated for eight (8) ITT subjects who had undergone 2-needle cannulation because six (6) subjects did not attend any long-term (6- or 12-month) visits, and two (2) subjects reverted to central venous catheter (CVC) dialysis.

Of the 102 subjects who had an AVF created during the study, seven (7) access sites (6.9%) were abandoned after a mean of 101 days. The main reasons for abandonment were total occlusion of 

{20}------------------------------------------------

the anastomosis or a target vessel that could not be reopened or central stenosis impeding the venous outflow.

| Parameter                                                     | Dialysis<br>Status at<br>Study<br>Enrollment | # of<br>Subjects<br>Evaluated<br>(N) | Mean<br>(days) | Standar<br>d<br>Deviation<br>(days) | Median<br>(days) | Min-Max<br>(days) |
|---------------------------------------------------------------|----------------------------------------------|--------------------------------------|----------------|-------------------------------------|------------------|-------------------|
| Time to First 2-<br>Needle<br>Cannulation for<br>Hemodialysis | Pre-Dialysis                                 | 16                                   | 162.9          | 86.6                                | 140.5            | 53-345            |
|                                                               | On Dialysis                                  | 55                                   | 100.2          | 51.9                                | 82.0             | 34-224            |
|                                                               | All<br>Evaluable<br>Subjects                 | 71                                   | 114.3          | 66.2                                | 88.0             | 34-345            |
| Estimated # of<br>Study Days AVF<br>Used for<br>Hemodialysis  | Pre-Dialysis                                 | 15                                   | 184.5          | 99.6                                | 203.0            | 10-319            |
|                                                               | On Dialysis                                  | 48                                   | 242.0          | 64.4                                | 247.5            | 84-471            |
|                                                               | All<br>Evaluable<br>Subjects                 | 63                                   | 228.3          | 77.4                                | 242.0            | 10-471            |
| Time to Access<br>Site<br>Abandonment1                        | N/A                                          | 7                                    | 101.4          | 125.5                               | 56.0             | 1-355             |

Table 14: Study AVF Utilization for Hemodialysis - 12 Months

1. Time to abandonment only calculated from the seven subjects whose access sites were abandoned.

Table 15 below shows CVC usage during the study. Of the 62 subjects who were on hemodialysis using the study AVF at the end of the study, nine subjects (14.5%) had no history of CVC use and went directly to hemodialysis using the Ellipsys AVF. A total of 48 subjects (77.4%) on hemodialysis with the study AVF at the end of the study had been on dialysis using a CVC at screening with a mean usage of 134 days before removal. Five subjects (8.1%) on hemodialysis with the study AVF at the end of the study required a CVC to start hemodialysis during the study, and all were initiated prior to 90 days and prior to maturation of the study AVF, with a mean CVC usage of 93.4 days.

| Hemodialysis<br>Status at 12<br>Months                   | Parameter                           | # of Subjects<br>Who Had<br>CVC Present<br>at Screening | # of Subjects<br>Who Had<br>CVC Initiated<br>During Study | No CVC Use<br>During Study |
|----------------------------------------------------------|-------------------------------------|---------------------------------------------------------|-----------------------------------------------------------|----------------------------|
| Subjects on<br>hemodialysis<br>using study<br>AVF (N=62) | # of Subjects                       | 48 (77.4%)                                              | 5 (8.1%)                                                  | 9 (14.5%)                  |
|                                                          | Mean (Median)<br>(study days used)2 | 134.4 (117.0)                                           | 93.4 (110.0)                                              | N/A                        |
|                                                          | SD<br>(study days used)             | 65.1…

---

**Source:** [https://fda-staging.innolitics.com/device/DEN170004](https://fda-staging.innolitics.com/device/DEN170004)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a De Novo](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).