BK251274 · Regen Lab, USA, LLC Jersey City, NJ 07302 · ORG · May 29, 2026 · Hematology
Device Facts
Record ID
BK251274
Device Name
RegenBMC® (RegenKit®-THT®-BMC)
Applicant
Regen Lab, USA, LLC Jersey City, NJ 07302
Product Code
ORG · Hematology
Decision Date
May 29, 2026
Decision
SESE
Regulation
21 CFR 864.9245
Device Class
Class 2
Indications for Use
The RegenBMC® system is indicated to be used intraoperatively at the point-of-care for the safe and rapid preparation of autologous platelet concentrate platelet-rich plasma (PRP) from a small sample of peripheral blood or a small sample of a mixture of peripheral blood and bone marrow. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics.
Device Story
RegenBMC® is a point-of-care system for preparing autologous platelet-rich plasma (PRP) from peripheral blood or a blood/bone marrow mixture; used intraoperatively by trained medical professionals. System consists of sterile vacuum glass tubes containing thixotropic separator gel and sodium citrate anticoagulant. Process: blood/marrow collected into tubes; centrifuged at 1500g for 9 minutes; platelets sequestered above gel layer; supernatant partially removed; cellular sediment resuspended by manual inversion. Resulting PRP collected via syringe and mixed with autograft/allograft bone to improve graft cohesion and handling properties. Device benefits include rapid preparation of autologous concentrate to enhance bone graft application in surgical settings. PRP must be used within 4 hours of collection.
Clinical Evidence
No clinical data. Evidence consists of bench testing: biocompatibility (ISO 10993-1), blood cell recovery/bioequivalence testing (comparing platelet, WBC, RBC, and neutrophil concentrations at 0 and 4 hours), and bone graft cohesion testing. Testing confirmed pH >= 5 and >50% platelet recovery.
Indicated for adult patients undergoing bone graft procedures requiring improved handling characteristics via autologous PRP preparation from peripheral blood or peripheral blood/bone marrow mixture. Safety/effectiveness not evaluated in children, pregnant, or lactating women.
Regulatory Classification
Identification
An automated blood cell separator is a device that uses a centrifugal or filtration separation principle to automatically withdraw whole blood from a donor, separate the whole blood into blood components, collect one or more of the blood components, and return to the donor the remainder of the whole blood and blood components. The automated blood cell separator device is intended for routine collection of blood and blood components for transfusion or further manufacturing use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is a guidance for industry and FDA staff entitled “Class II Special Controls Guidance Document: Automated Blood Cell Separator Device Operating by Centrifugal or Filtration Separation Principle.”
Predicate Devices
Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP System (BK210655)
{0}
# 510(K) SUMMARY
(a)(1). Submitted By:
RegenLab USA, LLC
95 Greene Street
Jersey City, NJ 07302
Contact Person:
Antoine Turzi
RegenLab USA LLC CEO
(b) (6)
(a)(2). Proprietary Name:
RegenBMC®
Common Name(s)
Platelet and plasma separator for bone graft handling
Classification Name:
21 CFR 864.9245 Automated blood cell separator
Regulatory Class:
Class II
Product Code:
ORG
(a)(3). Primary Predicate Device:
Primary Predicate: BK210655 – Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP System
Reference Device: BK090048 - RegenKit®-THT
(a)(4). Device Description
The RegenBMC® (RegenKit®-THT-BMC-3-10) is for collection of PRP from peripheral blood and bone marrow for improvement of bone graft handling characteristics. The RegenBMC® device uses thixotropic gel medical device vacuum tubes in conjunction with a clinical centrifuge. The RegenBMC® permits the rapid point-of-care preparation of platelet concentrate from either a small volume of the patient's peripheral blood or a mixture of the patient's peripheral blood and bone marrow. Up to 10mL per tube or 30mL per 3 tube kit of peripheral blood or a mixture of peripheral blood and bone marrow is collected from the patient into the medical device tubes and then spun in a centrifuge according to centrifuge operating instructions. The resulting platelet-rich plasma (PRP) is collected with a syringe and can be used to hydrate various bone grafts, such as demineralized cortical and cancellous grafts, for improved handling.
The RegenBMC® kits contain three (3) RegenTHT® 10 ml tubes (Figure 1) which are the main components of the kit. The RegenTHT® 10 ml tubes consist of a sterile pharmaceutical grade glass tube
1
{1}
under vacuum for blood or bone marrow collection. The glass tube contains a thixotropic separator gel (RegenGel-T) composed of a mixture of polymers. (Figure 2). The RegenTHT® 10 ml tubes also contain a sodium citrate anticoagulant solution above the separator gel (Figure 3). The tubes are closed with a bromobutyl stopper.

Figure 1. RegenBMC® (RK-THT-BMC-3-10)

| 1) | Glass tube |
| --- | --- |
| 2) | Stopper |
| 3) | Cap |
| 4) | Sodium Citrate Solution |
| 5) | Cell separator gel |
2
{2}
Figure 2. RegenBMC®

Figure 3. RegenTHT® Tube
If processing peripheral blood only, the peripheral blood sample collected from the patient is transferred into all three tubes and centrifuged in an appropriate centrifuge with specific parameters, as detailed in the device's instructions for use. If processing a mixture of peripheral blood + bone marrow, transfer the mixture into all three tubes. Centrifuge in an appropriate centrifuge with specific parameters, as detailed in the device's instructions for use.
After a portion of the plasma supernatant is removed from each tube [i.e. removal of approximately 3.5 mL of platelet-poor plasma (PPP) from each tube], the re-suspension of the cellular sediment into the remaining 2mL of plasma supernatant in each tube is performed by gently inverting the tube at least 20 times. The resulting 6mL (2mL from each of the three tubes) of platelet-rich plasma (PRP), derived from either peripheral blood or a mixture of peripheral blood and bone marrow aspirate is then ready to be collected into a single syringe using a blood transfer device (syringe and transfer device not included in this kit). The PRP preparation is then ready to be used by the physician to be mixed with autograft and/or allograft bone (to hydrate various bone grafts, such as demineralized cortical and cancellous grafts) prior to application to a bony defect for improved handling characteristics. The PRP should be used in this manner by the physician within four (4) hours from the time of the sample collection.
This 510(k) submission includes packaging testing, sterilization validation, shelf-life/stability validation, and biocompatibility testing according to ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. As an externally communicating – blood path, indirect device with a limited contact duration, the biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, material mediated pyrogenicity and hemocompatibility testing. This evaluation showed the device is non-cytotoxic, non-sensitizing, a non-irritant, no signs of acute systemic toxicity, non-pyrogenic and hemocompatible. Also included is bench testing, which includes blood cell concentration recovery testing, glass tube mechanical testing, bone graft cohesion testing and device equivalency testing, which supports a substantial equivalence determination.
3
{3}
The Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP System (BK210655) is the proposed predicate device. Evidence is provided to show the RegenBMC® is substantially equivalent to the predicate device. Based on the indications for use, technological characteristics, safety and performance testing, the RegenBMC® has been shown to be substantially equivalent to Arthrex ACP Max™ PRP System. The RegenKit®-THT (BK090048) device is also presented as a reference device, considering it has the same design and technological characteristics as the RegenBMC®.
### (a)(5). Indications for Use
The RegenBMC® system is indicated to be used intraoperatively at the point-of-care for the safe and rapid preparation of autologous platelet concentrate platelet-rich plasma (PRP) from a small sample of peripheral blood or a small sample of a mixture of peripheral blood and bone marrow. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics.
### (a)(6). Technological Characterizes
The RegenBMC® (RK-THT-BMC-3-10) is for collection of PRP from peripheral blood and bone marrow for improvement of bone graft handling characteristics. The RegenKit®-THT-BMC-3-10 permits the rapid point-of-care preparation of platelet concentrate from either a small volume of the patient's blood or a mixture of the patient's blood and bone marrow. The device is design to process up to 10mL per tube or 30mL per 3 tube kit of peripheral blood and/or mixture of peripheral blood and bone marrow which is collected from the patient into the medical device tubes. Peripheral blood or bone marrow is collected from the patient into the medical device tubes and then spun in a centrifuge according to centrifuge operating instructions. The resulting platelet-rich plasma (PRP) is collected with syringes and can be used to hydrate various bone grafts, such as demineralized cortical and cancellous grafts, for improved handling.
The Separator Gel used in RegenBMC® (RegenTHT® tube) is located at the bottom of the tube and remains stable with the following storage conditions: 41°F to 86°F.
Each RegenBMC® tube vacuum is designed to draw approximately 10mL of peripheral blood into each of the three (3) RegenBMC® vacuum tubes. If processing a mixture of peripheral blood and bone marrow, transfer the mixture into all three tubes. Following collection, the peripheral blood or mixture of peripheral blood and bone marrow aspirate is mixed with the anticoagulant by three (3) manual inversions to ensure adequate mixing of the preloaded sodium citrate anticoagulant with the sample. The RegenBMC® tubes are then centrifuged at a relative centrifugal force of 1500g for nine (9) minutes. Following centrifugation, the recovered platelets will be resting on top of the resolidified thixotropic gel layer (with the red blood cells sequestered beneath the gel). The RegenBMC® tube should be gently inverted no less than twenty (20) times, using a "see-saw" motion to resuspend the platelets into the plasma immediately before being transferred into a syringe.
### (b)(1). Substantial Equivalence: - Non-Clinical Evidence Performance Data
4
{4}
The Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP System (BK210655) is the predicate device. The subject device has the same intended use, similar performance characteristics, and has been evaluated against the same performance standards as the predicate device. The RegenBMC® device is biocompatible per ISO 10993-1.
Platelet-Rich Plasma Bioequivalence Testing Using 30mL of Peripheral Blood and a 30mL Mixture of Peripheral Blood and Bone Marrow showed that PRP produced from the RegenBMC® device is equivalent to the predicate device. The Equivalency Testing evaluated the concentrations of platelets, white blood cells (WBC), red blood cells (RBC), and neutrophils at 0 and 4 hours post collection. The Equivalency testing utilized a 20% margin to evaluate equivalence to predicate device. Testing also confirmed the RegenBMC® device has a pH of at least 5 with at least 50% platelet recovery.
Bone Graft Cohesion Testing was performed using PRP produced by the RegenBMC® device from both Peripheral blood and a combination of peripheral blood and bone marrow. Both PRPs improved bone graft handling characteristics, when compared to a saline control, including the ability of the graft to retain the shape of the syringe mold and hardness and handling properties.
Taken together, the data demonstrate that the RegenBMC® device produces PRP from a mixture of peripheral blood and bone marrow in a manner comparable to the predicate device. Supporting data also demonstrate the ability of the device to process PRP consistent with its intended use and improve bone graft handling characteristics. Observed differences between devices are consistent with known technological characteristics and do not alter the intended use or fundamental performance characteristics of the device, supporting a determination of substantial equivalence.
In addition, the subject device allows for equivalent platelet recovery percentage, when processing blood alone, and a higher platelet recovery percentage, when processing blood and bone marrow together, as compared to the predicate device. While these performance differences exist, there is no impact to the ability of the device to improve bone graft handling characteristics. Therefore, the subject device and the predicate device are substantially equivalent.
In addition, the RegenKit®-THT (BK090048) device is used as a reference device since it has the same components as the RegenBMC® device. The table below shows a comparison of the RegenBMC® device to the predicate and reference device to support substantial equivalence.
5
{5}
### Comparison of Technology to Predicate Device
| Device Name | RegenBMC® (Subject Device) | Arthrex ACP Max™ PRP System (Predicate Device) | RegenKit®-THT (Reference Device) | Comparison |
| --- | --- | --- | --- | --- |
| Regulatory Class | Class II | Class II | Class II | Same as Predicate and Reference Devices |
| 510(k) Number | BK251274 | BK210655 | BK090048 | Not Applicable |
| CFR Section | 21 CFR 864.9245 | 21 CFR 864.9245 | 21 CFR 880.5860 | Same as Predicate Device |
| FDA Product Code(s) | ORG | ORG | FMF | Same as Predicate Device |
| Manufacturer | RegenLab USA, LLC | Arthrex, Inc. | RegenLab SA | Different manufacturers are listed for the subject device and the predicate device |
| Intended Use/ Indications for Use | The RegenBMC™ system is indicated to be used intraoperatively at the point-of-care for the safe and rapid preparation of autologous platelet concentrate platelet-rich plasma (PRP) from a small sample of peripheral blood or a small sample of a mixture of peripheral blood and bone marrow. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. | The Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP system is indicated to be used intraoperatively at the point of care for the safe and rapid preparation of autologous platelet concentrate platelet rich plasma (PRP) from a small sample of peripheral blood or a small sample of a mixture of peripheral blood and bone marrow. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. | The RegenKit®-THT is designed to be used for the safe and rapid preparation of autologous platelet-rich plasma (PRP) from a small sample of blood at the patient's point of care. The PRP is mixed with autograft and/or allograft bone prior to application to an orthopaedic surgical site as deemed necessary by the clinical use requirements. RegenKit®-THT is for single use only. | Same as Predicate Device |
| Use Population | Adult populations that can undergo the sample collection process, as deemed safe and reasonable by their physician. | General care patients who are 18 years of age or older | Adult populations who do not have contraindications stated in the instructions for use. The safety and effectiveness have not been evaluated in children or in pregnant or lactating women. | Same as Predicate Device |
6
{6}
| Device Name | RegenBMC® (Subject Device) | Arthrex ACP Max™ PRP System (Predicate Device) | RegenKit®-THT (Reference Device) | Comparison |
| --- | --- | --- | --- | --- |
| Use Environment | Healthcare setting | Healthcare setting | Healthcare setting | Same as Predicate and Reference Devices |
| Intended User | Trained medical professional | Trained medical professional | Trained medical professional | Same as Predicate and Reference Devices |
| Prescription or OTC | Prescription Use | Prescription Use | Prescription Use | Same as Predicate and Reference Devices |
| Single Use | Yes | Yes | Yes | Same as Predicate and Reference Devices |
| Principles of Operation | Separation of liquids based on density | Separation of liquids based on density | Separation of liquids based on density | Same as Predicate and Reference Devices |
| Blood component separation method | Separator gel, anticoagulant solution and centrifugation | Centrifuge with two stage spin and anticoagulant solution | Separator gel, anticoagulant solution and centrifugation | While the blood component separation methods are different between the subject and predicate devices, the RegenBMC® device that has been confirmed biologically safe through biocompatibility testing and effective at producing PRP that improves bone graft handling through performance testing. Performance testing demonstrated that the device produces PRP from both peripheral blood and combination of peripheral blood and bone marrow and that the resulting PRP improves bone graft handling characteristics compared to saline, consistent with the device's intended use. A comparative performance study against the predicate Arthrex ACP Max System confirmed equivalence for key parameters including pH, red blood cell depletion rate, and ADP-activated platelet activation; differences observed in platelet |
7
{7}
| Device Name | RegenBMC® (Subject Device) | Arthrex ACP Max™ PRP System (Predicate Device) | RegenKit®-THT (Reference Device) | Comparison |
| --- | --- | --- | --- | --- |
| | | | | concentration, recovery, white blood cell concentration, and platelet aggregation are attributable to known technological differences between the devices and do not affect the intended use. Taken together, the data support a determination of substantial equivalence, as the observed performance differences do not alter the device's fundamental performance characteristics or its ability to produce PRP suitable for improving bone graft handling characteristics. |
| **Sterility** | The device is provided sterile by gamma irradiation | The device is provided sterile by ethylene oxide / gamma irradiation | The device is provided sterile by gamma irradiation | The sterilization is substantially equivalent as both the subject device and the predicate have the same SAL of 10^{-6}. Sterilization validation for the subject device has been completed. |
| **Sterility Assurance Level** | SAL 10^{-6} | SAL 10^{-6} | SAL 10^{-6} | Same as Predicate and Reference Devices |
| **Material – Device (not including OEM components)** | Tube: Medical grade glass (b) (4) – 125ml Stopper: Bromobutyl Tube Contents: Cell separator gel and Sodium Citrate Solution | Syringe and Piston: Polypropylene, TPE (Thermoplastic Elastomer), Silicone Fluid Cap: Polyethylene HD (High Density) Small Syringe: Polypropylene, Poly-Isoprene Rubber, Silicone Fluid Anticoagulant: Citrate Dextrose A Solution (ACD-A) | Tube: Medical grade glass (b) (4) – 125ml Stopper: Bromobutyl Cap: Medical Polypropylene Tube Contents: Cell separator gel and Sodium Citrate Solution | Similar materials. The RegenKit®-THT (BK090048) device is used as a reference device since it has already been cleared and uses the same components as the RegenBMC® device. |
8
{8}
| Device Name | RegenBMC® (Subject Device) | Arthrex ACP Max™ PRP System (Predicate Device) | RegenKit®-THT (Reference Device) | Comparison |
| --- | --- | --- | --- | --- |
| **Volume of Patient Blood Drawn** | 10mL to 30mL | 30mL to 90mL | 8mL to 10mL | Similar. The subject device extracts the same or less blood volume from the patient when compared to the predicate device. This difference does not present any new risks or questions on device safety and effectiveness. |
| **Volume of PRP /BMC Generated** | 6mL | 5.3mL | 4.72mL | Similar. Performance testing shows that the subject device and the predicate device are substantially equivalent. |
| **Time** | Must be used within 4 hours after blood draw | Must be used within 4 hours after blood draw | Must be used within 4 hours after blood draw | Same as Predicate and Reference Devices |
| **Packaging** | Kits are packaged in a single-use sterile blister, composed of PETG blister and Tyvek lid | Outer PETG tray with OEM components, Inner PETG tray sealed with Tyvek lid | Kits are packaged in a single-use sterile blister, composed of PETG blister and Tyvek lid | Same as Predicate and Reference Devices |
| **Shelf Life** | 24 months (2 years) | 24 months (2 years) | 24 months (2 years) | Same as Predicate and Reference Devices |
| **Performance Testing** | Biocompatibility – ISO 10993-1 Blood cell recovery equivalency testing Bone cohesion testing Sterility testing Shelf life/ Stability Testing | Biocompatibility – ISO 10993-1 Blood cell recovery equivalency testing Sterility testing Shelf life/ Stability Testing | Biocompatibility – ISO 10993-1 Sterility testing Shelf life/ Stability Testing | Same as Predicate Device |
9
{9}
# **(b)(2). Substantial Equivalence: - Clinical Evidence**
Clinical testing was not necessary for the determination of substantial equivalence.
# **(b)(3). Substantial Equivalence – Conclusions**
The RegenBMC® device is considered substantially equivalent to the Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP System (BK210655). Based on the indications for use, technological characteristics, and safety and performance testing, the RegenBMC® device has been shown to be appropriate for its intended use and is considered substantially equivalent to the Arthrex Autologous Conditioned Plasma (ACP) Max™ PRP System.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.