DEN140040 · Jan Medical, Inc. · POP · Aug 1, 2016 · Neurology
Device Facts
Record ID
DEN140040
Device Name
Nautilus BrainPulse 1000
Applicant
Jan Medical, Inc.
Product Code
POP · Neurology
Decision Date
Aug 1, 2016
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 882.1630
Device Class
Class 2
Indications for Use
The BrainPulse is intended for use on a patient’s head to non-invasively detect, amplify and capture the skull motion caused by pulsatile flow from the cardiac cycle. The BrainPulse is not indicated to aid in the diagnosis of neurological conditions, diseases, or disorders.
Device Story
BrainPulse Model 1100 measures cranial pulsatile movements via headset-mounted accelerometers (0.001–0.03 g range) and a forehead PPG sensor; includes ambient noise (SPL) sensor. System comprises headset, data collector, and computer. Data collector converts analog sensor signals to digital stream via Ethernet to computer. Software initiates/ends recordings and saves data files for third-party post-hoc review; no onboard data visualization. Used in clinical settings by trained technicians/clinicians. Provides objective physiological data on skull motion correlated with cardiac cycle; assists clinicians in observing measurement changes for clinical assessment. Benefits include non-invasive, stable, and repeatable capture of cranial motion.
Clinical Evidence
Six clinical studies (273 patients, 616 recordings) evaluated device ability to measure cranial motion due to cardiac pulsatile flow. Results demonstrated skull motion correlated with regular pulse related to cardiac cycle. Within-patient recordings showed no major variations. Two patients reported discomfort; no other adverse events reported. Bench testing confirmed stability, repeatability, and resolution above noise floor.
Technological Characteristics
Utilizes accelerometers to measure skull motion/acceleration. Requires biocompatibility for patient-contacting parts; electrical, thermal, and mechanical safety testing; electromagnetic compatibility (EMC) testing. Hardware and software must be fully characterized with verification, validation, and hazard analysis.
Indications for Use
Indicated for use on patients to non-invasively detect, amplify, and capture skull motion resulting from cardiac cycle pulsatile flow. Not indicated for diagnosis of neurological conditions, diseases, or disorders.
Regulatory Classification
Identification
A cranial motion measurement device is a prescription device that utilizes accelerometers to measure the motion or acceleration of the skull. These measurements are not to be used for diagnostic purposes.
Special Controls
In combination with the general controls of the FD&C Act, the Cranial Motion Measurement Device is subject to the following special controls:
- 1. The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
- a. Hardware specifications must be provided. Additionally, verification and validation testing as well as a hazard analysis must be performed.
- b. Software must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Additionally, software verification and validation testing as well as a hazard analysis must be performed.
- 2. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 3. The device must be designed and tested for electrical, thermal and mechanical safety and electromagnetic compatibility (EMC).
- 4. Clinical performance testing must demonstrate the accuracy, precision, stability, and repeatability of measuring cranial motion per the intended use in the intended use environment.
- 5. The labeling must include:
- The intended use population and the intended use environment. a.
- b. Any instructions technicians should convey to patients regarding the collection of cranial acceleration data to ensure device measurement accuracy, precision, stability, and repeatability.
- c. Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement.
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
(i) Hardware specifications must be provided. Additionally, verification and validation testing as well as a hazard analysis must be performed.
(ii) Software must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Additionally, software verification and validation testing as well as a hazard analysis must be performed.
(2) The device parts that contact the patient must be demonstrated to be biocompatible.
(3) The device must be designed and tested for electrical, thermal, and mechanical safety, and electromagnetic compatibility (EMC).
(4) Clinical performance testing must demonstrate the accuracy, precision, stability, and repeatability of measuring cranial motion per the intended use in the intended use environment.
(5) The labeling must include:
(i) The intended use population and the intended use environment.
(ii) Instructions for technicians to convey to patients regarding the collection of cranial acceleration data to ensure device measurement accuracy, precision, stability, and repeatability.
(iii) Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR BRAINPULSE, MODEL 1100
### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Cranial Motion Measurement Device. A cranial motion measurement device is a prescription device that utilizes accelerometers to measure the motion or acceleration of the skull. These measurements are not to be used for diagnostic purposes.
NEW REGULATION NUMBER: 21 CFR 882.1630
CLASSIFICATION: CLASS II
PRODUCT CODE: POP
#### BACKGROUND
DEVICE NAME: BRAINPULSE, MODEL 1100
SUBMISSION NUMBER: DEN140040
DATE OF DE NOVO: DECEMBER 23, 2014
- CONTACT: JAN MEDICAL, INC 110 Pioneer Way SUITE L MOUNTAIN VIEW, CA 94041
#### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
#### INDICATIONS FOR USE
The BrainPulse is intended for use on a patient's head to non-invasively detect, amplify and capture the skull motion caused by pulsatile flow from the cardiac cvcle. The BrainPulse is not indicated to aid in the diagnosis of neurological conditions, diseases, or disorders.
#### LIMITATIONS
For prescription use only.
The BrainPulse is not indicated for use on patients with a wound close to a sensor location.
The position of the subject has an impact on the BrainPulse recording. The recording may
{1}------------------------------------------------
be affected by whether the subject is sitting upright or is supine, whether the head is facing forward or to the side. Changes to the position cause changes to blood flow and this changes brain motion. As a result, if multiple recordings are obtained from the same subject, a consistent position between recordings will lead to more comparable recordings.
Patient respiration rate, heart rate, and head and body motion also may cause variation in the recording. The presence of atrial fibrillation, premature ventricular contractions, stroke, aneurysms, or intracranial bleeding in patients may also cause variations in the recording.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
As described above, the BrainPulse, Model 1100 (BrainPulse) is designed to measure skull motion caused by pulsatile blood flow. The BrainPulse measures these cranial pulsatile movements via an array of accelerometers placed on the scalp. The system consists of three main components: a headset, data collector, and computer.
Figure 1. BrainPulse system consisting of the data collector and battery, headset containing sensors, and computer.
The headset contains a forehead photoplethysmograph (PPG) sensor that measures the patient's pulse rate, a Sound Pressure Level (SPL) sensor for detecting ambient environment noise, and six accelerometers to detect the acceleration of the skull at six selected locations. These acceleration measurements typically fall in the range of 0.001 - 0.03 g. See Figure 2 below for a graphical representation of the proper accelerometer placement.
{2}------------------------------------------------
Image /page/2/Figure/0 description: The image shows three diagrams of a human skull. The first diagram shows a frontal view of the skull with green markers on the forehead and top of the head. The second diagram shows a side view of the skull with a red marker on the forehead and behind the ear. The third diagram shows a side view of the skull with the brain and blood vessels visible, and a blue marker on the back of the head.
Figure 2. BrainPulse accelerometer placement.
Image /page/2/Picture/2 description: The image shows two views of a white helmet on a white stand with a brown base. The helmet is round and has several attachments on the sides and top. The left view shows the side of the helmet, while the right view shows the back. The helmet appears to be made of plastic and has a smooth surface.
Figure 3. BrainPulse headset viewed from the front (left panel) and from the right (right panel).
Image /page/2/Picture/4 description: In the image, a person is wearing a white plastic headset. The headset has multiple adjustable straps and components that appear to be for medical or research purposes. The person's hair is dark and wet, and the headset is positioned on the back of their head. The background is a room with a desk and some equipment.
Figure 4. BrainPulse headset worn on a patient's head.
The data collector converts the analog signals from the headset sensors and provides a digital data stream via Ethernet cable to the computer. The computer is loaded with software that allows for the user to initiate and end recordings and to manage saved data files. The BrainPulse software is not capable of displaying the recorded data from the headset; rather the data are saved in multiple file formats that can be readily displayed using other third-party software for post-hoc review. Please refer to Figure 5 below for an example of what the acceleration measurements might look like when plotted using a third-party application.
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image shows a graph with two lines, one red and one green, plotted against each other. The x-axis is labeled 'points' and ranges from 0 to 4000, while the y-axis is labeled with a greater than sign '>'. The red and green lines oscillate, with the red line initially dipping below zero and the green line rising above zero. As the 'points' value increases, the oscillations of both lines dampen and converge towards zero.
- Figure 5. Sample plot of accelerometer data recording from the left and right temple accelerometers for a single heartbeat. This waveform is an average of data recorded across 45 heartbeats.
#### SUMMARY OF NONCLINICAL/BENCH STUDIES
The sponsor conducted a series of bench testing to demonstrate that the BrainPulse would perform as anticipated. Please refer to the sub-sections below for a discussion of each nonclinical test performed.
## BIOCOMPATIBILITY/MATERIALS
The BrainPulse contacts the patient's skin for typically no longer than a 30-minute recording session. The BrainPulse is categorized as a surface-contacting device with a limited duration of contact (less than 24 hours). In accordance with ISO 10993-1:2009, Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing within a Risk Management Process, the following biocompatibility testing was conducted on the BrainPulse. The biocompatibility evaluation was deemed adequate.
| Test | Purpose | Results |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Cytotoxicity Evaluation – L929<br>MEM Elution Test/L929 Neutral<br>Red Uptake Test | To assess the biological activity<br>of L-929 mouse fibroblast cells<br>(grown in culture) after exposure<br>to extracts prepared from the<br>completed BrainPulse system | Non-cytotoxic |
| Kligman Maximization Test | To estimate the potential for<br>sensitization of the BrainPulse<br>system extract using the guinea<br>pig as an animal model | Sensitization rate = 0%<br>Sensitization grade = “Weak” |
| Primary Skin Irritation Test | To estimate the potential to<br>produce primary skin irritation<br>after a single topical exposure to<br>the skin of New Zealand White<br>Rabbits | No signs of erythema or edema<br>Considered a Negligible Irritant |
| | | Table 1. Biocompatibility testing completed for the BrainPulse | | |
|--|--|----------------------------------------------------------------|--|--|
| | | | | |
{4}------------------------------------------------
| Test | Purpose | Results |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Intracutaneous Injection Test | To assess the irritating potential<br>of extracts of the BrainPulse<br>system to cause irritation to the<br>exposed part of the body | No difference between mean test<br>article score and mean control<br>score |
## SHELF LIFE/STERILITY
The BrainPulse is not provided sterile nor are any of the components to be sterilized by the end user. Cleaning and maintenance instructions are included in the device labeling.
The BrainPulse and its components do not have a stated shelf life as the products are not provided sterile. However, the device labeling states that the device has an expected service life of 5 years, which is typical for general electronic devices. Based on the nature of the system components, this estimation of product life is acceptable.
### ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY
The BrainPulse was tested in accordance with the following consensus standards and passed the following electromagnetic compatibility (EMC), electrical, mechanical, and thermal safety tests:
| Standard | Name |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC60601-1: 2005 +AM1: 2012 | Medical Electrical Equipment; Part 1: General Requirements for Safety |
| IEC60601-1-2: 2007 | Medical Electrical Equipment; Part 1-2: General Requirements for Safety - Section 2: Collateral standard: Electromagnetic compatibility - Requirements and tests. |
#### Table 2. EMC and electrical safety testing completed for the BrainPulse
#### SOFTWARE
Software for the device consisted of proprietary software. The software is consistent with a 'MODERATE' level of concern, as discussed in the FDA document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005. Appropriate documentation was provided as part of the de novo request.
#### PERFORMANCE TESTING - BENCH
The BrainPulse was subjected to a series of bench tests to assess its functional performance. These tests were performed on a final manufactured product. Table 3 summarizes the testing performed:
| | | | | | | Table 3. Bench testing completed for the BrainPulse |
|--|--|--|--|--|--|-----------------------------------------------------|
|--|--|--|--|--|--|-----------------------------------------------------|
| Test | Purpose | Results |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Accelerometer Measurement<br>Stability and Repeatability | To ensure acceleration<br>measurements made by the<br>BrainPulse device are stable<br>within a typical recording session | Passed; all within-session<br>recording segments demonstrated<br>stable correlation with a baseline<br>recording. ANOVA results did |
{5}------------------------------------------------
| Test | Purpose | Results |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Accelerometer Resolution | and are repeatable across multiple<br>recording sessions and multiple<br>operators<br>To ensure the expected changes<br>in acceleration measured by the<br>BrainPulse device are adequately<br>resolved by the BrainPulse and<br>are above the observed noise<br>floor | not demonstrate variation in the<br>recordings across multiple<br>sessions or operators<br>Passed; frequency analysis and<br>visual inspection demonstrate<br>signals of interest are resolved<br>above the observed noise floor<br>that confirms accelerometer<br>specifications |
| PPG Sensor Accuracy and<br>Precision | To ensure the BrainPulse device<br>accurately and precisely measures<br>heartrate based upon changes in<br>blood flow | Passed; visual comparison to<br>concurrent SpO2 (blood oxygen<br>saturation) sensor recordings<br>demonstrate adequate PPG sensor<br>performance |
| Hardware Verification | To ensure the accelerometers<br>adequately measure acceleration<br>across the range of expected<br>values during intended use of the<br>BrainPulse device<br><br>To ensure the Sound Pressure<br>Level (SPL) sensor adequately<br>measures ambient noise across a<br>frequency range relevant to the<br>BrainPulse device and in the<br>intended use environment of the<br>BrainPulse device<br><br>To ensure the Data Collector<br>battery charges and discharges<br>according to specification<br><br>To ensure the tablet system<br>properly interfaces with the Data<br>Collector and properly records<br>sensor data | Passed; calibration of<br>accelerometers confirms<br>operation according to<br>specification<br><br>Passed; successful measurement<br>of test signals in intended use<br>environment, compared to<br>reference measurement device<br><br>Passed; battery operates<br>according to specification<br><br>Passed; tablet passed all<br>functional requirements |
## SUMMARY OF CLINICAL INFORMATION
Summaries of six (6) clinical studies (4 completed or terminated, and 2 ongoing) were supplied to support a determination of a reasonable assurance of the safety and effectiveness of the BrainPulse. These studies were conducted both at centers within and outside the United States. While these clinical studies were performed on a wide variety of patient populations with varying neurological conditions, the BrainPulse has only been evaluated in the context of the indications for use; that is, the device's ability to measure cranial motion due to the pulsatile flow from the cardiac cycle.
In total, 616 successful recordings were evaluated from 273 patients across all six of the studies. All studies demonstrated the measured skull motion correlated with a regular pulse related to the cardiac cycle. Only some of the studies obtained multiple recordings per patient; however, no major variations in within-patient recordings were reported. These results were supported by the
{6}------------------------------------------------
stability and repeatability testing summarized in the "Performance Testing – Bench" section above.
Two patients from one of the studies complained of discomfort when wearing the headset. No other adverse events, complaints, or device issues or malfunctions were reported.
# LABELING
The user manual is consistent with the performance data and covers all the hazards and other clinically relevant information that may impact use of the device. The labeling satisfies the requirements of 21 CFR § 801.109 Prescription devices. The labeling for the BrainPulse includes:
- 1. The intended use population and the intended use environment.
- 2. Instructions technicians should convey to patients regarding the collection of cranial motion data.
- 3. Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement.
Because cranial motion data are currently not typically included in the clinical evaluation of patients, clinicians utilizing the measurement data provided by the BrainPulse should be physicians who have familiarized themselves with the labeling of the BrainPulse.
## RISKS TO HEALTH
Table 4 below identifies the risks to health that may be associated with use of Cranial Motion Measurement Device and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measure |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Adverse Tissue Reaction | Biocompatibility Evaluation<br>Labeling |
| Equipment Malfunction Leading to Injury to<br>User or Patient | Electrical, Mechanical and Thermal Safety<br>Testing<br>Electromagnetic Compatibility Testing<br>Labeling |
| Inaccurate Measurement | Clinical Performance Testing<br>Hardware and Software verification, validation<br>and hazard analysis<br>Electromagnetic Compatibility Testing<br>Labeling |
| Use Error | Hardware and Software verification, validation<br>and hazard analysis<br>Labeling |
Table 4. Identified Risks to Health and Mitigation Measures
{7}------------------------------------------------
## SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the Cranial Motion Measurement Device is subject to the following special controls:
- 1. The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
- a. Hardware specifications must be provided. Additionally, verification and validation testing as well as a hazard analysis must be performed.
- b. Software must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Additionally, software verification and validation testing as well as a hazard analysis must be performed.
- 2. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 3. The device must be designed and tested for electrical, thermal and mechanical safety and electromagnetic compatibility (EMC).
- 4. Clinical performance testing must demonstrate the accuracy, precision, stability, and repeatability of measuring cranial motion per the intended use in the intended use environment.
- 5. The labeling must include:
- The intended use population and the intended use environment. a.
- b. Any instructions technicians should convey to patients regarding the collection of cranial acceleration data to ensure device measurement accuracy, precision, stability, and repeatability.
- c. Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement.
## BENEFIT/RISK DETERMINATION
While the summaries of 6 completed, terminated, or ongoing clinical studies were provided, no formal clinical study was provided, requested, or deemed necessary for the BrainPulse based on the risks for the current indications for use. The probable risks of the device are based on risk analysis, bench and clinical testing that included assessments of accuracy and precision, as well as the stability and repeatability studies described above. Probable device-related adverse events include adverse tissue reaction, equipment malfunction leading to injury of the user or patient, inaccurate measurement, and use error leading to inaccurate measurement or patient discomfort. Based on the nonclinical and clinical information provided, the probability of each of these adverse events is low.
The probable benefits of the device are also based on the clinical summaries, bench testing, as well as the stability and repeatability studies described above. The BrainPulse measures skull motion correlated with the cardiac cycle. This measurement is stable and repeatable under
{8}------------------------------------------------
controlled procedures and in a controlled environment, allowing clinicians to observe notable differences or changes in the measurement that may be incorporated into their clinical assessment paradigm. It should be noted that the clinical data were only reviewed in the context of the indications for use; that is, the device's ability to measure skull motion due to the pulsatile flow from the cardiac cycle. Consequently, a demonstration of clinical diagnostic utility in specific patient populations was not required.
Additional factors to be considered in determining probable risks and benefits for the BrainPulse include:
- . Skull motion is a physiological measurement that is currently not typically included in the clinical evaluation of patients. Diagnostic utility of this additional physiological data has not been established for specific patient populations.
- . Patient respiration rate, heart rate, and head and body motion also may cause variation in the recording. The presence of atrial fibrillation, premature ventricular contractions, stroke, aneurysms or intracranial bleeding in patients may also cause variations in the recording. These sources of variation have been identified through risk analyses and preliminary clinical use; however, additional sources of variability remain unknown.
- Feedback from centers using the device in preliminary clinical studies has been positive in terms of ease of use.
## Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
## Benefit/Risk Conclusion
In conclusion, given the available information above, the data support that for use on a patient's head to non-invasively detect, amplify and capture the skull motion caused by pulsatile flow from the cardiac cycle, the probable benefits outweigh the probable risks for the BrainPulse. The device provides benefit, and the risks can be mitigated by the use of general controls and the identified special controls.
# CONCLUSION
The de novo request for the BrainPulse, Model 1100 is granted and the device is classified under the following:
Product Code: POP Device Type: Cranial Motion Measurement Device Class: Class II Regulation: 21 CFR 882.1630
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.