BK220668 · Alma Lasers, Inc. · QKL · Jun 12, 2023 · General, Plastic Surgery
Device Facts
Record ID
BK220668
Device Name
VorFat System
Applicant
Alma Lasers, Inc.
Product Code
QKL · General, Plastic Surgery
Decision Date
Jun 12, 2023
Decision
SESE
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VorFat System is intended for the closed-loop processing of lipoaspirate tissue in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue for aesthetic body contouring is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery. Only legally marketed accessory items, such as syringes, should be used with the system. Any fat concentrated with this system is to be transferred back to the patient without any additional manipulation.
Device Story
Closed-loop system for processing lipoaspirate; includes canister with filter and vortex mixer. Input: lipoaspirate harvested via legally marketed lipoplasty system. Operation: tissue enters collection canister; vortex mixer applies mechanical action to reduce fat cluster sizes; process preserves cell viability and tissue micro-architecture. Output: micronized fraction of adipose tissue transferred to syringes for autologous re-injection. Used in clinical/surgical settings by physicians. Benefits: provides homogenous, micronized adipose tissue for autologous transfer within operative procedure time.
Clinical Evidence
Bench testing only. Evaluated tissue cluster size (mm²), cell viability, and red blood cell content. Results demonstrated suitable characteristics for intended use and equivalence to predicate.
Technological Characteristics
Closed-loop system; 100mL canister filter; vortex mixer with sterile cover sleeve. Mechanical processing principle; no non-manual energy used for fragmentation. Canister is sterile (SAL 10^-6); mixer is multi-use with single-use sterile cover.
Indications for Use
Indicated for patients undergoing surgical procedures requiring autologous adipose tissue transfer for aesthetic body contouring across various specialties including orthopedic, arthroscopic, neurosurgery, gastrointestinal, urological, general, gynecological, thoracic, laparoscopic, and plastic/reconstructive surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}
# 510(k) Summary
## I. SUBMITTER
Owner/Submitter: Alma Lasers, Inc.
485 Half Day Rd Suite # 100
Buffalo Grove, Illinois, 60089Phone: 224-377-2019Contact Person: Jessica Rivera-MontejoDate Prepared: May 5, 2023
## II. DEVICE
Name of Device: VorFat System
Common or Usual Name: Lipoaspirate Washing System For Aesthetic Body Contouring
Classification Name: Suction Lipoplasty System
Regulation Number: 21 CFR 878.5040
Regulatory Class: Class II
Product Code: QKL
Classification Panel: General and Plastic Surgery
## III. PREDICATE DEVICE
Primary: The Lipogems System, K171135 / BK220712
## IV. DEVICE DESCRIPTION
The canister with filter (fat canister) of the VorFat System is a closed loop tissue collection system that includes a vacuum port, collection port, tissue port and lid. The canister with filter was previously cleared in K092482. It is used with legally marketed lipoplasty systems to separate, filter, and microfragment adipose tissue for re-injection. As the tissue is harvested from the patient, it enters the collection canister via the collection port in the canister lid. The vacuum port is connected to the standard waste canister, which is connected between the device and the legally marketed lipoplasty system. During harvest, waste is filtered out to the waste canister. After harvest, the collection canister is secured in the adapter attached to the mixer. The mixer applies a mechanical action to the tissue in the collection canister that allows for reduction of the fat cluster sizes. This gentle processing diminishes tissue damage, thereby preserving the cell viability and tissue micro-architecture of the adipose tissue. The product is a micronized fraction of fatty tissue that can be transferred to syringes via the tissue port for autologous fat re-injection. The canister filter is offered in a 100mL size.
{1}
The subject device VorFat System includes a canister, vortex mixer with canister adapter, and sterile cover for the mixer. The mechanical action of the Subject Device mixer produces a micronized fraction of the adipose tissue, and this technological characteristic is comparable to the predicate device K171135/BK220712. There is no additional non-manual energy delivered to aid in fragmentation of the adipose tissue.
## V. INDICATION FOR USE
The VorFat System is intended for the closed-loop processing of lipoaspirate tissue in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue for aesthetic body contouring is desired:
- orthopedic surgery,
- arthroscopic surgery,
- neurosurgery,
- gastrointestinal and affiliated organ surgery,
- urological surgery,
- general surgery,
- gynecological surgery,
- thoracic surgery,
- laparoscopic surgery,
- plastic and reconstructive surgery
Only legally marketed accessory items, such as syringes, should be used with the system. Any fat concentrated with this system is to be transferred back to the patient without any additional manipulation.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The VorFat System is substantially equivalent in terms of indications for use, technological characteristics, and performance characteristics to the predicate device. Tissue collection, separation and filtering for autologous reinjection is the technological principle for the predicate device. The VorFat System is substantially equivalent to the predicate device in terms of mechanical processing to produce a micronized fraction of tissue for autologous transfer using syringes.
The following Table provides a comparison and evidence of substantial equivalence of the subject device with the predicate device:
| Characteristic | Primary Predicate K171135 / BK220712 | Subject Device | Comparison |
| --- | --- | --- | --- |
| Device Name | The Lipogems System | VorFat System | N/A |
| Product Code | MUU / QKL | QKL | Same |
| Regulation | 21 CFR 878.5040 | 21 CFR 878.5040 | Same |
| Intended Use | Mechanical processing, filtering and transferring of autologous adipose tissue | Mechanical processing, filtering, and transferring of autologous adipose tissue | Same |
| Indications for Use | The Lipogems System is a sterile medical device intended for the closed-loop processing of lipoaspirate | The VorFat System is intended for the closed-loop processing of lipoaspirate tissue in medical | Same |
{2}
| Characteristic | Primary Predicate K171135 / BK220712 | Subject Device | Comparison |
| --- | --- | --- | --- |
| | tissue in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation. | procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue for aesthetic body contouring is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery. Only legally marketed accessory items, such as syringes, should be used with the system. Any fat concentrated with this system is to be transferred back to the patient without any additional manipulation. | |
| Prescription Use | Yes | Yes | Same |
| Design Features | The fat canister filter works with an aspiration system to separate, mechanically process, and filter adipose tissue for autologous reinjection. | The fat canister filter works with an aspiration system to separate, mechanically process, and filter adipose tissue for autologous reinjection. The fat canister is provided in the manufacturer's original packaging with original labeling. | Same |
| Operational Principle | Fat is aspirated into a syringe, and transferred to the Lipogems System device, where under constant gravity the fat is mixed with saline and undergoes micronization of the lipidic clusters through a sieve. The unit is shaken while the saline flows through, thus removing oil and blood residuals, and further reducing the clusters. This processing diminishes tissue disaggregation, thus preserving the cell and tissue micro-architecture of the adipose tissue. Finally, the concentrated adipose tissue clusters are passed out of the device into a collecting syringe for transferring back to the patient. | The VorFat System is a closed loop tissue collection system that includes a canister filter (fat canister) and vortex mixer, used with a legally marketed lipoplasty system to separate, filter, and microfragment adipose tissue for autologous reinjection. The canister is connected to a rapid stirrer/vortex mixer. The mechanical mixing action diminishes tissue disaggregation, thus preserving the cell and tissue micro-architecture of the adipose tissue. The result of processing with the VorFat System is a homogenous and micronized fraction of adipose tissue, which allows a patient's subcutaneous fat to be processed for autologous reinjection within the operative procedure time. | Substantially equivalent |
| Sterility | Canister: Sterile, SAL 10^{-6} Mixer: Shaken by hand by user | Canister: Sterile, SAL 10^{-6}, purchased from third party, previously cleared Mixer: Non-sterile (covered with sterile sleeve during use) | Same |
| Single Use? | Canister: Single Use | Canister: Single Use Mixer: Multi-use Cover sleeve: Single Use | Same |
{3}
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Bench Testing/Assessment
The following evaluations were performed for the VorFat System configuration to demonstrate substantial equivalence:
- Tissue Cluster Size Analysis (Adipose Aggregate/Cluster Size in mm²)
- Cell Viability
- Red Blood Cell Content
Both the subject and predicate devices had suitable characteristics for their indication for use.
## VIII. CONCLUSIONS
Based on the information contained within this submission, Alma Lasers, Inc. believes that the VorFat System is substantially equivalent to the identified predicate device and does not introduce any new or significantly modified risks, and no new questions of safety and efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.