The BD Leucocount™ Kit consists of BD Leucocount™ Reagent (propidium iodide fluorescent dye) and BD Trucount™ Tubes and is intended for use with the BD FACSCalibur™, BD FACSort™, BD FACScan™, BD FACSVia™, and BD FACSLyric™ flow cytometer systems, or for a flow cytometer equipped with a 488-nm laser able to threshold on propidium iodide fluorescence, for enumerating residual white blood cells (rWBCs) in leucoreduced blood products. For in vitro diagnostic use. The BD FACSLyric™ flow cytometer is intended for use as an in vitro diagnostic device for the following: Immunophenotyping using up to six fluorescence detection channels and two light scatter channels using a blue (488-nm) and a red (640-nm) laser. Enumeration of residual white blood cells (rWBCs) in leucoreduced blood products. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.
Device Story
BD Leucocount™ Kit uses propidium iodide fluorescent dye to stain DNA in nucleated white blood cells; BD Trucount™ tubes with internal reference beads enable absolute cell counting. Samples are prepared manually and acquired on BD FACSLyric™ flow cytometer. BD Leucocount™ Assay Module for BD FACSuite™ Clinical software automates data processing; identifies rWBC and bead populations via gating; calculates absolute rWBC counts based on bead events and sample volume. Used in clinical laboratories by trained personnel. Output includes lab reports with rWBC counts (cells/μL) and total counts per pack. Assists clinicians in verifying leucoreduction quality in blood products; ensures patient safety by minimizing residual WBCs in transfusions.
Clinical Evidence
Bench testing and clinical method comparison performed. Method comparison study (n=502 evaluable specimens) across four sites compared BD Leucocount™ on BD FACSLyric™ vs. BD FACSVia™; acceptance criteria met. Precision, repeatability, linearity, LoB (0 cells/μL), LoQ (0.7 cells/μL), carryover, and inter-site reproducibility studies conducted per CLSI standards (EP05-A3, EP06-Ed2, EP17-A2, EP09c).
Technological Characteristics
Flow cytometric assay using propidium iodide (nucleic acid dye) and Trucount™ beads. BD FACSLyric™ system uses blue (488-nm) and red (640-nm) lasers. Detection: 586/42 filter (blue laser) for reagent, 660/10 filter (red laser) for beads. Software: BD FACSuite™ Clinical with BD Leucocount™ Assay Module. Manual sample preparation. Connectivity: Standalone or networked clinical system.
Indications for Use
Indicated for enumeration of residual white blood cells (rWBCs) in leucoreduced blood products (red blood cells and platelets) using flow cytometry.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
BD FACSVia™ System with BD Leucocount™ Reagent Assay (BK150307)
{0}
## ATTACHMENT 36 – 510(k) SUMMARY
This 510(k) summary is being provided in accordance with the requirements of 21 CFR 807.92.
Date of Summary: April 24, 2023
### 1. Submitted By
Becton, Dickinson and Company
BD Biosciences
2350 Qume Drive
San Jose, CA 95131-1807 USA
Contact: Jonathan Lee
Senior Regulatory Affairs Specialist
Telephone: (415) 680-5626
Email: (b) (6) @bd.com
### 2. Device
Trade Name/Device Name: BD Leucocount™ Kit;
BD FACSuite™ Clinical BD Leucocount™ Assay Module;
BD FACSLyric™ Flow Cytometer
Classification: Class II
Device Classification: Flow Cytometric Reagents and Accessories
Regulation Description: Automated Differential Cell Counter
Regulation Medical Specialty: Hematology
Product Code: OYE
Regulation Number: 21 CFR 864.5220
### 3. Predicate Device
Predicate
Trade Name/Device Name: BD FACSVia™ System with BD Leucocount™ Reagent Assay
510(k) Number: BK150307
Classification: Class II
Device Classification: Flow Cytometric Reagents and Accessories
Regulation Description: Automated Differential Cell Counter
Regulation Medical Specialty: Hematology
Product Code: OYE
Regulation Number: 21 CFR 864.5220
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 1 of 10
{1}
Predicate
| Trade Name/Device Name: | BD FACSLyric™ Flow Cytometer |
| --- | --- |
| 510(k) Number: | K201814 |
| Classification: | Class II |
| Device Classification: | Flow Cytometric Reagents and Accessories |
| Regulation Description: | Automated Differential Cell Counter |
| Regulation Medical Specialty: | Hematology |
| Product Code: | OYE |
| Regulation Number: | 21 CFR 864.5220 |
#### 4. Device Description
BD Leucocount™ Kit is a flow cytometric in vitro diagnostic assay based on a nucleic acid fluorescent dye. The purpose of this 510(k) is to add use of the cleared BD Leucocount™ Kit with the BD FACSLyric™ flow cytometer system.
In addition, the intended use statement of the BD FACSLyric™ Flow Cytometer will be updated to support use with the BD Leucocount™ Kit.
To use the BD Leucocount™ Kit on the BD FACSLyric™ flow cytometer system, the following system components are required:
- BD Leucocount™ Kit
- BD Leucocount™ Assay Module for BD FACSuite™ Clinical application
- BD FACSLyric™ flow cytometer system
- BD Leucocount™ RBC Control, BD Leucocount™ PLT Control, or BD Leucocount™ Combo Control
- BD FACSLink™ (optional)
To enable use of the BD Leucocount™ Kit with the BD FACSLyric™ flow cytometer system, BD Leucocount™ Assay Module has been developed.
#### 5. Indications for Use
##### **BD Leucocount™ Kit**
The BD Leucocount™ Kit consists of BD Leucocount™ Reagent (propidium iodide fluorescent dye) and BD Trucount™ Tubes and is intended for use with the BD FACSCalibur™, BD FACSort™, BD FACScan™, BD FACSVia™, and BD FACSLyric™ flow cytometer systems, or for a flow cytometer equipped with a 488-nm laser able to threshold on propidium iodide fluorescence, for enumerating residual white blood cells (rWBCs) in leucoreduced blood products.
For in vitro diagnostic use.
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 2 of 10
{2}
### **BD FACSLyric™ Flow Cytometer**
The BD FACSLyric™ flow cytometer is intended for use as an in vitro diagnostic device for the following:
- Immunophenotyping using up to six fluorescence detection channels and two light scatter channels using a blue (488-nm) and a red (640-nm) laser.
- Enumeration of residual white blood cells (rWBCs) in leucoreduced blood products.
It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.
## **6. Substantial Equivalence Comparison between Subject and Predicate Devices**
Table 1 provides a summary list of the subject devices and their corresponding predicate device.
**Table 1. Subject and Predicate Devices**
| Subject Device | Predicate Device |
| --- | --- |
| BD Leucocount™ Kit (with BD FACSuite™ Clinical BD Leucocount™ Assay Module) | BD FACSVia™ System with BD Leucocount™ Reagent Assay (BK150307) |
| Subject Device | Predicate Device |
| BD FACSLyric™ Flow Cytometer | BD FACSLyric™ Flow Cytometer (K201814) |
### **BD Leucocount™ Kit – Summary of Device Comparison**
Table 2 provides a comparison between the subject BD Leucocount™ Kit and its predicate device.
The subject BD Leucocount™ Kit is the same as its predicate device, as follows:
- Same indications for use: for enumerating residual white blood cells in leucoreduced blood products.
- Same components in the BD Leucocount™ Kit (BD Leucocount™ Reagent and BD Trucount™ Tubes).
- Same assay principle.
- Same specimen types, specimen stability, cell population for analysis, and sample preparation method.
The subject BD Leucocount™ Kit differs from its predicate device, as follows:
- BD FACSLyric™ Flow Cytometer has been added for use with the subject device.
- An assay module (BD Leucocount™ Assay Module) has been developed to enable BD Leucocount™ Kit to be used with BD FACSuite™ Clinical application on BD FACSLyric™ Flow Cytometer.
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 3 of 10
{3}
### BD FACSLyric™ Flow Cytometer – Summary of Device Comparison
Table 3 provides a comparison between the subject BD FACSLyric™ Flow Cytometer and its predicate device.
The only difference between the subject BD FACSLyric™ Flow Cytometer and its predicate is that the intended use statement will be updated to support use with the BD Leucocount™ Kit. All other aspects of the BD FACSLyric™ Flow Cytometer remain the same as described in the most recent BD FACSLyric™ Flow Cytometer 510(k) clearance K201814 (predicate).
**Table 2. Comparison Between BD Leucocount™ Kit and its Predicate**
| Feature/Attribute | Subject Device | Predicate Device |
| --- | --- | --- |
| | **BD Leucocount™ Kit (with BD Leucocount™ Assay Module, for use with BD FACSLyric™ Flow Cytometer)** | **BD Leucocount™ Reagent Assay (BD Leucocount™ Kit) for use with BD FACSVia™ System (BK150307)** |
| **Device Classification and Product Code** | Regulation Description: Flow Cytometric Reagents and Accessories Regulatory Class: II Regulation Number: 21 CFR 864.5220 Product Code: OYE | |
| **Intended Use / Indications for Use** | The BD Leucocount™ Kit consists of BD Leucocount™ Reagent (propidium iodide fluorescent dye) and BD Trucount™ Tubes and is intended for use with the BD FACSCalibur™, BD FACSort™, BD FACScan™, BD FACSVia™, **and BD FACSLyric™ flow cytometer systems**, or for a flow cytometer equipped with a 488-nm laser able to threshold on propidium iodide fluorescence, for enumerating residual white blood cells (rWBCs) in leucoreduced blood products. For in vitro diagnostic use. | BD Leucocount™ Kit consists of BD Leucocount™ reagent (propidium iodide fluorescent dye) and BD Trucount™ tubes and is intended for use with the BD FACSCalibur™, BD FACSort™, BD FACScan™, and BD FACSVia™ flow cytometer systems, or for a flow cytometer equipped with a 488-nm argon ion laser able to threshold on FL2, for enumerating residual white blood cells (rWBCs) in leucoreduced blood products. For in vitro diagnostic use. |
| **Reagents** | BD Leucocount™ Kit • BD Leucocount™ Reagent • BD Trucount™ Tubes | |
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 4 of 10
{4}
| Feature/Attribute | Subject Device BD Leucocount™ Kit (with BD Leucocount™ Assay Module, for use with BD FACSLyric™ Flow Cytometer) | Predicate Device BD Leucocount™ Reagent Assay (BD Leucocount™ Kit) for use with BD FACSVia™ System (BK150307) |
| --- | --- | --- |
| Detection/Assay Principle | The BD Leucocount™ Reagent contains propidium iodide fluorescent dye. Propidium iodide is a nucleic acid dye which, when used with RNAse, stains only cellular DNA. White blood cells are nucleated cells that contain DNA and are therefore stained with the dye. Non-nucleated particles (including platelets and red blood cells) do not stain with this reagent. BD Trucount™ Tubes contain beads that act as an internal reference to accurately determine the absolute count of residual white blood cells (rWBCs). Appropriate samples are combined with the lyophilized bead pellet in the BD Trucount™ Tube before staining. After staining rWBCs, samples are acquired on a flow cytometer. Absolute rWBC counts are determined by using a simple calculation based on bead number and sample volume. | |
| Specimen Type | Leucoreduced blood products, including: - Leucoreduced red blood cells - Leucoreduced platelets | |
| Sample Volume | 100 μL | |
| Sample Preparation | Manual | |
| Instrument and Software Platform | **For BD FACSLyric™ Flow Cytometer:** - BD FACSLyric™ Flow Cytometer using BD FACSuite™ Clinical application with BD Leucocount™ Assay Module - System adjusts/optimizes photomultiplier tube (PMT) voltages | **For BD FACSVia™ Flow Cytometer:** - BD FACSVia™ Flow Cytometer using BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file - Fixed PMT voltages |
| Software - Inputs | **For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module:** - Flow cytometry data. - Lot and expiry information for BD Trucount™ beads and setup beads. - **Lot and expiry information for reagents.** - Sample tube ID. - May also input the pack volume. | **For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file:** - Flow cytometry data. - Lot and expiry information for BD Trucount™ beads and setup beads. - Sample tube ID. - May also input the pack volume. |
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 5 of 10
{5}
| Feature/Attribute | Subject Device BD Leucocount™ Kit (with BD Leucocount™ Assay Module, for use with BD FACSLyric™ Flow Cytometer) | Predicate Device BD Leucocount™ Reagent Assay (BD Leucocount™ Kit) for use with BD FACSVia™ System (BK150307) |
| --- | --- | --- |
| **Software - Outputs** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: • BD Leucocount™ Lab Report • Assay Setup Report • Instrument QC Report | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: • BD Leucocount™ Lab Report • Instrument QC Report |
| **Software – Operator/User** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: • Clinical laboratory personnel. | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: • Clinical laboratory personnel. |
| **Software – Operator/User Actions** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: • Analysis gates/regions are automatically applied to the flow cytometry dot plots. The user inspects the dot plots, and the gates/regions may be adjusted by the user as needed. • Reports generated by the software with BD Leucocount™ Assay Module are reviewed by the user. | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: • Analysis gates/regions are automatically applied to the flow cytometry dot plots. The user inspects the dot plots, and the gates/regions may be adjusted by the user as needed. • Reports generated by the software with BD Leucocount™ assay test definition file are reviewed by the user. |
| **Software – Acquisition Stopping Criteria** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: • 10,000 BD Trucount™ beads events collected. • Acquisition time limit: 5 minutes. | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: • 10,000 BD Trucount™ beads events collected. • Acquisition time limit: 5 minutes. |
| **Software – Gated Analysis Events/ Populations** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: • Residual white blood cells (rWBCs) • BD Trucount™ Beads | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: • Residual white blood cells (rWBCs) • BD Trucount™ Beads |
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 6 of 10
{6}
| Feature/Attribute | Subject Device | Predicate Device |
| --- | --- | --- |
| | **BD Leucocount™ Kit (with BD Leucocount™ Assay Module, for use with BD FACSLyric™ Flow Cytometer)** | **BD Leucocount™ Reagent Assay (BD Leucocount™ Kit) for use with BD FACSVia™ System (BK150307)** |
| **Software – rWBC Analysis Gate Type** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: - The rWBC population is identified by a default pre-defined gate template/region that is automatically set. The gate may be adjusted by the user as needed. | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: - The rWBC population is identified by a default pre-defined gate template/region that is automatically set. The gate may be adjusted by the user as needed. |
| **Software – BD Trucount™ Beads Analysis Gate Type** | For BD FACSuite™ Clinical Application with BD Leucocount™ Assay Module: - The BD Trucount™ Beads population is identified **with an algorithm gate** that is automatically set and adjusted. The gate may be adjusted further by the user as needed. | For BD FACSVia™ Clinical software with BD Leucocount™ assay test definition file: - The BD Trucount™ Beads population is identified by a default pre-defined gate template/region that is automatically set. The gate may be adjusted by the user as needed. |
| **Instrument Configuration** | For BD FACSLyric™ Flow Cytometer: - Equipped to detect forward scatter, side scatter and up to six fluorescence channels for IVD use. - BD Leucocount™ reagent (propidium iodide) uses the blue laser's detector with the 586/42 filter. - BD Trucount™ absolute counting beads use the red laser's detector with the 660/10 filter. | For BD FACSVia™ Flow Cytometer: - Equipped to detect forward scatter, side scatter and up to two fluorescence channels for IVD use. - BD Leucocount™ reagent (propidium iodide) uses the blue laser's detector with the 585/40 filter. - BD Trucount™ absolute counting beads use the blue laser's detector with the 533/30 filter. |
| **Specimen Stability Age of Blood (AOB) Age of Stain (AOS)** | Prepare and run the samples within 48 hours following leucoreduction. For specimens stained within 24 hours of leucoreduction, acquire the samples within 24 hours of staining. For specimens stained within 48 hours of leucoreduction, acquire the samples within 60 minutes of staining. | |
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 7 of 10
{7}
| Feature/Attribute | Subject Device BD Leucocount™ Kit (with BD Leucocount™ Assay Module, for use with BD FACSLyric™ Flow Cytometer) | Predicate Device BD Leucocount™ Reagent Assay (BD Leucocount™ Kit) for use with BD FACSVia™ System (BK150307) |
| --- | --- | --- |
| Number of Tubes per Assay | 1 Tube | |
| Measurement Range | For BD FACSLyric™ Flow Cytometer: 0.7 – 350 rWBCs/μL | For BD FACSVia™ Flow Cytometer: 0 – 350 rWBCs/μL |
| Results Reporting | Software-assisted report generation. The lab report provides the rWBC count in cells/μL. The lab report also provides the total rWBC count per pack, if the pack volume is entered. | |
Table 3. Comparison Between BD FACSLyric™ Flow Cytometer and its Predicate
| Feature/Attribute | Subject Device BD FACSLyric™ Flow Cytometer | Predicate Device BD FACSLyric™ Flow Cytometer (K201814) |
| --- | --- | --- |
| Intended Use / Indications for Use | The BD FACSLyric™ flow cytometer is intended for use as an in vitro diagnostic device for the following: • Immunophenotyping using up to six fluorescence detection channels and two light scatter channels using a blue (488-nm) and a red (640-nm) laser. • Enumeration of residual white blood cells (rWBCs) in leucoreduced blood products. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument. | The BD FACSLyric™ flow cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to six fluorescence detection channels and two light scatter channels using a blue (488-nm) and a red (640-nm) laser. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument. |
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 8 of 10
{8}
## 7. Performance Data
The following performance studies were conducted to support the substantial equivalency determination.
**Table 4. BD Leucocount™ Kit Bench Performance Studies**
| Study | Standard | Objective | Results |
| --- | --- | --- | --- |
| Within-Site Precision Using Control Materials | CLSI EP05-A3 | To evaluate the precision performance of BD Leucocount™ Kit on the BD FACSLyric™ Flow Cytometer. | All acceptance criteria were met. |
| Repeatability Evaluation Using Clinical Specimens | CLSI EP05-A3 | To verify repeatability performance of BD Leucocount™ Kit on the BD FACSLyric™ Flow Cytometer through testing clinical specimens. | All acceptance criteria were met. |
| Linearity | CLSI EP06-Ed2 | To evaluate the linear range of the BD Leucocount™ Kit using BD FACSLyric™ Flow Cytometer, for measuring residual White Blood Cell (rWBC) absolute counts. | The linear range for rWBC absolute counts using BD Leucocount™ Kit on BD FACSLyric™ Flow Cytometer was established based on the acceptance criteria. |
| Limit of Blank | CLSI EP17-A2 | To evaluate Limit of Blank of BD Leucocount™ Kit with BD FACSLyric™ Flow Cytometer. | The Limit of Blank (LoB) is 0 cells/μL for rWBC absolute counts. |
| Limit of Quantitation | CLSI EP17-A2 | To evaluate Limit of Quantitation of BD Leucocount™ Kit with BD FACSLyric™ Flow Cytometer. | The Limit of Quantitation (LoQ) is 0.7 cells/μL for rWBC absolute counts. |
| Interfering Substances | CLSI EP07-Ed3 CLSI EP37-Ed1 | To assess the performance of BD Leucocount™ Kit for use with BD FACSLyric™ Flow Cytometer in the presence of potential endogenous and exogenous interfering substances. | There was no detectable interference for all the tested substances, with the exception of triglycerides, a known interferent that causes light scattering changes. This interfering condition caused by lipemic specimens will be reflected in the instructions for use. |
| Sample and Reagent Carryover | CLSI H26-A2 | To evaluate sample and reagent carryover using BD Leucocount™ Kit with BD FACSLyric™ Flow Cytometer. | All acceptance criteria were met. |
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 9 of 10
{9}
| Study | Standard | Objective | Results |
| --- | --- | --- | --- |
| BD FACSLyric™ Manual Versus Loader Acquisition Equivalency | CLSI EP09c | To determine the method bias of the BD Leucocount™ Kit for samples acquired on the BD FACSLyric™ Flow Cytometer using BD FACS™ Universal Loader, versus samples that are acquired manually. | All acceptance criteria were met. |
| Anticoagulants | CLSI EP07-Ed3 | To assess the compatibility of the BD Leucocount™ Kit with various anticoagulants used in blood products collection and storage. | All tested anticoagulants (ACD, CPD, CP2D, CPDA, EDTA, 4% Sodium Citrate, Heparin) met the study acceptance criteria. |
**Table 5. BD Leucocount™ Kit Performance Studies Conducted at Clinical Sites**
| Study | Standard | Objective | Results |
| --- | --- | --- | --- |
| Method Comparison | CLSI EP09c | To determine the method bias between the subject device (BD Leucocount™ Kit on BD FACSLyric™ Flow Cytometer) versus the predicate device (BD Leucocount™ Kit on BD FACSVia™ Flow Cytometer). | A total of 520 specimens were enrolled across four testing sites. 502 specimens were evaluable. 18 specimens were non-evaluable. The acceptance criteria were met. |
| Inter-Site Reproducibility | CLSI EP05-A3 | To assess inter-site reproducibility of BD Leucocount™ Kit on BD FACSLyric™ Flow Cytometer. | The results of inter-site reproducibility for the BD Leucocount™ Kit on the BD FACSLyric™ Flow Cytometer demonstrated that the variability across three sites met study acceptance criteria. |
| Specimen Stability | CLSI EP25-A | To support the sample stability and stained sample stability claims for leucoreduced red blood cells and leucoreduced platelets. | The acceptance criteria were met. The results from the study support the specimen stability claims. |
## 8. Conclusion
The subject device, BD Leucocount™ Kit (with BD Leucocount™ Assay Module, to enable use with the BD FACSLyric™ Flow Cytometer), is substantially equivalent to the predicate device, BD Leucocount™ Kit (for use with the BD FACSVia™ Flow Cytometer).
BD Biosciences Traditional 510(k)
BD Leucocount™ Kit
Attachment 36 (510(k) Summary)
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.