Retrospective clinical CSF samples; Patient medical records/case report forms (CRFs); Routine clinical diagnostic findings (MRI, bacterial culture, in-house PCR)
Retrospective clinical samples were used to supplement the prospective clinical study to evaluate the sensitivity of the device by comparing results against a comparator method (two PCR/bi-directional sequencing assays).
Retrospective collection of preselected positive samples; Study Period: 2004-2013
Patients with signs and symptoms of HSV CNS infection; Sample Size: 55; Number of Sites: 4
Two PCR/bi-directional sequencing assays
Sensitivity (Positive Percent Agreement)
Indications for Use
The Focus Diagnostics Simplexa™ HSV 1 & 2 Direct is intended for use on the 3M Integrated Cycler instrument for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infections of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infections of the CNS. Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions. The assay is not intended for use as a donor screening test. The assay is for professional use only. The Positive Control is intended to be used as a control with the Simplexa™ HSV 1 & 2 Direct. This control is not intended for use with other assays or systems.
Device Story
Real-time PCR assay for direct amplification, detection, and differentiation of HSV-1 and HSV-2 DNA from unprocessed CSF; no nucleic acid extraction required. System includes Simplexa™ HSV 1 & 2 Direct reagents, 3M Integrated Cycler, and Direct Amplification Disc (DAD). User adds 50 µL reaction mix and 50 µL specimen to DAD wells; instrument performs thermocycling and fluorometric detection. Software automatically calculates results (Detected/Not Detected/Invalid/EC500). Used in clinical laboratories by trained professionals. Output aids diagnosis of CNS HSV infections; negative results do not rule out infection. Enables rapid diagnosis, facilitating prompt initiation of disease-specific antiviral therapy.
Clinical Evidence
Clinical study of 219 CSF samples (prospective and retrospective) compared device to two PCR/bidirectional sequencing assays. Prospective study (n=164): HSV-1 PPA 100% (3/3), NPA 98.8% (159/161); HSV-2 PPA 85.7% (6/7), NPA 99.4% (156/157). Retrospective study (n=55): HSV-1 PPA 100% (13/13), HSV-2 PPA 100% (42/42). Bench testing included LoD, reproducibility, interference, and carry-over studies.
Technological Characteristics
Real-time PCR; targets conserved regions of HSV-1 and HSV-2 DNA polymerase genes. Uses bi-functional fluorescent probe-primers (HSV-1: CFR610; HSV-2: FAM; DNA IC: Q670). Direct Amplification Disc (DAD) consumable with microfluidic channels and laser-activated valves. Instrument: 3M Integrated Cycler with Integrated Cycler Studio Software (v5.0+). Standalone system; no nucleic acid extraction.
Indications for Use
Indicated for qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) from patients suspected of CNS HSV infection. For professional, prescription use only.
Regulatory Classification
Identification
A herpes simplex virus nucleic acid-based assay for central nervous system infections is a qualitative in vitro diagnostic device intended for the detection and differentiation of HSV-1 and HSV-2 in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infections of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infections of the CNS. Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
Special Controls
(b) Classification. Class II (special controls). The special controls for this device are:
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including primer design and selection.
(ii) Detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (limit of detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carryover, and cross contamination. Documentation must include reagent and sample stability recommendations.
(iii) Detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to the results of two polymerase chain reaction methods followed by bidirectional sequencing.
(iv) Documentation of an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.
(v) Quality assurance protocols and detailed documentation for device software, including standalone software applications and hardware-based devices that incorporate software.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed explanation of the interpretation of results and acceptance criteria.
(ii) A limiting statement indicating that negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Simplexa™ HSV 1 & 2 Direct
## DECISION SUMMARY
## A. 510(k) Number:
K133621
## B. Purpose for Submission:
De novo request for evaluation of automatic class III designation for the Simplexa™ HSV 1 & 2 Direct
## C. Measurand:
Target DNA sequences from conserved regions of the HSV-1 and HSV-2 DNA polymerase genes.
## D. Type of Test:
A real-time Polymerase Chain Reaction (PCR) for the direct amplification, detection and differentiation of HSV-1 and HSV-2 DNA from unprocessed CSF samples without nucleic acid extraction.
# E. Applicant:
Focus Diagnostics, Inc.
## F. Proprietary and Established Names:
Simplexa™ HSV 1 & 2 Direct
## G. Regulatory Information:
- 1. Regulation: 21 CFR 866.3307
- 2. Classification: Class II (special controls)
- 3. Product code: PGH
- 4. Panel: Microbiology (83)
{1}------------------------------------------------
## H. Intended Use:
- 1. Intended use(s):
#### Simplexa™ HSV 1 & 2 Direct
The Focus Diagnostics Simplexa™ HSV 1 & 2 Direct is intended for use on the 3M Integrated Cycler instrument for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infections of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infections of the CNS.
Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
The assay is not intended for use as a donor screening test. The assay is for professional use only.
The Positive Control is intended to be used as a control with the Simplexa™ HSV 1 & 2 Direct. This control is not intended for use with other assays or systems.
- 2. Indication(s) for use:
Same as intended use
- 3. Special conditions for use statement(s):
The Simplexa" HSV 1 & 2 Direct is for prescription use only in accordance with 21 CFR 801.109.
- 4. Special instrument requirements:
To be used with the 3MTM Integrated Cycler with Integrated Cycler Studio Software version 5.0 or higher
### I. Device Description:
### Simplexa™ HSV 1 & 2 Direct
The Simplexa™ HSV 1 & 2 Direct system is a real-time PCR that enables the direct amplification, detection and differentiation of HSV-2 DNA from unprocessed CSF samples without nucleic acid extraction. The system consists of the Simplexa™ HSV 1 & 2 Direct, the 3M Integrated Cycler (with 3M Integrated Cycler Studio Software), the Direct Amplification Disc and associated accessories.
In the SimplexaTM HSV 1 & 2 Direct, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify HSV-1, HSV-2 and DNA internal control (IC) targets. Well conserved regions of the HSV-1 and HSV-2 DNA polymerase genes are targeted to identify HSV-1 and HSV-2 DNA respectively in the specimen. The IC
{2}------------------------------------------------
is used to detect PCR failure and/or inhibition.
The 3M Integrated Cycler is a real-time PCR thermocycler which uses real-time fluorometric detection to identify targets within the sample wells. The instrument is controlled by an external computer running the Integrated Cycler Studio Software.
The Simplexa™ HSV 1 & 2 Direct assay reaction takes place in the DAD consumable. The DAD consumable is compartmentalized into 8 separate wedges and up to 8 separate samples or controls may be processed on each disc. Each wedge can be used only once, however, the disc may be reused until all wedges have been utilized. Each wedge contains sample and reagent input wells, microfluidic channels and laser activated valves to control the fluid flow. and a reaction chamber. The disc is specifically designed to meter the amount of reagent (Reaction Mix) and sample that are placed into specific wells in the disc. To start processing a patient sample, a foil seal is lifted and the user adds 50 µL of Reaction Mix to the reagent input well using a fixed volume pipette. Next, the user adds 50 uL of unextracted specimen to the sample input well.
The SimplexaTM HSV 1 & 2 Direct kit contains reagents for 24 reactions. Each vial contains sufficient material for a single reaction. The kit contains the Reaction Mix, the Simplexa™ HSV 1 & 2 Direct Barcode Card with assay specific information, and the Package Insert.
| Kit<br>Component | Contents | | | | |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------|------------------|-------------------------|
| Simplexa™<br>HSV 1 & 2<br>Direct Reaction<br>Mix | DNA polymerase, buffer, dNTPs, template DNA (IC), dye-labeled<br>fluorescent probe-primers specific for detection of HSV-1, HSV-2, and the<br>DNA IC. | | | | |
| | Target | Probe<br>Fluorophore<br>(Dye) | Excitation<br>(nm) | Emission<br>(nm) | Targeted Gene |
| | HSV-1 | CFR610 | 590 | 610 | HSV-1 DNA<br>polymerase |
| | HSV-2 | FAM | 495 | 520 | HSV-2 DNA<br>polymerase |
| | DNA<br>IC | Q670 | 644 | 670 | NA |
| Simplexa™<br>HSV 1 & 2 Kit<br>Barcode Card | Assay specific parameters. | | | | |
Component Description
The Positive Control contains ten vials of Inactivated HSV-1 and HSV-2.
{3}------------------------------------------------
## J. Standard/Guidance Documents Referenced:
Guidance for Industry and FDA Staff: Administrative Procedures for CLIA Categorization, May 7, 2008
Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k), August 12, 2005
Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices, September 9, 1999
Guidance for Industry: Cybersecurity for Networked Medical Devices Containing Off-The-Shelf (OTS) Software, July 18, 2007
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005
General Principles of Software Validation: Final Guidance for Industry and FDA Staff, January 11, 2002
# K. Test Principle:
The Simplexa™ HSV 1 & 2 Direct system is a real-time PCR for the direct amplification, detection and differentiation of HSV-1 and HSV-2 DNA from unprocessed CSF specimens without nucleic acid extraction. The system consists of the Simplexa™ HSV 1 & 2 Direct, the 3M Integrated Cycler (with 3M Integrated Cycler Studio Software), the Direct Amplification Disc and associated accessories.
In the Simplexa™ HSV 1 & 2 Direct assay reaction, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify HSV-1. HSV-2 and IC targets. Well conserved regions of the HSV-1 and HSV-2 DNA polymerase genes are targeted to identify HSV-1 and HSV-2 DNA, respectively, in the specimen. An internal control is used to detect PCR failure and/or inhibition. The principle of the test and the procedural steps are summarized below.
## Specimen Collection:
The specimen type is CSF. Specimens should be transported on ice and stored at 2 to 8°C for up to 7 days post collection. If there is delay for more than 7 days before processing, the specimens should be stored at -70°C.
### Real-Time PCR Instrument Setup:
The 3M Integrated Cycler Operator Manual provides details on how to configure the 3M Integrated Cycler Studio Software, to add an assay definition, and to set up and analyze runs on the 3M Integrated Cycler.
Direct Amplification Disc Loading and Real-Time PCR Amplification:
- 1. Select samples that need to be tested.
- 2. Thaw Reaction Mix vials at room temperature (approximate range 18 to 25°C). Thaw one Reaction Mix vial for each sample or control to be tested.
{4}------------------------------------------------
- 3. Scan the barcode on the Simplexa™ HSV 1 & 2 Direct Reaction Mix vial or barcode card.
- 4. Scan the disc barcode on the Direct Amplification Disc (DAD).
- 5. Scan or type in each sample identifier.
- 6. For one wedge at a time, peel the adhesive foil back to expose the Sample (SAMPLE) and Reaction (R) wells without completely removing the adhesive foil cover (Figures 1 & 2). A void touching the under side of the foil that will be in contact with the wells and disc surface.
- 7. Ensure that the Reaction Mix is completely thawed. Briefly spin down the tubes as needed. (Do not vortex the Reaction Mix).
- Use the fixed volume pipette to transfer 50 uL of the Reaction Mix into the Reaction (R) 8. well.
- 9. Use the fixed volume pipette to transfer 50 µL of sample or control; pipette sample or control into the Sample well (SAMPLE).
- 10. Cover the wedge sealing the wells with the peeled adhesive foil, pressing down firmly near the edge of the wedge. If the original foil is torn, do not load the wells in the wedge. Instead load another wedge.
- 11. Tear off the tab portion of the foil cover along the perforation.
- 12. Repeat steps 6 to 11 for the next sample(s).
- 13. Load the sealed Direct Amplification Disc into the 3M Integrated Cycler and start the run.
Image /page/4/Figure/11 description: The image contains two figures showing a disc with pre-use foil lifted from sample and reaction wells. Figure 1 shows the disc with the foil lifted from wedge #3, revealing the numbers 2, 3, and 4 on the disc. Figure 2 shows a close-up of the sample and reaction wells, labeled as "SAMPLE" and "R" respectively, with the number 4 visible on the foil.
# Interpretation of Results:
Upon completion of the run, the Integrated Cycler Studio Software automatically calculates and displays results. The display presents the user with a separate report box for each of the analytes as well as the IC. The IC can be reported as "valid" or "invalid".
The software reports one of four possible outcomes for each of HSV-1 and HSV-2 after a run
{5}------------------------------------------------
is completed for each sample ID entered: "Detected". "Not Detected". "Invalid" or "EC500".
- "Detected" result indicates that HSV-1 and/or HSV-2 DNA were detected in the patient a. sample tested (whether the DNA IC was detected or not detected)
- b. "Not Detected" result indicates that HSV-1 and/or HSV-2 DNA were not detected in the patient sample tested whereas the DNA IC was detected.
- c. "Invalid" result points to the inability to detect presence or absence of HSV-1 and/or HSV-2 DNA in the patient sample. This result may be due to:
- 1. DNA Internal Control (DNA IC) failure, or
- 2. Failure to detect sufficient specimen.
If an invalid result occurs, the sample needs to be re-tested per the instructions provided in the device Package Insert.
- d. "EC500" result points to a data quality error for the particular viral analyte(s). The software was unable to determine a valid amplification for that analyte(s). The sample should be re-tested.
## L. Performance Characteristics (if/when applicable):
- 1. Analytical performance:
- a. Reproducibility
The reproducibility study evaluated the device's inter-laboratory, inter-assay, and intra-assay reproducibility for high negative, low positive (approximately equal to the Limit of Detection (LoD)) and moderately positive (approximately 4 times LoD) samples for HSV-1 (McIntyre strain) and HSV-2 (G strain) and a positive control. The testing panel consisted of a total of 6 samples: a high negative pool containing both HSV-1 and HSV-2 targets, a low positive and a moderately positive sample separately for each of HSV-1 and HSV-2, and a positive control containing both HSV-1 and HSV-2 targets. The samples (except for the positive control) were generated by spiking viral stock into CSF that was screened to be negative for both analytes. For each sample panel member, two different operators performed two runs per day with 3 replicates per run, for five days at each of the three sites. This provided a total of 90 (3 sites x 5 days x 2 runs (each by different operator) x 3 replicates) replicates per sample panel member over five non-consecutive days. Three sites assessed the device's inter-laboratory reproducibility and inter/intra-assay reproducibility. Combined results for all sites and results stratified by site are presented in the tables below.
{6}------------------------------------------------
| | Site - 1 | | | Site - 2 | | | Site - 3 | | | | |
|-----------------------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|------------|--------------|-----------------------------------------------------|-----------------------|
| Sample | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | Total %<br>Agreement<br>with<br>Expected<br>Results | 95% CI |
| HSV-1 Low<br>Positive | 100.0%<br>(30/30) | 36.9 | 2.0 | 100.0%<br>(30/30) | 35.4 | 2.2 | 100.0%<br>(30/30) | 36.6 | 2.7 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% |
| HSV-1<br>Medium<br>Positive | 100.0%<br>(30/30) | 34.6 | 1.6 | 100.0%<br>(30/30) | 33.2 | 1.2 | 100.0%<br>(30/30) | 34.3 | 1.8 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% |
| HSV-2 Low<br>Positive | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(90/90)a | 95.9%<br>to<br>100.0% |
| HSV-2<br>Medium<br>Positive | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(90/90)a | 95.9%<br>to<br>100.0% |
| High<br>Negative | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(30/30)a | NA | NA | 100.0%<br>(90/90)a | 95.9%<br>to<br>100.0% |
| Positive<br>Control | 100.0%<br>(30/30) | 30.7 | 1.5 | 100.0%<br>(30/30) | 30.6 | 1.2 | 100.0%<br>(30/30) | 30.2 | 1.4 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% |
| Total<br>Agreement | 100.0% (180/180) | | | 100.0% (180/180) | | | 100.0% (180/180) | | | 100.0%<br>(540/540) | 99.3%<br>to<br>100.0% |
ª Expected Results for HSV-2 Low Positive, HSV-2 Medium Positive and High Negative samples are "Negative" for
HSV-1.
{7}------------------------------------------------
| | Site - 1 | | | Site - 2 | | | Site - 3 | | | | |
|-----------------------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|------------|--------------|-----------------------------------------------------|-----------------------|
| Sample | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | Total %<br>Agreement<br>with<br>Expected<br>Results | 95% CI |
| HSV-1 Low<br>Positive | 100.0%<br>(30/30)b | NA | NA | 100.0%<br>(30/30)b | NA | NA | 100.0%<br>(30/30)b | NA | NA | 100.0%<br>(90/90)b | 95.9%<br>to<br>100.0% |
| HSV-1<br>Medium<br>Positive | 100.0%<br>(30/30)b | NA | NA | 100.0%<br>(30/30)b | NA | NA | 100.0%<br>(30/30)b | NA | NA | 100.0%<br>(90/90)b | 95.9%<br>to<br>100.0% |
| HSV-2 Low<br>Positive | 96.7%<br>(29/30) | 38.1 | 2.9 | 90.0%<br>(27/30) | 38.4 | 3.2 | 83.3%<br>(25/30) | 38.3 | 2.4 | 90.0%<br>(81/90)b | 82.1%<br>to<br>94.6% |
| HSV-2<br>Medium<br>Positive | 100.0%<br>(30/30) | 35.0 | 1.3 | 100.0%<br>(30/30) | 34.6 | 1.8 | 100.0%<br>(30/30) | 35.0 | 1.4 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% |
| High<br>Negative | 93.3%<br>(28/30)b | 39.7 | 0.2 | 96.7%<br>(29/30)b | 38.4 | NA | 96.7%<br>(29/30)b | 39.1 | NA | 95.6%<br>(86/90)b | 89.1%<br>to<br>98.3% |
| Positive<br>Control | 100.0%<br>(30/30) | 30.1 | 0.9 | 100.0%<br>(30/30) | 30.6 | 1.1 | 100.0%<br>(30/30) | 30.0 | 1.2 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% |
| Total<br>Agreement | 98.3% (177/180) | | | 97.8% (176/180) | | | 96.7% (174/180) | | | 97.6%<br>(527/540) | 95.9%<br>to<br>98.6% |
b Expected Results for HSV-1 Low Positive, HSV-1 Medium Positive and High Negative samples are "Negative" for HSV-2.
# Reproducibility – DNA IC
| Sample | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | Site - 1 | | | Site - 2 | | | Site - 3 | | | Total %<br>Agreement<br>with<br>Expected<br>Results | 95% CI |
|-----------------------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|------------|--------------|-----------------------------------------------|-----------------------|--------------|-----------------------------------------------------|-----------------------|
| | | | | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | %<br>Agreement<br>with<br>Expected<br>Results | Avg.<br>Ct | Total<br>%CV | | |
| HSV-1 Low<br>Positive | 100.0%<br>(30/30) | 29.6 | 0.7 | 100.0%<br>(30/30) | 30.0 | 1.7 | 100.0%<br>(30/30) | 29.1 | 0.6 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% | | | |
| HSV-1<br>Medium<br>Positive | 100.0%<br>(30/30) | 29.6 | 0.8 | 100.0%<br>(30/30) | 30.1 | 1.2 | 100.0%<br>(30/30) | 29.1 | 0.6 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% | | | |
| HSV-2 Low<br>Positive | 100.0%<br>(30/30) | 29.4 | 0.7 | 100.0%<br>(30/30) | 30.1 | 1.0 | 100.0%<br>(30/30) | 29.1 | 0.7 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% | | | |
| HSV-2<br>Medium<br>Positive | 100.0%<br>(30/30) | 29.4 | 0.9 | 100.0%<br>(30/30) | 30.0 | 1.3 | 100.0%<br>(30/30) | 29.2 | 1.6 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% | | | |
| High<br>Negative | 100.0%<br>(30/30) | 29.4 | 0.8 | 100.0%<br>(30/30) | 30.1 | 1.3 | 100.0%<br>(30/30) | 29.2 | 1.3 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% | | | |
| Positive<br>Control | 100.0%<br>(30/30) | 29.6 | 1.0 | 100.0%<br>(30/30) | 30.0 | 1.1 | 100.0%<br>(30/30) | 29.2 | 0.8 | 100.0%<br>(90/90) | 95.9%<br>to<br>100.0% | | | |
| Total<br>Agreement | | | | 100.0% (180/180) | | | 100.0% (180/180) | | | 100.0% (180/180) | | | 100.0%<br>(540/540) | 99.3%<br>to<br>100.0% |
The positive control Inter-Lot precision was evaluated to assess if there was lot to lot
{8}------------------------------------------------
variability of the positive control. The study assessed the reproducibility of 3 different lots of positive control Pack by testing each lot of control with one lot of Reaction Mix, using one instrument and one operator over 3 consecutive days. Each lot of positive control was run with 2 replicates per run, 2 runs per day for 3 days, thereby providing a total 12 replicates per lot. The testing was performed at Focus Diagnostics. Results of the study were analyzed for overall mean, overall (total) variability, variability between day (inter-day), variability between run (inter-run), variability between lot (inter-lot) and the residual. The variability is presented in terms of standard deviation (SD) and/or percent coefficient of variation (%CV) of Ct. The inter-lot variability component is summarized in the table below.
| Quantitative Summary of PC Inter-Lot Reproducibility | | | | | | | | | | | | |
|------------------------------------------------------|----|---------|-----------|-----|-----------|-----|-----------|-----|---------------|-----|-------|-----|
| | | | Inter-Day | | Inter-Run | | Inter-Lot | | Intra-Run/Lot | | Total | |
| Analyte | N | Mean Ct | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| HSV-1 | 36 | 29.4 | 0.00 | 0.0 | 0.00 | 0.0 | 0.11 | 0.4 | 0.11 | 0.4 | 0.16 | 0.5 |
| HSV-2 | 36 | 29.8 | 0.00 | 0.0 | 0.00 | 0.0 | 0.12 | 0.4 | 0.13 | 0.5 | 0.18 | 0.6 |
| IC | 36 | 29.6 | 0.00 | 0.0 | 0.08 | 0.3 | 0.08 | 0.3 | 0.19 | 0.7 | 0.22 | 0.8 |
b. Linearity/assay reportable range:
Not applicable.
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Positive Control Stability:
The positive control for this product is a blend of inactivated HSV-1 and HSV-2 provided in single use aliquots frozen at -20°C. The positive control has been assigned an expiration date of 12 months at -20℃.
To assess the room temperature stability of positive control, aliquots of the positive control were thawed and kept at room temperature for up to 24 hours, and were then evaluated.
#### Room-temperature stability of Positive Control
{9}------------------------------------------------
| | HSV-2 (FAM) | | HSV-1 (CFR610) | | IC (Q670) | |
|--------------------------------|-------------|----------------------|----------------|----------------------|-----------|----------------------|
| Time (hrs) at room temperature | Mean Ct | Change from time '0' | Mean Ct | Change from time '0' | Mean Ct | Change from time '0' |
| 0 | 30.6 | | 30.7 | | 30.1 | |
| 8 hours | 30.5 | -0.33% | 30.9 | 0.65% | 30.3 | 0.66% |
| 0 | 30.8 | | 31.3 | | 30.2 | |
| 24 hours | 30.8 | 0.00% | 31.0 | -0.96% | 30.2 | 0.00% |
A negative percent difference indicates earlier Ct detection after room temperature storage. As shown in the table above, no significant change in detection of each target analyte was observed when positive control was stored on the bench top at room temperature for 24 hours. Based on these results, the positive control is stable for at least 24 hours at room temperature.
#### d. Assay cut-off:
The Simplexa™ HSV 1 & 2 Direct cut-off was determined during feasibility and verification studies. During verification, all studies were performed to a maximum of 45 cycles of amplification for informational use, but the data were calculated with a cut-off of Ct=40 for HSV-1 and HSV-2 targets. When the verification data were analyzed, there were some results that prompted the Ct cut-off for HSV-2 to be re-set from Ct=40 to Ct=42. The LoD verification study detected HSV-2 between a Ct of 40 and 41 for four replicates at HSV-2 concentrations below the established HSV-2 LoD concentration. No HSV-2 LoD replicates (at or below LoD concentration) were detected beyond Ct=41. The HSV-1 cut-off was not adjusted from Ct=40 because no HSV-1 LoD replicates from the LoD verification study were detected at a Ct>40. Based on the results of the verification studies, the final cut-offs were set at Ct=40 for HSV-1 and at Ct=42 for HSV-2. The PCR cycling protocol was reduced to 42 cycles because all target cut-offs are at Ct<42. The Ct cut-offs and cycling protocol is stored in the assay definition for the Simplexa™ HSV 1 & 2 Direct.
### e. Detection limit (LoD)
The Limit of Detection (LoD) was determined for the Simplexa™ HSV 1 & 2 Direct using viral stocks of HSV-1 and HSV-2. The samples used for the study were prepared using a total of four strains of HSV-1 (McIntyre and HF strains) and HSV-2 (G and MS strains) that were individually diluted into bulk negative human CSF matrix. Each strain of HSV used in the study consisted of verified stock material that had been regrown and retitered from the original source. The study consisted of multiple runs for each strain to evaluate the LoD of Simplexa™ HSV 1 & 2 Direct and all samples were stored overnight at -20°C or below prior to LoD testing.
{10}------------------------------------------------
For each of the HSV-1 McIntyre and HSV-2 G strains, screening and confirmation testing was consolidated into 32 DAD runs using two lots of Simplexa™ HSV 1 & 2 Direct across four instruments. For each of the two targets, the designated strains were spiked into negative human CSF matrix and serially diluted into 2-fold serial dilutions to encompass concentrations around the theoretical LoD. In total, each dilution was assayed in 32 replicates across the 32 DAD runs, where each run consisted of one replicate of each dilution. Per target strain, each instrument underwent eight runs in total, where four of the runs used material from one designated Simplexa™ HSV 1 & 2 Direct lot and the other four runs used material from a second designated Simplexa™ HSV 1 & 2 lot, resulting in 32 runs per strain.
The LoDs for the remaining two strains (HSV-1 HF strain and HSV-2 MS strain) were determined from screening runs that provided tentative LoDs for both the HSV-1 and HSV-2 strains and confirmation runs to confirm the respective LoDs. To determine the tentative LoDs, the designated strains were spiked into negative human CSF matrix and serially diluted to their approximate theoretical LoDs. For each of these two strains, at least three replicates of each dilution were tested and the lowest concentration at which all replicates were detected was taken as the tentative LoD. For the confirmation of tentative LoD, each designated strain was spiked into negative human CSF matrix at the concentration of the tentative LoD. The samples for each strain were prepared in singlicate and assayed in 32 replicates across multiple disc runs using a single lot of Simplexa™ HSV 1 & 2 Direct. The LoD was determined as the lowest concentration for which ≥ 95% of the replicates were detected.
| Virus<br>Strain | LoD<br>(TCID50/mL) | Qualitative<br>Results at LoD<br>(#Detected/#Total) | Mean Ct ± SD<br>(from Detected<br>Replicates only) |
|-------------------|--------------------|-----------------------------------------------------|----------------------------------------------------|
| HSV-1<br>McIntyre | 5 | 31/32 | 37.1 ± 1.17 |
| HSV-1 HF | 40 | 31/32 | 36.9 ± 0.90 |
| HSV-2 G | 1.25 | 31/32 | 38.3 ± 0.74 |
| HSV-2 MS | 20 | 32/32 | 37.2 ± 1.03 |
Limit of Detection Summary
#### Analytical Reactivity f.
The analytical reactivity of the Simplexa™ HSV 1 & 2 Direct was evaluated using different strains of HSV-1 that were not used to determine the LoD for the assay. No additional strains were available for HSV-2. Ouantified viral material was spiked into negative CSF using a single dilution and assayed in triplicate. The Simplexa™ HSV 1 & 2 Direct was able to detect both strains of HSV-1 virus at 20 TCID« /mL.
{11}------------------------------------------------
| Viral Strain | Concentration<br>[TCID50/mL] | Qualitative Results<br>(#Detected/#Total) | | |
|--------------|------------------------------|-------------------------------------------|-----|-----|
| HSV-1 KOS | 20 | 3/3 | 0/3 | 3/3 |
| HSV-1 KOS | 20 | 2/3 | 0/3 | 3/3 |
| HSV-1 F | 40 | 3/3 | 0/3 | 3/3 |
| HSV-1 F | 80 | 3/3 | 0/3 | 3/3 |
## Analytical Reactivity with Additional Viral Strains
### g. Analytical specificity:
## Cross-Reactivity:
The analytical specificity of the Simplexa™ HSV 1 & 2 Direct was evaluated by testing 51 potential cross-reactants that are closely related, cause similar clinical symptoms, or may be present in CSF. The potential cross-reactants were spiked into negative CSF and assayed in triplicate. No cross-reactivity was observed as summarized in the following table.
| Cross-Reactant | Tested<br>Concentration | Qualitative Result<br>(#Detected/#Total) | |
|-----------------------------------------------|------------------------------|------------------------------------------|-------|
| | | HSV-1 | HSV-2 |
| Baseline | Not Applicable | 0/20 | 0/20 |
| Adenovirus (type 1) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Adenovirus (type 7A) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| BKV | 1.00 X 105 copies/mL | 0/3 | 0/3 |
| Citrobacter freundii | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Citrobacter koseri | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Cryptococcus neoformans | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Cytomegalovirus (AD169) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Dengue (Type1) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Encephalomyocarditis virus | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Enterobacter aerogenes | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Enterovirus 71 | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Epstein Barr virus (B95-8) | 1.00 X 105 copies/mL | 0/3 | 0/3 |
| Cross-Reactant | Tested<br>Concentration | Qualitative Result<br>(#Detected/#Total) | |
| | | HSV-1 | HSV-2 |
| Escherichia coli | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Haemophilus influenzae | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Haemophilus influenzae type b<br>(MinnA) | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Hepatitis B | 1.00 X 105 IU/mL | 0/3 | 0/3 |
| Hepatitis C | 1.00 X 105 IU/mL | 0/3 | 0/3 |
| HIV1 (type IIIB) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| HIV2 (type NIHZ) | Not Available* | 0/3 | 0/3 |
| Human herpesvirus 6 | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Human herpesvirus 7 (Type SB) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Human herpesvirus 8 | 1.00 X 105 copies/mL | 0/3 | 0/3 |
| Influenza A/California/7/2009<br>NYMC x-179-A | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Influenza B/Florida/02/2006 | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| JCV (MAD-4) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Klebsiella pneumoniae | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| La Crosse encephalitis | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Listeria monocytogenes | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Measles | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Mumps | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Mycobacterium tuberculosis<br>(Genomic DNA) | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Naegleria fowleri | 1.00 X 104 cells/mL | 0/3 | 0/3 |
| Neisseria meningitides (serogroup A) | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Parainfluenza Virus 1 | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Parainfluenza Virus 2 | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Parainfluenza Virus 3 | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Parvovirus B19 | 1.00 X 105 IU/mL | 0/3 | 0/3 |
| Poliovirus (Type 3) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Proteus mirabilis (Z050) | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Cross-Reactant | Tested<br>Concentration | Qualitative Result<br>(#Detected/#Total) | |
| | | HSV-1 | HSV-2 |
| Pseudomonas aeruginosa | 1.00 X 106 copies/mL | 0/3 | 0/3 |
| Rabies | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Rhinovirus (Type 1A) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Rotavirus (Type Wa) | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Rubella | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| St. Louis Encephalitis | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
| Staphylococcus aureus COL | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Streptococcus agalactiae | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Streptococcus pneumoniae Z022; 19F | 1.00 X 106 cfu/mL | 0/3 | 0/3 |
| Toxoplasma gondii | 1.00 X 106<br>tachyzoites/mL | 0/3 | 0/3 |
| Varicella zoster virus | 1.00 X 105 copies/mL | 0/3 | 0/3 |
| West Nile Virus | 1.00 X 105<br>TCID50/mL | 0/3 | 0/3 |
# Simplexa™ HSV 1 & 2 Direct Cross-Reactivity Results
{12}------------------------------------------------
{13}------------------------------------------------
h. Interference studies:
# Interfering Substances:
The effect of potentially interfering substances that may be present in CSF specimens was evaluated at the concentrations indicated in the table below. A total of 7 potentially interfering substances was tested in a low positive HSV-1 and HSV-2 sample (4 times LoD) in CSF matrix and assayed in triplicate. No interference was observed.
| Potential Interferent | Interferent<br>Concentration | Qualitative Result (#Detected/#Total) | | |
|-------------------------------|------------------------------|---------------------------------------|-------|--------|
| | | HSV-1 | HSV-2 | DNA IC |
| Albumin (protein) | 10 mg/mL | 3/3 | 3/3 | 3/3 |
| Casein (protein) | 10 mg/mL | 3/3 | 3/3 | 3/3 |
| Hemoglobin | 0.625 mg/mL | 3/3 | 3/3 | 3/3 |
| White Blood Cells | 5.5 x 108 WBC/mL | 3/3 | 3/3 | 3/3 |
| Antiviral Drug (Acyclovir) | 2.5 mg/mL | 3/3 | 3/3 | 3/3 |
| Topical Antiseptic (Betadine) | 5% (v/v) | 3/3 | 3/3 | 3/3 |
| Whole Blood | 10% (v/v) | 3/3 | 3/3 | 3/3 |
Competitive Interference:
{14}------------------------------------------------
Competitive interference was studied to evaluate the effects of clinically relevant coinfections with each of the analytes detected by the Simplexa™ HSV 1 & 2 Direct. The study assessed whether a high concentration of one virus in the specimen could potentially affect the SimplexaTM HSV 1 & 2 Direct performance for the other viral target present at low level sample was contrived at approximately 4 times LoD for each target, (HSV-1 McIntryre strain and HSV-2 G strain), and a baseline Ct was determined for each sample. Each potential concomitant infecting virus was spiked into the low level specimen and assayed in triplicate. Competitive interference was observed in samples with a very high concentration of HSV-1 virus and a low concentration of HSV-2 virus. Baseline sample results are also shown below.
| Baseline<br>(Low Concentration) | | Competitive Interferent<br>(High Concentration) | | Qualitative Results<br>(#Detected/#Total) | | |
|---------------------------------|------------------------------|-------------------------------------------------|------------------------------|-------------------------------------------|-----------|-----------|
| Strain | Concentration<br>[TCID50/mL] | Strain | Concentration<br>[TCID50/mL] | HSV-<br>1 | HSV-<br>2 | DNA<br>IC |
| HSV-1 McIntyre | 20 | - | - | 5/5 | 0/5 | 5/5 |
| HSV-1 McIntyre | 20 | HSV-2 G | 10000 | 3/3 | 3/3 | 3/3 |
| HSV-2 G | 5 | - | - | 0/5 | 5/5 | 5/5 |
| HSV-2 G | 5 | HSV-1<br>McIntyre | 20000 | 8/8 | 1/8 | 8/8 |
| HSV-2 G | 5 | HSV-1<br>McIntyre | 10000 | 8/8 | 6/8 | 8/8 |
| HSV-2 G | 5 | HSV-1<br>McIntyre | 5000 | 3/3 | 3/3 | 3/3 |
SimplexaTM HSV 1 & 2 Direct Competitive Interference Results
Inhibition by Other Microorganisms:
The Simplexa™ HSV 1 & 2 Direct was evaluated by testing the ability to identify HSV-1 and HSV-2 viruses when other potentially inhibitory organisms are present. The panel of 51 potentially inhibitory organisms was individually spiked into a pool with a low concentration (approximately 4 times LoD) of HSV-1 and HSV-2 in CSF.
{15}------------------------------------------------
Each microorganism sample was initially tested in triplicate and, if any one of the replicates was "Not Detected" for either the HSV-1 or the HSV-2 targets, then five additional replicates were tested to confirm if any inhibition was caused by the microorganism. If the majority (>4/8) replicates were "Not Detected", then an inhibitory effect would be determined. None of the microorganisms caused >4/8 of the replicates to be "Not Detected". One of 8 replicates of JCV (MAD-4) and Rabies were "Not Detected" for HSV-2 and 2/8 replicates of Dengue (Type1) were "Not Detected" for HSV-2. No inhibition by other organisms was observed for either HSV-1 or HSV-2.
| No. | Microorganism | Tested Concentration | Qualitative Result<br>(#Detected/#Total) | |
|-----|-----------------------------------------------|---------------------------|------------------------------------------|--------|
| | | | HSV-1 | HSV-2 |
| 1 | Baseline | Not Applicable | 20/20 | 20/20 |
| 2 | Adenovirus (type 1) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 3 | Adenovirus (type 7A) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 4 | BKV | 1.00 X 105 copies/mL | 3/3 | 3/3 |
| 5 | Citrobacter freundii | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 6 | Citrobacter koseri | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 7 | Cryptococcus neoformans | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 8 | Cytomegalovirus (AD169) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 9 | Dengue (Type1) | 1.00 X 105 TCID50/mL | 8/8 | **6/8 |
| 10 | Encephalomyocarditis virus | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 11 | Enterobacter aerogenes | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 12 | Enterovirus 71 | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 13 | Epstein Barr virus (B95-8) | 1.00 X 105 copies/mL | 3/3 | 3/3 |
| 14 | Escherichia coli | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 15 | Haemophilus influenzae | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 16 | Haemophilus influenzae type b<br>(MinnA) | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 17 | Hepatitis B | 1.00 X 105 IU/mL | 3/3 | 3/3 |
| 18 | Hepatitis C | 1.00 X 105 IU/mL | 3/3 | 3/3 |
| 19 | HIV1 (type IIIB) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 20 | HIV2 (type NIHZ) | Not Available* | 3/3 | 3/3 |
| 21 | Human herpesvirus 6 | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 22 | Human herpesvirus 7 (Type SB) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 23 | Human herpesvirus 8 | 1.00 X 105 copies/mL | 3/3 | 3/3 |
| 24 | Influenza A/California/7/2009<br>NYMC x-179-A | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 25 | Influenza B/Florida/02/2006 | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 26 | JCV (MAD-4) | 1.00 X 105 TCID50/mL | 8/8 | ***7/8 |
| 27 | Klebsiella pneumoniae | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| No. | Microorganism | Tested Concentration | Qualitative Result<br>(#Detected/#Total) | |
| | | | HSV-1 | HSV-2 |
| 28 | LA Crosse encephalitis | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 29 | Listeria monocytogenes | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 30 | Measles | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 31 | Mumps | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 32 | Mycobacterium tuberculosis<br>(Genomic DNA) | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 33 | Naegleria fowleri | 1.00 X 104 cells/mL | 3/3 | 3/3 |
| 34 | Neisseria meningitides<br>(serogroup A) | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 35 | Parainfluenza Virus 1 | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 36 | Parainfluenza Virus 2 | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 37 | Parainfluenza Virus 3 | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 38 | Parvovirus B19 | 1.00 X 105 IU/mL | 3/3 | 3/3 |
| 39 | Poliovirus (Type 3) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 40 | Proteus mirabilis (Z050) | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 41 | Pseudomonas aeruginosa | 1.00 X 106 copies/mL | 3/3 | 3/3 |
| 42 | Rabies | 1.00 X 105 TCID50/mL | 8/8 | ***7/8 |
| 43 | Rhinovirus (Type 1A) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 44 | Rotavirus (Type Wa) | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 45 | Rubella | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 46 | St. Louis Encephalitis | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
| 47 | Staphylococcus aureus COL | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 48 | Streptococcus agalactiae | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 49 | Streptococcus pneumoniae<br>Z022; 19F | 1.00 X 106 cfu/mL | 3/3 | 3/3 |
| 50 | Toxoplasma gondii | 1.00 X 106 tachyzoites/mL | 3/3 | 3/3 |
| 51 | Varicella zoster virus | 1.00 X 105 copies/mL | 3/3 | 3/3 |
| 52 | West Nile Virus | 1.00 X 105 TCID50/mL | 3/3 | 3/3 |
Simplexa™ HSV 1 & 2: Inhibition by Other Microorganisms
{16}------------------------------------------------
*Quantified material was not available to test; instead the vendor provided a culture fluid with a known Ct value. The site was directed to dilute the stock to a relevant Ct value; 1:200 dilution factor. ** 1/3 and 1/5 replicates were "Not Detected" for HSV-2 during initial and confirmatory testing respectively.
***1/3 replicate was "Not Detected" for HSV-2 during initial testing.
#### Sample stability studies i.
The objective of this study was to evaluate the difference between results from the Simplexa™ HSV 1 & 2 Direct testing of fresh (never frozen, unextracted and tested within 72 hours of preparation) specimens versus frozen (unextracted and tested after
{17}------------------------------------------------
2 freeze-thaw steps) specimens. A panel of contrived samples consisting of 120 pairs (fresh and frozen) of spiked specimens prepared in negative human CSF matrix was used for this study; 60 of the pairs spanned the analytical range for HSV-1 and the other 60 pairs spanned the range for HSV-2. Sixty percent of the samples (36/60) were spiked to be close to LoD so that 18 of the samples were 1-1.5 times LoD and 18 samples were spiked at 4 times LoD. The remaining 40% of the samples (24/60) were spiked across the assay range. One lot of Simplexa ™ HSV 1 & 2 Direct was used for the study. The frozen samples were stored at -20°C for a minimum of 5 days, after which they were thawed for at least 1 hour and refrozen for >24 hours and then thawed again for testing. The HSV-1 (McIntyre) and HSV-2 G strains were used in the study.
No significant storage effect on the samples stored fresh was detected when compared to the results from the same samples frozen. Below is the summary of results from the Fresh vs. Frozen Study of the Simplexa™ HSV 1 & 2 Direct along with the concentration [TCID50/mL] of the virus strains. In the SimplexaTM HSV 1 & 2 Direct Fresh vs. Frozen study, 100% (60/60) qualitative agreement was observed with 95% CI (94.0% to 100.0%) between the fresh and frozen states for both HSV-1 and HSV-2 targets. Regression analysis was performed to estimate quantitative bias (in terms of Ct) between Ct values from the paired fresh and frozen samples. The regression plots are shown below. The 95% confidence intervals for intercept and slope of both targets indicate statistically non-significant bias between Ct values for fresh and frozen samples. These results support the testing of frozen samples to establish performance characteristics of the assay.
| HSV-1 Sample Distribution | |
|---------------------------------------------------------|-----------------------------------------|
| HSV-1 (McIntyre)<br>Spiked Concentration<br>[TCID50/mL] | Number of<br>Replicates<br>(% of Total) |
| 5 (LoD) | 18 (30%) |
| 20 (4XLoD) | 18 (30%) |
| 78.125 | 4 (6.7%) |
| 156.25 | 4 (6.7%) |
| 312.5 | 4 (6.7%) |
| 625 | 4 (6.7%) |
| 1250 | 4 (6.7%) |
| 2500 | 4 (6.7%) |
| Total | 60 (100%) |
{18}------------------------------------------------
Image /page/18/Figure/0 description: The image contains a Passing-Bablok regression fit with a confidence bound, showing the relationship between fresh and frozen HSV1 Ct values. The regression line equation is P-B Reg Line = -1.54 + 1.05 * X, with a 95% confidence interval for the intercept of (-3.1146, 0.1000) and for the slope of (1.0000, 1.0976), based on a sample size of n=60. Additionally, there is a table showing the HSV-2 sample distribution, detailing the number of replicates and their percentage of the total for various spiked concentrations of HSV-2 (G strain), ranging from 1.875 (1.5XLoD) to 2500 [TCID50/mL].
Spiked Virus-HSV1 Passing-Bablok Regression Fit with Confidence Bound
{19}------------------------------------------------
Spiked Virus-HSV2 Passing-Bablok Regression Fit with Confidence Bound
Image /page/19/Figure/1 description: The image is a scatter plot that shows the correlation between fresh and frozen HSV2 Ct values. The x-axis represents fresh HSV2 Ct, and the y-axis represents frozen HSV2 Ct. The plot includes a Passing-Bablok regression line with the equation P-B Reg Line = -0.91 + 1.04 * X, along with 95% confidence intervals for the intercept (-2.6000, 0.2000) and slope (1.0000, 1.0889), with a sample size of n=60. Different symbols represent varying concentrations of TCID50/mL.
#### j. Stability studies
A real-time stability was completed over a 14 month time period supporting a shelflife claim of 12 months for the Simplexa™ HSV 1 & 2 Direct and the positive control.
#### k. Carry-over Contamination
The amplification carry-over for the Simplexa™ assays including the DirectSimplexa™ HSV 1 & 2 Direct Simplexa™ HSV 1 & 2 Direct was assessed from the Simplexa™ Flu A/B & RSV Direct (K120413). The carry-over study in K120413 can be applied to the Simplexa™ HSV 1 & 2 Direct as carry-over contamination is not expected to be analyte specific and there have been no hardware changes to the instrument since the K120413 clearance. The carry-over study in K120413 searched for the presence of contamination in negative samples that were adjacent to high positive samples. The study was designed by alternately placing high positive and negative samples on each disc. A total of 60 negative samples and 60 high positive samples were tested across 17 runs using 4 different instruments. The
{20}------------------------------------------------
sample size of 60 was chosen to provide an acceptance criterion of > 90% for the lower bound of the two-sided 95% Confidence Interval for the observed negative rate of the negative samples. This acceptance criterion would be met if no more than one of the 60 negative samples gave a "Detected" result indicative of carry-over contamination. All 60 negative samples gave a result of "Not Detected" for all three target viruses in the K120413 study with a negative rate of 100% and a lower bound of the two-sided 95% CI of 94.0%. Therefore, no evidence of carryover was observed.
## 2. Comparison studies:
- a. Method comparison with predicate device:
Not applicable. Refer to the Clinical Studies section of this document.
- b. Matrix comparison:
Not applicable.
- 3. Clinical studies:
A total of 219 CSF samples were prospectively or retrospectively collected from patients with signs and symptoms of Herpes Simplex Virus (HSV) central nervous system (CNS) infection at eight external sites. All sites sent the frozen samples to Focus Diagnostics. Two aliquots were prepared from each sample and the samples were then blinded for testing. One aliquot was sent to an external testing with the SimplexaTM HSV 1 & 2 Direct. Five external sites, which included four of the collection sites, performed testing with the Simplexa™ HSV 1 & 2 Direct. The other aliquot was retained at Focus Diagnostics for the comparator PCR/sequencing method.
Results from the Simplexa™ HSV 1 & 2 Direct were compared to results from two PCR/bi-directional sequencing assays (comparator). A positive result by the comparator is defined as a positive by at least one of the two PCR/bidirectional sequencing assays. For a result to be defined as negative by the comparator, both of the two PCR/bidirectional sequencing assays should yield negative results. The samples were tested using two PCR reactions targeting two distinct regions of the HSV genome followed by analysis with two bi-directional sequencing assays. Three hundred (300) base-pair dideoxy DNA sequencing assays were validated for two different HSV gene targets. SYBR Green PCR using extracted patient sample DNA was used to test all samples. Amplicons from positive samples were used for bidirectional sequencing. Sequence requirements had a phred quality score ≥20 (for each base) and a continuous read length >200 and a 2x coverage >180. Sequence alignments were analyzed against the GenBank database using the BLAST search program to determine if the patient sample was positive for HSV-1 or HSV-2.
Neural imaging/clinical impression results documented in a case report form (CRF) as
{21}------------------------------------------------
determined by the patient's attending physician and other clinical information such as chemistries, bacterial culture, MRI/CT scans and in-house PCR results were collected for all patients. The Final Diagnosis takes into consideration all of the laboratory findings and the results of a PCR test performed locally along with clinical presentation of the patient. No dual positive (both HSV-1 Positive and HSV-2 Positive) samples were found by the Simplexa HSV1&2 Direct and the comparator algorithm. The table below shows the distribution of the different clinical parameters collected from patients who were confirmed positive and negative by the results from the two PCR/bi-directional sequencing assays.
| Parameter | Patients which<br>are positive by<br>the comparator<br>for HSV<br>(Number and %) | Patients which<br>are negative by<br>the comparator<br>for HSV<br>(Number and %) |
|---------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| MRI results suggestive of HSV infection | 18/65 (27.7%) | 9/154 (5.8%) |
| MRI results not suggestive of HSV infection | 43/65 (66.2%) | 121/154 (78.6%) |
| MRI results not available | 4/65 (6.2%) | 24/154 (15.6%) |
| Final Diagnosis positive for HSV infection. | 61/65 (93.8%) | 33/154 (21.4%) |
| Final Diagnosis negative for HSV infection | 4/65 (6.2%) | 121/154 (78.6%) |
| Bacteria culture positive | 2/65 (3.1%) | 4/154 (2.6%) |
| Bacteria culture negative | 63/65 (96.9%) | 145/154 (94.2%) |
| Bacteria culture not performed | 0/65 (0.0%) | 5/154 (3.2%) |
{22}------------------------------------------------
| The following tables summarize the results of the Simplexa™ HSV 1 & 2 Direct assay's | |
|---------------------------------------------------------------------------------------|--|
| agreement with the comparator (results of the 2 PCR/bidirectional sequencing assays). | |
| HSV-1: Prospective Samples | | | | HSV-2: Prospective Samples | | | |
|--------------------------------------------------------------------------------|------------------------------------------|-----------------|-------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-----------------|-------|
| Simplexa™<br>Results | Comparator: Two PCR/SEQ | | | Simplexa™<br>Results | Comparator: Two PCR/SEQ | | |
| | Detected | Not<br>Detected | Total | | Detected | Not<br>Detected | Total |
| Detected | 3 | 2* | 5 | Detected | 6 | 1## | 7 |
| Not Detected | 0 | 159 | 159 | Not Detected | 1# | 156 | 157 |
| Total | 3 | 161 | 164 | Total | 7 | 157 | 164 |
| PPA | 100.0% (3/3)<br>95% CI: 43.8 to 100.0% | | | PPA | 85.7% (6/7)<br>95% CI: 48.7 to 97.4% | | |
| NPA | 98.8% (159/161)<br>95% CI: 95.6 to 99.7% | | | NPA | 99.4% (156/157)<br>95% CI: 96.5 to 99.9% | | |
| *1/2 sample had a final diagnosis of meningitis<br>from the chart information. | | | | #Sample was collected from patient with a final<br>diagnosis of meningitis from the chart<br>information.<br>##Sample was collected from patient with a final<br>diagnosis of meningitis from the chart<br>information. | | | |
Note: PPA: Positive Percent Agreement; NPA: Negative Percent Agreement
Fifty five (55) retrospective/preselected HSV positive (as determined by the collection sites) samples from four sites were collected between 2004 and 2013 and confirmed positive by the two PCR/bi-directional sequencing assays. These retrospective samples were used to supplement the prospective study to evaluate the sensitivity of the Simplexa™ HSV 1 & 2 Direct. The results are shown in the table below.
{23}------------------------------------------------
| Retrospective/Preselected Positive Samples –<br>Positive Percent Agreement (PPA) | | |
|----------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Simplexa™<br>Results | Comparator: Two PCR/SEQ | |
| | HSV-1 | HSV-2 |
| HSV-1 | 13 | 0 |
| HSV-2 | 0 | 42 |
| Not Detected | 0 | 0 |
| Total | 13 | 42 |
| | HSV-1 | HSV-2 |
| PPA | 100.0% (13/13)<br>95% CI: 77.2 to<br>100.0% | 100.0% (42/42)<br>95% CI: 91.6 to<br>100.0% |
# 4. Expected Values
The observed expected values of HSV-1 and HSV-2 in the prospective population of the Simplexa™ HSV 1 & 2 Direct clinical study varied between institutions, and are shown in the tables below.
| Sample<br>Population | Sample<br>Collection<br>Site | No. of<br>Samples | Expected Values Based on<br>Simplexa™ HSV1/2 Direct | | |
|----------------------|------------------------------|-------------------|-----------------------------------------------------|-----------------|--------------------|
| | | | HSV-1 | HSV-2 | Not<br>Detected |
| Prospective | 1 | 31 | - | 6.5%<br>(2/31) | 93.5%<br>(29/31) |
| | 7 | 24 | 4.2%<br>(1/24) | - | 95.8%<br>(23/24) |
| | 8 | 94 | 4.3%<br>(4/94) | 5.3%<br>(5/94) | 90.4%<br>(85/94) |
| | 9 | 15 | - | - | 100.0%<br>(15/15) |
| | All | 164 | 3.0%<br>(5/164) | 4.3%<br>(7/164) | 92.7%<br>(152/164) |
{24}------------------------------------------------
| Age<br>Category | Prospective - Number of Samples Positive by Simplexa™ HSV1/2 Direct | | | | | | | | | | | |
|--------------------------|---------------------------------------------------------------------|-------|-----------------|-----|--------|-------|-----------------|-----|-------|-------|-----------------|-----|
| | Combined | | | | Female | | | | Male | | | |
| | HSV-1 | HSV-2 | Not<br>Detected | All | HSV-1 | HSV-2 | Not<br>Detected | All | HSV-1 | HSV-2 | Not<br>Detected | All |
| From birth<br>to 1 month | 1 | 1 | 9 | 10 | | | 6 | 6 | | 1 | 3 | 4 |
| >1 month to<br>2 years | 1 | 1 | 25 | 26 | | | 8 | 8 | | 1 | 17 | 18 |
| >2 years to<br>12 years | | | 9 | 9 | | | 6 | 6 | | | 3 | 3 |
| >12 years to<br>21 years | 1 | 1 | 11 | 12 | 1 | | 4 | 5 | | | 7 | 7 |
| >21 years to<br>60 years | 4 | 4 | 65 | 73 | 3 | 1 | 35 | 39 | 1 | 3 | 30 | 34 |
| >60 years | 1 | | 33 | 34 | 1 | | 20 | 21 | | | 13 | 13 |
| All | 5 | 7 | 152 | 164 | 4 | 2 | 79 | 85 | 1 | 5 | 73 | 79 |
{25}------------------------------------------------
## M. Instrument Names:
Integrated Cycler with Integrated Cycler Studio Software version 5.0 or higher (3M)
## N. System Description:
- 1. Modes of Operation: Batch
- 2. Specimen Identificati…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.