BK230980 · Becton, Dickinson & Company Franklin Lakes, NJ · GIM · Apr 17, 2024 · Clinical Chemistry
Device Facts
Record ID
BK230980
Device Name
BD Vacutainer® K2EDTA Blood Collection Tubes
Applicant
Becton, Dickinson & Company Franklin Lakes, NJ
Product Code
GIM · Clinical Chemistry
Decision Date
Apr 17, 2024
Decision
SESE
Regulation
21 CFR 862.1675
Device Class
Class 2
Attributes
Pediatric
Indications for Use
BD Vacutainer® K₂EDTA Blood Collection Tubes are plastic, evacuated, sterile, single use, in vitro diagnostic medical devices. They are intended to be used by trained healthcare professionals for the collection, containment, preservation, and transport of human venous blood specimens used for in vitro diagnostic testing. BD Vacutainer® K₂EDTA Blood Collection Tubes may be used for immunohematology testing, such as ABO blood grouping and Rh typing. The performance characteristics of these tubes have not been established for immunohematology testing in general; therefore, users must validate the use of these tubes for their specific assay-instrument/reagent system combinations and specimen storage conditions.
Device Story
Evacuated plastic tube containing spray-dried K₂EDTA anticoagulant; used for collection, containment, preservation, and transport of human venous blood. Operated by trained healthcare professionals in clinical settings. Tube features Hemogard™ safety closure to reduce user blood exposure. Input is venous blood; output is a contained specimen for in vitro diagnostic testing. Device facilitates clinical decision-making by ensuring specimen integrity for immunohematology assays. Benefits include reduced risk of breakage compared to glass and enhanced safety via closed-system design.
Clinical Evidence
Clinical equivalence demonstrated via comparative testing of blood collected in subject vs. comparator devices. Evaluated immunohematology parameters (ABO grouping/Rh typing) in a cohort representative of the US population. Additional studies confirmed within-tube stability, shelf-life performance, and repeatability/reproducibility. All results met pre-determined acceptance criteria.
Indicated for all patient populations requiring venous blood collection for in vitro diagnostic testing, specifically immunohematology testing including ABO blood grouping and Rh typing.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}
BD Vacutainer® K₂EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
## 5 510(K) SUMMARY
### 5.1 Device Name:
BD Vacutainer® K₂EDTA Blood Collection Tubes
### 5.2 Summary Preparation Date:
November 1, 2024
### 5.3 Submitted by:
Becton, Dickinson and Company
1 Becton Drive
Franklin Lakes, NJ 07417-1885
Phone: (201) 847-6800
### 5.4 Contact:
Alexandra Kirby
Staff Regulatory Affairs Specialist
Email: Alexandra.Kirby@bd.com
Phone: (b) (6)
### 5.5 Proprietary Names:
BD Vacutainer® K₂EDTA Blood Collection Tubes
### 5.6 Common or Usual Names:
EDTA Blood Collection Tube
### 5.7 Regulatory Information:
Classification Name: Tubes, Vacuum Sample, With Anticoagulant
Classification Regulation: 21 CFR § 862.1675
Regulatory Class: Class II
Product Code: GIM
### 5.8 Predicate Device(s):
BK050036 BD Vacutainer® Plus K₂EDTA Tubes
### 5.9 Device Establishment
Becton, Dickinson and Company
18
{1}
BD Vacutainer® K₂EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
**5.10 Registration Number:**
2243072
**5.11 Performance Standards:**
- ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
- ANSI AAMI ISO 11137-1:2006/(R)2015 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2019)]
- ANSI AAMI ISO 11137-2:2013/(R)2019 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
- ANSI/AAMI/ISO 11137-3:2017 Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine control
- ANSI/AAMI/ISO 11737-1:2018 Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products
- ANSI/AAMI/ISO 11737-2:2019 Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ANSI AAMI ST67:2019 Sterilization of health care products - Requirements and guidance for selecting a sterility assurance level (SAL) for products labeled "sterile"
- EN ISO 14971:2019 Medical Devices – Application of risk management to medical devices
**5.12 Intended Use**
BD Vacutainer® K₂EDTA Blood Collection Tubes are plastic, evacuated, sterile, single use, *in vitro* diagnostic medical devices. They are intended to be used by trained healthcare professionals for the collection, containment, preservation, and transport of human venous blood specimens used for *in vitro* diagnostic testing.
BD Vacutainer® K₂EDTA Blood Collection Tubes may be used for immunohematology testing, such as ABO blood grouping and Rh typing.
The performance characteristics of these tubes have not been established for immunohematology testing in general; therefore, users must validate the use of these tubes for their specific assay-instrument/reagent system combinations and specimen storage conditions.
19
{2}
BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
### 5.13 Device Description
BD Vacutainer® K$_{2}$ EDTA Blood Collection Tubes are used for collection, containment, preservation, and transport of human venous blood specimens in a closed tube. The evacuated blood collection tube consists of a Vacutainer® Hemogard™ Closure Assembly, a plastic tube and EDTA additive. The EDTA anticoagulant is spray coated in the dipotassium (K$_{2}$) form. The EDTA additive prevents specimen coagulation.
The plastic tubes are closed with the Vacutainer® Hemogard™ Closure Assembly which consists of a rubber stopper and protective plastic shield. The Hemogard™ Closure Assembly, introduced in 1995, is intended to help reduce user exposure to blood. All stopper/closures are color coded to reflect additive type; the closures included in this submission are lavender to indicate the presence of the EDTA additive.
The plastic tube is manufactured from Polyethylene terephthalate (PET). These plastic tubes were introduced in 1990. Plastic tubes enhance user safety and disposal because of the reduced risk of tube breakage and the availability of incineration as a method of disposal.
### 5.14 Substantial Equivalence
The subject and predicate device are substantially equivalent as described in Table 2.
20
{3}
BD Vacutainer® K2EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
**Table 2. Substantial Equivalence Comparison**
| Characteristic | Subject Device BD Vacutainer® K2EDTA Blood Collection Tubes | Predicate Device BD Vacutainer® Plus K2EDTA Tubes BK050036 | Comments |
| --- | --- | --- | --- |
| Indication for use | BD Vacutainer® K2EDTA Blood Collection Tubes are plastic, evacuated, sterile, single use, *in vitro* diagnostic medical devices. They are intended to be used by trained healthcare professionals for the collection, containment, preservation, and transport of human venous blood specimens used for *in vitro* diagnostic testing. BD Vacutainer® K2EDTA Blood Collection Tubes may be used for immunohematology testing, such as ABO blood grouping and Rh typing. The performance characteristics of these tubes have not been established for immunohematology testing in general; therefore, users must validate the use of these tubes for their specific assay-instrument/reagent system combinations and specimen storage conditions. | BD Vacutainer® Plus Serum Tubes, BD Vacutainer® SST™ Plus Tubes, BD Vacutainer® SST™ Glass Tubes, and BD Vacutainer® SST™ II Advance Tubes may be used for routine blood donor screening and diagnostic testing of serum for infectious disease. The performance characteristics of these tubes have not been established for infectious disease testing in general; therefore, users must validate the use of these tubes for their specific assay-instrument/reagent system combinations and specimen storage conditions. BD Vacutainer® Plus Serum and BD Vacutainer® Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening. The performance characteristics of these tubes have not been established for immunohematology testing in general; therefore, users must validate the use of these tubes for their specific assay-instrument/reagent system combinations and specimen storage conditions. | The proposed indication for use is the same as the predicate with regards to the claim for immunohematology testing and the performance characteristics disclaimer, with the exception of more explicitly stating the intended user (trained healthcare professional) and removing blood donor screening. The subject device indication for use provides additional clarity on certain information consistent with current best practices. Therefore, the subject and predicate devices have the same intended use. |
| Intended Population | General Use – all populations | General Use – all populations | Identical. |
21
{4}
# BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
| Characteristic | Subject Device BD Vacutainer® K_{2}EDTA Blood Collection Tubes | Predicate Device BD Vacutainer® Plus K_{2}EDTA Tubes BK050036 | Comments |
| --- | --- | --- | --- |
| Evacuated Blood Collection Tube | Yes | Yes | Identical. |
| Clot/Anti- coagulation | Anti-coagulation | Anti-coagulation | Identical. |
| Additive Type | K_{2}EDTA | K_{2}EDTA | Identical. |
| Additive Quantity | 1.8 mg/mL | 1.8 mg/mL | Identical. |
| Tube Dimensions (mm) | 13x75, 13x100, 16x100 | 13x75, 13x100, 16x100 | Identical. |
| Draw Volume | 2, 3, 4, 6, 10 mL | 2, 3, 4, 6, 10 mL | Identical. |
| Tube Material | Plastic | Plastic | Identical. |
| Tube Closure | Hemogard™ Safety Closure | Conventional or Hemogard™ Safety Closure | Only Hemogard™ Safety Closure versions are included in this 510(k) submission. This difference does not raise different questions of safety or effectiveness. |
| Stopper Fabrication | Compression Molded Rubber | Compression Molded Rubber | Identical |
| Hemogard™ Shield fabrication | Injection Molded Plastic | Injection Molded Plastic | Identical. |
| Additive Dispense | Spray Dry (K_{2}EDTA) | Spray Dry (K_{2}EDTA) | Identical. |
| Tube Evacuation | Vacuum Chamber | Vacuum Chamber | Identical. |
| Unit Labeling | Printed paper label | Printed paper label or imprinted on tube | Changing from paper label and imprinted label options to paper label only does not raise different questions of safety or effectiveness. |
| Sterilization Method | Gamma Irradiation | Gamma Irradiation | Identical. |
| Sterility Assurance Level (SAL) | ≤ 10^{-3} | ≤ 10^{-3} | Identical. |
| Shelf-Life | 12-16 months | 15-24 months | Differences in shelf-life are based on the data available at the time of submission for each tube configuration and do not raise different questions of safety or effectiveness. |
| Packaging | Unit – tube and closure Shelf – shrink wrapped polystyrene tray Case – corrugated cardboard | Unit – tube and closure Shelf – shrink wrapped polystyrene tray Case – corrugated cardboard | Identical. |
22
{5}
BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
The following discussion on substantial equivalence is pursuant to FDA Guidance dated . July 28, 2014, the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications.
#### 5.14.1.1 *Intended Use/Indications for Use*
The predicate 510(k) for BD Vacutainer® Plus K$_{2}$EDTA Tubes (BK050036) refers to an immunohematology intended use in addition to use for blood donor screening. The proposed indication for use is the same as the predicate with regards to the claim for immunohematology testing and the performance characteristics disclaimer, with the exception of more explicitly stating the intended user (trained healthcare professional) and removing blood donor screening. The subject device indication for use provides additional clarity on certain information consistent with current best practices; e.g., the statement, “such as ABO blood grouping and Rh typing.” This additional terminology is consistent with what was cleared under BK980011 for BD Vacutainer® Plus K$_{2}$EDTA Tubes, and more recently under BK210649 for BD Vacutainer® ACD Blood Collection Tubes.
As with the predicate device, the subject device will be used for all patient populations in settings where a venous sample is collected by a trained healthcare professional.
Therefore, using BK050036 as the predicate satisfies the requirement to have an appropriate predicate where both the subject and predicate devices have the same intended use.
#### 5.14.1.2 *Technological Characteristics*
Both the subject and predicate tubes cleared under BK050036 have similar technological characteristics. They both use K$_{2}$EDTA anticoagulant that has been spray dried onto a sterile evacuated plastic tube and closed with a Hemogard™ closure.
Differences between the subject and predicate and justification that such differences do not raise different questions of safety and effectiveness are described below:
**Tube Closure:** Only Hemogard™ Safety Closure versions are included in this 510(k) submission. Limiting the tube closure to only one version does not raise different questions of safety and effectiveness.
**Unit Labelling:** Changing from paper label and imprinted label options to a paper label only does not raise different questions of safety or effectiveness.
**Shelf-Life:** Regarding the varied shelf-life, the shelf-life is established based on available shelf-life data so difference in shelf-life do not raise different questions of safety and effectiveness. Testing was completed on the subject device and demonstrated that the product requirements were met.
### 5.15 Performance Testing – Bench Summary
Non-clinical performance testing was conducted following defined protocols and with established acceptance criteria to evaluate the following attributes of the BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes at time-zero and over the proposed shelf-life: Draw Volume,
23
{6}
BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes
BD Life Sciences, Integrated Diagnostic Solutions, Specimen Management
Becton, Dickinson and Company
X-Value, Second Stopper Pullout, Stopper/Shield Separation, Stopper Leakage, Tube Leakage, Resistance to Breakage during Drop Testing, and Resistance to Breakage During Centrifugation. Additionally, Ship Testing was conducted to assess the functional performance of the packaging materials.
All bench testing was completed on final, finished devices and conducted on a minimum of three unique lots of product to assess potential sources of lot-to-lot variability. Each test was also completed on a subset of product sterilized in excess of the maximum specified irradiation dosage. Testing over shelf-life was conducted over 13-20 months of accelerated aging, during which the devices were subjected to storage at 40°C and 50% Relative Humidity (RH) to accelerate the aging process according to the Arrhenius equation, and over 13-20 months of real time aging, during which the devices were subjected to storage at 25°C and 50% RH. Test intervals were selected as at least 16 months K$_{2}$EDTA products with a 15-month shelf-life; at least 17 months for K$_{2}$EDTA products with a 16-month shelf-life; at least 13 months for K$_{2}$EDTA products with a 12-month shelf-life. Real time aging studies were completed to confirm all accelerated aging data, and both sets of aging results met the predetermined acceptance criteria.
### **5.16 Performance Testing – Animal Summary**
No animal studies were performed in support of this submission.
### **5.17 Performance Testing – Clinical Summary**
Clinical testing was conducted on blood collected in both the subject devices (BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes) and legally marketed comparator devices to demonstrate Clinical Equivalence. Additional clinical testing was completed to evaluate Within-Tube Stability, Shelf-Life Performance, and Repeatability/Reproducibility. Clinical testing was conducted for a representative panel of immunohematology test parameters, with the distribution of ABO grouping/Rh typing subjects selected to be approximately representative of the general US population.
Results based on pre-determined acceptance criteria demonstrated the BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes are suitable for use in immunohematology testing.
### **5.18 Conclusion**
The technical performance characteristics of the subject device are unchanged. The BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes and the predicate devices have the same intended use and any differences in technological characteristic do not raise different questions of safety and effectiveness. Non-Clinical and Clinical Performance Testing support the determination that the subject BD Vacutainer® K$_{2}$EDTA Blood Collection Tubes continue to perform as intended. Based on information provided in this submission, the subject device is substantially equivalent to the predicate devices.
24
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.