BK230806 · Immucor, Inc. Norcross, GA · MMH · Apr 25, 2023 · Hematology
Device Facts
Record ID
BK230806
Device Name
ImmuLINK (v3.0)
Applicant
Immucor, Inc. Norcross, GA
Product Code
MMH · Hematology
Decision Date
Apr 25, 2023
Decision
SESE
Regulation
21 CFR 864.9165
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ImmuLINK is software intended for use in a blood banking environment as an aid in interfacing and managing data between blood bank instruments, Blood Establishment Computer Software, and Laboratory Information Systems. ImmuLINK includes the following features and/or functions: • It provides the ability to create and print reports of tests and results. • It supports the ability to compare current results with previous results. • It supports the automatic ordering of reflex tests. • It provides a tool so that test results produced by a testing platform can be remotely validated prior to delivery to an LIS. Results can also be viewed remotely and edited prior to delivery to an LIS. • It provides bi-directional communication between an LIS and instrument platform(s). • It provides a tool to assist with antibody exclusion for Capture-R results that is based upon the exclusion (rule-out) approach recommended in the Technical Manual (AABB).
Device Story
ImmuLINK (v3.0) is a blood bank software interface; connects automated/semi-automated instruments (e.g., Galileo Neo, NEO Iris, Echo, AIS) with LIS and Blood Establishment Computer Software. Inputs: patient/donor sample data, test results from instruments. Processing: data integration, remote validation, bi-directional communication, and new Antibody ID feature. Output: validated test results, customizable reports, antibody exclusion suggestions. Used in blood banks/reference labs by laboratory personnel. Antibody ID feature uses AABB Technical Manual-based rules to suggest antibody assignments or identify need for further investigation; excludes antibodies using antigen-positive cells. Benefits: streamlined workflow, remote result validation, standardized antibody rule-out process, improved regulatory compliance.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by technological comparison and performance testing of the software's rule-based algorithm against established AABB Technical Manual guidelines.
Technological Characteristics
Software-based data management system; thin-client architecture; supports multi-instrument connectivity and concurrent users. Features include real-time delta checking, reflex ordering, and rules-based decision processing. Algorithm is rule-based, derived from AABB Technical Manual (18th Ed) for antibody exclusion. Connectivity: bi-directional LIS/instrument interface. Data storage and retrieval capabilities included.
Indications for Use
Indicated for use by trained laboratory personnel in a blood banking environment to interface and manage data between blood bank instruments, Blood Establishment Computer Software, and Laboratory Information Systems, including antibody exclusion (rule-out) for Capture-R results.
Regulatory Classification
Identification
Blood establishment computer software (BECS) is a device used in the manufacture of blood and blood components to assist in the prevention of disease in humans by identifying ineligible donors, by preventing the release of unsuitable blood and blood components for transfusion or for further manufacturing into products for human treatment or diagnosis, by performing compatibility testing between donor and recipient, or by performing positive identification of patients and blood components at the point of transfusion to prevent transfusion reactions. This generic type of device may include a BECS accessory, a device intended for use with BECS to augment the performance of the BECS or to expand or modify its indications for use.
Special Controls
*Classification.* Class II (special controls). The special controls for these devices are:(1) Software performance and functional requirements including detailed design specifications (
*e.g.,* algorithms or control characteristics, alarms, device limitations, and safety requirements).(2) Verification and validation testing and hazard analysis must be performed.
(3) Labeling must include:
(i) Software limitations;
(ii) Unresolved anomalies, annotated with an explanation of the impact on safety or effectiveness;
(iii) Revision history; and
(iv) Hardware and peripheral specifications.
(4) Traceability matrix must be performed.
(5) Performance testing to ensure the safety and effectiveness of the system must be performed, including when adding new functional requirements (
*e.g.,* electrical safety, electromagnetic compatibility, or wireless coexistence).
{0}
Traditional 510(k) Premarket Notification
IMMUCOR
## 7.0 510(k) Summary
### Date Prepared
April 14, 2023
### 510(k) Owner
Immucor, Inc.
3130 Gateway Drive
Norcross, Georgia 30071
Establishment Registration Number: 1034569
### Contact Information
Name of Contact: Steven Appel
Phone Number: 770-440-2051
Fax Number: 770-242-8930
### Modified Device Information
| Device Classification Name | Blood Establishment Computer Software And Accessories |
| --- | --- |
| Tradename | ImmuLINK (v3.0) |
| Device Name | Blood Establishment Computer Software And Accessories |
| Regulation Number | 864.9165 |
| Classification Product Code | MMH |
| Classification Advisory Committee | Hematology |
| Review Advisory Committee | Hematology |
### Predicate Device Information
| Tradename | ImmuLINK, V.1.7 |
| --- | --- |
| Company | Immucor, Inc. |
| 510(K) Number | BK150333 |
| Decision Date | 11/19/2015 |
| Regulation Number | 864.9165 |
| Classification Product Code | MMH |
| Tradename | Bio-Rad IH-AbID Integrated Antibody Identification |
| --- | --- |
| Company | Bio-Rad Medical Diagnostics GmbH Industriestrasse 1 |
| 510(K) Number | BK220769 |
| Decision Date | 12/7/2022 |
| Regulation Number | 864.9165 |
| Classification Product Code | MMH |
Modification to ImmuLINK
Addition of Antibody ID Feature
57 of 102
{1}
Traditional 510(k) Premarket Notification
IMMUCOR
## Device Description
ImmuLINK (ImmuLINK®|manage combined with ImmuLINK®|archive) is a software package designed to provide an interface between the operator, blood bank automated (and semi-automated) instruments and Blood Establishment Computer Software, such as the Laboratory Information System (LIS). ImmuLINK|manage, combined with ImmuLINK|archive, allows the integration of sample, donor, or patient data that has been collected from multiple sources and provides data management tools to create customizable reports from the diverse data.
ImmuLINK|manage, combined with ImmuLINK|archive, is configurable to interface with multiple instrument platforms including (but not limited to) Immucor's Galileo Neo, NEO Iris, Galileo Echo and Echo Lumena Blood Bank Systems, and BioArray's Array Imaging System (AIS) semi-automated instrument and associated products.
ImmuLINK|manage is used as a remote validation tool so that the result of a test produced by a testing platform can be validated prior to delivery to an LIS. This delivery is characterized by the transmission of an information package. The result validation requires an electronic signature for approval and can be performed remotely without having to be in the laboratory where the result was produced. Results can also be viewed and edited prior to the ImmuLINK|manage validation.
ImmuLINK|manage has bi-directional features, whereby test orders can be retrieved from the LIS and then transmitted to the appropriate instrument platform(s).
ImmuLINK|manage includes an Antibody ID feature which provides review of Master List for Capture-R products combined with instrument results and optional antibody exclusion (rule-out) capabilities.
ImmuLINK|archive is used to create and view result archives, which are stored in a secure environment.
### Relationship between Instruments and ImmuLINK|manage
The diagram below summarizes the functional relationships between instruments and ImmuLINK|manage, using the example of two (2) blood bank instruments:
Modification to ImmuLINK
Addition of Antibody ID Feature
58 of 102
{2}
Traditional 510(k) Premarket Notification
IMMUCOR

# ImmuLINK Antibody ID Feature
Antibody ID (also referred to as Panel ID) allows viewing master list information, executing the Antibody ID algorithm evaluation, assigning an antibody to the result, and printing the results on a master list template.
The Antibody ID algorithm for applying rules as defined in AABB manual for rule out/in of antibodies in blood samples. The algorithm will provide user with suggested antibody assignments or antibodies that require additional investigation.
The algorithm available with Antibody ID is a first step in the antibody identification process. The algorithm will exclude using one double dose example of an antigen where the genetic background is known. The algorithm rules were derived using the AABB Technical Manual, 18th Edition, Chapter 16 – “Identification of Antibodies to Red Cell Antigens”.
In cases like Lewis and P1 blood groups, the antigenic presentation cannot be determined. To accommodate this the algorithm will use either two or three antigen positive cells that do not react with the sample to exclude. Immucor personnel can configure ImmuLINK to use two or three cells according to customer policy or practice. In the presence of anti-D, C and E can be excluded on single dose cells (r'r, and r"r). This exclusion is included in the algorithm calculation of ImmuLINK.
# Intended Use
ImmuLINK is software intended for use in a blood banking environment as an aid in interfacing and managing data between blood bank instruments, Blood Establishment Computer Software, and Laboratory Information Systems. ImmuLINK includes the following features and/or functions:
- It provides the ability to create and print reports of tests and results.
- It supports the ability to compare current results with previous results.
- It supports the automatic ordering of reflex tests.
- It provides a tool so that test results produced by a testing platform can be
Modification to ImmuLINK
Addition of Antibody ID Feature
59 of 102
{3}
Traditional 510(k) Premarket Notification
IMMUCOR
remotely validated prior to delivery to an LIS. Results can also be viewed remotely and edited prior to delivery to an LIS.
- It provides bi-directional communication between an LIS and instrument platform(s).
- It provides a tool to assist with antibody exclusion for Capture-R results that is based upon the exclusion (rule-out) approach recommended in the Technical Manual (AABB).
### Technological Comparison to Predicate Devices
Below is a summary of the technological characteristics of modified device compared to the predicate devices.
| Characteristic / Feature | Predicate Device: ImmuLINK (v.1.7) | Modified Device: ImmuLINK (v3.0) |
| --- | --- | --- |
| Classification | II | SAME |
| Product Code | MMH | SAME |
| Regulation Number | 21 CFR 864.9165 | SAME |
| Multi-Instrument Connectivity | Capable of interfacing with more than one instrument at a time | SAME |
| Thin-client | A computer or a computer program which depends heavily on some other computer (its server) to fulfill its traditional computation | SAME |
| Multiple, Concurrent Users | More than one end user | SAME |
| LIS Backup | Continuous results production while your LIS is unavailable | SAME |
| Integrated QC | Customer is able to perform the qc using the same run controls and software will check the status of the qc for each instrument | SAME |
| Data Storage | Able to store data | SAME |
| Data Retrieval | Able to retrieve data | SAME |
| Data Mining | Able to turn data into information through analysis | SAME |
| Serology Results | Capable of displaying serology based results | SAME |
| Molecular Results | Capable of displaying molecular based results | SAME |
| Regulatory Compliance | Documentation control for better regulatory compliance | SAME |
| Specimen Routing | Sample processing for better turnaround time | SAME |
| User-defined rules | Customer customizable rules based on workflow and lab practices | SAME |
| Rules-based decision processing | Yes | SAME |
Modification to ImmuLINK
Addition of Antibody ID Feature
60 of 102
{4}
Traditional 510(k) Premarket Notification
| Characteristic / Feature | Predicate Device: ImmuLINK (v.1.7) | Modified Device: ImmuLINK (v3.0) |
| --- | --- | --- |
| Real-time delta checking | Current results are checked against previous results (when available) in real time | SAME |
| Reflex ordering | Automatic ordering of follow-up test based on previous results | SAME |
| Previous results | Allow for storage of prior results | SAME |
| Color coding using rules logic | Ability to display information in various colors based on outcome of predefined rules | SAME |
| Filtering | Ability to display information based on end user profile | SAME |
| Reactions Images Storage | Storage of result images | SAME |
| Images Review | Provides viewing of result images | SAME |
| Instrument Log details | Captures and displays logs from instruments | SAME |
| Intended Use | ImmuLINK is software intended for use in a blood banking environment as an aid in interfacing and managing data between blood bank instruments, Blood Establishment Computer Software, and Laboratory Information Systems. ImmuLINK includes the following features and/or functions: • It provides the ability to create and print reports of tests and results. • It supports the ability to compare current results with previous results. • It supports the automatic ordering of reflex tests. • It provides a tool so that test results produced by a testing platform can be remotely validated prior to delivery to an LIS. Results can also be viewed remotely and edited prior to delivery to an LIS. • It provides bi-directional communication between an LIS and instrument platform(s). | ImmuLINK is software intended for use in a blood banking environment as an aid in interfacing and managing data between blood bank instruments, Blood Establishment Computer Software, and Laboratory Information Systems. ImmuLINK includes the following features and/or functions: • It provides the ability to create and print reports of tests and results. • It supports the ability to compare current results with previous results. • It supports the automatic ordering of reflex tests. • It provides a tool so that test results produced by a testing platform can be remotely validated prior to delivery to an LIS. Results can also be viewed remotely and edited prior to delivery to an LIS. • It provides bi-directional communication between an LIS and instrument platform(s). • It provides a tool to assist with antibody exclusion for Capture-R results that is based upon the exclusion (rule-out) approach recommended in the *Technical Manual* (AABB). |
| Characteristic / Feature | Bio-Rad IH-AbID Integrated Antibody Identification | Modified Device: ImmuLINK (v3.0) |
| --- | --- | --- |
| Classification | II | SAME |
| Product Code | MMH | SAME |
| Regulation Number | 21 CFR 864.9165 | SAME |
Modification to ImmuLINK
Addition of Antibody ID Feature
61 of 102
{5}
Traditional 510(k) Premarket Notification
IMMUCOR
| Characteristic / Feature | Bio-Rad IH-AbID Integrated Antibody Identification | Modified Device: ImmuLINK (v3.0) |
| --- | --- | --- |
| Tests managed | Antibody Screening Antibody Identification | SAME |
| User interface | Duplicates the existing antibody test procedures used in blood reference laboratories in a graphical electronic form. | SAME |
| Use environment | Blood bank software used by reference laboratories and transfusion services—an application with functionality specifically for blood establishments | SAME |
| Interpretation | Makes no decision about which antibodies are present. It helps the technologist select the best red cells for testing, analyze which antibodies are ruled out and decide which antibodies are present. | SAME |
| Reagent data | Reagent data is updated by the submitter from data sheets provided with each new lot of reagent cells distributed by the original manufacturers. | SAME |
| Displayed results | Shows which antibodies are ruled out based on negative test results for both homozygous and heterozygous positive antigens on reagent cells, and displays a count of these rule-outs. | SAME |
| Results management | The software's results are repeatable so a reviewer may check the technologists' work and decisions. | SAME |
| Audit trail | Audit trail of processing steps and user actions | SAME |
Modification to ImmuLINK
Addition of Antibody ID Feature
62 of 102
{6}
Traditional 510(k) Premarket Notification
IMMUCOR®
| Characteristic / Feature | Bio-Rad IH-AbID Integrated Antibody Identification | Modified Device: ImmuLINK (v3.0) |
| --- | --- | --- |
| Intended Use | IH-AbID Integrated Antibody Identification Software is an optional software module for IH- Com that provides the user with guidance and information on the identification of antibodies to red blood cell antigens subsequent to the immunohematology testing with the IH-system. For use by trained laboratory personnel, in a blood banking environment | ImmuLINK is software intended for use in a blood banking environment as an aid in interfacing and managing data between blood bank instruments, Blood Establishment Computer Software, and Laboratory Information Systems. ImmuLINK includes the following features and/or functions: • It provides the ability to create and print reports of tests and results. • It supports the ability to compare current results with previous results. • It supports the automatic ordering of reflex tests. • It provides a tool so that test results produced by a testing platform can be remotely validated prior to delivery to an LIS. Results can also be viewed remotely and edited prior to delivery to an LIS. • It provides bi-directional communication between an LIS and instrument platform(s). • It provides a tool to assist with antibody exclusion for Capture-R results that is based upon the exclusion (rule-out) approach recommended in the Technical Manual (AABB). |
### Basis for Claim of Substantial Equivalence
The modified ImmuLINK is substantially equivalent to the predicate devices relative to technological characteristics of the software devices.
This Traditional 510(k) is submitted to modify legally a marketed, predicate device.
Fundamental scientific technology of the proposed device is unchanged from the legally marketed, predicate. There are no significant differences between the modified device and the predicate as related to the Intended Use or Principle of Operation, except for the addition to the Intended Use statement that it provides antibody exclusion (rule-out) capabilities for Capture-R results [with the Antibody ID feature].
Immucor concludes, based on the indications for use, technological characteristics, and performance testing, that modified ImmuLINK with Antibody ID is safe and effective for the intended use described above.
Modification to ImmuLINK
Addition of Antibody ID Feature
63 of 102
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.