DEN110011 · Sensory Medical, Inc. · OVP · Dec 18, 2013 · Physical Medicine
Device Facts
Record ID
DEN110011
Device Name
SYMPHONY DEVICE
Applicant
Sensory Medical, Inc.
Product Code
OVP · Physical Medicine
Decision Date
Dec 18, 2013
Decision
DENG
Submission Type
Post-NSE
Regulation
21 CFR 882.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The purpose of the Symphony Device is to improve the quality of sleep in patients with primary Restless Legs Syndrome (RLS) through the use of vibratory counter-stimulation.
Device Story
Symphony Device provides electrically powered mechanical vibration to legs; intended to improve sleep quality in primary Restless Legs Syndrome (RLS) patients. Device operates via vibratory counter-stimulation; mechanism intended to mitigate RLS symptoms. Used by patients; prescription-only. Output is mechanical vibration; healthcare providers use device to manage RLS-related sleep disturbances. Benefits include improved sleep quality for RLS patients. Safety relies on non-clinical testing of vibration parameters (frequency, amplitude, acceleration), electrical/thermal safety, and biocompatibility of patient-contacting components.
Clinical Evidence
Two prospective, multicenter, randomized trials (SMI-001, SMI-002) evaluated safety and effectiveness in primary RLS patients. SMI-001 (n=77) compared device to auditory sham; SMI-002 (n=81) compared to light-emitting sham. Primary endpoint (IRLS score) showed no significant difference. Secondary endpoints (MOS-I, MOS-II sleep scales) showed statistically significant improvement in SMI-001 (p=0.0117, p=0.0230); SMI-002 showed mean improvement but lacked statistical significance. Adverse events included worsening of RLS symptoms (up to 10.3% in Symphony group), leg cramps, and soreness; all resolved post-discontinuation.
Technological Characteristics
Non-sterile, reusable foam pad with six embedded electric motors. Materials: polyurethane foam, lycra inner cover, cotton outer cover. Operates via electronic controller with variable intensity settings (1-9). Complies with IEC 60601-1, EN 55011, and EN 61000-4 series for electrical safety and EMC. Firmware-controlled vibration cycle.
Indications for Use
Indicated for patients with primary Restless Legs Syndrome (RLS) to improve sleep quality. Contraindicated in patients with deep venous thrombosis (DVT) within the last 6 months, leg skin disorders (e.g., eczema, psoriasis, cellulitis, non-healing wounds), or secondary RLS.
Regulatory Classification
Identification
A vibratory counter-stimulation device is a prescription device that provides electrically powered mechanical vibration to improve the quality of sleep in patients with primary Restless Legs Syndrome.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. The special controls for this device are:(1) Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
(2) If the device contains software or firmware, appropriate verification, validation, and hazard analysis must be performed.
(3) The elements of the device that contact the patient must be assessed to be biocompatible.
(4) Non-clinical testing data (including vibration frequency, amplitude, and acceleration) must demonstrate that the device performs as intended under anticipated conditions of use.
(5) Labeling must include:
(i) Specific information pertinent to use of the device by the intended patient population and the treatment regimen;
(ii) Warning to only use the device on normal, intact, clean, healthy skin;
(iii) Warning to not use the device if the user has leg skin disorders, such as eczema, psoriasis, cellulitis, non-healing wounds;
(iv) Warning to discontinue use if Restless Leg Syndrome symptoms worsen; and
(v) Instructions for end users to contact the device manufacturer and MedWatch in case they experience any adverse events when using this device.
In combination with the general controls of the FD&C Act, the Vibratory Counter-Stimulation device is subject to the following special controls:
- 1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
- 2. If the device contains software or firmware, appropriate verification, validation, and hazard analysis must be performed.
- 3. The elements of the device that contact the patient must be assessed to be biocompatible.
- 4. Non-clinical testing data (including vibration frequency, amplitude and acceleration) must demonstrate that the device performs as intended under anticipated conditions of use.
- 5. Labeling must include:
- a. Specific information pertinent to use of the device by the intended patient population and the treatment regimen.
- Warning to only use the device on normal, intact, clean, healthy skin. b.
- Warning to not use the device if the user has leg skin disorders, such as eczema, C. psoriasis, cellulitis, non-healing wounds.
- d. Warning to discontinue use if restless leg syndrome symptoms worsen.
- Instructions for end users to contact the device manufacturer and MedWatch in e. case they experience any adverse events when using this device.
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR SYMPHONY DEVICE
### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Vibratory counter-stimulation device. A vibratory counter-stimulation device is a prescription device that provides electrically powered mechanical vibration to improve the quality of sleep in patients with primary Restless Legs Syndrome.
### NEW REGULATION NUMBER: 882.5895
CLASSIFICATION: CLASS II
PRODUCT CODE: OVP
### BACKGROUND
DEVICE NAME: SYMPHONYTM DEVICE
SUBMISSION NUMBER: DEN110011
DATE OF DE NOVO: JULY 13, 2011
- CONTACT: SENSORY MEDICAL, INC. FRED BURBANK, MD, CEO 1235 Puerta del Sol, #500 SAN CLEMENTE, CA 92673
#### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
#### INDICATIONS FOR USE
The purpose of the Symphony Device is to improve the quality of sleep in patients with primary Restless Legs Syndrome (RLS) through the use of vibratory counter-stimulation.
### LIMITATIONS
For prescription use only.
Limitations on device use are also achieved through the following statements included in the Instructions for Use Manual:
{1}------------------------------------------------
Contraindication:
- . This device should not be used on patients who have been diagnosed with deep venous thrombosis (DVT) in either leg during the last 6 months because of the known potential to dislodge or break up the clot and cause a pulmonary embolism.
## Warnings:
The Symphony Device is only intended for use on patients with primary Restless Legs Syndrome (RLS). It is not intended to be used by people who:
- have leg skin disorders such as eczema, psoriasis, cellulitis, non-healing wounds; or .
- . have secondary RLS.
Due to the potential for the device to cause worsening of symptoms in some patients, the patient labeling should include a warning to discontinue device use and contact a doctor if symptoms worsen.
### PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The Symphony Device is a non-sterile, reusable, cloth-covered foam pad that contains six embedded electric motors that swing eccentric weights to produce vibration throughout the pad. The device is designed to be placed under the user's legs while sleeping. When the device is activated, an electronic controller runs a 35 minute preset vibration cycle. The controller allows the patient to control the intensity level of the vibration for the first 30 minutes of the cycle. In the last five minutes, the controller slowly ramps down the vibration intensity of the device and then automatically shuts the device off at the end of the timed cycle. One repeat cycle can be initiated by the patient.
The controller contains the on/off function and has variable mechanical counter-stimulation intensity settings (1 to 9). The motor amplitude and frequency (vibration) can be set to different levels of intensity by means of a knob. The polyurethane foam pad is covered by an inner lycra cover and an outer cover. The outer cover on the Symphony™ Device is made of cotton. The device will be offered in three sizes: Small, Medium, and Large.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
### BIOCOMPATIBILITY/MATERIALS
The Symphony is intended to only contact intact skin for a limited duration (< 24 hours). The patient-contacting, outer cover of the device is made of cotton. In lieu of biocompatibility testing, the sponsor provided a justification that the identical material has a demonstrated long history of safe use in legally marketed devices with the same type and duration of contact.
### SHELF LIFE/STERILITY
The Symphony is a non-sterile, reusable device. It is intended only for external use and the user manual includes appropriate cleaning instructions for the removable outer cover.
{2}------------------------------------------------
The device does not have a stated shelf life, which, based upon the nature of the device components, is acceptable.
## ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY
The Symphony device was tested for and found to be in compliance with the following standards for electromagnetic compatibility and electrical safety:
| Standard | Title |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | Medical electrical equipment – Part 1: General requirements for safety. |
| EN 55011 | Limits and methods of measurement of radio disturbance characteristics of industrial scientific & medical (ISM) radio-frequency equipment |
| EN 61000-4-2 | Electromagnetic compatibility (EMC) – Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test |
| EN 61000-4-3 | Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
| EN 61000-4-4 | Electromagnetic compatibility (EMC) – Part 4-4: Testing and measurement techniques – Electrical fast transient/ burst immunity test |
| EN 61000-4-5 | Electromagnetic compatibility (EMC) – Part 4-5: Testing and measurement techniques – Surge immunity test |
| EN 61000-4-6 | Electromagnetic compatibility (EMC) – Part 4-6: Testing and measurement techniques – Immunity to conducted disturbances, induced by radio frequency fields |
### SOFTWARE
Software for the device consisted of firmware that controls the vibration cycle of the device. The software was reviewed and the provided documentation was found adequate and consistent with a 'MINOR' level of concern, as discussed in the FDA document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005.
# PERFORMANCE TESTING - BENCH
Device outputs - Testing was conducted to characterize the device outputs at eachintensity setting on the controller. The mean, minimum, and maximum values of acceleration, velocity, displacement (amplitude), and frequency (rpm and Hz) were within specifications for each setting.
Device durability and safety - The following testing was conducted to characterize the Symphony device's ability to perform safely when subjected to different use scenarios: life cycle testing, decibel level, pad surface temperature, motor housing deformation, drop test, bend test, strain relief and electrical cable tensile strength, and shaft tensile strength.
{3}------------------------------------------------
# SUMMARY OF CLINICAL INFORMATION
Two clinical trials were conducted to assess the safety and effectiveness of counter-stimulation with the Symphony Device in the treatment of patients with primary RLS. The primary difference between the two studies was the design of the sham device. The first study (SMI-001) used a sound-emitting (auditory) sham in the control group, while the second study (SMI-002) used a light-emitting sham control. Both studies had the same primary and secondary endpoints.
# 1. SMI-001
Design: This was a prospective, 1:1 randomized, multicenter (5 sites) trial which assessed the safety and effectiveness of the Symphony Device in the reduction of symptoms associated with RLS by comparison to an auditory sham control device. Subjects were adults with moderate to severe primary RLS involving only the legs, with a minimum baseline score of 15 points on the International Restless Legs Scale (IRLS) scale and symptoms occurring at least 15 nights per month. Patients with secondary RLS, other sleep disorders, disorders involving the legs (e.g., cellulitis, deep vein thrombosis) or taking unapproved medications to treat their RLS were excluded. If subjects were on an approved RLS medication, they were required to be on stable doses during the trial. Patients used the device for at least one 35 minute cycle each night. The primary endpoint was the mean change from baseline in the IRLS total score to week 1, 2, 3 and 4. The safety endpoint was a descriptive analysis of adverse events from both groups. Secondary endpoints included the mean change from baseline to 4 weeks in the Medical Outcomes Study (MOS) Sleep Scale scores and the Johns Hopkins Restless Legs Syndrome Quality of Life questionnaire (RLSQOL). The MOS inventory is a patient-reported questionnaire that evaluates sleep disturbance. The MOS sleep problem index I (MOS-I) contains 6 of the 12 inventory questions, while the MOS sleep problem index II (MOS-II) contains an additional 3 questions (9 of the 12). The scale measures sleep difficulty based on recall about sleep quality during the four weeks prior to taking the test.
Results: A total of 77 subjects were enrolled in the study at a 1:1 ratio (39 used Symphony, 38 used the sham device). Although there were no statistically significant differences between active and sham groups for the primary endpoint (IRLS score), there were statistically significant improvements in the mean MOS-I scores (-16.76 for Symphony vs -5.09 for sham. p=0.0117) and MOS-II scores (-15.83 for Symphony vs -5.34 for sham, p=0.0230).
Adverse Events: There were 7 AEs in the Symphony group (17.9%) and 1 in the sham group (2.6%) that were at least possibly device-related. There were an additional 2 AEs in the Symphony group and 3 AEs in the sham group that were not device-related. All devicerelated AEs resolved after use of the device was stopped, without medical intervention or clinical sequelae. One AE (cramping sensation in both legs) was graded as moderate and all 7 remaining AEs were graded as mild. The most common AE was worsening of RLS symptoms, which occurred in 4 subjects with the Symphony device (10.3%) and 1 subject with the sham device (2.6%). Worsening of RLS symptoms resolved within 3 weeks of discontinuation of device use without additional intervention. Other device-related AEs
{4}------------------------------------------------
reported in the Symphony group included non-RLS leg cramps (2 AEs, 5.1%) and leg soreness (1 AE, 2.6%).
# 2. SMI-002
Design: This was a prospective, 2:1 randomized, multicenter (4 sites) trial which assessed the safety and effectiveness of the Symphony Device in the reduction of symptoms associated with RLS by comparison to a light-emitting sham control device. The inclusion criteria, study design, and endpoints were very similar to SMI-001, with only minor differences.
Results: A total of 81 subjects were enrolled in the study at a 2:1 ratio (52 used Symphony, 29 used the sham device). Although there were no statistically significant differences between active and sham groups in any of the primary or secondary endpoints, there were mean improvements in the MOS-I scores (-10.13 for Symphony vs -5.48 for sham, p=0.2358) and MOS-II scores (-11.41 for Symphony vs -7.36 for sham, p=0.3040) that did not achieve statistical significance.
Adverse Events (AEs): There were 5 AEs in the Symphony group (9.6%) and 2 in the sham group (6.9%) that were at least possibly device-related. There were an additional 2 AEs that were not device-related, both in the Symphony group. All device-related AEs in the Symphony group were categorized as worsening of RLS symptoms and resolved within 3 weeks of discontinuation of device use without additional intervention. One of these AEs was graded as moderate and the remaining 4 were graded as mild. In the sham group, 1 AE was categorized as worsening of RLS symptoms (mild), which resolved after stopping use of the sham device. The other AE was pain below the knee (moderate), which resolved with use of a heating pad.
# LABELING
The Symphony User Manual is consistent with the clinical data and covers all the hazards and other clinically relevant information that may impact use of the device. The labeling is sufficient and satisfies the requirements of 21 CFR § 801.109 Prescription devices. Because the clinical data indicate that approximately 10% of patients using the Symphony device experience worsening of RLS symptoms, the patient labeling includes instructions to discontinue device use and contact a doctor if symptoms worsen.
# RISKS TO HEALTH
The table below identifies the risks to health that are associated with use of vibratory counterstimulation devices and the measures necessary to mitigate these risks.
{5}------------------------------------------------
| Identified Risk | Mitigation Method |
|---------------------------------------------|------------------------------------------------------|
| Pain, discomfort, worsening of RLS symptoms | Non-clinical Testing<br>Software Testing<br>Labeling |
| Electrical shock | Electrical Safety Testing<br>Labeling |
| Burns | Electrical and Thermal Safety Testing<br>Labeling |
| Adverse skin reactions | Biocompatibility Assessment<br>Labeling |
| Interference with other medical devices | Electromagnetic Compatibility Testing<br>Labeling |
### SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the Vibratory Counter-Stimulation device is subject to the following special controls:
- 1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
- 2. If the device contains software or firmware, appropriate verification, validation, and hazard analysis must be performed.
- 3. The elements of the device that contact the patient must be assessed to be biocompatible.
- 4. Non-clinical testing data (including vibration frequency, amplitude and acceleration) must demonstrate that the device performs as intended under anticipated conditions of use.
- 5. Labeling must include:
- a. Specific information pertinent to use of the device by the intended patient population and the treatment regimen.
- Warning to only use the device on normal, intact, clean, healthy skin. b.
- Warning to not use the device if the user has leg skin disorders, such as eczema, C. psoriasis, cellulitis, non-healing wounds.
- d. Warning to discontinue use if restless leg syndrome symptoms worsen.
- Instructions for end users to contact the device manufacturer and MedWatch in e. case they experience any adverse events when using this device.
In addition, this is a prescription device and must comply with 21 CFR 801.109.
### BENEFIT/RISK DETERMINATION
The risks of the device are based on data collected in the clinical studies described above. No serious device-related adverse events were reported in the clinical performance data. The majority of adverse events were minor and all resolved on their own within 3 weeks after discontinuation of device use. Based on this information, along with a demonstrated history of safe use of devices with similar technology (i.e., therapeutic massagers), the risk associated with this device is considered low.
{6}------------------------------------------------
The probable benefits of the device are also based on data collected in two clinical studies as described above. Although the IRLS scores do not support that the device is effective for improvement of RLS symptoms in general, the data do indicate an improvement in the specific area of sleep quality. The SMI-001 clinical study demonstrated a statistically significant improvement in the 4-week mean MOS-I and MOS-II sleep scores compared to the control. Although the results of the SMI-002 study did not reach statistical significance, the mean reduction in MOS-I and MOS-II scores was greater than with the sham device. Overall, this indicates that there is clinical benefit for patients with primary RLS.
Additional factors to be considered in determining probable risks and benefits for the Symphony device include: (1) there are no legally marketed devices available for patients with RLS. There are approved drug treatments, but these have known adverse events that are more common and more severe than those seen in the Symphony trials. (2) Although RLS is a chronic disease, longterm safety and effectiveness data are not available. The clinical data is limited to 4 weeks in duration.
Given the available information above, the data support that for improving the quality of sleep in patients with primary Restless Legs Syndrome (RLS), the probable benefits outweigh the probable risks for the Symphony™ device. The device risks can be mitigated by the use of general and the identified special controls.
### CONCLUSION
The de novo for the Symphony device is granted and the device is classified under the following:
Product Code: OVP Device Type: Vibratory counter-stimulation device Class: Class II Regulation: 21 CFR 882.5895
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.