DEN130047 · Ergon Medical , Ltd. · PIA · Mar 20, 2015 · Gastroenterology, Urology
Device Facts
Record ID
DEN130047
Device Name
PROLONG
Applicant
Ergon Medical , Ltd.
Product Code
PIA · Gastroenterology, Urology
Decision Date
Mar 20, 2015
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 876.5025
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Prolong™ is an over-the-counter device under 21 CFR Subpart C that is indicated for use as part of a climax control program for males who suffer from premature ejaculation. It is designed to increase the time between arousal and ejaculation through the use of the training sessions using the stimulating vibratory effects of the device on the penis as part of the “start-stop” technique. Prolong is intended to be available as an over-the-counter device.
Device Story
Prolong™ is a vibratory device for home use by males suffering from premature ejaculation; operates as a training aid for the “start-stop” technique; device provides stimulating vibratory effects to the penis; intended to increase time between arousal and ejaculation; patient-operated; no clinical data provided; safety relies on biocompatibility, electrical, thermal, and mechanical durability testing.
Clinical Evidence
Open-label clinical audit of 33 enrolled subjects (23 completed 6-week study). Primary endpoint: change in IELT/masturbation ejaculation time. Results: median increase of 3 minutes in ejaculation time. Secondary endpoints: GRISS, BAI, and BDI-II scores showed improvement in sexual satisfaction and reduction in anxiety/depression. No adverse events reported. Comparison study (Zamar 2012) showed Prolong (n=58) delayed ejaculation to 8.8 minutes vs 2.6 minutes for CBT.
Technological Characteristics
Vibratory stimulation device; requires biocompatibility testing for patient-contacting materials; requires electrical, thermal, and electromagnetic compatibility (EMC) safety testing; requires mechanical durability testing to withstand forces during use.
Indications for Use
Indicated for males suffering from premature ejaculation as part of a climax control program to increase time between arousal and ejaculation.
Regulatory Classification
Identification
A vibrator for climax control of premature ejaculation is used for males who suffer from premature ejaculation. It is designed to increase the time between arousal and ejaculation using the stimulating vibratory effects of the device on the penis.
Special Controls
In combination with the general controls of the Food Drug & Cosmetic Act, the Vibrator for Climax Control of Premature Ejaculation is subject to the following special controls:
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. The special controls for this device are:(1) The labeling must include specific instructions regarding the proper placement and use of the device.
(2) The portions of the device that contact the patient must be demonstrated to be biocompatible.
(3) Appropriate analysis/testing must demonstrate electromagnetic compatibility safety, electrical safety, and thermal safety of the device.
(4) Mechanical safety testing must demonstrate that the device will withstand forces encountered during use.
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR PROLONG
### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Vibrator for climax control of premature ejaculation. A vibrator for climax control of premature ejaculation is used for males who suffer from premature ejaculation. It is designed to increase the time between arousal and ejaculation using the stimulating vibratory effects of the device on the penis.
NEW REGULATION NUMBER: 21 CFR 876.5025
CLASSIFICATION: II
PRODUCT CODE: PIA
### BACKGROUND
DEVICE NAME: PROLONGTM
SUBMISSION NUMBER: DEN130047
DATE OF DE NOVO: NOVEMBER 21, 2013
CONTACT: ERGON MEDICAL LTD. c/o Barry Pearce, Shotwell & Carr, Inc. 25 BARKER CLOSE, FISHBOURNE, CHICHESTER WEST SUSSEX, PO 18 8BJ, U.K.
#### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
#### INDICATIONS FOR USE
Prolong is used as part of a climax control program for males who suffer from premature ejaculation. It is designed to increase the time between arousal and ejaculation through the use of the training sessions using the stimulating vibratory effects of the device on the penis as part of the "start-stop" technique. Prolong is intended to be available as an over-thecounter device.
#### LIMITATIONS
Limitations on device use are also achieved through the following statements included in the Instructions for Use Manual:
### Warnings:
{1}------------------------------------------------
If you suffer psychological problems or physical symptoms other than early ejaculation such as passing a lot of urine, you must consult your doctor. Excessive use will not necessarily give better results and can injure your skin. Do not use on inflamed or broken skin. Not for internal use. Do not swallow. Keep out of sight and reach of children. Intended for personal use only. Do not share. Use only the lubricant supplied with device. Do not use device close to sensitive equipment which may be affected by electromagnetic interference. No serviceable parts. Store in a cool, dry place.
Do not use for more than 30 minutes at a time and no more than 3 times per week.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The PROLONG™ device consists of the PROLONG™ vibrating device and a 4-ounce bottle of water-based lubricant. The lubricant was FDA cleared under K013086. The PROLONG™ vibrating device includes a white silicone body and a vibrator unit as shown in Figure 1. The vibrator unit comprises a 3-Volt coin (pancake) motor connected to a battery pack of two battery cells (Type PX625A) shrink wrapped together and an Acrylonitrile Butadiene Styrene (ABS) slide switch unit to activate the motor. The device should be disposed of at the end of the six week treatment.
Image /page/1/Picture/5 description: The image shows two different perspective views of a computer mouse. The mouse is white with black outlines. The left image shows the top of the mouse with many small oval-shaped holes, while the right image shows the side of the mouse with a scroll wheel.
Figure 1: PROLONG™ Vibrating Device
## SUMMARY OF NONCLINICAL/BENCH STUDIES
### BIOCOMPATIBILITY/MATERIALS
The patient-contacting materials of the PROLONG™ device are silicon rubber and Acrylonitrile Butadiene Styrene (ABS). Samples of the final finished device were subjected to cytotoxicity, sensitization and irritation testing according to ISO 10993-5 for cvtotoxicity and ISO 10993-10 for irritation and skin sensitization. The extracts were determined to be non-cytotoxic, non-sensitized and produced a negligible irritation response.
## SHELF LIFE/STERILITY
{2}------------------------------------------------
Prolong is a non-sterile, reusable and single-user device. It is intended only for external use and the instructions for use include appropriate cleaning instructions for the Prolong device. The accompanying FDA-cleared lubricant (K013086) has a two-year shelf life. The sponsor performed battery testing to demonstrate that the shelf life of the battery is 3 vears.
### ELECTROMAGNETIC COMPATIBILITY (EMC) AND ELECTRICAL SAFETY
Testing of the battery and motor against the requirements of IEC 60601-1, EN 60601-1-2, 55014-1, 61000-4-2, 61000-4-3, 61000-6-1, 61000-6-3 and 61000-4-8 was performed. The results of the testing confirmed that the device complies with all applicable parts of the EMC standards. Testing to EU standards for EMC was determined to be sufficient to also satisfy the requirements in the FDA-recognized IEC versions of the standards. Electrical and thermal safety testing was conducted in accordance with standard ANSI/AAMI ES60601-1. All measurements and test results were within the acceptable limits specified by the safety standard.
## PERFORMANCE TESTING - BENCH
Mechanical Safety - Pull testing and mechanical durability testing were conducted to demonstrate that the device will withstand forces encountered during use.
Battery Testing - The durability test was conducted to demonstrate the battery capacity to sustain a minimum of six-week use of the subject devices nearing or at the end of their three-year shelf life. The test units were subject to battery tests that were conducted over a 6-week period with the devices being switched on and off. The devices performed satisfactorily and met the criteria for use as recommended in the instructions for use.
## SUMMARY OF CLINICAL INFORMATION
A single site, open-label clinical audit/evaluation of Prolong was conducted to evaluate the safety and effectiveness of the device in subjects with premature ejaculation. The study used the Prolong together with an FDA-cleared lubricant manufactured by Trigg Laboratories, Inc., California, or the CE Marked (but not FDA cleared) Prolong lubricant manufactured by Megasol Cosmetic GmbH. Föhren. Germany. The Megasol lubricant is currently supplied with the Prolong in Europe. Only the Trigg lubricant will be supplied with the Prolong in the US. Subjects were allocated openly in an alternate sequential manner to receive either the Trigg lubricant or the Prolong lubricant. The Prolong was used in conjunction with the lubricant at least three times per week for a total of six weeks.
A total of 39 subjects were screened of which 33 were enrolled in the study. Of the 16 subjects who received Prolong lubricant. 10 subjects completed the six week treatment period with follow-up at every time point; of the 16 subjects who received Trigg water-based lubricant, 12 subjects completed the six week treatment period with follow-up at every time point.
## Study Objectives:
{3}------------------------------------------------
- · To measure and assess the improvement in performance in terms of ejaculation time during the intra-vaginal ejaculatory latency time (ILET) or masturbation following six weeks of treatment with the Prolong and lubricant;
- · To assess the reduction in distress in subjects with premature ejaculation following treatment;
- · To assess the improvement of parameters of sexual satisfaction following treatment;
- · To assess adverse side effects during use of the Prolong and lubricant, if used as instructed.
After an initial screening, visits involved completion of several questionnaires such as Golombock Rust Inventory of Sexual Satisfaction (GRISS), Beck Anxiety Inventory (BAI), and Beck Depression Inventory (BDI-II).
Physical examination included vital signs and blood hematology, such as complete blood count and biochemistry screen, including random blood glucose, thyroid and liver function tests, and urine screening for drugs of abuse, including alcohol. Eligible subjects with written informed consent received the Prolong, lubricant, demonstration video, stop-watch, and a six-week diary card to record ejaculation time from erection.
Subjects returned for further follow-up assessment at the end of the six-week treatment, when they were again asked to complete the questionnaires and a 10-point Likert satisfaction scale. Their diary cards recorded all individual times to ejaculation from erection during intercourse or masturbation and any side/adverse effects, if any occurred, for each of the six weeks of treatment.
| Inclusion | Exclusion |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Male subject $\ge$ 18 and $\le$ 60 years of age | Organic cause ED (substance misuse, pelvic trauma/surgery) |
| Patients with a Diagnostic and Statistical<br>Manual of Mental Disorders, 4th Ed.,<br>Text Revision (DSM-IV-TR) criteria for<br>premature ejaculation | Diabetes |
| Subjects willing and able to provide<br>written informed consent to participate in<br>the treatment audit/evaluation | Prostatitis |
| Subjects able to attend follow-up | Abnormal screening blood test<br>Active substance misuse |
| | Patient meets BDI-II for depression |
## Inclusion/Exclusion Criteria:
### Outcome Measures:
- · Comparison of pre-and post-treatment scores in measurements of BDI, BAI, and GRISS at end of treatment (week 6)
{4}------------------------------------------------
- · Change in latency period to ejaculation on a weekly basis using stop-watch measurement of IELT or during masturbation
- · Drop-out rates
- · Subjective satisfaction with treatment on a 10-point Likert scale.
## Statistical Methods:
The data were evaluated using descriptive statistics including mean, median, 95% confidence interval (CI) and change from pre-treatment (baseline) to end of treatment for both treatment groups. No formal statistical comparison between treatment groups was planned or conducted.
# Results:
Thirty-three subjects were enrolled in the study and 23 were still using the device at week 6 with a median increase in time to ejaculation for these 23 subjects of 3 minutes.
A comparison of pre-treatment and end-of-treatment scores for completed BAI, BDI-II and GRISS questionnaires showed an overall improvement in both treatment groups for distress such as anxiety, depression, and parameters of sexual satisfaction such as sexual function and sexual relationship, respectively, at the end of the study. Six-week use of the Prolong with lubricant was associated with an improvement in impotence, premature ejaculation, avoidance, dissatisfaction, and infrequency in both treatment groups.
There were no reported (0%) adverse events (AEs) for the Prolong hand-held device in either of the two lubricant subject groups who completed the six week treatment period. Eleven subjects dropped out of the study; none were related to treatment side effect/adverse events. A 10-point Likert satisfaction survey showed high levels of satisfaction with of use of the Prolong and lubricants.
# Clinical Conclusions:
- No adverse events occurred when either the Trigg water-based lubricant or Prolong lubricant were used with the Prolong.
- · Trigg water-based and Prolong lubricants, when used with the Prolong for a minimum of three times per week for six weeks, were effective in the treatment of premature ejaculation. There were 33 subjects enrolled in the study and 23 were still using the device at week 6 with a median increase in time to ejaculation for these 23 subjects of 3 minutes.
- · A 10-point Likert satisfaction survey of all subjects who responded to treatment showed high levels of satisfaction with ease of use and effectiveness of the Prolong and lubricants.
# Comparison with Cognitive Behavioral Therapy (CBT):
The start-stop and squeeze techniques incorporated with Cognitive Behavioral Therapy (CBT) are the gold standard training program for the treatment of premature ejaculation, consisting of using a device such as a desensitizing ring daily for no more than 30 minutes for up to six weeks. A study by Zamar (2012) using Prolong in 58 subjects was found to delay time to eiaculation from 48 seconds (0.8 minutes) to 8 minutes and 48 seconds (8.8 minutes) on average in 61% of men, compared to 2 minutes and 36 seconds (2.6 minutes) in 40% of men having the CBT with a difference on average of 6 minutes and 12 seconds (6.2 minutes) between the two groups.
{5}------------------------------------------------
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of vibrator for climax control of premature ejaculation and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measure |
|---------------------------------------------------------|---------------------------------------------------|
| Pain or discomfort due to misuse of<br>device | Labeling |
| Burns | Electrical and Thermal Safety Testing<br>Labeling |
| Electrical Shock | Electrical Safety Testing<br>Labeling |
| Adverse skin reactions | Biocompatibility Testing |
| Patient injury due to device breakage or<br>failure | Mechanical Safety Testing<br>Labeling |
| Interference with other devices/electrical<br>equipment | Electromagnetic Compatibility Testing<br>Labeling |
### SPECIAL CONTROLS
In combination with the general controls of the Food Drug & Cosmetic Act, the Vibrator for Climax Control of Premature Ejaculation is subject to the following special controls:
- (1) The labeling must include specific instructions regarding the proper placement and use of the device.
- (2) The portions of the device that contact the patient must be demonstrated to be biocompatible.
- (3) Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC) safety, electrical safety, and thermal safety of the device.
- (4) Mechanical safety testing must demonstrate that the device will withstand forces encountered during use.
### BENEFIT/RISK DETERMINATION
Probable benefits were demonstrated in two studies presented by the sponsor.
### Prolong Device (Two Lubricants Tested)
- Six-week use of the Prolong device with lubricant was associated with an improvement in ● impotence, premature ejaculation, avoidance, dissatisfaction, infrequency in both treatment groups. The study results showed that 33 subjects were enrolled in the study and 23 were still using the device at week 6 with a median increase in time to ejaculation for these 23 subjects of 3 minutes. The subjects were followed up to six months post treatment and found to have stable or further improved delayed times to ejaculation.
{6}------------------------------------------------
## Prolong Device versus Cognitive Behavioral Therapy (CBT) Study
- The device showed the delay of time to ejaculation from 48 seconds (0.8 minutes) to 8 minutes and 48 seconds (8.8 minutes) on average in 61% of men, compared to 2 minutes and 36 seconds (2.6 minutes) in 40% of men having the CBT with an average difference of 6 minutes and 12 seconds (6.2 minutes) between the two groups.
The risks of the Prolong device are based on nonclinical laboratory data as well as data collected in the clinical study described above. No side effects were reported during the six-week treatment period.
In conclusion, given the available information above, the data provides support for a climax control program using the Prolong device for males who suffer from premature ejaculation, where the probable benefits outweigh the probable risks for the Prolong. The device provides some benefit and the risks can be mitigated by the use of general and the identified special controls.
# CONCLUSION
The de novo for the Prolong™ is granted and the device is classified under the following:
Product Code: PIA Device Type: Vibrator for climax control of premature ejaculation Class: Class II Regulation: 21 CFR 876.5025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.