Multichem ID-HIVp24 is intended for use as a qualitative quality control serum to monitor the precision of laboratory testing procedures for the HIV-1 p24 antigen detected by the heterogeneous immunoassay systems listed in the Instructions for Use. These products are not intended to replace manufacturer's recommended controls provided in their package insert. Quality Control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.
Device Story
Multichem ID-HIVp24 is a liquid, ready-to-use, third-party quality control serum; composed of disease state human serum diluted with negative human serum, preservatives, and stabilizers. Designed for use in clinical laboratories to monitor the precision of HIV-1 p24 antigen immunoassay testing procedures. Supplied in barcoded vials for automated platform application. Laboratory personnel use the control to verify system performance; it does not replace manufacturer-provided controls. The device provides a stable reference to detect systematic deviations in assay performance, supporting good laboratory practices.
Clinical Evidence
Bench testing only. Precision study (20 days, 240 replicates) and reproducibility study (5 days, 225 replicates) conducted on the cobas e801 instrument. Precision results showed low variability (within-run %CV 1.74%, between-run %CV 1.99%). Reproducibility study across three sites showed low %CV across all sources of variability (within-run 1.41%, between-day 1.51%, between-site 9.68%, between-QC lots 9.55%).
Technological Characteristics
Liquid, serum-based quality control; contains HIV-1 p24 antigen, preservatives, and stabilizers. Ready-to-use, barcoded vials. Storage at 2–8°C. Shelf life 13 months. No energy source or software algorithm; manual/automated laboratory assay control.
Regulatory Classification
Identification
An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.
Special Controls
An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”
{0}

**CLINICAL
DIAGNOSTICS**
**Science for a Safer World**
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21CFR 807.92.
| **Applicant Name** | Techno-path Manufacturing Ltd. |
| --- | --- |
| **Applicant Address** | Fort Henry Business Park, Ballina Co. Tipperary V94 FF1P Ireland |
### Device Identification
| **Trade Name** | Multichem ID-HIVp24 |
| --- | --- |
| **Common Name** | N/A |
| **Classification Name** | Assayed quality control material for clinical microbiology assays |
| **Regulation Number** | 21 CFR 866.3920 |
| **Product code** | QCH |
### Device Substantial Equivalence
| **Predicate Device** | |
| --- | --- |
| Predicate Trade Name | BioRad VIROTROL® HIV-1 Ag |
| Predicate Common Name | N/A |
| Predicate Classification Name | Kit, quality control for blood banking reagents |
| Regulation Number | 21 CFR 864.9650 |
| Product code | KSF |
| Predicate 510(k) Number | BK930033 |
### 1. Device Description
Multichem ID-HIV p24 is supplied as a single level, liquid control that is ready-to-use with barcodes that are designed for application on to the targeted platform. It consists of disease state human serum diluted with negative human serum.
Multichem ID-HIV p24 is a liquid stable, third-party quality control used to monitor the precision of laboratory testing procedures and is an integral part of good laboratory practices. A control is provided to allow performance monitoring of the test system. The Multichem ID-HIV p24 Quality Control has been prepared from human plasma to which preservatives and stabilizers have been added. Multichem ID-HIV p24 is intended for use as a qualitative quality control serum to monitor the precision of laboratory testing procedures for the HIV-1 p24 antigen detected by the heterogeneous immunoassay systems listed in the package insert. These products are not intended to replace manufacturer’s recommended controls provided in their package insert.
Techno-path Manufacturing Limited. Registered in Ireland No. 419288. Registered Office: Fort Henry Business Park, Ballina, Co. Tipperary, Ireland.
Technopath Clinical Diagnostics is a registered business name of Techno-path Manufacturing Limited. LGC Clinical Diagnostics, Maine Standards,
SeraCare, Technopath Clinical Diagnostics, and The Native Antigen Company are trading names of LGC Limited.
lgcclinicaldiagnostics.com
Page 1 of 4
{1}
LGC
CLINICAL DIAGNOSTICS
Science for a Safer World
# Intended Use/Indications for Use
Multichem ID-HIVp24 is intended for use as a qualitative quality control serum to monitor the precision of laboratory testing procedures for the HIV-1 p24 antigen detected by the heterogeneous immunoassay systems listed in the Instructions for Use. These products are not intended to replace manufacturer's recommended controls provided in their package insert. Quality Control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.
# 2. Comparison with Predicate Device
Table 2. General Device Characteristic Similarities
| | Subject Device: | Predicate Device: BK930033 |
| --- | --- | --- |
| | Multichem ID-HIVp24 | VIROTROL HIV-1 Ag |
| Intended Use | Multichem ID-HIVp24 is intended for use as a qualitative quality control to monitor the precision of laboratory testing procedures for the HIV-1 p24 antigen detected by heterogeneous immunoassay systems. These products are not intended to replace manufacturer's recommended controls provided in their package insert. Quality Control materials should be used in accordance with local, state, federal regulations, and accreditation requirements. | VIROTROL HIV-1 Ag is intended for use as an unassayed reactive quality control with in vitro assay procedures for determination of antigens to Human Immunodeficiency Virus Type 1 (HIV-1). This product is intended to provide a means of estimating precision and has the potential for detecting systematic deviations from specific laboratory testing procedures. |
| Form | Liquid | Same |
| Storage | 2 - 8°C Single use only; discard after use. | Same |
| Matrix | Serum | Same |
| Assay Target Analyte | HIV-1 p24 Antigen | HIV-1 p24 antigen |
TECHNOPATH CLINICAL DIAGNOSTICS
lgcclinicaldiagnostics.com
Techno-path Manufacturing Limited. Registered in Ireland No. 419288. Registered Office: Fort Henry Business Park, Ballina, Co. Tipperary, Ireland. Technopath Clinical Diagnostics is a registered business name of Techno-path Manufacturing Limited. LGC Clinical Diagnostics, Maine Standards, SeraCare, Technopath Clinical Diagnostics, and The Native Antigen Company are trading names of LGC Limited.
Page 2 of 4
{2}
LGC
CLINICAL DIAGNOSTICS
Science for a Safer World
Table 3. General Device Characteristic Differences
| | Subject Device: Multichem ID-HIVp24 | Predicate Device: BK930033 VIROTROL HIV-1 Ag |
| --- | --- | --- |
| Open Vial Stability | 30 days at 2 – 8°C | 60 days at 2 – 8°C |
| Shelf Life | 13 months at 2 – 8°C | 18 months at 2 – 8°C |
### 3. Non-Clinical Performance Evaluation
A summary of nonclinical performance testing associated with the Multichem ID-HIVp24 Quality Control is as follows:
#### 4.1 Precision
A 20-day precision study with two runs per day and two replicates per run using three lots of the Multichem ID-HIVp24 quality control was performed on the cobas e801 instrument by a single operator which resulted in 240 replicates (80 replicates per each QC lot). The data obtained were analyzed for the components of Cut-off Index (COI) variation within-run (repeatability), and between-runs. As demonstrated in Table 4a the variability between-run, and within-run was low.
Table 4a. Precision of Multichem ID-HIVp24 on the cobas e801 instrument
| Sample | N | Mean COI | Within-Run (Repeatability) | | Between-Run | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | % CV | SD | % CV |
| HIV p24 | 240 | 5.65 | 0.09 | 1.74 | 0.11 | 1.99 |
N= number of replicates tested, COI = Cut-off Index, SD = Standard Deviation, %CV = percent Coefficient of Variation.
#### 4.2 Reproducibility
A five-day reproducibility study was performed using three lots of the Multichem ID-HIVp24 quality control, with five replicates per lot on the cobas e801 instrument at three different sites, which resulted in a total of 225 replicates (75 replicates per each QC lot). The data obtained were analyzed for the components of Cut-off Index (COI) variation within-run, between-days, between-sites, and between the three QC lots. As demonstrated in Table 4b the percent coefficient of variation (%CV) was low between the various sources of variability.
Table 4b. Reproducibility of Multichem ID-HIV p24 on the cobas e801 instrument
| Sample | N | Mean COI | Within-Run | | Between-Day | | Between-Site | | Between-QC Lots | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| HIV p24 | 225 | 6.08 | 0.87 | 1.41 | 0.87 | 1.51 | 0.58 | 9.68 | 0.58 | 9.55 |
N= number of replicates tested, COI = Cut-off Index, SD = Standard Deviation, %CV = percent Coefficient of Variation.
TECHNOPATH
CLINICAL DIAGNOSTICS
lgcclinicaldiagnostics.com
Techno-path Manufacturing Limited. Registered in Ireland No. 419288. Registered Office: Fort Henry Business Park, Ballina, Co. Tipperary, Ireland. Technopath Clinical Diagnostics is a registered business name of Techno-path Manufacturing Limited. LGC Clinical Diagnostics, Maine Standards, SaraCare, Technopath Clinical Diagnostics, and The Native Antigen Company are trading names of LGC Limited.
Page 3 of 4
{3}
LGC CLINICAL DIAGNOSTICS
Science for a Safer World
#### 4. Conclusions
The Multichem ID-HIVp24 is a serum-based quality control reagent used for monitoring the performance of immunoassay procedures for the qualitative detection of HIV-1 p24 antigen. It is comprised of ready-to-use, barcoded vials that have an expiration of 13 months when stored at 2 – 8°C. The Multichem ID-HIVp24 has the same technology, similar intended use and the data presented demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
lgcclinicaldiagnostics.com
Techno-path Manufacturing Limited. Registered in Ireland No. 419288. Registered Office: Fort Henry Business Park, Ballina, Co. Tipperary, Ireland.
Technopath Clinical Diagnostics is a registered business name of Techno-path Manufacturing Limited. LGC Clinical Diagnostics, Maine Standards, SeraCare, Technopath Clinical Diagnostics, and The Native Antigen Company are trading names of LGC Limited.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.