# ORTHO VISION® Analyzer and ORTHO VISION® Max Analyzer (BK251238)

_Ortho-Clinical Diagnostics, Inc. · KSZ · Oct 10, 2025 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/BK251238

## Device Facts

- **Applicant:** Ortho-Clinical Diagnostics, Inc.
- **Product Code:** [KSZ](/productcode/KSZ.md)
- **Decision Date:** Oct 10, 2025
- **Decision:** SESE
- **Regulation:** 21 CFR 864.9175
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

ORTHO VISION® / ORTHO VISION® Max Analyzer is an instrument designed to automate in vitro immunohematology testing of human blood utilizing ID-MTS™ gel card technology. ORTHO VISION® / ORTHO VISION® Max Analyzer automates test processing functions including liquid pipetting, reagent handling, incubation, centrifugation, reaction grading and interpretation, and data management requirements using cards and digital image processing. ORTHO VISION® / ORTHO VISION® Max Analyzer can be used as a standalone instrument or interfaced to the customer's Laboratory Information System (LIS).

## Device Story

The ORTHO VISION/VISION Max Analyzer is a reusable, automated instrument for immunohematology testing. It processes human blood samples (whole blood, plasma, serum, or red cell suspensions) using ID-MTS gel card technology. The device automates liquid pipetting, reagent handling, incubation, centrifugation, and reaction grading. It utilizes digital image processing to capture high-resolution color images of gel card microtubes. The system interprets reaction results based on pre-defined software rules and manages data, with optional LIS integration. The modification adds the MTS DAT Card to the test menu, requiring a new punch tool to open microtube foil seals. The device is used in clinical laboratory settings by trained personnel. By automating the DAT process, the system provides standardized, objective interpretation of red cell sensitization, aiding in the diagnosis of hemolytic disease states when used alongside patient history and other clinical indicators.

## Clinical Evidence

Clinical study conducted at four sites with 2,778 random patient/donor samples. Comparison between MTS™ DAT Card and tube method showed 99% negative percent agreement (NPA) for both Anti-IgG and Anti-C3b,-C3d. Low positive percent agreement (PPA) in random samples (64% for Anti-IgG, 56% for Anti-C3b,-C3d) was attributed to low prevalence of positive samples. Supplemental testing of 203 well-characterized contrived positive samples demonstrated 100% agreement, supporting clinical utility for suspected hemolytic disease states.

## Technological Characteristics

Automated immunohematology analyzer using column agglutination technology (CAT) with ID-MTS™ gel cards. Features include digital image capture, automated pipetting, incubation, and centrifugation. Connectivity via LIS or standalone. Software-based rule-driven interpretation. Modification includes a mechanical punch tool for MTS™ DAT Card foil seals.

## Regulatory Identification

An automated blood grouping and antibody test system is a device used to group erythrocytes (red blood cells) and to detect antibodies to blood group antigens.

## Predicate Devices

- ORTHO VISION® Analyzer ([BK210557](/device/BK210557.md))
- ORTHO VISION® Max Analyzer ([BK210564](/device/BK210564.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Ortho Clinical Diagnostics, Inc.  
1001 US Highway 202  
Raritan, NJ 08869

Premarket Notification  
Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
Registration/FEI No. 2250051  
510(k) Summary---

## 510(k) Summary

**510(k) #:** BK251238

**Preparation Date:** June 11, 2025

### Contact Details

21 CFR 807.92(a)(1)

**Applicant Name:** Ortho Clinical Diagnostics, Inc.

**Applicant Address:** 1001 US Highway 202, Raritan, NJ 08869

**Applicant Contact Telephone:** (b) (6)

**Applicant Contact:** Christopher Kolz, PhD

**Applicant Contact Email:** christopher.kolz@quidelortho.com

**Correspondent Name:** Ortho Clinical Diagnostics, Inc.

**Correspondent Address:** 1001 US Highway 202, Raritan, NJ 08869

**Correspondent Contact Telephone:** 585-453-4253

**Correspondent Contact:** Darlene Phillips, RAC

**Correspondent Contact Email:** darlene.phillips@quidelortho.com

### Device Name

21 CFR 807.92(a)(2)

**Device Trade Name:**

ORTHO VISION® Analyzer

ORTHO VISION Max® Analyzer

**Common Name:** ORTHO VISION® / ORTHO VISION® Max

**Classification Name:** System, Test, Automated Blood Grouping and Antibody

---SP01088

Page 1 of 8

{1}

Ortho Clinical Diagnostics, Inc.  
1001 US Highway 202  
Raritan, NJ 08869

Premarket Notification  
Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
Registration/FEI No. 2250051  
510(k) Summary

**Regulation Number:** 864.9175

**Product Code(s):** KSZ

### Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

|  Predicate # | Predicate Trade Name | Product Code  |
| --- | --- | --- |
|  BK210557 | ORTHO VISION® Analyzer | KSZ  |
|  BK210564 | ORTHO VISION® Max Analyzer | KSZ  |

### Device Description Summary

21 CFR 807.92(a)(4)

ORTHO VISION® / VISION® Max Analyzer is a reusable multi-patient instrument designed to automate in vitro immunohematology testing of human blood utilizing ID-MTS™ gel card technology. ORTHO VISION® / VISION® Max Analyzer automates test processing functions including liquid pipetting, reagent handling, incubation, centrifugation, reaction grading and interpretation, and data management requirements using cards and digital image processing. ORTHO VISION® / VISION® Max Analyzer can be used as a standalone instrument or interfaced to the customer's Laboratory Information System (LIS).

The modification to the ORTHO VISION® Analyzer and ORTHO VISION® Max Analyzer is for the addition of the Immunohematology Anti-Human Globulin Anti-IgG (Rabbit)/Anti-C3b,-C3d (Murine Monoclonal)/DAT Control, MTS™ DAT Card (referred as the MTS™ DAT Card). The MTS™ DAT Card includes qualitative tests in the ID-MTS™ column agglutination technology (CAT) format for the differentiation of (patient or donor) human red blood cells sensitized in vivo by IgG type immunoglobulins or complement C3b and / or C3d components using the ORTHO® Workstation, ORTHO VISION®, and ORTHO VISION® Max platforms.

The addition of the MTS™ DAT Card test to the ORTHO VISION® Analyzer and ORTHO VISION® Max Analyzer test menu requires a modification to the analyzer Assay Data Disk (ADD). The ADD is a data file which contains processing parameters used by the analyzer software. There are no changes required to the analyzer software or hardware for this ADD modification. However, a new punch tool has been added to open the foil seal on the microtubes of the MTS™ DAT Card required for each sample to be tested. MTS™ Gel Cards have punch tools specific to microtubes that need to be accessed for pipetting sample required for the specific test to be performed.

SP01088

Page 2 of 8

{2}

Ortho Clinical Diagnostics, Inc.  
 1001 US Highway 202  
 Raritan, NJ 08869

Premarket Notification  
 Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
 Registration/FEI No. 2250051  
 510(k) Summary

## Intended Use/Indications for Use

21 CFR 807.92(a)(5)

ORTHO VISION® / ORTHO VISION® Max Analyzer is an instrument designed to automate in vitro immunohematology testing of human blood utilizing ID-MTS™ gel card technology. ORTHO VISION® / ORTHO VISION® Max Analyzer automates test processing functions including liquid pipetting, reagent handling, incubation, centrifugation, reaction grading and interpretation, and data management requirements using cards and digital image processing. ORTHO VISION® / ORTHO VISION® Max Analyzer can be used as a standalone instrument or interfaced to the customer's Laboratory Information System (LIS).

## Comparison to Predicate Devices

21 CFR 807.92(a)(6)

There are no changes required to the analyzer software or hardware for the addition of the MTS™ DAT Card to the test menu.

|  Device Characteristic | Predicate Devices ORTHO VISION® Analyzer (BK210557) ORTHO VISION® Max Analyzer (BK210564) | Modified Devices ORTHO VISION® Analyzer (BK210557) ORTHO VISION® Max Analyzer (BK210564) [Addition of the MTS™ DAT Card to the instrument test menu]  |
| --- | --- | --- |
|  Intended Use | ORTHO VISION® / ORTHO VISION® Max Analyzer is an instrument designed to automate in vitro immunohematology testing of human blood utilizing ID-MTS™ gel card technology. ORTHO VISION® / ORTHO VISION® Max Analyzer automates test processing functions including liquid pipetting, reagent handling, incubation, centrifugation, reaction grading and interpretation, and data management requirements using cards and digital image processing. ORTHO VISION® / ORTHO VISION® Max Analyzer can be used as a standalone instrument or interfaced to the customer's Laboratory Information System (LIS). | Same  |

SP01088

Page 3 of 8

{3}

Ortho Clinical Diagnostics, Inc.

1001 US Highway 202

Raritan, NJ 08869

Premarket Notification

Traditional 510(k) ORTHO VISION/VISION Max Analyzer

Registration/FEI No. 2250051

510(k) Summary

|  Device Characteristic | Predicate Devices ORTHO VISION® Analyzer (BK210557) ORTHO VISION® Max Analyzer (BK210564) | Modified Devices ORTHO VISION® Analyzer (BK210557) ORTHO VISION® Max Analyzer (BK210564) [Addition of the MTS™ DAT Card to the instrument test menu]  |
| --- | --- | --- |
|  Classification | Name: Automated blood grouping and antibody test system Device Class: II Product Code: KSZ | Same  |
|  Test Method | Column agglutination technology (CAT) using ID-MTS™ Gel Cards and reagents | Same  |
|  Sample Types | • Centrifuged whole blood (anticoagulated) • Plasma or serum • Packed red blood cells • 3-5% red cell suspension (pre-diluted patient/donor) • 0.8% red cell suspension (pre-diluted patient/donor) | Same  |
|  Reaction Grading | Digital Image Capture and Analysis for reaction grading | Same  |
|  Result Interpretation | According to pre-defined rules as part of the software | Same  |
|  Measurement | Qualitative | Same  |
|  Image | High resolution color image | Same  |
|  Barcode Symbologies | • Codabar • ISBT 128 • Code 128 • Code 3 and 9 • Code 2 and 5 (Interleaved) | Same  |
|  Interface | Bidirectional Laboratory Information System (LIS) or standalone | Same  |
|  Automated Analysis | Instrument automates test processing functions including liquid pipetting, reagent handling, incubation, centrifugation, reaction grading and interpretation, and data management requirements. | Same  |
|  Test Menu | MTS™ DAT Card is not available | Addition of the MTS™ DAT Card to the test menu  |
|  Punch Tools | MTS™ DAT Card punch tool is not available | Addition of the MTS™ DAT Card punch tool  |

SP01088

Page 4 of 8

{4}

Ortho Clinical Diagnostics, Inc.  
1001 US Highway 202  
Raritan, NJ 08869

Premarket Notification  
Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
Registration/FEI No. 2250051  
510(k) Summary---

## Non-Clinical and Clinical Tests Summary

21 CFR 807.92(b)

### Non-Clinical Performance

Nonclinical tests were performed to support the safety, effectiveness, and performance of the MTS™ DAT Card on the ORTHO VISION® / VISION® Max Analyzer.

Testing demonstrated that environmental conditions, extreme to nominal ORTHO VISION® / ORTHO VISION® Max Analyzer specifications of temperature and humidity, have no impact on the assay performance of the MTS™ DAT Card. For both test platforms, repeatability within run and between analyzers across two validation lots of MTS™ DAT Cards resulted in 100% agreement and over 99% concordance at the lower bound of the 95% confidence limit. Testing also demonstrated that the current MTS™ Gel Card serological timeout restrictions for Anti-Human Globulin formulations used for direct antiglobulin testing (DAT) may be applied to the MTS™ DAT Card for use on the ORTHO VISION® and ORTHO VISION® Max Analyzers.

Assay performance of the Anti-IgG, Anti-C3b, -C3d, and DAT Control formulations is acceptable with samples collected in the anticoagulant recommended for DAT testing, ethylenediaminetetraacetic acid (EDTA), and processed on the ORTHO VISION® / VISION® Max Analyzer. Sample stability for patient and donor samples drawn into EDTA was confirmed to be consistent with the MTS™ Gel Card customer Instructions for Use, maintaining the recommendation to run samples within 24 hours of collection. Additionally, MTS™ DAT Card assay performance using hemolyzed samples was evaluated to confirm that hemolysis may interfere with the test interpretation when run on the ORTHO VISION® / VISION® Max Analyzer. The results confirmed that testing with heavily hemolyzed samples may result in interference with test interpretation. MTS™ DAT Card assay performance was also evaluated to demonstrate that cell washing does not affect test results when run on the ORTHO VISION® / VISION® Max Analyzer. There was 100% concordance between Test results and paired Control results. Assay performance was not compromised by washing cells with saline three times.

Internal testing was performed to verify acceptable agreement between the MTS™ DAT Card and comparator tube methods (ORTHO® Anti-IgG and BioClone® Anti-C3b, -C3d) when used on the ORTHO VISION® Analyzer and the ORTHO VISION® Max Analyzer.

---SP01088

Page 5 of 8

{5}

Ortho Clinical Diagnostics, Inc.  
 1001 US Highway 202  
 Raritan, NJ 08869

Premarket Notification  
 Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
 Registration/FEI No. 2250051  
 510(k) Summary

## Conclusion

All nonclinical testing results for the use of the MTS™ DAT Card on the ORTHO VISION® / VISION® Max Analyzer met their acceptance criteria. The successful nonclinical testing demonstrates that each MTS™ DAT Card formulation run on the ORTHO VISION® / VISION® Max Analyzer is substantially equivalent to the legally marketed predicate devices for the defined indications for use.

## Clinical Performance

Clinical equivalency testing was performed at four sites (two external and two internal sites) that routinely perform immunohematology testing. Random clinical specimens including donors and patients (N=2778) were tested on the ORTHO VISION® / VISION® Max Analyzer.

Random sample results were assessed on a microtube-to-tube test basis using a paired sample comparison between the comparator tube method and the gel card microtube under evaluation. For reaction grades to be concordant between methods either both results had to be negative, or both had to be positive (any reaction grade 1+ through 4+). For the tube test method, a weak reaction was considered positive when comparing to the gel card microtube. The combined results from all sites are summarized in the following table. Percent agreement indicates concordance between the two assays and does not indicate which method gave the correct result. The results below do not reflect testing to resolve initial discrepant results between methods.

## Anti-IgG and Anti-C3b,-C3d Microtube Concordance

|  Random Clinical Specimens  |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  All Sites Combined | Total |   |   | Positive |   |   | Negative  |   |   |
|  Test | N | Percent Agreement | Lower Bound of 95% CI | N* | Percent Agreement | Lower Bound of 95% CI | N | Percent Agreement | Lower Bound of 95% CI  |
|  Anti-IgG | 2778 | 98% | 98% | 28 | 64% | 47% | 2750 | 99% | 98%  |
|  Anti-C3b,-C3d | 2778 | 99% | 99% | 9 | 56% | 25% | 2769 | 99% | 99%  |
|  Control | All samples demonstrated negative results in the Control microtube.  |   |   |   |   |   |   |   |   |

SP01088

Page 6 of 8

{6}

Ortho Clinical Diagnostics, Inc.  
1001 US Highway 202  
Raritan, NJ 08869

Premarket Notification  
Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
Registration/FEI No. 2250051  
510(k) Summary---

The combined total percent agreement and negative percent agreement (NPA) at the lower bound of the 95% confidence interval met the NPA requirement for direct antiglobulin tests of ≥ 95% agreement at the lower bound of the 95% confidence interval when testing random patient and donor samples on the ORTHO VISION® / VISION® Max Analyzer and in all microtube formulations.

The positive percent agreement (PPA) was 56% for the Anti-C3b, -C3d microtube formulation and 64% for the Anti-IgG microtube formulation for random patient and donor samples. There was a low frequency (1.3% to 1.4%) of random DAT positive patient and donor samples across the full study. If all random positive samples were concordant, the lower bound of the 95% CI would still not meet the minimal requirement of ≥ 95%.

Due to the small sample size of random positives, IgG and C3 contrived positive samples were used to support intended use for positive results. 203 well-characterized, contrived positive samples were tested and met a 100% point estimate agreement on the ORTHO VISION® and VISION® Max Analyzers. Testing of contrived positive samples coated with clinically significant IgG antibodies and components of complement across a range of reaction strengths further supports acceptable performance of the MTS™ DAT Card reagents.

Taking into consideration the clinical utility of the Direct Antiglobulin Test, as well as the fact that the DAT result is used in conjunction with patient history and other prognostic indicators, there is no expectation that the low PPA resulting from random donor and patient sampling encountered in the MTS™ DAT Card clinical trials would translate to a negative impact on patient care when used as intended for suspected hemolytic disease states and not as a random diagnostic immunohematology test.

The clinical evaluation study of the MTS™ DAT Card showed acceptable performance in the intended use environment.

## Basis for Claim of Substantial Equivalence

This Traditional 510(k) is submitted to modify legally marketed devices (predicates). Ortho Clinical Diagnostics, Inc. is the holder of the 510(k)s for the predicate devices. The Indications for Use of the proposed devices are unchanged from the legally marketed devices (predicates). The intended use of the modified devices, as described in the labeling, has not changed as a result of the modifications. The fundamental scientific technology of the proposed devices is unchanged from the legally marketed devices

---SP01088

Page 7 of 8

{7}

Ortho Clinical Diagnostics, Inc.  
1001 US Highway 202  
Raritan, NJ 08869

Premarket Notification  
Traditional 510(k) ORTHO VISION/VISION Max Analyzer  
Registration/FEI No. 2250051  
510(k) Summary---

(predicates). There are no significant differences between the modified instruments and the predicates related to the Intended Use or Principle of Operation. With the exception of the addition of the MTS™ DAT Card to the test menu, the predicate and modified devices are identical.

The modified ORTHO VISION® and VISION® Max Analyzers are as safe and effective as the currently marketed predicates under BK210557 and BK210564, respectively.

---SP01088

Page 8 of 8

---

**Source:** [https://fda-staging.innolitics.com/device/BK251238](https://fda-staging.innolitics.com/device/BK251238)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).