← Product Code [KSZ](/productcode/KSZ) · BK251183

# Erytra (BK251183)

_Diagnostic Grifols, S.A. Barcelona, Spain · KSZ · Jun 17, 2025 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/BK251183

## Device Facts

- **Applicant:** Diagnostic Grifols, S.A. Barcelona, Spain
- **Product Code:** [KSZ](/productcode/KSZ.md)
- **Decision Date:** Jun 17, 2025
- **Decision:** SESE
- **Regulation:** 21 CFR 864.9175
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The Erytra is a fully automated high-throughput analyzer designed to automate in vitro immunohematological testing of human blood utilizing DG Gel 8 card technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), the Erytra automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing.

## Device Story

Erytra is a fully automated high-throughput analyzer for gel immunohematology testing; processes human blood samples (plasma, serum, red blood cells) using DG Gel 8 cards. System automates sample/reagent loading, incubation, centrifugation, and digital image processing of test results. Operated by laboratory personnel in clinical/transfusion settings; interfaces with LIS for data management. Device performs ABO/Rh grouping, antigen typing, antibody screening/identification/titration, and crossmatching. Automated serial dilution method supports titration up to 1/2048; supports single point and range titration techniques. Standardizes analytical steps to eliminate manual handling errors, improve traceability, and reduce operator exposure to biohazards. Output provides objective interpretation of immunohematological results, facilitating clinical decision-making for transfusion compatibility and patient blood management.

## Clinical Evidence

Bench testing only. In-house method comparison study (n=243 serial dilution titrations, n=76 range titrations, n=356 single dilution tests) compared Erytra against manual serial double dilution methods. Results showed high concordance (97.4%–100% agreement) for Anti-Human Globulin and Neutral Gel cards. Reproducibility/repeatability study (6 samples x 3 analyzers x 5 days x 2 runs x 2 replicates) confirmed consistent performance. All acceptance criteria (≥95% agreement) met, with discrepancies resolved in favor of the subject device.

## Technological Characteristics

Automated high-throughput analyzer; uses DG Gel 8 card technology. Features 2 pipetting units, 4 independent incubators (12 cards each), 2 independent centrifuges (12 cards each). Barcode scanning (Interleaved 2 of 5, Code 3 of 9, Codabar, EAN/JAN, Code 128). Connectivity via LIS interface. Software-driven digital image processing for result interpretation. 7-year useful life. Sterilization/materials not specified.

## Regulatory Identification

An automated blood grouping and antibody test system is a device used to group erythrocytes (red blood cells) and to detect antibodies to blood group antigens.

## Predicate Devices

- Erytra ([BK220751](/device/BK220751.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra

5 - 1

# **CH2.03 – 510(k) Summary**

# **Contact Details:**

Name Diagnostic Grifols S.A.
Address Passeig Fluvial, 24, Parets del Valles
Barcelona, 08150, Spain
Establishment Registration 3002772505
Number:
Contact Person: Elvira Estapé Egea
Regulatory Affairs Manager
elvira.estape@grifols.com
Phone: (34) 670-922-237
Date of Summary: June 13th, 2025

# **Device Name:**

Trade Name: Erytra
Classification Name: Automated Blood Grouping and Antibody Test System
Device Class: II
Product Code: KSZ
Regulation Number: 21 CFR 864.9175

# **Legally Marketed Predicate Device:**

Trade Name: Erytra
Classification Name: Automated Blood Grouping and Antibody Test System
510(k) Number: BK220751
Device Class: II
Product Code: KSZ
Regulation Number: 21 CFR 864.9175
Clearance Letter November 1, 2022

# **Device Description:**

The Erytra is designed to automate all necessary operations and procedures to process gel immunohematology tests, allowing laboratories to:

1. Create test profiles and optimize profile implementation in the shortest time and with the most accurate results possible.
2. Increase process safety and traceability by eliminating possible identification and transcription errors.

{1}

Diagnostic Grifols, S.A.

510(k) Premarket Notification

Erytra

5 - 2

3. Increase analytical reliability by standardizing all steps, eliminating possible handling, and processing errors, and interpreting the results with objective criteria.
4. Reduce the danger of contamination for Operators by reducing Operator interaction with the samples and reagents during the analytical process. Operator interaction is limited to the loading and unloading of the analyzer.

In addition, the Erytra adapts to the needs and differing operational workflows in immunohematology laboratories, donation centers, transfusion centers, and clinical testing laboratories, as well as different work rhythms (routine, emergency) and the flow of samples processed over different shifts.

The Erytra automates the following gel immunohematology tests:

- ABO Red Cell and Serum Grouping.
- Rh(D) Typing.
- Antigen Typing.
- Antibody detection.
- Antibody identification.
- Antibody titration.
- Direct Antiglobulin test.
- Compatibility Tests (Crossmatching).

### Indications for Use:

The Erytra is a fully automated high-throughput analyzer designed to automate in vitro immunohematological testing of human blood utilizing DG Gel 8 card technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests.

As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), the Erytra automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing.

{2}

Diagnostic Grifols, S.A.  
510(k) Premarket Notification  
Erytra

5 - 3

# **Comparison to Predicate Device:**

|  Parameter | Predicate Device Diagnostic Grifols S.A. Erytra (BK220751) | Subject Device Diagnostic Grifols S.A. Erytra  |
| --- | --- | --- |
|  **General**  |   |   |
|  **Indications for Use Statement** | The Erytra is a fully automated high-throughput analyzer designed to automate *in vitro* immunohematological testing of human blood utilizing DG Gel 8 card technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), the Erytra automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing. | The Erytra is a fully automated high-throughput analyzer designed to automate *in vitro* immunohematological testing of human blood utilizing DG Gel 8 card technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), the Erytra automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing.  |
|  **Classification** | II | II  |
|  **Product Code** | KSZ | KSZ  |
|  **Regulation number** | 21 CFR 864.9175 | 21 CFR 864.9175  |
|  **Common name** | Automated Blood Grouping and Antibody Test System | Automated Blood Grouping and Antibody Test System  |
|  **Test performed** | - ABO Red Cell and Serum Grouping - Rh(D) Typing - Antigen Typing - Antibody detection - Antibody identification - Antibody titration - Direct Antiglobulin test - Compatibility Tests (Crossmatching) | - ABO Red Cell and Serum Grouping - Rh(D) Typing - Antigen Typing - Antibody detection - Antibody identification - Antibody titration (including Single Point and Range Titration techniques) - Direct Antiglobulin test - Compatibility Tests (Crossmatching)  |
|  **Primary components** | Analyzer Computer Software Optional hand-held bar code scanner Optional printer | Analyzer Computer Software Optional hand-held bar code scanner Optional printer  |
|  **Specimen Types** | Plasma, Serum and Red Blood Cells. | Plasma, Serum and Red Blood Cells.  |
|  **Reagents** | Erytra Eflexis is used with Diagnostic Grifols DG Gel 8 cards, Medion Grifols Diagnostics Reagent Red Blood Cells and Validated Antisera Reagents. | Erytra Eflexis is used with Diagnostic Grifols DG Gel 8 cards, Medion Grifols Diagnostics Reagent Red Blood Cells and Validated Antisera Reagents.  |

{3}

Diagnostic Grifols, S.A.  
510(k) Premarket Notification  
Erytra

5 - 4

|  Parameter | Predicate Device Diagnostic Grifols S.A. Erytra (BK220751) | Subject Device Diagnostic Grifols S.A. Erytra  |
| --- | --- | --- |
|  **Positive identification of samples and reagents** | Yes | Yes  |
|  **Throughput** | 67 samples (ABO/Rh cards) per hour, including forward & reverse group | 67 samples (ABO/Rh cards) per hour, including forward & reverse group  |
|  **Useful life** | 7 years, considering 8 hours a day of continuous operation at maximum throughput (using ABO/Rh DG Gel 8 cards), 250 days a year. | 7 years, considering 8 hours a day of continuous operation at maximum throughput (using ABO/Rh DG Gel 8 cards), 250 days a year.  |
|  **Hardware**  |   |   |
|  **Reagent Red Cell suspension** | Maintained by Rotation Movement | Maintained by Rotation Movement  |
|  **Barcode Symbologies** | - Interleaved 2 of 5. - Code 3 of 9. - Codabar. - EAN 8 / EAN 13 / JAN 8 / JAN 13. - Codes 128 A, B & C (ISBT 128). - Others under configuration. | - Interleaved 2 of 5. - Code 3 of 9. - Codabar. - EAN 8 / EAN 13 / JAN 8 / JAN 13. - Codes 128 A, B & C (ISBT 128). - Others under configuration.  |
|  **Sample loading capacity** | 96 tubes distributed in four drawers. | 96 tubes distributed in four drawers.  |
|  **Reagent loading capacity** | 54 vials (36 of them agitated) distributed in two drawers. | 54 vials (36 of them agitated) distributed in two drawers.  |
|  **Sample/Reagent Dispensing (pipetting) Unit** | 2 | 2  |
|  **Card loading capacity** | 400 cards | 400 cards  |
|  **Incubator** | 4 independent incubators for 12 cards each. | 4 independent incubators for 12 cards each.  |
|  **Centrifuge** | 2 independent centrifuges for 12 cards each. | 2 independent centrifuges for 12 cards each.  |
|  **System solution and waste containers** | 1 container for Grifols Wash Solution A 1 container for Grifols Wash Solution B 2 containers for waste solution 1 container for processed DG Gel 8 card disposal Option to double the capacity of Grifols Wash Solution A and B with the liquid waste drainage configuration activated | 1 container for Grifols Wash Solution A 1 container for Grifols Wash Solution B 2 containers for waste solution 1 container for processed DG Gel 8 card disposal Option to double the capacity of Grifols Wash Solution A and B with the liquid waste drainage configuration activated  |

{4}

Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra

5 - 5

The differences between the Erytra and its predicate device, the Erytra (BK220751), are related to the Antibody Titration technique. The Erytra uses an automated serial dilution method that allows for the preparation of dilutions up to 1/2048. Additionally, the Erytra supports both single point and range titration techniques within antibody titration tests, when required by the user, in comparison to the predicate device. These differences do not raise new concerns of safety and effectiveness.

**Performance:**

All new risks and hazard analysis related to the automated serial dilution method, and single dilution and range titration techniques, have been performed and documented per ISO 14971 guidelines.

An in-house method comparison study was conducted with three Erytra, and with DG Gel 8 Anti-IgG (Rabbit) and DG Gel 8 Neutral cards, and 0.8% Reagent Red Blood Cells. The following methods were used as comparative methods:

- Serial dilution titration technique: the current FDA approved automated single direct dilution method was used as comparative for titers up to 64, and the manual serial double dilution method was used as comparative for titers higher than 64.
- Single dilution and range titration techniques: the manual serial double dilution method was used as comparative for high range titrations.

A total of 243 titrations were compared for serial dilution titration techniques and out of 76 samples were used for range titration techniques comparison. Additionally, a total of 356 single dilution tests were compared with the manual method.

The percent agreements and the lower limits of 95% one-side confidence interval (CI) for Erytra and antibody titration technique are indicated in the following table:

|  Antibody Titration | N° of samples | Percent Agreement (Lower 95% CI)  |
| --- | --- | --- |
|  **Concordances Obtained for Anti-Human Globulin**  |   |   |
|  Serial Dilution Technique | 115 | 100% (97.4%)  |
|  Ranges Technique | 74 | 100% (96.0%)  |
|  Single Point | 201 | 100% (98.5%)  |
|  **Concordances Obtained for the Neutral Gel**  |   |   |
|  Serial Dilution Technique | 128 | 100% (97.7%)  |
|  Ranges Technique | 75 | 100% (96.1%)  |
|  Single Point | 155 | 97.4% (94.2%)  |

{5}

Diagnostic Grifols, S.A.  
510(k) Premarket Notification  
Erytra

5 - 6

The percent agreement (estimated with a 95% CI) between the three antibody titration techniques performed by Erytra and with the respective comparative methods met all the acceptance criteria previously established as $\geq 95\%$ for DG Gel 8 Anti-IgG (Rabbit) and for DG Gel 8 Neutral cards. The discrepancies found caused the single point antibody titration technique for the Neutral Gel to not meet the acceptance criteria. However, the 4 discrepancies were resolved in favor of the Erytra.

The results obtained in the Method comparison study supported the conclusion that the Erytra including the new automated serial dilution method, and single dilution and range titration techniques, yielded equivalent results to the comparative methods using FDA licenses reagents and instruments.

In addition, an internal reproducibility and repeatability study was conducted. The study was done using three Erytra in accordance to the following profile: 6 Samples x 3 Erytra x 5 days x 2 runs (am/pm) x 2 replicates.

The study demonstrated that the Erytra consistently obtained the expected results and the reproducibility and repeatability of the results for these techniques was deemed demonstrated.

In summary, the results of the in-house evaluation supported the conclusion that the Erytra including the new automated serial dilution method, and single dilution and range titration techniques, can perform antibody titrations with its DG Gel 8 cards and Medion Grifols Diagnostics Reagent Red Blood Cells, safely and effectively. In addition, it is concluded that the Erytra is substantially equivalent when compared to the results obtained using the same DG Gel 8 cards and Medion Grifols Diagnostics Reagent Red Blood Cells tested by the comparative methods.

### **Conclusions:**

Diagnostic Grifols S.A. concludes, based on all the information submitted and discussed in this submission and in this summary that the Erytra, when used for the defined indications for use, performs as well as or better than the legally marketed predicate device the Erytra (BK220751). In addition, all the requirements for a product to be marketed in the United States have been demonstrated.

---

**Source:** [https://fda-staging.innolitics.com/device/BK251183](https://fda-staging.innolitics.com/device/BK251183)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).