BK251157 · B.T.I. Biotechnology Institute, S.L. Miñano (Álava), Spain · ORG · Oct 3, 2025 · Hematology
Device Facts
Record ID
BK251157
Device Name
ENDORET® Kit
Applicant
B.T.I. Biotechnology Institute, S.L. Miñano (Álava), Spain
Product Code
ORG · Hematology
Decision Date
Oct 3, 2025
Decision
SESE
Regulation
21 CFR 864.9245
Device Class
Class 2
Indications for Use
The ENDORET® Kit is designed to be used for the safe and rapid preparation of autologous platelet-rich-plasma (PRP) from a small sample of peripheral blood at the patient’s point of care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics.
Device Story
ENDORET® Kit facilitates point-of-care preparation of autologous platelet-rich-plasma (PRP); utilizes vacuum-based blood collection and centrifugation to separate blood components by density; includes collection tubes, needles, plasma transfer device, fractionation tubes, and activator agent. Healthcare professionals operate the system; PRP output is mixed with bone graft material to improve handling for bony defect applications. Benefits include rapid, safe autologous PRP preparation. Device components are medical-grade polymers, elastomers, rubbers, and stainless steel.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence supported by bench testing, including sterilization validation (ISO 11137-1, ISO 11137-2, ISO 17665-1), biocompatibility (ISO 10993 series), sterile barrier/shelf-life validation (ISO 11607-1, ASTM standards), and comparative performance testing of PRP parameters (platelet concentration, yield, pH, aggregation, activation, and clot integrity).
Technological Characteristics
Materials: medical grade polymers, elastomers, rubbers, stainless steel. Principle: density-based blood fractionation via centrifugation. Energy: external centrifuge (sold separately). Connectivity: none. Sterilization: irradiation, ethylene oxide, or autoclave (SAL 10⁻⁶). Shelf-life: 18 months. Standards: ISO 11137-1/2, ISO 17665-1, ISO 10993-4/5/10/11, ISO 11607-1, ASTM F88, F1886, F1929, F1140, F2096.
Indications for Use
Indicated for patients requiring autologous platelet-rich-plasma (PRP) preparation from peripheral blood for mixing with autograft or allograft bone to improve handling characteristics during application to bony defects.
Regulatory Classification
Identification
An automated blood cell separator is a device that uses a centrifugal or filtration separation principle to automatically withdraw whole blood from a donor, separate the whole blood into blood components, collect one or more of the blood components, and return to the donor the remainder of the whole blood and blood components. The automated blood cell separator device is intended for routine collection of blood and blood components for transfusion or further manufacturing use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is a guidance for industry and FDA staff entitled “Class II Special Controls Guidance Document: Automated Blood Cell Separator Device Operating by Centrifugal or Filtration Separation Principle.”
{0}
# 510(k) SUMMARY – BK251157
## 1. SUBMITTER INFORMATION
| Name and address: | B.T.I. Biotechnology Institute, S.L. Parque Tecnológico de Álava C/Leonardo Da Vinci 14 01510 Miñano (Álava), Spain |
| --- | --- |
| Phone: | (+34) 945 297 030 |
| Fax: | (+34) 945 297 031 |
| Contact person: | David Fernández, Regulatory Manager |
| Date prepared: | September 18, 2025 |
## 2. DEVICE NAME AND CLASSIFICATION
| Name of device: | ENDORET® Kit |
| --- | --- |
| Regulation Number: | 21 CFR 864.9245 |
| Regulation Description: | Automated blood cell separator |
| Regulatory Class: | Class II |
| Product Code: | ORG |
| Classification Panel: | Hematology |
## 3. PREDICATE AND REFERENCE DEVICES
Primary predicate device:
- BK150294 ENDORET® Kit, B.T.I. Biotechnology Institute, S.L.
Reference device:
- BK130049 ENDORET® Kit, B.T.I. Biotechnology Institute, S.L.
## 4. INDICATIONS FOR USE
The ENDORET® Kit is designed to be used for the safe and rapid preparation of autologous platelet-rich-plasma (PRP) from a small sample of peripheral blood at the patient’s point of care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics.
## 5. DEVICE DESCRIPTION
The ENDORET® Kit includes sterile, single use components to draw blood by vacuum from at a patient’s point-of-care and separate the blood components by centrifugation to create a platelet-rich plasma (PRP) output, that is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. The ENDORET® Kit is intended to be used under the supervision of a healthcare professional.
The ENDORET® kit includes blood collection tubes with anticoagulant, blood collection needles, Plasma Transfer Device to collect the PRP fraction in sterile fractionation tubes, and an activator agent that is added to the PRP prior to its application. This ENDORET® Kit includes modified or additional components.
1
{1}
The materials of the ENDORET® Kit components are medical grade polymers, elastomers, rubbers and stainless steel suitable for use in medical devices. These components are provided sterile by irradiation, ethylene oxide or autoclave.
Additional equipment used for blood centrifugation is available and sold separately to the end-user in order to operate the device as intended.
## 6. PERFORMANCE DATA
Non-clinical testing information submitted in order to demonstrate substantial equivalence of the proposed device include:
- Sterilization validation for subject device components, provided sterile, to a sterility assurance level (SAL) of 10⁻⁶, according to ISO 11137-1, ISO 11137-2 and ISO 17665-1.
- Biocompatibility testing for new or modified device components, according to ISO 10993-4, ISO 10993-5, ISO 10993-10, and ISO10993-11.
- Sterile Barrier/Shelf-Life validation according to ISO 11607-1, ASTM F88, ASTM F1886/F1886M, ASTM 1929, ASTM F1140/F1140M and ASTM F2096.
- Comparative performance bench testing, including measurements of the following parameters:
- Platelet concentration factor
- Platelet yield
- pH
- platelet aggregation
- platelet activation
- clot integrity of the composite clot formed with bone graft material and PRP.
The results obtained in the study demonstrate substantial equivalence performance between the PRP obtained with subject and predicate ENDORET® kit.
No clinical data were included in this submission.
## 7. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE DISCUSSION
The proposed device is substantially equivalent in Indications for Use, operating principle and design to the primary predicate and reference devices identified above. Table 1 below compares the subject device to the predicate and reference devices indications for use and technological characteristics.
The subject ENDORET® Kit is substantially equivalent to the ENDORET® kit currently cleared in BK150294 in design, materials, manufacturing processes, sterilization and shelf-life. Reference device ENDORET® kit, cleared in BK130294 is in support of fractionation tube included in selected subject device models.
## 8. CONCLUSION
The subject device components have identical Indications for use and equivalent technological characteristics of primary predicate and reference device identified above. The subject device has identical intended use and shares basic design features with the already cleared ENDORET® Kits. The subject device, predicate device and reference device include components to obtain a Platelet-rich Plasma output by centrifugation. The subject device includes modified or additional components that do not raise new concerns of safety and effectiveness.
The information included in this submission demonstrates that the subject ENDORET® Kit is substantially equivalent to the identified predicate and reference devices.
2
{2}
**Table 1.** Substantial Equivalence – Indications for use and technological characteristics comparison
| FEATURE | SUBJECT DEVICE | PREDICATE DEVICE | REFERENCE DEVICE | ASSESSMENT |
| --- | --- | --- | --- | --- |
| **Name** | **ENDORET® Kit** | **ENDORET® Kit (BK150294)** | **ENDORET® Kit (BK130049)** | (not applicable) |
| **Product Classification** | Product Code: ORG Regulation No.: 21 CFR 864.9245 Device Class II | Product Code: ORG Regulation No.: 21 CFR 864.9245 Device Class II | Product Code: ORG Regulation No.: 21 CFR 864.9245 Device Class II | Identical. |
| **Indications for Use (STATEMENT)** | The ENDORET® Kit is designed to be used for the safe and rapid preparation of autologous platelet-rich-plasma (PRP) from a small sample of peripheral blood at the patient's point of care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. | The ENDORET® KIT is designed to be used for the safe and rapid preparation of autologous platelet-rich-plasma (PRP) from a small sample of peripheral blood at the patient's point of care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. | The system is designed to be used for the safe and rapid preparation of autologous platelet-rich-plasma (PRP) from a small sample of peripheral blood at the patient's point of care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. | Identical to BK150294. Very similar wording is included in BK130049, for same intended use. |
| **Principle of operation** | Separation of blood fractions by density. | Separation of blood fractions by density. | Separation of blood fractions by density. | Identical. |
| **Blood Component separation method** | Centrifugation | Centrifugation | Centrifugation | Identical. |
| **Device components** | Blood collection tubes, anticoagulated. | Blood collection tubes, anticoagulated. | Blood collection tubes, anticoagulated. | Equivalent. Very similar or identical components included in predicate and reference device are included in subject device. Identical anticoagulant and activator agents are used in subject, predicate and reference devices. The additional components inclusion does not modify the intended use of the device. |
| | Blood collection set. | Blood collection set. | Blood collection set. | |
| | Plasma Transfer Device. | Plasma Transfer Device. | Plasma Transfer Device. | |
| | Fractionation tubes (evacuated and non-evacuated). | Fractionation tubes (non-evacuated). | Fractionation tubes (evacuated). | |
| | ENDORET Activator. | ENDORET Activator. | ENDORET Activator. | |
| | Syringe for Activator. | Syringe for Activator. | Syringe for Activator. | |
| **Sterilization** | Additional caps, connectors and needles. | - | - | Identical |
| | Device components provided sterile. | Device components provided sterile. | Device components provided sterile. | |
| **Shelf-life** | 18 months | 12 months | 12 months | Equivalent. Shelf-life analyses performed and provided in this submission support this extended shelf life for subject device. |
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.