Elecsys Syphilis; PreciControl Syphilis; PreciControl Release Syphilis

BK230839 · Roche Diagnostics Indianapolis, IN · Jul 31, 2023

Device Facts

Record IDBK230839
Device NameElecsys Syphilis; PreciControl Syphilis; PreciControl Release Syphilis
ApplicantRoche Diagnostics Indianapolis, IN
Decision DateJul 31, 2023
DecisionSESE

Indications for Use

Elecsys Syphilis is an in vitro immunoassay for the qualitative detection of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma. Elecsys Syphilis is intended to screen individual human donors, including volunteer donors of whole blood and blood components. This test is also intended to be used to screen organ, tissue and cell donors, when donor samples are obtained while the donor's heart is still beating. It is not intended for use on cord blood specimens. The electrochemiluminescence immunoassay "ECLIA" is intended for use with the cobas pro serology solution equipped with cobas e 801 analytical unit.

Device Story

Double antigen sandwich immunoassay; uses streptavidin-coated microparticles and biotinylated/ruthenium-labeled recombinant TP-specific antigens. Input: human serum or plasma samples. Operation: electrochemiluminescence immunoassay (ECLIA) on cobas pro serology solution (cobas e 801 analytical unit). Software compares sample signal to calibrated cutoff value. Output: qualitative reactive/non-reactive result (>= 1.0 COI). Used in clinical laboratories/blood centers; operated by trained laboratory personnel. Benefits: rapid screening (18 min) for syphilis antibodies in donor populations to ensure blood/tissue safety.

Clinical Evidence

Clinical study evaluated 4,556 donor specimens for specificity and 552 positive specimens for sensitivity. Specificity was 99.91% (95% CI: 99.77%–99.97%). Sensitivity was 100% (95% CI: 99.31%–100%) across primary, secondary, latent, and uncharacterized syphilis infections. Reproducibility study (n=270) across 3 sites showed minor variability.

Technological Characteristics

Double antigen sandwich immunoassay; streptavidin-coated microparticles; biotinylated and ruthenium-labeled recombinant TP-specific antigens. Detection: electrochemiluminescence (ECLIA). Platform: cobas pro serology solution (cobas e 801). Sample types: serum, Li-heparin, K2/K3-EDTA, CPD, Na-citrate plasma. Connectivity: integrated with cobas pro system. Reagent stability: 16 weeks on-board.

Indications for Use

Indicated for qualitative detection of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma for screening individual human donors (whole blood/blood components) and organ, tissue, and cell donors (beating heart). Not for use on cord blood.

Predicate Devices

Submission Summary (Full Text)

{0} # Elecsys Syphilis ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. In accordance with 21 CFR 807.87, Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k). The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for the Elecsys Syphilis. Page 1 {1} | **Submitter Name** | Roche Diagnostics | | --- | --- | | **Address** | 9115 Hague Road P.O. Box 50416 Indianapolis, IN 46250-0457 | | **Contact** | Kelly Brennan Phone: (b) (6) Email: (b) (6) @roche.com Secondary Contact: Reanna Toney Phone: (b) (6) Email: (b) (6) @roche.com | | **Date Prepared** | May 03, 2023 | | **Proprietary Name** | Elecsys Syphilis PreciControl Syphilis PreciControl Release Syphilis | | **Common Name** | Treponema pallidum serological reagents | | **Classification Name** | Test, Donor, Syphilis | | **Product Codes, Regulation Numbers** | MYR, 21 CFR 866.3830 | | **Predicate Devices** | Beckman Coulter PK TP System | | **Establishment Registration** | Roche Diagnostics GmbH Mannheim, Germany: 9610126 Roche Diagnostics GmbH Penzberg, Germany: 9610529 Roche Diagnostics Indianapolis, IN United States: 1823260. | Page 2 {2} # 1. DEVICE DESCRIPTION Elecsys Syphilis is a double antigen sandwich immunoassay with streptavidin microparticles, biotinylated recombinant TP-specific antigen labeled with a ruthenium complex and electrochemiluminescence detection. The results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration. Results greater than or equal to 1.0 COI are considered reactive for Treponema pallidum antibody. The test system contains the calibrators intended for use with the system. ## 1.1. Elecsys Syphilis The reagent working solutions (the cobas e pack) include: Reagent cobas e pack (kit placed on the analytical unit) - M Streptavidin-coated microparticles, 1 bottle, 14.1 mL: Streptavidin-coated microparticles 0.72 mg/mL; preservative. - R1 TP-specific recombinant antigens (E. coli)~biotin, 1 bottle, 19.7 mL: Biotinylated TP-specific recombinant antigens (E. coli) 0.7 mg/L, MESb) buffer 50 mmol/L, pH 6.5; preservative. - R2 TP-specific recombinant (E. coli)~Ru(bpy)2+, 1 bottle, 19.7 mL: TP-specific recombinant antigens (E. coli) labeled with ruthenium complex 0.7 mg/L, MES buffer 50 mmol/L, pH 6.5; preservative. b) MES = 2-morpholino-ethane sulfonic acid Calibrators (packed with the reagent cobas e pack) - SYPHB Cal1 Non-reactive calibrator 1 (lyophilized), 2 vials each of 1.0 mL: Human serum, non-reactive for anti-TP antibodies; preservative. Page 3 {3} • SYPHB Cal2 Reactive calibrator 2 (lyophilized), 2 vials each of 1.0 mL: Human serum, reactive for anti-TP antibodies; preservative. ### 1.2. PreciControl Syphilis The PreciControl Syphilis is a lyophilized control serum based on human serum, containing one control level reactive for anti-Treponema pallidum antibodies, and one control level non-reactive for anti-Treponema pallidum antibodies. ### 1.3. PreciControl Release Syphilis The PreciControl Release Syphilis is a lyophilized control serum based on human serum, reactive for anti-Treponema pallidum antibodies. The control is used to validate the cobas pro serology solution and to release sample results for the Elecsys Syphilis immunoassay. ## 2. INDICATIONS FOR USE ### 2.1. Elecsys Syphilis Elecsys Syphilis is an in vitro immunoassay for the qualitative detection of total antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma. Elecsys Syphilis is intended to screen individual human donors, including volunteer donors of whole blood and blood components. This test is also intended to be used to screen organ, tissue and cell donors, when donor samples are obtained while the donor's heart is still beating. It is not intended for use on cord blood specimens. The electrochemiluminescence immunoassay "ECLIA" is intended for use with the cobas pro serology solution equipped with cobas e 801 analytical unit. ### 2.2. PreciControl Syphilis PreciControl Syphilis is used for quality control of the Elecsys Syphilis immunoassay on cobas pro serology solution. Page 4 {4} ## 2.3. PreciControl Release Syphilis PreciControl Release Syphilis is used to validate the cobas pro serology solution and to release the sample results for the Elecsys Syphilis immunoassay. The recovery of the release control with Roche specified limits ensures the specified sensitivity of the assay under customer site conditions. The release control is tested at user-defined intervals with a maximum span of every 300 samples or 350 determinations and must be tested in order to release the test results. For release control values that fall outside the defined limits, samples measured before a failed release control are flagged as invalid by the cobas pro serology controller and need to be repeated. Reactive results will not be invalidated by a failed release control and must be retested in duplicate. Page 5 {5} ### 3. TECHNOLOGICAL CHARACTERISTICS **Table 1: Technical Characteristics Comparison Table between Elecsys Syphilis and Beckman Coulter PK TP System** | Feature | Candidate Device Elecsys Syphilis | Predicate Device Beckman Coulter PK TP on the PK7300 System (BK060062) | Predicate Device Newmarket TP HA REAGENT on the PK7400 System (BK180301) | | --- | --- | --- | --- | | Intended Use | Elecsys Syphilis is an in vitro immunoassay for the qualitative detection of total antibodies (IgG and IgM) to *Treponema pallidum* in human serum and plasma. Elecsys Syphilis is intended to screen individual human donors, including volunteer donors of whole blood and blood components. This test is also intended to be used to screen organ, tissue and cell donors, when donor samples are obtained while the donor's heart is still beating. It is not intended for use on cord blood specimens. The electrochemiluminescence immunoassay 'ECLIA' is intended for use with the **cobas pro** serology solution equipped with **cobas e** 801 analytical unit. | Qualitative screening of blood and plasma donors for the detection of *Treponema pallidum* IgG and IgM antibodies to syphilis in human serum or EDTA plasma using the Beckman Coulter PK7200 and/or PK7300 Automated Microplate Systems. | PK7400 TP HA REAGENT is intended for the qualitative screening of blood and plasma donors for the detection of *Treponema pallidum* IgG and IgM antibodies to syphilis in human serum, EDTA plasma or CPDA plasma using the Beckman Coulter PK7400 Automated Microplate System. This assay is not intended for diagnostic use. | | Assay Method | Double antigen sandwich assay | Hemagglutination | Hemagglutination | | Detection Method | electrochemiluminescence immunoassay 'ECLIA' | Photometric | Photometric | | Applications/Test Time | 18 minutes | 60 minutes | 60 minutes | | Instrument Platform | **cobas pro** serology solution | PK7300 | PK7400 | | Sample Type | Human serum, plasma | Human serum, plasma | Human serum, plasma | | Sample Anticoagulants | Li-heparin, K_{2}-EDTA, K_{3}-EDTA, CPD and Na-citrate | EDTA | EDTA, CPDA | | Controls | PreciControl Syphilis PreciControl Release Syphilis | PK TP System Control Set | PK7400 TP HA Controls | Page 6 {6} | Feature | Candidate Device Elecsys Syphilis | Predicate Device Beckman Coulter PK TP on the PK7300 System (BK060062) | Predicate Device Newmarket TP HA REAGENT on the PK7400 System (BK180301) | | --- | --- | --- | --- | | Reagent Stability | Unopened at 2-8 °C = up to the stated expiration date On the **cobas e** 801 analytical unit = 16 weeks | Unopened at 2-8 °C = up to the stated expiration date | 18 months when stored unopened at 2-8 °C | #### 4. NON-CLINICAL PERFORMANCE EVALUATION The following performance data are provided in support of the substantial equivalence determination: - Analytical sensitivity - Sample handling and preparation - Analytical specificity - Cross-Reactivity - Stability All performance specifications were met. #### 4.1. Analytical Sensitivity – Limit of Blank (LoB) and Limit of Detection (LoD) The Limit of Blank (LoB) and Limit of Detection (LoD) were determined in accordance with CLSI EP17-A2. The limit of blank (LoB): LoB for the Elecsys Syphilis assay was estimated using five analyte-free serum and five analyte-free plasma samples measured with one reagent kit lot in duplicate determination over three days and six runs on one cobas e 801 analytical unit. Sixty measured values of analyte-free samples were obtained. The LoB for serum was determined to be 0.115 COI and the LoB for plasma was determined to be 0.0883 COI. Results for LoB for serum and plasma met the acceptance criteria. Page 7 {7} The limit of detection (LoD): LoD for the Elecsys Syphilis assay was estimated using five serum and five plasma samples with low-analyte concentration (approximately four times the LoB) measured with one lot of Elecsys Syphilis in duplicate determination over three days in six runs on one cobas e 801 analytical unit. Sixty measured values of samples with low analyte concentration were obtained. The LoD for serum was determined to be 0.128 COI and the LoD for plasma was determined to be 0.107 COI. Results for LoD for serum and plasma met the acceptance criteria. ## 4.2. Sample handling and preparation Only the following specimens listed were tested and found acceptable. Serum and Li-heparin, K₂-EDTA, K₃-EDTA, CPD and Na-citrate plasma collected using standard sampling tubes. Serum and Li-Heparin, K₂-EDTA plasma collected in tubes containing separating gel. Stable on-the-clot for 7 days at 15-30°C, 14 days at 2-8°C. Do not freeze samples on-the-clot. Samples off-the-clot are stable for 7 days at 15-30°C, 14 days at 2-8°C and 12 months at -20°C (± 5°C). Samples off-the-clot may be frozen up to 4 times. ## 4.3. Analytical Specificity ### 4.3.1. Endogenous Interferences The effect of the following endogenous substances on assay performance was tested with the Elecsys Syphilis assay on the cobas e 801 analytical unit. The interfering agents were tested using native or spiked serum samples. For each interfering substance five sample replicates of three human serum samples (anti-Treponema non-reactive, reactive, and high reactive) were tested with one reagent kit lot. The mean recovery (absolute deviation or percent recovery) was calculated for each sample compared to the expected (reference) value. Interferences were tested up to the listed concentrations and no impact on results was observed. Page 8 {8} | Compound | Concentration tested | | --- | --- | | Bilirubin | ≤ 753 μmol/L or ≤ 44 mg/dL | | Hemoglobin | ≤ 0.311 mmol/L or ≤ 500 mg/dL | | Intralipid | ≤ 2000 mg/dL | | Biotin | ≤ 4912 nmol/L or ≤ 1200 ng/mL | | Albumin | ≤ 7.0 g/dL | Additionally, fifty samples with naturally elevated levels of triglycerides (lipemia), bilirubin, hemoglobin (hemolyzed), albumin and rheumatoid factor, respectively, were tested with the Elecsys Syphilis assay on the cobas e 801 analytical unit. For each interfering substance 10 human serum samples with naturally elevated levels of the interferent were tested in single determinations with one reagent kit lot; no false reactive results were found. ### 4.3.2. Drug Interference Seventeen common therapeutic drugs were tested for potential interference. Each drug was tested spiked into one non-reactive sample and one anti-treponemal reactive sample. Samples were tested five-fold determination and compared to unspiked serum (reference). Each drug was found to be non-interfering at the claimed concentrations. | Compound | Tested Concentration (mg/L) | | --- | --- | | Acetylcystein | 150 | | Acetylsalicylic Acid | 30 | | Ampicillin-Na | 75 | | Ascorbic acid | 52.5 | | Cefoxitin | 750 | | Doxycycline | 18 | | Heparin | 3300 IU/L | | Levodopa | 7.5 | | Methyldopa | 22.5 | | Metronidazole | 123 | | Rifampicin | 48 | | Acetaminophen | 156 | | Cyclosporine | 1.8 | | Ibuprofen | 219 | | Theophylline | 60 | | Phenylbutazone | 321 | | Itraconazole | 30 | Page 9 {9} ### 4.3.3. High Dose Hook Effect A study was conducted to evaluate whether high dose hook effect can result in false non-reactive results using Elecsys Syphilis. Three high titer reactive samples were diluted in non-reactive serum in 12 dilution steps to generate a dilution series that covers the range from low-reactive to high reactive cut-off values. The diluted samples were measured on one cobas e 801 analytical unit in three-fold determination with one lot of Elecsys Syphilis. No false non-reactive result due to high dose hook effect was found with the Elecsys Syphilis assay. ### 4.4. Cross-Reactivity A total of 170 samples containing potentially interfering factors were tested with the Elecsys Syphilis on the cobas e 801 analytical unit comprising specimens: - containing antibodies against HAV, HBV, HCV, HIV, HTLV-I/II, CMV, EBV, HSV-1/2, Rubella - containing Antinuclear Antibodies (ANA) and elevated titers of rheumatoid factor - containing antibodies against Escherichia coli, Toxoplasma gondii and Borelia burgdorferi - After vaccination against influenza - From pregnant women and Multiparous Pregnancies Results showed no interference from the above agents. ### 4.5. Stability Kit stability (unopened): Kit stability/shelf-life was determined using three production lots of the Elecsys Syphilis cobas e pack stored refrigerated at 2-8°C, using 5 samples (non-reactive and reactive samples) and the PreciControls (PC SYPH1 B and PC SYPH2 B) in duplicate on the cobas e 801 analytical unit. The stability data supports Roche Diagnostic's claims as reported in the package labeling. Page 10 {10} On-board (In-use) stability: An on-board stability study was performed for the Elecsys Syphilis cobas e pack to determine the time the reagent kit can be kept on-board the cobas e 801 analytical unit once opened. The study was performed with two non-reactive specimens, three Treponema antibody reactive specimens (COI ranged from 1 to 14.9) and the PreciControls (PC SYPH1 B and PC SYPH2 B). Samples were tested in duplicate after storage on-board for the specified time. Percent recovery for each sample was calculated compared to the unstressed Elecsys Syphilis cobas e pack. The stability data supports Roche Diagnostic's claims as reported in the package labeling. PreciControl Stability (Shelf-life): Kit stability/shelf-life was determined using three production lots of the PreciControl Syphilis stored refrigerated at 2-8°C. PreciControls (PC SYPH1 B and PC SYPH2 B) were tested in triplicate on the cobas e 801 analytical unit. The stability data supports Roche Diagnostic's claims as reported in the package labeling. PreciControl On-board (In-use) Stability: On-board (In-use) stability was determined using one production lot of the PreciControl Syphilis. PreciControls (PC SYPH1 B and PC SYPH2 B) were reconstituted and tested in duplicate on the cobas e 801 analytical unit. Reconstituted controls were stored at 20-25 °C and retested in duplicate at several time points. The stability data supports Roche Diagnostic's claims as reported in the package labeling. ### 4.6. Precision Precision was determined using Elecsys reagents, pooled human sera and controls in a protocol (EP05-A3) of the CLSI (Clinical and Laboratory Standards Institute): 2 runs per day in duplicate each for 21 days (n = 84). Precision of the Elecsys Syphilis assay demonstrated minor variability from run to run, day to day and between reagent lots. The following results were obtained: | Sample | Mean (COI) | Repeatability SD (COI) | Repeatability % CV | Between run SD (COI) | Between run % CV | | --- | --- | --- | --- | --- | --- | | HSP 01 c) | 0.125 | 0.002 | 1.5 | 0.000 | 0.4 | | HSP 02 | 0.888 | 0.018 | 2.0 | 0.016 | 1.8 | | HSP 03 | 1.09 | 0.017 | 1.6 | 0.011 | 1.0 | | HSP 04 | 4.11 | 0.098 | 2.4 | 0.079 | 1.9 | Page 11 {11} | Sample | Mean (COI) | Repeatability SD (COI) | Repeatability % CV | Between run SD (COI) | Between run % CV | | --- | --- | --- | --- | --- | --- | | HSP 05 | 6.88 | 0.198 | 2.9 | 0.073 | 1.1 | | HSP 06 | 15.8 | 0.395 | 2.5 | 0.216 | 1.4 | | HSP 07 | 16.4 | 0.395 | 2.4 | 0.158 | 1.0 | | PC SYPH1 B d) | 0.0951 | 0.001 | 1.1 | 0.000 | 0.0 | | PC SYPH2 B | 5.90 | 0.126 | 2.1 | 0.087 | 1.5 | c) HSP = human specimens d) PC = PreciControl | Sample | Mean (COI) | Between day SD (COI) | Between day % CV | Intermediate precision SD (COI) | Intermediate precision % CV | | --- | --- | --- | --- | --- | --- | | HSP 01 c) | 0.125 | 0.001 | 0.6 | 0.002 | 1.7 | | HSP 02 | 0.888 | 0.012 | 1.3 | 0.026 | 3.0 | | HSP 03 | 1.09 | 0.016 | 1.5 | 0.026 | 2.4 | | HSP 04 | 4.11 | 0.000 | 0.0 | 0.126 | 3.1 | | HSP 05 | 6.88 | 0.133 | 1.9 | 0.249 | 3.6 | | HSP 06 | 15.8 | 0.357 | 2.3 | 0.574 | 3.6 | | HSP 07 | 16.4 | 0.332 | 2.0 | 0.540 | 3.3 | | PC SYPH1 B d) | 0.0951 | 0.001 | 0.8 | 0.001 | 1.4 | | PC SYPH2 B | 5.90 | 0.025 | 0.4 | 0.155 | 2.6 | Page 12 {12} ## 5. EXTERNAL (CLINICAL) TESTING ### 5.1. Reproducibility A study was performed based on guidance from CLSI EP05-A3 (n = 270). Testing was conducted at 3 external sites using 3 lots of the Elecsys Syphilis assay and 1 lot of PreciControl Syphilis on the **cobas pro** serology solution. Panel members and controls were tested in 2 runs per day for 5 days with 3 sample replicates per run. The reproducibility of the Elecsys Syphilis assay demonstrated minor variability from run to run, day to day and between reagent lots. The results for the Elecsys Syphilis assay are presented in the following tables. | Sample | Mean (COI) | Repeatability SD (COI) | Repeatability % CV | Between run SD (COI) | Between run % CV | | --- | --- | --- | --- | --- | --- | | HSP 01 a) | 1.63 | 0.030 | 1.84 | 0.012 | 0.715 | | HSP 02 | 10.7 | 0.186 | 1.74 | 0.098 | 0.916 | | PC SYPH1 B b) | 0.085 | 0.001 | 1.19 | 0.000 | 0.000 | | PC SYPH2 B | 5.05 | 0.074 | 1.47 | 0.048 | 0.959 | a) HSP = human specimen b) PC = PreciControl | Sample | Mean (COI) | Between day SD (COI) | Between day % CV | Intermediate precision SD (COI) | Intermediate precision % CV | | --- | --- | --- | --- | --- | --- | | HSP 01 | 1.63 | 0.021 | 1.31 | 0.039 | 2.37 | | HSP 02 | 10.7 | 0.107 | 0.999 | 0.235 | 2.21 | | PC SYPH1 B | 0.085 | 0.001 | 0.814 | 0.001 | 1.44 | | PC SYPH2 B | 5.05 | 0.054 | 1.07 | 0.104 | 2.05 | | Sample | Mean (COI) | Between site SD (COI) | Between site % CV | Between lot SD (COI) | Between lot % CV | | --- | --- | --- | --- | --- | --- | | HSP 01 | 1.63 | 0.006 | 0.374 | 0.031 | 1.90 | | HSP 02 | 10.7 | 0.046 | 0.432 | 0.097 | 0.911 | | PC SYPH1 B | 0.085 | 0.001 | 1.03 | 0.001 | 1.75 | Page 13 {13} | Sample | Mean (COI) | Between site SD (COI) | Between site % CV | Between lot SD (COI) | Between lot % CV | | --- | --- | --- | --- | --- | --- | | PC SYPH2 B | 5.05 | 0.017 | 0.345 | 0.125 | 2.47 | | Sample | Mean (COI) | Reproducibility SD (COI) | Reproducibility % CV | | --- | --- | --- | --- | | HSP 01 | 1.63 | 0.050 | 3.06 | | HSP 02 | 10.7 | 0.259 | 2.43 | | PC SYPH1 B | 0.085 | 0.002 | 2.49 | | PC SYPH2 B | 5.05 | 0.163 | 3.23 | ## 5.2. Clinical Specificity A total of 4556 specimens (2052 serum and 2504 plasma specimens) were tested at three sites with the Elecsys Syphilis assay on the cobas pro serology solution and FDA-cleared screening assays, including the FDA-cleared confirmatory assays. Out of the 4556 specimens, 4548 specimens were interpreted with an initial status of negative. The results of the algorithm and supplemental testing indicated that four specimens were repeatedly reactive with a negative status, and four specimens were repeatedly reactive with a positive status. The overall specificity performance of the Elecsys Syphilis assay was 99.91% with a 95% CI of 99.77% to 99.97%. The initial and repeat reactive rates were 0.10 % (2/2052) for the serum specimens and 0.24 % (6/2054) for the plasma specimens | Specimen Category | Number tested | Number IR a) (% of total) | Number RR b) (% of total) | Number by Supplemental Testing (% of RR) | Specificity (%) (95 % CI) | | --- | --- | --- | --- | --- | --- | | Volunteer blood donors - serum | 2052 | 2 (0.10) | 2 (0.10) | 2 (100) | 100 (2050/2050) (99.81, 100) | | Volunteer blood | 2504 | 6 | 6 | 2 | 99.84 | Page 14 {14} | Specimen Category | Number tested | Number IR a) (% of total) | Number RR b) (% of total) | Number by Supplemental Testing (% of RR) | Specificity (%) (95 % CI) | | --- | --- | --- | --- | --- | --- | | donors - plasma | | (0.24) | (0.24) | (33.3) | (2498/2502) (99.59, 99.94) | | Total donors | 4556 | 8 (0.18) | 8 (0.18) | 4 (50.0) | 99.91 (4548/4552) (99.7, 99.97) | a) IR = initially reactive b) RR = repeatedly reactive ### 5.3. Clinical Sensitivity Five hundred and fifty-two specimens (201 serum, 351 plasma) were tested for evaluating the sensitivity of the Elecsys Syphilis assay on the cobas pro serology solution, using three reagent kit lots distributed among three test-sites. All 552 specimens showing a positive syphilis specimen status were congruently repeatedly reactive with the Elecsys Syphilis assay. The sensitivity was calculated to be 100 % (552/552) with a 95 % confidence interval of 99.31 % to 100 %. Combined (serum and plasma) | Specimen category | Specimen status Positive | Number RR | Sensitivity (%) (95% CI) | | --- | --- | --- | --- | | Latent infection | 97 | 97 | 100 (96.19, 100) | | Primary infection | 76 | 76 | 100 (95.19, 100) | | Secondary infection | 64 | 64 | 100 (94.34, 100) | | Uncharacterized syphilis infection | 315 | 315 | 100 (98.80, 100) | | Total | 552 | 552 | 100 (99.31, 100) | Page 15 {15} # Serum | Specimen category | Specimen status Positive | Number RR | Sensitivity (%) (95% CI) | | --- | --- | --- | --- | | Latent infection | 70 | 70 | 100 (94.80, 100) | | Primary infection | 50 | 50 | 100 (92.87, 100) | | Secondary infection | 63 | 63 | 100 (94.25, 100) | | Uncharacterized syphilis infection | 18 | 18 | 100 (82.41, 100) | | Total | 201 | 201 | 100 (98.12, 100) | # Plasma | Specimen category | Specimen status Positive | Number RR | Sensitivity (%) (95% CI) | | --- | --- | --- | --- | | Latent infection | 27 | 27 | 100 (87.54, 100) | | Primary infection | 26 | 26 | 100 (87.13, 100) | | Secondary infection | 1 | 1 | 100 (20.65, 100) | | Uncharacterized syphilis infection | 297 | 297 | 100 (98.72, 100) | | Total | 351 | 351 | 100 (98.92, 100) | Page 16 {16} ## 6. CONCLUSIONS The results of non-clinical analytical and clinical performance studies demonstrate that the Elecsys Syphilis test for use on the **cobas pro** system when equipped with the **cobas e 801** analytical unit is as safe, as effective, and performs as well as the predicate device. Page 17
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...