“HsingChi” LipoSelector Disposable Fat Collection System (LS-1200)
BK251197 · HsingChi Medical Ltd. TAIWAN · QKL · Dec 23, 2025 · General, Plastic Surgery
Device Facts
Record ID
BK251197
Device Name
“HsingChi” LipoSelector Disposable Fat Collection System (LS-1200)
Applicant
HsingChi Medical Ltd. TAIWAN
Product Code
QKL · General, Plastic Surgery
Decision Date
Dec 23, 2025
Decision
SESE
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The “HsingChi” LipoSelector Disposable Fat Collection System (LS-1200) is a sterile medical device intended for the processing of lipoaspirate tissue for the purpose of transferring autologous adipose tissue for aesthetic body contouring (lipofilling) in applications including plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be reimplanted, the harvested fat is only to be used without any additional manipulation.
Device Story
Closed-system disposable canister (1200cc) for processing lipoaspirate; used in healthcare facilities by trained medical staff. Device connects to external liposuction system via tubing; aspirates fat into canister. Internal filter basket (3.0mm) captures large tissue strands; 300μm mesh filter separates impurities (blood, broken fat) via bottom port. Decanting process (8-10 mins) separates tumescent fluid/water, drained through side port. Device does not contact patient directly. Output is processed autologous fat for immediate reimplantation without further manipulation. Disposable design eliminates cleaning/sterilization burden and reduces infection risk.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including biocompatibility (ISO 10993), EO sterilization validation (ISO 11135), package integrity (ASTM F1980, F1886, F1140, F88), and performance testing (fat filtration, container strength, chemical analysis).
Technological Characteristics
Closed-system canister (1200cc) made of PETG, HDPE, PP, and PC. Features double-filter design: 3.0mm filter basket and 300μm mesh filter. EO sterilized. Single-use. Vacuum-operated via external suction machine (max 900mbar).
Indications for Use
Indicated for patients undergoing autologous adipose tissue transfer for aesthetic body contouring in various surgical specialties including plastic, reconstructive, neuro, gastrointestinal, urological, general, orthopedic, gynecological, thoracic, and laparoscopic surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}
# 510(k) Summary - BK251197
1. **Type of Submission:** Traditional
2. **Date of Summary:** December 22, 2025
### 3. Submitter information
**Name:** HsingChi Medical Ltd.
**Address:** No. 10, Ln. 82, Shijian St., Beitou Dist., Taipei City
11276, Taiwan (R.O.C.)
**Phone:** (b) (6)
**Correspondent:** Ting-Hui Lee (matt@hsingchi.com)
### 4. Identification of the subject device
**Trade Name:** 'HsingChi' LipoSelector Disposable Fat
Collection System
**Common Name:** Lipoaspirate washing system for aesthetic
body contouring
**Review Panel:** General & Plastic Surgery
**Regulation Number:** 21 CFR 878.5040
**Regulation Description:** Suction lipoplasty system
**Product Code:** QKL
**Device Class:** Class II
### 5. Identification of the predicate device
**510(k) Number:** BK220691, Formally K101713
**Predicate Trade Name:** LipoCollector II
**Applicant:** Human Med AG
**Regulation Number:** 21 CFR 878.5040
**Regulation Description:** Suction lipoplasty system
**Product Code:** QKL
**Device Class:** Class II
{1}
## 6. Indications for use
The “HsingChi” LipoSelector Disposable Fat Collection System (LS-1200) is a sterile medical device intended for the processing of lipoaspirate tissue for the purpose of transferring autologous adipose tissue for aesthetic body contouring (lipofilling) in applications including plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be reimplanted, the harvested fat is only to be used without any additional manipulation.
## 7. Device description
The “HsingChi” LipoSelector Disposable Fat Collection System (hereafter LipoSelector) is designed for fat harvesting and filtering purpose.
The main structure of LipoSelector: Canister (1200c.c), filter basket, side port and filter mesh, it does not contact with the patient body directly.
LipoSelector is a completely closed system and undergoes ethylene oxide (EO) sterilization, ensuring that the entire liposuction process is free from contamination.
The function principle of LipoSelector in medical institutions typically involves the use a liposuction system to aspirated fat from a patient's body, which is then collected and processed in a canister. The filtration gap of the filter basket is 3.0 mm, larger tissue strands of the aspirate are caught in the filter basket. Turn off the suction machine when you stop the harvesting, then proceed with the decant procedure (around 8-10 mins) until tumescent and fat separation. The tumescent fluid and water are then drained through the side port. The 300 μm filter mesh is fine enough to effectively further separate impurities such as broken fat and blood through the bottom port.
Its disposable nature allows each patient to have their own LipoSelector, eliminating the risk of infection. This also spares nursing staff from the burden of cleaning and sterilizing the fat collection, simplifying the liposuction process.
{2}
### 8. Non-clinical Testing
A series of safety and performance tests were conducted on the subject device, "HsingChi" LipoSelector Disposable Fat Collection System.
# ➤ Biocompatibility testing
The biocompatibility evaluation for the subject device was in accordance with the FDA Biocompatibility guidance (Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process").
We have executed the following test. Test results meet the acceptance criteria.
- Biological evaluation plan and report (ISO10993-1)
- In vitro Cytotoxicity Test (ISO10993-5)
- Guinea Pig Skin Sensitization Study (ISO10993-10, ISO10993-12)
- White Rabbit Intracutaneous Irritation Test (ISO10993-10, ISO10993-12)
- Acute Systemic Injection Study (ISO10993-11, ISO10993-12)
- Pyrogen Study in White Rabbits (ISO10993-11, ISO10993-12, USP<151>)
- Hemolysis Test (ISO10993-4, ASTM F756, ASTM F619)
# ➤ EO Sterilization validation report
A series of tests, including bioburden test, EO/ECH/EG residual test, sterility test has been conducted to validate the EO sterilization of the subject device. Test results showed the validation meet the requirements of ISO 11135 and the achievement of the required sterility assurance level (SAL) of 10⁻⁶.
# ➤ Package integrity testing report
A series of tests was conducted on the subject device after 3 years of accelerated aging to validate package integrity, including visual inspection, package burst test, package creep test, dye penetration test, seal strength test, sterility test has been conducted to validate the package integrity. Test results showed the visual inspection package seal and package integrity of the sterile barrier system can meet the requirements, in accordance with ASTM F1980-16 and ASTM F1886/F1886M. And the package burst test, package creep test, dye penetration test and seal strength test of the sterile barrier system can meet the requirements of relating standards, including ASTM F1140/F1140M-2013, ASTM F88/F88M-15, and ISO11737-2: 2019.
# ➤ Package integrity testing during shipping and handling
{3}
Validation of package integrity during shipping and handling was conducted in accordance with ASTM D4332-22 and ASTM D4169 requirements. A series of tests were conducted on the subject device, including visual inspection, package burst test, package creep test, dye penetration test, seal strength test, sterility test to validate the package integrity. Test results showed the visual inspection of the sterile barrier system can meet the requirements, in accordance with ASTM F1886/F1886M-16. And the package burst test, package creep test, dye penetration test, seal strength test and sterility test of the sterile barrier system can meet the requirements of relating standards, including ASTM F1140/F1140M-13(R2020), ASTM F1929-23, ASTM F88/F88M-23, and ISO11737-2: 2019.
# ➤ **Reliability and performance testing**
A series of safety and tests were conducted on the subject device. Testing are conducted to support the shelf life of three years when stored at 25 ± 2 °C, RH of 45%.
- Performance test of Fat filtration
- Performance test of Visible
- Performance test of Air Leakage
- Performance test of Filtering
- Performance test of Connector
- Performance test of Foam Preventing
- Test for Container Strength
- Physical and Chemical Analysis
(e.g. pH-value, Residue on evaporation, Heavy metals, Phthalates content test, Reducing (oxidizable) matter, UV absorption)
All the test results demonstrate “HsingChi” LipoSelector Disposable Fat Collection System meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.
# 9. Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
{4}
## 10. Substantial equivalence comparison
“HsingChi” LipoSelector Disposable Fat Collection System submitted in 510(k) files is substantially equivalent in intended use, safety and performance to the FDA cleared LipoCollector II (BK220691). Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
| Item | Subject device | Predicate device | Substantial equivalence determination |
| --- | --- | --- | --- |
| Proprietary name | “HsingChi” LipoSelector Disposable Fat Collection System | LipoCollector II | |
| 510(k) No. | BK251197 | BK220691 | |
| Product code | QKL | QKL | Same. |
| Product class | 2 | 2 | Same. |
| Indications for use | The “HsingChi” LipoSelector Disposable Fat Collection System (LS-1200) is a sterile medical device intended for the processing of lipoaspirate tissue for the purpose of transferring autologous adipose tissue for aesthetic body contouring (lipofilling) in applications including plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be reimplanted, the harvested fat is only to be used without any additional manipulation. | The **LipoCollector II** is for harvesting, filtering and transferring of autologous fat tissue for aesthetic body contouring. | Similar. |
| User population | It is designed to be operated by trained medical staff. | It is designed to be operated by trained medical staff. | Same. |
{5}
| Item | Subject device | Predicate device | Substantial equivalence determination |
| --- | --- | --- | --- |
| Proprietary name | “HsingChi” LipoSelector Disposable Fat Collection System | LipoCollector II | |
| 510(k) No. | BK251197 | BK220691 | |
| Environment of use | It is intended for use in healthcare facilities. | It is intended for use in healthcare facilities. | Same. |
| Interaction with patient / Anatomical location of use | It will not contact with patient directly. The anatomical location will be determined by the medical staff based on clinical needs. | It will not contact with patient directly. The anatomical location will be determined by the medical staff based on clinical needs. | Same. |
| Design principles | Utilizing vacuum (liposuction devices) to aspirate adipose tissue from the patient and subsequently transport the adipose tissue through a tube into a collection container that contains a filtering mechanism to allow fluids to pass, but retains the adipose tissue within the container. They all consist of a polymeric tissue collection container with a filter unit within the chamber, and all have connection ports on the superior ends of the collection container for the attachment of suction tubing. | Utilizing vacuum (liposuction devices) to aspirate adipose tissue from the patient and subsequently transport the adipose tissue through a tube into a collection container that contains a filtering mechanism to allow fluids to pass, but retains the adipose tissue within the container. They all consist of a polymeric tissue collection container with a filter unit within the chamber, and all have connection ports on the superior ends of the collection container for the attachment of suction tubing. | Same. |
| Filter design | Double filter | Double filter | Same. |
| Mesh filter | 300 μm | Three sizes of mesh filters are supplied with the LipoCollector II (200 μm, 250 μm and 315 μm). | Different. The difference doesn’t raise any new issues of substantial equivalence. |
{6}
| Item | Subject device | Predicate device | Substantial equivalence determination |
| --- | --- | --- | --- |
| Proprietary name | “HsingChi” LipoSelector Disposable Fat Collection System | LipoCollector II | |
| 510(k) No. | BK251197 | BK220691 | |
| **Limitation of suction machine** | Maximum vacuum 900mbar | Not setting the maximum vacuum | Different. We have conducted testing to validate the strength of the collection container is not imploding, crack. The difference doesn't raise any new issues of substantial equivalence. |
| **Material** | PETG, HDPE, PP, PC | Made by polycarbonates (PC), stainless steel | Different. Both devices are Fulfill ISO10993 requirement. The difference doesn't raise any new issues of substantial equivalence. |
| **Sterilization method** | EO Sterilization | EO Sterilization | Same. |
| **Volume Range** | 1,200ml | 1,000ml | Different. The difference doesn't raise any new issues of substantial equivalence. |
| **Properties** | Single use only and closed system | Reusable | Different. |
{7}
| Item | Subject device | Predicate device | Substantial equivalence determination |
| --- | --- | --- | --- |
| Proprietary name | “HsingChi” LipoSelector Disposable Fat Collection System | LipoCollector II | |
| 510(k) No. | BK251197 | BK220691 | |
| | | | The difference doesn't raise any new issues of substantial equivalence. |
### 11. Similarity and difference
The “HsingChi” LipoSelector Disposable Fat Collection System has been compared with the LipoCollector II (BK220691). The subject device has the same intended use, environment of use, user population, interaction with patient / anatomical location of use, sterilization method, design principles as the predicate devices.
Although the reuse type, size of mesh filter, volume range and limitation of suction machine are different between the subject device and predicate devices. The subject device has undergone a series of testing, and the results passed the acceptance criteria defined by the testing standards used. It has proven the differences between the subject device and the predicate device do not raise any new issues on safety and effectiveness. The subject device is substantially equivalent to the predicate devices as it claims.
### 12. Conclusion
The subject device has tested safety and performance, and the test results complied with the test requests. Therefore, the differences between subject device and predicate device do not raise any problems of safety or effectiveness. We, HsingChi Medical Ltd., believe that “HsingChi” LipoSelector Disposable Fat Collection System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.